{"id":{"repo_id":"unt","oai_identifier":"info:ark/67531/metadc9786"},"canonical_url":"https://search.ndltd.org/etd/unt/info:ark/67531/metadc9786","repository":{"repo_id":"unt","name":"University of North Texas","base_url":"https://digital.library.unt.edu/oai/"},"display":{"title":"Errors in skin temperature measurements.","abstract":"Numerical simulation is used to investigate the accuracy of a direct-contact device for measuring skin-surface temperature. A variation of thermal conductivity of the foam has greater effect on the error rather than a variation of the blood perfusion rate. For a thermal conductivity of zero, an error of 1.5 oC in temperature was identified. For foam pad conductivities of 0.03 and 0.06 W/m-oC, the errors are 0.5 and 0.15 oC. For the transient study, with k=0 W/m-oC, it takes 4,900 seconds for the temperature to reach steady state compared with k=0.03 W/m-oC and k=0.06 W/m-oC where it takes 3,000 seconds. The configuration without the foam and in presence of an air gap between the skin surface and the sensor gives the most uniform temperature profile.","abstract_html":"Numerical simulation is used to investigate the accuracy of a direct-contact device for measuring skin-surface temperature. A variation of thermal conductivity of the foam has greater effect on the error rather than a variation of the blood perfusion rate. For a thermal conductivity of zero, an error of 1.5 oC in temperature was identified. For foam pad conductivities of 0.03 and 0.06 W/m-oC, the errors are 0.5 and 0.15 oC. For the transient study, with k=0 W/m-oC, it takes 4,900 seconds for the temperature to reach steady state compared with k=0.03 W/m-oC and k=0.06 W/m-oC where it takes 3,000 seconds. The configuration without the foam and in presence of an air gap between the skin surface and the sensor gives the most uniform temperature profile.","abstract_has_math":false,"creators":["Dugay, Murielle"],"institution":"University of North Texas","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":["Boetcher, Sandra","Traum, Matthew J.","Boubekri, Nourredine"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2008,"date_issued":"2008-12","date_published":"2008-12","updated_at":"2026-11-19T08:45:59Z","subjects":["simulation","thermal conductivity","bioheat transport","skin-surface temperature","heat conduction","Body temperature -- Measurement.","Medical thermography."],"languages":["English"],"rights":["Public","Copyright","Dugay, Murielle","Copyright is held by the author, unless otherwise noted. All rights reserved."],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier","label":"Identifier","values":["oclc: 437349288","https://digital.library.unt.edu/ark:/67531/metadc9786/","ark: ark:/67531/metadc9786"],"render_values":[{"text":"oclc: 437349288","href":null,"code":true},{"text":"https://digital.library.unt.edu/ark:/67531/metadc9786/","href":"https://digital.library.unt.edu/ark:/67531/metadc9786/","code":true},{"text":"ark: ark:/67531/metadc9786","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://doi.org/10.12794/metadc9786","outbound_label":"DOI","outbound_source":"dc:identifier","record_url":"https://doi.org/10.12794/metadc9786","record_label":"Repository record","record_kind":"doi","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digital.library.unt.edu/oai/?verb=GetRecord&metadataPrefix=oai_dc&identifier=info%3Aark%2F67531%2Fmetadc9786","prefix":"oai_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Boetcher, Sandra","Traum, Matthew J.","Boubekri, Nourredine"]},{"key":"dc:creator","label":"Author","values":["Dugay, Murielle"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2008-12"]},{"key":"dc:publisher","label":"Institution","values":["University of North Texas"]},{"key":"dc:type","label":"Dc Type","values":["Thesis or Dissertation"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["simulation","thermal conductivity","bioheat transport","skin-surface temperature","heat conduction","Body temperature -- Measurement.","Medical thermography."]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["English"]},{"key":"dc:rights","label":"Dc Rights","values":["Public","Copyright","Dugay, Murielle","Copyright is held by the author, unless otherwise noted. All rights reserved."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["oclc: 437349288","doi: 10.12794/metadc9786","https://digital.library.unt.edu/ark:/67531/metadc9786/","ark: ark:/67531/metadc9786"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Numerical simulation is used to investigate the accuracy of a direct-contact device for measuring skin-surface temperature. A variation of thermal conductivity of the foam has greater effect on the error rather than a variation of the blood perfusion rate. For a thermal conductivity of zero, an error of 1.5 oC in temperature was identified. For foam pad conductivities of 0.03 and 0.06 W/m-oC, the errors are 0.5 and 0.15 oC. For the transient study, with k=0 W/m-oC, it takes 4,900 seconds for the temperature to reach steady state compared with k=0.03 W/m-oC and k=0.06 W/m-oC where it takes 3,000 seconds. The configuration without the foam and in presence of an air gap between the skin surface and the sensor gives the most uniform temperature profile."]},{"key":"dc:format","label":"Dc Format","values":["Text"]},{"key":"dc:title","label":"Title","values":["Errors in skin temperature measurements."]}]}],"canonical_facts":{"dc:contributor":["Boetcher, Sandra","Traum, Matthew J.","Boubekri, Nourredine"],"dc:creator":["Dugay, Murielle"],"dc:date":["2008-12"],"dc:description":["Numerical simulation is used to investigate the accuracy of a direct-contact device for measuring skin-surface temperature. A variation of thermal conductivity of the foam has greater effect on the error rather than a variation of the blood perfusion rate. For a thermal conductivity of zero, an error of 1.5 oC in temperature was identified. For foam pad conductivities of 0.03 and 0.06 W/m-oC, the errors are 0.5 and 0.15 oC. For the transient study, with k=0 W/m-oC, it takes 4,900 seconds for the temperature to reach steady state compared with k=0.03 W/m-oC and k=0.06 W/m-oC where it takes 3,000 seconds. The configuration without the foam and in presence of an air gap between the skin surface and the sensor gives the most uniform temperature profile."],"dc:format":["Text"],"dc:identifier":["oclc: 437349288","doi: 10.12794/metadc9786","https://digital.library.unt.edu/ark:/67531/metadc9786/","ark: ark:/67531/metadc9786"],"dc:language":["English"],"dc:publisher":["University of North Texas"],"dc:rights":["Public","Copyright","Dugay, Murielle","Copyright is held by the author, unless otherwise noted. All rights reserved."],"dc:subject":["simulation","thermal conductivity","bioheat transport","skin-surface temperature","heat conduction","Body temperature -- Measurement.","Medical thermography."],"dc:title":["Errors in skin temperature measurements."],"dc:type":["Thesis or Dissertation"]},"updated_at":"2026-11-19T08:45:59Z"}