{"id":{"repo_id":"usf","oai_identifier":"oai:digitalcommons.usf.edu:etd-10898"},"canonical_url":"https://search.ndltd.org/etd/usf/oai:digitalcommons.usf.edu:etd-10898","repository":{"repo_id":"usf","name":"University of South Florida","base_url":"https://digitalcommons.usf.edu/do/oai/"},"display":{"title":"Design of Acoustic Wave Sensor System for Detection of BTEX in Gas Phase","abstract":"Benzene, toluene, ethylbenzene, and xylene are classified as volatile organic compounds (VOC), which if present at higher concentrations can cause acute and/or chronic health effects. To protect people from VOCs, the occupational safety and health agency (OSHA) established short term and eight-hour workday exposure limits. The short time levels of exposure for BTEX materials are 5 ppm, 150 ppm, 125 ppm, and 150 ppm, for benzene, toluene, ethylbenzene, and xylene respectively. The eight-hour workday exposure for these same materials are 1 ppm for benzene, 200 ppm for toluene, 100 ppm for ethylbenzene and 100 ppm for xylene. Although there are several laboratory instruments, for example GS, MS, PID, that are used for measuring these materials in a laboratory environment. This work shows that acoustic wave sensors should also be considered for monitoring, detecting, and measuring these types of materials. The sensors are not only sensitive, reliable, and easy to operate, but they are also suitable for providing real time measurements.For this thesis work we used a quartz crystal microbalance (QCM) coated with a polymer solvent, and diluted setup that allowed us to successfully detect approximate 500 ppm of these analytes in a gas phase. In our apparatus system, a controlled amount of BTEX was released into a nitrogen flow stream that brought the mixture of Nitrogen and BTEX to the sensor. The concentration of analytes was controlled through a gastight syringe attached to a syringe pump. The nitrogen flow was controlled by a LabVIEW program and mass flow controller. The presence of the plasticizer in the polymer coating increased the sensitivity of the sensor to the tested analytes. We will also show a design of a surface acoustic wave (SAW) test device capable for gas sensing. Using this fixture and a SAW sensor with Rayleigh waves and a polymer coated delay line, we were able to detect 55 ppm of benzene in a gas phase.","abstract_html":"Benzene, toluene, ethylbenzene, and xylene are classified as volatile organic compounds (VOC), which if present at higher concentrations can cause acute and/or chronic health effects. To protect people from VOCs, the occupational safety and health agency (OSHA) established short term and eight-hour workday exposure limits. The short time levels of exposure for BTEX materials are 5 ppm, 150 ppm, 125 ppm, and 150 ppm, for benzene, toluene, ethylbenzene, and xylene respectively. The eight-hour workday exposure for these same materials are 1 ppm for benzene, 200 ppm for toluene, 100 ppm for ethylbenzene and 100 ppm for xylene. Although there are several laboratory instruments, for example GS, MS, PID, that are used for measuring these materials in a laboratory environment. This work shows that acoustic wave sensors should also be considered for monitoring, detecting, and measuring these types of materials. The sensors are not only sensitive, reliable, and easy to operate, but they are also suitable for providing real time measurements.For this thesis work we used a quartz crystal microbalance (QCM) coated with a polymer solvent, and diluted setup that allowed us to successfully detect approximate 500 ppm of these analytes in a gas phase. In our apparatus system, a controlled amount of BTEX was released into a nitrogen flow stream that brought the mixture of Nitrogen and BTEX to the sensor. The concentration of analytes was controlled through a gastight syringe attached to a syringe pump. The nitrogen flow was controlled by a LabVIEW program and mass flow controller. The presence of the plasticizer in the polymer coating increased the sensitivity of the sensor to the tested analytes. We will also show a design of a surface acoustic wave (SAW) test device capable for gas sensing. Using this fixture and a SAW sensor with Rayleigh waves and a polymer coated delay line, we were able to detect 55 ppm of benzene in a gas phase.","abstract_has_math":false,"creators":["Mitevska, Veselinka"],"institution":"University of South Florida","degree_name":"MS in Materials Science and Engineering (M.S.M.S.E)","degree_level":"Thesis","degree_discipline":"Chemical Engineering","degree_department":null,"school":null,"contributors":["Venkat R. Bhethanabotla, Ph.D.","Scott Campbell, Ph.D.","John Kuhn, Ph.D."],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2021,"date_issued":"2021-10-28T07:00:00Z","date_published":"2021-10-28T07:00:00Z","updated_at":"2026-11-19T08:48:23Z","subjects":["QCM","SAW","polymer film","sorption","frequency shift","Materials Science and Engineering"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.usf.edu/etd/9701","outbound_label":"Repository record","outbound_source":"dc:identifier","record_url":"https://digitalcommons.usf.edu/etd/9701","record_label":"Repository record","record_kind":"record","record_source":"dc:identifier","download_url":null,"download_label":null,"download_source":null,"access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digitalcommons.usf.edu/do/oai/?verb=GetRecord&metadataPrefix=oai_etdms&identifier=oai%3Adigitalcommons.usf.edu%3Aetd-10898","prefix":"oai_etdms"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Venkat R. Bhethanabotla, Ph.D.","Scott Campbell, Ph.D.","John Kuhn, Ph.D."]},{"key":"dc:creator","label":"Author","values":["Mitevska, Veselinka"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2021-10-28T07:00:00Z"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Chemical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Thesis","M.S.M.S.E."]},{"key":"thesis:degree_name","label":"Degree Name","values":["MS in Materials Science and Engineering (M.S.M.S.E)"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of South Florida"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["QCM","SAW","polymer film","sorption","frequency shift","Materials Science and Engineering"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.usf.edu/etd/9701"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Benzene, toluene, ethylbenzene, and xylene are classified as volatile organic compounds (VOC), which if present at higher concentrations can cause acute and/or chronic health effects. To protect people from VOCs, the occupational safety and health agency (OSHA) established short term and eight-hour workday exposure limits. The short time levels of exposure for BTEX materials are 5 ppm, 150 ppm, 125 ppm, and 150 ppm, for benzene, toluene, ethylbenzene, and xylene respectively. The eight-hour workday exposure for these same materials are 1 ppm for benzene, 200 ppm for toluene, 100 ppm for ethylbenzene and 100 ppm for xylene. Although there are several laboratory instruments, for example GS, MS, PID, that are used for measuring these materials in a laboratory environment. This work shows that acoustic wave sensors should also be considered for monitoring, detecting, and measuring these types of materials. The sensors are not only sensitive, reliable, and easy to operate, but they are also suitable for providing real time measurements.For this thesis work we used a quartz crystal microbalance (QCM) coated with a polymer solvent, and diluted setup that allowed us to successfully detect approximate 500 ppm of these analytes in a gas phase. In our apparatus system, a controlled amount of BTEX was released into a nitrogen flow stream that brought the mixture of Nitrogen and BTEX to the sensor. The concentration of analytes was controlled through a gastight syringe attached to a syringe pump. The nitrogen flow was controlled by a LabVIEW program and mass flow controller. The presence of the plasticizer in the polymer coating increased the sensitivity of the sensor to the tested analytes. We will also show a design of a surface acoustic wave (SAW) test device capable for gas sensing. Using this fixture and a SAW sensor with Rayleigh waves and a polymer coated delay line, we were able to detect 55 ppm of benzene in a gas phase."]},{"key":"dc:title","label":"Title","values":["Design of Acoustic Wave Sensor System for Detection of BTEX in Gas Phase"]}]}],"canonical_facts":{"dc:contributor":["Venkat R. Bhethanabotla, Ph.D.","Scott Campbell, Ph.D.","John Kuhn, Ph.D."],"dc:creator":["Mitevska, Veselinka"],"dc:date":["2021-10-28T07:00:00Z"],"dc:description":["Benzene, toluene, ethylbenzene, and xylene are classified as volatile organic compounds (VOC), which if present at higher concentrations can cause acute and/or chronic health effects. To protect people from VOCs, the occupational safety and health agency (OSHA) established short term and eight-hour workday exposure limits. The short time levels of exposure for BTEX materials are 5 ppm, 150 ppm, 125 ppm, and 150 ppm, for benzene, toluene, ethylbenzene, and xylene respectively. The eight-hour workday exposure for these same materials are 1 ppm for benzene, 200 ppm for toluene, 100 ppm for ethylbenzene and 100 ppm for xylene. Although there are several laboratory instruments, for example GS, MS, PID, that are used for measuring these materials in a laboratory environment. This work shows that acoustic wave sensors should also be considered for monitoring, detecting, and measuring these types of materials. The sensors are not only sensitive, reliable, and easy to operate, but they are also suitable for providing real time measurements.For this thesis work we used a quartz crystal microbalance (QCM) coated with a polymer solvent, and diluted setup that allowed us to successfully detect approximate 500 ppm of these analytes in a gas phase. In our apparatus system, a controlled amount of BTEX was released into a nitrogen flow stream that brought the mixture of Nitrogen and BTEX to the sensor. The concentration of analytes was controlled through a gastight syringe attached to a syringe pump. The nitrogen flow was controlled by a LabVIEW program and mass flow controller. The presence of the plasticizer in the polymer coating increased the sensitivity of the sensor to the tested analytes. We will also show a design of a surface acoustic wave (SAW) test device capable for gas sensing. Using this fixture and a SAW sensor with Rayleigh waves and a polymer coated delay line, we were able to detect 55 ppm of benzene in a gas phase."],"dc:identifier":["https://digitalcommons.usf.edu/etd/9701"],"dc:subject":["QCM","SAW","polymer film","sorption","frequency shift","Materials Science and Engineering"],"dc:title":["Design of Acoustic Wave Sensor System for Detection of BTEX in Gas Phase"],"thesis:degree_discipline":["Chemical Engineering"],"thesis:degree_level":["Thesis","M.S.M.S.E."],"thesis:degree_name":["MS in Materials Science and Engineering (M.S.M.S.E)"],"thesis:institution_name":["University of South Florida"]},"updated_at":"2026-11-19T08:48:23Z"}