{"id":{"repo_id":"washington","oai_identifier":"oai:digital.lib.washington.edu:1773/26148"},"canonical_url":"https://search.ndltd.org/etd/washington/oai:digital.lib.washington.edu:1773/26148","repository":{"repo_id":"washington","name":"University of Washington","base_url":"https://digital.lib.washington.edu/server/oai/request"},"display":{"title":"Low Power IC Design for Health Sensing","abstract":"This work proposes a new time-domain integration method to realize Electrochemical Impedance Spectroscopy (EIS). Unlike traditional EIS systems using Fast Fourier Transform (FFT) or Frequency Response Analyzer (FRA) based on a clean quadrature sinusoid stimulus, A single low-frequency, low-amplitude sinusoid stimulus is proposed by using a sinusoid DAC which fully eliminates the analog filter. The response current is integrated by a switched capacitor circuit which is synchronized by the proposed sinusoid DAC. The integration output is sampled and digitized by an 8-bit SAR ADC. The phase and amplitude information of target impedance can be extracted from the time-domain data. The error caused by harmonic generation in the sinusoid DAC can be suppressed by the time-domain integration. The (1×1.1) mm^2 prototype is fabricated in a 130nm CMOS process. It consumes 10µA from a 1.2V supply. Compared with the existing EIS systems, the test results show that the system can obtain a reasonable error with much smaller area and much lower power consumption.","abstract_html":"This work proposes a new time-domain integration method to realize Electrochemical Impedance Spectroscopy (EIS). Unlike traditional EIS systems using Fast Fourier Transform (FFT) or Frequency Response Analyzer (FRA) based on a clean quadrature sinusoid stimulus, A single low-frequency, low-amplitude sinusoid stimulus is proposed by using a sinusoid DAC which fully eliminates the analog filter. The response current is integrated by a switched capacitor circuit which is synchronized by the proposed sinusoid DAC. The integration output is sampled and digitized by an 8-bit SAR ADC. The phase and amplitude information of target impedance can be extracted from the time-domain data. The error caused by harmonic generation in the sinusoid DAC can be suppressed by the time-domain integration. The (1×1.1) mm^2 prototype is fabricated in a 130nm CMOS process. It consumes 10µA from a 1.2V supply. Compared with the existing EIS systems, the test results show that the system can obtain a reasonable error with much smaller area and much lower power consumption.","abstract_has_math":false,"creators":["Gu, Jingren"],"institution":null,"degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Otis, Brian"],"committee_chairs":[],"committee_members":[],"year":null,"date_issued":"","date_published":null,"updated_at":"2026-11-18T20:03:20Z","subjects":["Contact Lens; Electrochemical Impedance Spectroscopy; Low power IC design; Sinusoid DAC","Electrical engineering","Biomedical engineering"],"languages":["en_US"],"rights":["Copyright is held by the individual authors."],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier.other","label":"Dc Identifier Other","values":["Gu_washington_0250O_13516.pdf"],"render_values":[{"text":"Gu_washington_0250O_13516.pdf","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":"https://hdl.handle.net/1773/26148","outbound_label":"Handle","outbound_source":"dc:identifier.uri","record_url":"https://hdl.handle.net/1773/26148","record_label":"Repository record","record_kind":"handle","record_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://digital.lib.washington.edu/server/oai/request?verb=GetRecord&metadataPrefix=dim&identifier=oai%3Adigital.lib.washington.edu%3A1773%2F26148","prefix":"dim"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Otis, Brian"]},{"key":"dc:creator","label":"Author","values":["Gu, Jingren"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2014-10-13T16:57:48Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2015-12-14T17:55:52Z"]},{"key":"dc:date.submitted","label":"Dc Date Submitted","values":["2014"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Contact Lens; Electrochemical Impedance Spectroscopy; Low power IC design; Sinusoid DAC"]},{"key":"dc:subject.other","label":"Dc Subject Other","values":["Electrical engineering","Biomedical engineering"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["en_US"]},{"key":"dc:rights","label":"Dc Rights","values":["Copyright is held by the individual authors."]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.other","label":"Dc Identifier Other","values":["Gu_washington_0250O_13516.pdf"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1773/26148"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Thesis (Master's)--University of Washington, 2014"]},{"key":"dc:description.abstract","label":"Abstract","values":["This work proposes a new time-domain integration method to realize Electrochemical Impedance Spectroscopy (EIS). Unlike traditional EIS systems using Fast Fourier Transform (FFT) or Frequency Response Analyzer (FRA) based on a clean quadrature sinusoid stimulus, A single low-frequency, low-amplitude sinusoid stimulus is proposed by using a sinusoid DAC which fully eliminates the analog filter. The response current is integrated by a switched capacitor circuit which is synchronized by the proposed sinusoid DAC. The integration output is sampled and digitized by an 8-bit SAR ADC. The phase and amplitude information of target impedance can be extracted from the time-domain data. The error caused by harmonic generation in the sinusoid DAC can be suppressed by the time-domain integration. The (1×1.1) mm^2 prototype is fabricated in a 130nm CMOS process. It consumes 10µA from a 1.2V supply. Compared with the existing EIS systems, the test results show that the system can obtain a reasonable error with much smaller area and much lower power consumption."]},{"key":"dc:embargo.terms","label":"Dc Embargo Terms","values":["Restrict to UW for 1 year -- then make Open Access"]},{"key":"dc:format.mimetype","label":"Dc Format Mimetype","values":["application/pdf"]},{"key":"dc:title","label":"Title","values":["Low Power IC Design for Health Sensing"]},{"key":"others:access_status","label":"Others Access Status","values":["open.access"]}]}],"canonical_facts":{"dc:contributor.advisor":["Otis, Brian"],"dc:creator":["Gu, Jingren"],"dc:date.accessioned":["2014-10-13T16:57:48Z"],"dc:date.available":["2015-12-14T17:55:52Z"],"dc:date.submitted":["2014"],"dc:description":["Thesis (Master's)--University of Washington, 2014"],"dc:description.abstract":["This work proposes a new time-domain integration method to realize Electrochemical Impedance Spectroscopy (EIS). Unlike traditional EIS systems using Fast Fourier Transform (FFT) or Frequency Response Analyzer (FRA) based on a clean quadrature sinusoid stimulus, A single low-frequency, low-amplitude sinusoid stimulus is proposed by using a sinusoid DAC which fully eliminates the analog filter. The response current is integrated by a switched capacitor circuit which is synchronized by the proposed sinusoid DAC. The integration output is sampled and digitized by an 8-bit SAR ADC. The phase and amplitude information of target impedance can be extracted from the time-domain data. The error caused by harmonic generation in the sinusoid DAC can be suppressed by the time-domain integration. The (1×1.1) mm^2 prototype is fabricated in a 130nm CMOS process. It consumes 10µA from a 1.2V supply. Compared with the existing EIS systems, the test results show that the system can obtain a reasonable error with much smaller area and much lower power consumption."],"dc:embargo.terms":["Restrict to UW for 1 year -- then make Open Access"],"dc:format.mimetype":["application/pdf"],"dc:identifier.other":["Gu_washington_0250O_13516.pdf"],"dc:identifier.uri":["http://hdl.handle.net/1773/26148"],"dc:language.iso":["en_US"],"dc:rights":["Copyright is held by the individual authors."],"dc:subject":["Contact Lens; Electrochemical Impedance Spectroscopy; Low power IC design; Sinusoid DAC"],"dc:subject.other":["Electrical engineering","Biomedical engineering"],"dc:title":["Low Power IC Design for Health Sensing"],"dc:type":["Thesis"],"others:access_status":["open.access"]},"updated_at":"2026-11-18T20:03:20Z"}