{"id":{"repo_id":"umn","oai_identifier":"oai:conservancy.umn.edu:11299/277998"},"canonical_url":"https://search.ndltd.org/etd/umn/oai:conservancy.umn.edu:11299/277998","repository":{"repo_id":"umn","name":"University of Minnesota","base_url":"https://conservancy.umn.edu/server/oai/request"},"display":{"title":"Creating a custom calibration for the portable X-ray fluorescence instrument to identify iron formation using seawater signatures","abstract":"Iron formations are chemical sedimentary rocks that formed via precipitation from seawater between ~3.7-1.8 Ga. These rocks are hypothesized to retain seawater geochemical signatures as REE+Y trends, which can be obtained by analytical geochemistry, if the samples contain minimal detrital contamination and limited metamorphic overprinting. Current analytical geochemical methods, such as ICP-MS/-OES, are costly and time consuming, especially when samples do not contain the geochemical signatures of interest. Additionally, these methods require destroying sample first, which results in the loss of textural and petrographic data. However, in the last two decades the pXRF has been marketed as a reliable, rapid, and non-destructive instrument able to produce quality data, nearly on par with traditional geochemistry (i.e., ICP-MS/-OES), particularly when used as an initial screening tool for sample selection. The main goal of this study was to determine if the pXRF could identify seawater signatures in iron formation and distinguish them from detrital contamination using internal and custom calibrations. If the pXRF could detect REE+Y concentrations, then detecting seawater signature trends locked in iron formation could be plausible in the field. Additionally, if the pXRF could measure REE+Y seawater signatures in iron formation, then large-scale field geochemical mapping of in situ units could be possible. This study tested this hypothesis using the Bruker Tracer 5g pXRF and its internal calibrations along with one custom iron formation calibration (Bruker, 2019). Initial testing of the pXRF’s internal calibrations used four representative elements (Fe, Si, Al, and La) of iron formation geochemistry to understand instrument functionality. Even though the pXRF demonstrated precision and sensitivity to chemical changes, preliminary tests showed that the internal calibrations could not reliably detect the REE+Y representative element La, therefore a custom calibration was required. This iron formation specific calibration was created using Bruker (2019) EasyCal software and 24 Biwabik Iron Formation samples as powdered internal standards. These internal standards collectively contained the four representative elements and 35 additional elements including all of the REE+Y. After creating the calibration, the powdered internal standards, associated billets, and core were analyzed using the pXRF custom calibration. The pXRF iron formation calibration was able to detect the four interested elements and majority of the other elements within two standard deviations of the ICP-MS/-OES concentrations for the same samples across the three material types (powdered, billet, core). The final test looked at the accuracy and sensitivity of the instrument to detect key elements of the seawater signature in different rock samples of known and unknown geochemistry. Samples with measured ICP-MS/-OES concentrations showed that pXRF measurements for rare elements consistently high compared to standard digestion ICP-MS/OES concentrations, despite being within two deviations of ICP-MS/-OES values. The tests also showed that the calibration had difficulty detecting low concentration elements. Future corrections for non-powdered samples may be able to produce pXRF data that more closely approaches values obtained via digestion chemistry approaches. In summary, the geochemical sensitivity tests demonstrated that the iron formation calibration could differentiate seawater signature bearing iron formation from non-iron formation units by comparing the general seawater signature trends. The key outcome of the study is the demonstration that a custom calibration can be used to detect trace elements in iron-rich samples that are powered, or have flat, fresh-cut surfaces with minimal irregularities. This key result opens the possibility of detailed, mm-scale trace element mapping of fresh cut drill core.","abstract_html":"Iron formations are chemical sedimentary rocks that formed via precipitation from seawater between ~3.7-1.8 Ga. These rocks are hypothesized to retain seawater geochemical signatures as REE+Y trends, which can be obtained by analytical geochemistry, if the samples contain minimal detrital contamination and limited metamorphic overprinting. Current analytical geochemical methods, such as ICP-MS/-OES, are costly and time consuming, especially when samples do not contain the geochemical signatures of interest. Additionally, these methods require destroying sample first, which results in the loss of textural and petrographic data. However, in the last two decades the pXRF has been marketed as a reliable, rapid, and non-destructive instrument able to produce quality data, nearly on par with traditional geochemistry (i.e., ICP-MS/-OES), particularly when used as an initial screening tool for sample selection. The main goal of this study was to determine if the pXRF could identify seawater signatures in iron formation and distinguish them from detrital contamination using internal and custom calibrations. If the pXRF could detect REE+Y concentrations, then detecting seawater signature trends locked in iron formation could be plausible in the field. Additionally, if the pXRF could measure REE+Y seawater signatures in iron formation, then large-scale field geochemical mapping of in situ units could be possible. This study tested this hypothesis using the Bruker Tracer 5g pXRF and its internal calibrations along with one custom iron formation calibration (Bruker, 2019). Initial testing of the pXRF’s internal calibrations used four representative elements (Fe, Si, Al, and La) of iron formation geochemistry to understand instrument functionality. Even though the pXRF demonstrated precision and sensitivity to chemical changes, preliminary tests showed that the internal calibrations could not reliably detect the REE+Y representative element La, therefore a custom calibration was required. This iron formation specific calibration was created using Bruker (2019) EasyCal software and 24 Biwabik Iron Formation samples as powdered internal standards. These internal standards collectively contained the four representative elements and 35 additional elements including all of the REE+Y. After creating the calibration, the powdered internal standards, associated billets, and core were analyzed using the pXRF custom calibration. The pXRF iron formation calibration was able to detect the four interested elements and majority of the other elements within two standard deviations of the ICP-MS/-OES concentrations for the same samples across the three material types (powdered, billet, core). The final test looked at the accuracy and sensitivity of the instrument to detect key elements of the seawater signature in different rock samples of known and unknown geochemistry. Samples with measured ICP-MS/-OES concentrations showed that pXRF measurements for rare elements consistently high compared to standard digestion ICP-MS/OES concentrations, despite being within two deviations of ICP-MS/-OES values. The tests also showed that the calibration had difficulty detecting low concentration elements. Future corrections for non-powdered samples may be able to produce pXRF data that more closely approaches values obtained via digestion chemistry approaches. In summary, the geochemical sensitivity tests demonstrated that the iron formation calibration could differentiate seawater signature bearing iron formation from non-iron formation units by comparing the general seawater signature trends. The key outcome of the study is the demonstration that a custom calibration can be used to detect trace elements in iron-rich samples that are powered, or have flat, fresh-cut surfaces with minimal irregularities. This key result opens the possibility of detailed, mm-scale trace element mapping of fresh cut drill core.","abstract_has_math":false,"creators":["Prue, Ann Marie"],"institution":null,"degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2025,"date_issued":"2025-08","date_published":"2025-08","updated_at":"2026-11-19T09:52:05Z","subjects":["Master of Science","Master of Science in Geological Sciences","Department of Earth and Environmental Sciences","Swenson College of Science and Engineering","University of Minnesota Duluth","Plan As (thesis-based master's degrees)"],"languages":["en"],"rights":[],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.citation","label":"Dc Identifier Citation","values":["Prue, Ann Marie. (2025). Creating a custom calibration for the portable X-ray fluorescence instrument to identify iron formation using seawater signatures. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/277998."],"render_values":[{"text":"Prue, Ann Marie. (2025). Creating a custom calibration for the portable X-ray fluorescence instrument to identify iron formation using seawater signatures. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/277998.","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/11299/277998","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/11299/277998","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"oai_pmh_record":{"url":"https://union.ndltd.org/oai-pmh/?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Aunion.ndltd.org%3Arepo%3Aumn%3Aoai%3Aconservancy.umn.edu%3A11299%2F277998","prefix":"oai_dc"},"source_record":{"url":"https://conservancy.umn.edu/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Aconservancy.umn.edu%3A11299%2F277998","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:creator","label":"Author","values":["Prue, Ann Marie"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2026-02-03T19:56:22Z"]},{"key":"dc:date.issued","label":"Date","values":["2025-08"]},{"key":"dc:type","label":"Dc Type","values":["Thesis or Dissertation"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Master of Science","Master of Science in Geological Sciences","Department of Earth and Environmental Sciences","Swenson College of Science and Engineering","University of Minnesota Duluth","Plan As (thesis-based master's degrees)"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.citation","label":"Dc Identifier Citation","values":["Prue, Ann Marie. (2025). Creating a custom calibration for the portable X-ray fluorescence instrument to identify iron formation using seawater signatures. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/277998."]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://hdl.handle.net/11299/277998"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["University of Minnesota M.S. thesis. August 2025. Major: Geological Sciences. Advisor: Latisha Brengman. 1 computer file (PDF); xvi, 242 pages."]},{"key":"dc:description.abstract","label":"Abstract","values":["Iron formations are chemical sedimentary rocks that formed via precipitation from seawater between ~3.7-1.8 Ga. These rocks are hypothesized to retain seawater geochemical signatures as REE+Y trends, which can be obtained by analytical geochemistry, if the samples contain minimal detrital contamination and limited metamorphic overprinting. Current analytical geochemical methods, such as ICP-MS/-OES, are costly and time consuming, especially when samples do not contain the geochemical signatures of interest. Additionally, these methods require destroying sample first, which results in the loss of textural and petrographic data. However, in the last two decades the pXRF has been marketed as a reliable, rapid, and non-destructive instrument able to produce quality data, nearly on par with traditional geochemistry (i.e., ICP-MS/-OES), particularly when used as an initial screening tool for sample selection. The main goal of this study was to determine if the pXRF could identify seawater signatures in iron formation and distinguish them from detrital contamination using internal and custom calibrations. If the pXRF could detect REE+Y concentrations, then detecting seawater signature trends locked in iron formation could be plausible in the field. Additionally, if the pXRF could measure REE+Y seawater signatures in iron formation, then large-scale field geochemical mapping of in situ units could be possible. This study tested this hypothesis using the Bruker Tracer 5g pXRF and its internal calibrations along with one custom iron formation calibration (Bruker, 2019). Initial testing of the pXRF’s internal calibrations used four representative elements (Fe, Si, Al, and La) of iron formation geochemistry to understand instrument functionality. Even though the pXRF demonstrated precision and sensitivity to chemical changes, preliminary tests showed that the internal calibrations could not reliably detect the REE+Y representative element La, therefore a custom calibration was required. This iron formation specific calibration was created using Bruker (2019) EasyCal software and 24 Biwabik Iron Formation samples as powdered internal standards. These internal standards collectively contained the four representative elements and 35 additional elements including all of the REE+Y. After creating the calibration, the powdered internal standards, associated billets, and core were analyzed using the pXRF custom calibration. The pXRF iron formation calibration was able to detect the four interested elements and majority of the other elements within two standard deviations of the ICP-MS/-OES concentrations for the same samples across the three material types (powdered, billet, core). The final test looked at the accuracy and sensitivity of the instrument to detect key elements of the seawater signature in different rock samples of known and unknown geochemistry. Samples with measured ICP-MS/-OES concentrations showed that pXRF measurements for rare elements consistently high compared to standard digestion ICP-MS/OES concentrations, despite being within two deviations of ICP-MS/-OES values. The tests also showed that the calibration had difficulty detecting low concentration elements. Future corrections for non-powdered samples may be able to produce pXRF data that more closely approaches values obtained via digestion chemistry approaches. In summary, the geochemical sensitivity tests demonstrated that the iron formation calibration could differentiate seawater signature bearing iron formation from non-iron formation units by comparing the general seawater signature trends. The key outcome of the study is the demonstration that a custom calibration can be used to detect trace elements in iron-rich samples that are powered, or have flat, fresh-cut surfaces with minimal irregularities. This key result opens the possibility of detailed, mm-scale trace element mapping of fresh cut drill core."]},{"key":"dc:title","label":"Title","values":["Creating a custom calibration for the portable X-ray fluorescence instrument to identify iron formation using seawater signatures"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:creator":["Prue, Ann Marie"],"dc:date.accessioned":["2026-02-03T19:56:22Z"],"dc:date.issued":["2025-08"],"dc:description":["University of Minnesota M.S. thesis. August 2025. Major: Geological Sciences. Advisor: Latisha Brengman. 1 computer file (PDF); xvi, 242 pages."],"dc:description.abstract":["Iron formations are chemical sedimentary rocks that formed via precipitation from seawater between ~3.7-1.8 Ga. These rocks are hypothesized to retain seawater geochemical signatures as REE+Y trends, which can be obtained by analytical geochemistry, if the samples contain minimal detrital contamination and limited metamorphic overprinting. Current analytical geochemical methods, such as ICP-MS/-OES, are costly and time consuming, especially when samples do not contain the geochemical signatures of interest. Additionally, these methods require destroying sample first, which results in the loss of textural and petrographic data. However, in the last two decades the pXRF has been marketed as a reliable, rapid, and non-destructive instrument able to produce quality data, nearly on par with traditional geochemistry (i.e., ICP-MS/-OES), particularly when used as an initial screening tool for sample selection. The main goal of this study was to determine if the pXRF could identify seawater signatures in iron formation and distinguish them from detrital contamination using internal and custom calibrations. If the pXRF could detect REE+Y concentrations, then detecting seawater signature trends locked in iron formation could be plausible in the field. Additionally, if the pXRF could measure REE+Y seawater signatures in iron formation, then large-scale field geochemical mapping of in situ units could be possible. This study tested this hypothesis using the Bruker Tracer 5g pXRF and its internal calibrations along with one custom iron formation calibration (Bruker, 2019). Initial testing of the pXRF’s internal calibrations used four representative elements (Fe, Si, Al, and La) of iron formation geochemistry to understand instrument functionality. Even though the pXRF demonstrated precision and sensitivity to chemical changes, preliminary tests showed that the internal calibrations could not reliably detect the REE+Y representative element La, therefore a custom calibration was required. This iron formation specific calibration was created using Bruker (2019) EasyCal software and 24 Biwabik Iron Formation samples as powdered internal standards. These internal standards collectively contained the four representative elements and 35 additional elements including all of the REE+Y. After creating the calibration, the powdered internal standards, associated billets, and core were analyzed using the pXRF custom calibration. The pXRF iron formation calibration was able to detect the four interested elements and majority of the other elements within two standard deviations of the ICP-MS/-OES concentrations for the same samples across the three material types (powdered, billet, core). The final test looked at the accuracy and sensitivity of the instrument to detect key elements of the seawater signature in different rock samples of known and unknown geochemistry. Samples with measured ICP-MS/-OES concentrations showed that pXRF measurements for rare elements consistently high compared to standard digestion ICP-MS/OES concentrations, despite being within two deviations of ICP-MS/-OES values. The tests also showed that the calibration had difficulty detecting low concentration elements. Future corrections for non-powdered samples may be able to produce pXRF data that more closely approaches values obtained via digestion chemistry approaches. In summary, the geochemical sensitivity tests demonstrated that the iron formation calibration could differentiate seawater signature bearing iron formation from non-iron formation units by comparing the general seawater signature trends. The key outcome of the study is the demonstration that a custom calibration can be used to detect trace elements in iron-rich samples that are powered, or have flat, fresh-cut surfaces with minimal irregularities. This key result opens the possibility of detailed, mm-scale trace element mapping of fresh cut drill core."],"dc:identifier.citation":["Prue, Ann Marie. (2025). Creating a custom calibration for the portable X-ray fluorescence instrument to identify iron formation using seawater signatures. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/277998."],"dc:identifier.uri":["https://hdl.handle.net/11299/277998"],"dc:language.iso":["en"],"dc:subject":["Master of Science","Master of Science in Geological Sciences","Department of Earth and Environmental Sciences","Swenson College of Science and Engineering","University of Minnesota Duluth","Plan As (thesis-based master's degrees)"],"dc:title":["Creating a custom calibration for the portable X-ray fluorescence instrument to identify iron formation using seawater signatures"],"dc:type":["Thesis or Dissertation"],"others:access_status":["open.access"]},"updated_at":"2026-11-19T09:52:05Z"}