{"id":{"repo_id":"ucl","oai_identifier":"oai:eprints.ucl.ac.uk.OAI2:1354095"},"canonical_url":"https://search.ndltd.org/etd/ucl/oai:eprints.ucl.ac.uk.OAI2:1354095","repository":{"repo_id":"ucl","name":"University College London","base_url":"https://discovery.ucl.ac.uk/cgi/oai2"},"display":{"title":"Development of endovascular stent-grafts based on a nanocomposite polymer","abstract":"Objective: To use a novel nanocomposite polymer based on polyhedral oligomeric silsesquioxane-poly(carbonate-urea)urethane with superior bio-mimetic properties in the development of endovascular stent-grafts. Methods: A self-expanding and sutureless aortic stent-graft was developed using nanocomposite polymer bonded to Nitinol. A new aortic model was designed for physiological assessment of stent-grafts and compliance and viscoelasticity were measured and compared with porcine aortas as control. The stent-grafts (n=4) were fatigue tested using in vitro accelerated model for 400-million cycles equivalent to 10-years in human body and compared with zero-cycled control. A curved and conformable stent-graft was developed for thoracic aorta and aortic arch. Compliance and stiffness index of the thoracic stent-graft were measured in vitro and compared to FDA-approved Gore TagTM stent-graft based on ePTFE (expanded polytetrafluoroethylene). MRI compatibility of the thoracic stent-graft was assessed by analysing signal attenuation and velocity measurements (flux) and compared to FDA-approved Medtronic ValiantTM. Results: The stent-graft had expanded diameter of 31.1 mm and was successfully collapsed to 6.5 mm to achieve delivery profile similar to current devices. The thoracic stent-graft had uniform graft thickness of 150.7±6.6 µg and conformed to the curvature of aortic arch. The new aortic model was significantly more compliant than porcine aortas with no significant difference in elastic stiffness. All the stent-grafts successfully completed accelerated pulsatile fatigue testing. Scanning electron microscopy images confirmed uniform surface topography. There was no loss of tensile strength, or compliance and no evidence of thermo-mechanical degradation in the nanocomposite polymer. Compliance of the thoracic stent-graft was significantly better compared with ePTFE stent-graft (3.3±0.61 vs. 2.3±0.95 %/mm Hg x 10-2; P=0.0003). On MRI, there was no significant signal attenuation and no significant difference in flux between Valiant and nanocomposite polymer stent-grafts (102±2.27 vs. 99.8±2.4 ml/sec; P=0.33). Conclusions: A new endovascular stent-graft based on novel design and nanocomposite polymer with properties of compliance, viscoelasticity, anti-thrombogenicity and MRI compatibility has been developed. Sutureless technology with new biocompatible material bonded to Nitinol stents proved to be robust with no separation over accelerated 10-year cycle. These stent-grafts have the potential to address poor long-term durability, thrombogenicity, and compliance mismatch associated with present generation devices and reduce reintervention rate.","abstract_html":"Objective: To use a novel nanocomposite polymer based on polyhedral oligomeric silsesquioxane-poly(carbonate-urea)urethane with superior bio-mimetic properties in the development of endovascular stent-grafts. Methods: A self-expanding and sutureless aortic stent-graft was developed using nanocomposite polymer bonded to Nitinol. A new aortic model was designed for physiological assessment of stent-grafts and compliance and viscoelasticity were measured and compared with porcine aortas as control. The stent-grafts (n=4) were fatigue tested using in vitro accelerated model for 400-million cycles equivalent to 10-years in human body and compared with zero-cycled control. A curved and conformable stent-graft was developed for thoracic aorta and aortic arch. Compliance and stiffness index of the thoracic stent-graft were measured in vitro and compared to FDA-approved Gore TagTM stent-graft based on ePTFE (expanded polytetrafluoroethylene). MRI compatibility of the thoracic stent-graft was assessed by analysing signal attenuation and velocity measurements (flux) and compared to FDA-approved Medtronic ValiantTM. Results: The stent-graft had expanded diameter of 31.1 mm and was successfully collapsed to 6.5 mm to achieve delivery profile similar to current devices. The thoracic stent-graft had uniform graft thickness of 150.7±6.6 µg and conformed to the curvature of aortic arch. The new aortic model was significantly more compliant than porcine aortas with no significant difference in elastic stiffness. All the stent-grafts successfully completed accelerated pulsatile fatigue testing. Scanning electron microscopy images confirmed uniform surface topography. There was no loss of tensile strength, or compliance and no evidence of thermo-mechanical degradation in the nanocomposite polymer. Compliance of the thoracic stent-graft was significantly better compared with ePTFE stent-graft (3.3±0.61 vs. 2.3±0.95 %/mm Hg x 10-2; P=0.0003). On MRI, there was no significant signal attenuation and no significant difference in flux between Valiant and nanocomposite polymer stent-grafts (102±2.27 vs. 99.8±2.4 ml/sec; P=0.33). Conclusions: A new endovascular stent-graft based on novel design and nanocomposite polymer with properties of compliance, viscoelasticity, anti-thrombogenicity and MRI compatibility has been developed. Sutureless technology with new biocompatible material bonded to Nitinol stents proved to be robust with no separation over accelerated 10-year cycle. These stent-grafts have the potential to address poor long-term durability, thrombogenicity, and compliance mismatch associated with present generation devices and reduce reintervention rate.","abstract_has_math":false,"creators":["Desai, MY"],"institution":"University College London","degree_name":null,"degree_level":"Doctoral","degree_discipline":null,"degree_department":null,"school":null,"contributors":["Seifalian, AM","Hamilton, G"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2012,"date_issued":"2012-08","date_published":"2012-08","updated_at":"2026-11-26T03:28:40Z","subjects":[],"languages":["en","eng"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://discovery.ucl.ac.uk/id/eprint/1354095/","outbound_label":"Repository record","outbound_source":"dc:identifier.uri","record_url":"https://discovery.ucl.ac.uk/id/eprint/1354095/","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aucl%3Aoai%3Aeprints.ucl.ac.uk.OAI2%3A1354095","prefix":"oai_dc"},"source_record":{"url":"https://discovery.ucl.ac.uk/cgi/oai2?verb=GetRecord&metadataPrefix=uketd_dc&identifier=oai%3Aeprints.ucl.ac.uk.OAI2%3A1354095","prefix":"uketd_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Seifalian, AM","Hamilton, G"]},{"key":"dc:creator","label":"Author","values":["Desai, MY"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2012-08-28"]},{"key":"dc:date.issued","label":"Date","values":["2012-08"]},{"key":"dc:publisher.department","label":"Dc Publisher Department","values":["Research Department of General Surgery"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University College London"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["Doctoral"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["en","eng"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://discovery.ucl.ac.uk/id/eprint/1354095/"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Objective: To use a novel nanocomposite polymer based on polyhedral oligomeric silsesquioxane-poly(carbonate-urea)urethane with superior bio-mimetic properties in the development of endovascular stent-grafts. Methods: A self-expanding and sutureless aortic stent-graft was developed using nanocomposite polymer bonded to Nitinol. A new aortic model was designed for physiological assessment of stent-grafts and compliance and viscoelasticity were measured and compared with porcine aortas as control. The stent-grafts (n=4) were fatigue tested using in vitro accelerated model for 400-million cycles equivalent to 10-years in human body and compared with zero-cycled control. A curved and conformable stent-graft was developed for thoracic aorta and aortic arch. Compliance and stiffness index of the thoracic stent-graft were measured in vitro and compared to FDA-approved Gore TagTM stent-graft based on ePTFE (expanded polytetrafluoroethylene). MRI compatibility of the thoracic stent-graft was assessed by analysing signal attenuation and velocity measurements (flux) and compared to FDA-approved Medtronic ValiantTM. Results: The stent-graft had expanded diameter of 31.1 mm and was successfully collapsed to 6.5 mm to achieve delivery profile similar to current devices. The thoracic stent-graft had uniform graft thickness of 150.7±6.6 µg and conformed to the curvature of aortic arch. The new aortic model was significantly more compliant than porcine aortas with no significant difference in elastic stiffness. All the stent-grafts successfully completed accelerated pulsatile fatigue testing. Scanning electron microscopy images confirmed uniform surface topography. There was no loss of tensile strength, or compliance and no evidence of thermo-mechanical degradation in the nanocomposite polymer. Compliance of the thoracic stent-graft was significantly better compared with ePTFE stent-graft (3.3±0.61 vs. 2.3±0.95 %/mm Hg x 10-2; P=0.0003). On MRI, there was no significant signal attenuation and no significant difference in flux between Valiant and nanocomposite polymer stent-grafts (102±2.27 vs. 99.8±2.4 ml/sec; P=0.33). Conclusions: A new endovascular stent-graft based on novel design and nanocomposite polymer with properties of compliance, viscoelasticity, anti-thrombogenicity and MRI compatibility has been developed. Sutureless technology with new biocompatible material bonded to Nitinol stents proved to be robust with no separation over accelerated 10-year cycle. These stent-grafts have the potential to address poor long-term durability, thrombogenicity, and compliance mismatch associated with present generation devices and reduce reintervention rate."]},{"key":"dc:format","label":"Dc Format","values":["application/pdf"]},{"key":"dcterms:accessrights","label":"Dcterms Accessrights","values":["public"]},{"key":"dc:title","label":"Title","values":["Development of endovascular stent-grafts based on a nanocomposite polymer"]}]}],"canonical_facts":{"dc:contributor":["Seifalian, AM","Hamilton, G"],"dc:creator":["Desai, MY"],"dc:date":["2012-08-28"],"dc:date.issued":["2012-08"],"dc:description.abstract":["Objective: To use a novel nanocomposite polymer based on polyhedral oligomeric silsesquioxane-poly(carbonate-urea)urethane with superior bio-mimetic properties in the development of endovascular stent-grafts. Methods: A self-expanding and sutureless aortic stent-graft was developed using nanocomposite polymer bonded to Nitinol. A new aortic model was designed for physiological assessment of stent-grafts and compliance and viscoelasticity were measured and compared with porcine aortas as control. The stent-grafts (n=4) were fatigue tested using in vitro accelerated model for 400-million cycles equivalent to 10-years in human body and compared with zero-cycled control. A curved and conformable stent-graft was developed for thoracic aorta and aortic arch. Compliance and stiffness index of the thoracic stent-graft were measured in vitro and compared to FDA-approved Gore TagTM stent-graft based on ePTFE (expanded polytetrafluoroethylene). MRI compatibility of the thoracic stent-graft was assessed by analysing signal attenuation and velocity measurements (flux) and compared to FDA-approved Medtronic ValiantTM. Results: The stent-graft had expanded diameter of 31.1 mm and was successfully collapsed to 6.5 mm to achieve delivery profile similar to current devices. The thoracic stent-graft had uniform graft thickness of 150.7±6.6 µg and conformed to the curvature of aortic arch. The new aortic model was significantly more compliant than porcine aortas with no significant difference in elastic stiffness. All the stent-grafts successfully completed accelerated pulsatile fatigue testing. Scanning electron microscopy images confirmed uniform surface topography. There was no loss of tensile strength, or compliance and no evidence of thermo-mechanical degradation in the nanocomposite polymer. Compliance of the thoracic stent-graft was significantly better compared with ePTFE stent-graft (3.3±0.61 vs. 2.3±0.95 %/mm Hg x 10-2; P=0.0003). On MRI, there was no significant signal attenuation and no significant difference in flux between Valiant and nanocomposite polymer stent-grafts (102±2.27 vs. 99.8±2.4 ml/sec; P=0.33). Conclusions: A new endovascular stent-graft based on novel design and nanocomposite polymer with properties of compliance, viscoelasticity, anti-thrombogenicity and MRI compatibility has been developed. Sutureless technology with new biocompatible material bonded to Nitinol stents proved to be robust with no separation over accelerated 10-year cycle. These stent-grafts have the potential to address poor long-term durability, thrombogenicity, and compliance mismatch associated with present generation devices and reduce reintervention rate."],"dc:format":["application/pdf"],"dc:identifier.uri":["https://discovery.ucl.ac.uk/id/eprint/1354095/"],"dc:language":["en","eng"],"dc:publisher.department":["Research Department of General Surgery"],"dc:publisher.institution":["University College London"],"dcterms:accessrights":["public"],"dc:title":["Development of endovascular stent-grafts based on a nanocomposite polymer"],"dc:type":["Thesis"],"dc:type.qualificationlevel":["Doctoral"]},"updated_at":"2026-11-26T03:28:40Z"}