{"id":{"repo_id":"whiterose","oai_identifier":"oai:etheses.whiterose.ac.uk:17952"},"canonical_url":"https://search.ndltd.org/etd/whiterose/oai:etheses.whiterose.ac.uk:17952","repository":{"repo_id":"whiterose","name":"White Rose University Consortium","base_url":"https://etheses.whiterose.ac.uk/cgi/oai2"},"display":{"title":"Comparison of Multiple Degrees of Electrode Surface Roughness: Impedance, Charge Storage Capacity, Biofouling and Biocompatibility","abstract":"Devices called neural electrodes are generally used to record and / or stimulate neural activity whilst extracellularly interfacing with neurons. The ultimate goal of the field is to develop neural electrodes that can continuously record and / or stimulate neural activity for long periods of time (10+ years). Electrodes require low electrode impedance to enable good signal to noise ratio and a high capacity for chemically stable injection of charge (CSC) for stimulating neural activity. Electrode function typically deteriorates after a few months in-vivo due two factors: biofouling and local cellular response (glial scarring). Biofouling is theorised to be an accumulation of multiple proteins at the electrode surface. However this has only been backed up experimentally by single protein models in literature. Electrodes need to be able to combat the effects of biofouling and glial scarring. This study uses nanometre scale roughened gold (Au) electrodes. Electrode roughening has previously been shown to lower impedance, increase CSC and reduce glial response and the effects of biofouling. We compared multiple degrees of roughness with the aim of finding the optimal degree for improving impedance, CSC, biofouling and cellular response. We found that surface roughening increased impedance and only increased CSC for two only degrees of roughness. To find the optimal degree of roughness across conditions, we suspect a larger range of roughness may be necessary to lower impedance with our fabrication technique. This study is the first to use a multiprotein biofouling model. Contrary to literature, we found that incubation with protein decreased impedance, likely due to protein-protein interactions not accounted for in single protein models. Biocompatibility was improved for two degrees of Au roughness. Roughening of SU8, a polymer used to surround the electrodes, decreased biocompatibility. We also used artificial cerebral spinal fluid (aCSF) as an electrolyte, which is more chemically similar to in-vivo than commonly used phosphate buffered saline solution (PBS). The use aCSF as a medium was significant as the measures in aCSF were different from that in PBS.","abstract_html":"Devices called neural electrodes are generally used to record and / or stimulate neural activity whilst extracellularly interfacing with neurons. The ultimate goal of the field is to develop neural electrodes that can continuously record and / or stimulate neural activity for long periods of time (10+ years). Electrodes require low electrode impedance to enable good signal to noise ratio and a high capacity for chemically stable injection of charge (CSC) for stimulating neural activity. Electrode function typically deteriorates after a few months in-vivo due two factors: biofouling and local cellular response (glial scarring). Biofouling is theorised to be an accumulation of multiple proteins at the electrode surface. However this has only been backed up experimentally by single protein models in literature. Electrodes need to be able to combat the effects of biofouling and glial scarring. This study uses nanometre scale roughened gold (Au) electrodes. Electrode roughening has previously been shown to lower impedance, increase CSC and reduce glial response and the effects of biofouling. We compared multiple degrees of roughness with the aim of finding the optimal degree for improving impedance, CSC, biofouling and cellular response. We found that surface roughening increased impedance and only increased CSC for two only degrees of roughness. To find the optimal degree of roughness across conditions, we suspect a larger range of roughness may be necessary to lower impedance with our fabrication technique. This study is the first to use a multiprotein biofouling model. Contrary to literature, we found that incubation with protein decreased impedance, likely due to protein-protein interactions not accounted for in single protein models. Biocompatibility was improved for two degrees of Au roughness. Roughening of SU8, a polymer used to surround the electrodes, decreased biocompatibility. We also used artificial cerebral spinal fluid (aCSF) as an electrolyte, which is more chemically similar to in-vivo than commonly used phosphate buffered saline solution (PBS). The use aCSF as a medium was significant as the measures in aCSF were different from that in PBS.","abstract_has_math":false,"creators":["Elms, Mungo Arthur"],"institution":"University of Leeds","degree_name":"M.Sc by research","degree_level":"masters","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Chakrabarty, Samit","Steenson, Paul"],"committee_chairs":[],"committee_members":[],"year":2017,"date_issued":"2017-02","date_published":"2017-02","updated_at":"2026-11-18T22:22:33Z","subjects":[],"languages":["eng"],"rights":[],"rights_urls":[],"identifier_entries":[],"person_identifier_entries":[]},"links":{"outbound_url":null,"outbound_label":null,"outbound_source":null,"record_url":null,"record_label":null,"record_kind":null,"record_source":null,"download_url":"https://etheses.whiterose.ac.uk/id/eprint/17952/1/Mungo%20Elms%20Masters%20By%20Research%20Thesis%20FINAL.pdf","download_label":"Download PDF","download_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://etheses.whiterose.ac.uk/cgi/oai2?verb=GetRecord&metadataPrefix=uketd_dc&identifier=oai%3Aetheses.whiterose.ac.uk%3A17952","prefix":"uketd_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Chakrabarty, Samit","Steenson, Paul"]},{"key":"dc:creator","label":"Author","values":["Elms, Mungo Arthur"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2017-02"]},{"key":"dc:date.issued","label":"Date","values":["2017-02"]},{"key":"dc:publisher.commercial","label":"Dc Publisher Commercial","values":["University of Leeds"]},{"key":"dc:publisher.department","label":"Dc Publisher Department","values":["Institute of Membrane and Systems Biology (Leeds)"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University of Leeds"]},{"key":"dc:relation.isreferencedby","label":"Dc Relation Isreferencedby","values":["https://etheses.whiterose.ac.uk/id/eprint/17952/"]},{"key":"dc:relation.references","label":"Dc Relation References","values":["Alexander, C.K. & Sadiku, M.N.O., 2001. Fundamentals of Electric Circuits 3rd ed., New York: McGraw-Hill Higher Education. Altuna, A. et al., 2013. SU-8 based microprobes for simultaneous neural depth recording and drug delivery in the brain. Lab on a chip, 13(7), pp.1422–30. Altuna, A. et al., 2012. SU-8 based microprobes with integrated planar electrodes for enhanced neural depth recording. Biosensors and Bioelectronics. Altuna, A., Berganzo, J. & Fernández, L.J., 2015. Polymer SU-8-Based Microprobes for Neural Recording and Drug Delivery. Frontiers in Materials, 2(June), pp.1–5. Aurian-Blajeni, B., Kimball, A.G. & Robblee, L.S., 1987. Correlation Between Charge Storage Capacity and Morphology. Journal of The Electrochemical Society, pp.2637–2638. Babcock, A. a. et al., 2003. Chemokine Expression by Glial Cells Directs Leukocytes to Sites of Axonal Injury in the CNS. J. Neurosci., 23(21), pp.7922–7930. Balaban, N.Q. et al., 2001. Force and focal adhesion assembly: a close relationship studied using elastic micropatterned substrates. Nature cell biology, 3(5), pp.466–472. Barrese, J.C. et al., 2013. Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates. Journal of Neural Engineering, 10(6), p.66014. Bershadsky, A.D. et al., 2006. Assembly and mechanosensory function of focal adhesions: Experiments and models. European Journal of Cell Biology, 85(3-4), pp.165–173. Bhandari, R., Negi, S. & Solzbacher, F., 2011. A novel surface modification method to achieve low impedance neural microelectrode arrays. 2011 16th International Solid-State Sensors, Actuators and Microsystems Conference, TRANSDUCERS’11, pp.2196–2199. Bidola, E.D., Bulhoes, L.O.S. & Rocha-Filho, R.C., 1994. Pt / HCl04 Interface CPE: Influence of Surface Roughness and Electrolyte Concentration. Electrochimica Acta, 39(5), pp.763–769. Mungo Elms 200688934 79 Biggs, M.J.P., Richards, R.G. & Dalby, M.J., 2010. Nanotopographical modification: a regulator of cellular function through focal adhesions. Nanomedicine: Nanotechnology, Biology, and Medicine, 6(5), pp.619–633. Biran, R., Martin, D.C. & Tresco, P.A., 2005. Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays. Experimental Neurology, 195(1), pp.115–126. Bjornsson, C.S. et al., 2006. Effects of insertion conditions on tissue strain and vascular damage during neuroprosthetic device insertion. Journal of neural engineering, 3(3), pp.196–207. Boehler, C. et al., 2015. Iridium Oxide ( IrOx ) serves as adhesion promoter for conducting polymers on neural microelectrodes. IEEE, 49, pp.22–24. Bosman, F.T. & Stamenkovic, I., 2003. Functional structure and composition of the extracellular matrix. Journal of Pathology, 200(4), pp.423–428. Brownson, D.A.C. & Banks, C.E., 2014. The Handbook of Graphene Electrochemistry 1st ed., London: Springer. Buffo, A., Rolando, C. & Ceruti, S., 2010. Astrocytes in the damaged brain: Molecular and cellular insights into their reactive response and healing potential. Biochemical Pharmacology, 79(2), pp.77–89. Burridge, K., 1988. Focal Adhesions: Transmembrane Junctions Between The Extracellular Matrix And The Cytoskeleton. Annual Review of Cell and Developmental Biology, 4(1), pp.487–525. Castagnola, E. et al., 2015. Pedot-CNT-coated low-impedance, ultra-flexible, and brain-conformable micro-ECoG arrays. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 23(3), pp.342–350. Cellot, G. et al., 2016. PEDOT:PSS interfaces support the development of neuronal synaptic networks with reduced neuroglia response in vitro. Frontiers in neuroscience, 9(521). Chapman, C.A.R. et al., 2016. Mechanisms of Reduced Astrocyte Surface Coverage in Cortical Neuron-Glia Co-cultures on Nanoporous Gold Surfaces. Cellular and Molecular Mungo Elms 200688934 80 Bioengineering, 9(3), pp.433–442. Chapman, C.A.R. et al., 2015. Nanoporous gold as a neural interface coating: Effects of topography, surface chemistry, and feature size. ACS Applied Materials and Interfaces, 7(13), pp.7093–7100. Chen, H. et al., 2011. Multilayered polypyrrole-coated carbon nanotubes to improve functional stability and electrical properties of neural electrodes. Journal of Physical Chemistry C, 115(13), pp.5492–5499. Chen, Z. et al., 2005. Impedance immunosensor based on receptor protein adsorbed directly on porous gold film. Analytica Chimica Acta, 553(1-2), pp.190–195. Chesler, M., 2008. Regulation and Modulation of pH in the Brain. Brain, pp.1183–1221. Cho, S.H. et al., 2008. Biocompatible SU-8-based microprobes for recording neural spike signals from regenerated peripheral nerve fibers. IEEE Sensors Journal, 8(11), pp.1830–1836. Chung, T., Wang, J.Q., Wang, J., et al., 2015. Electrode modifications to lower electrode impedance and improve neural signal recording sensitivity. Journal of Neural Engineering, 12(5), p.056018. Chung, T., Wang, J.Q., Cao, B., et al., 2015. Electrode modifications to lower electrode impedance and improve neural signal recording sensitivity. Journal of Neural Engineering, 12(5), p.056018. Cogan, S.F., 2008. Neural stimulation and recording electrodes. Annual review of biomedical engineering, 10, pp.275–309. Cohen, M. et al., 2004. Spatial and temporal sequence of events in cell adhesion: From molecular recognition to focal adhesion assembly. ChemBioChem, 5(10), pp.1393–1399. Compton, R.G. & Banks, C.E., 2011. Understanding Voltammetry 1st ed., Singapore: World Scientific. Cragg, P., Patterson, L. & Purves, M., 1977. The pH of brain extracellular fluid in the cat. The Journal of Physiology, 272(1), p.137. Mungo Elms 200688934 81 Cui, X. et al., 2003. In vivo studies of polypyrrole/peptide coated neural probes. Biomaterials, 24(5), pp.777–787. Curtis, A.S.G. et al., 2001. Substratum nanotopography and the adhesion of biological cells. Are symmetry or regularity of nanotopography important? Biophysical Chemistry, 94(3), pp.275–283. Daikhin, L., Kornyshev, a. & Urbakh, M., 1996. Double-layer capacitance on a rough metal surface. Physical Review E, 53(6), pp.6192–6199. Daikhin, L.I., Kornyshev, a. a. & Urbakh, M., 1997. Double layer capacitance on a rough metal surface: Surface roughness measured by “Debye ruler.” Electrochimica Acta, 42(19), pp.2853–2860. Debye, P., 1936. Dielectric properties of pure liquids. Chemical Reviews, 19, pp.171–182. Desai, S.A. et al., 2010. Improving impedance of implantable microwire multi-electrode arrays by ultrasonic electroplating of durable platinum black. Frontiers in neuroengineering, 3, p.5. Devore, J.L., Farnum, N.R. & Doi, J.A., 1999. Applied Statistics for Engineers and Scientists 3rd ed. M. Taylor, L. Wheel, & A. Coppola, eds., Stamford: Cengage Learning. Di, L. et al., 2011. Protein adsorption and peroxidation of rat retinas under stimulation of a neural probe coated with polyaniline. Acta Biomaterialia, 7(10), pp.3738–3745. Diener, A. et al., 2005. Control of focal adhesion dynamics by material surface characteristics. Biomaterials, 26(4), pp.383–392. Doña Rodríguez, J.M., Herrera Melián, J. a. & Pérez Peña, J., 2000. Determination of the Real Surface Area of Pt Electrodes by Hydrogen Adsorption Using Cyclic Voltammetry. Journal of Chemical Education, 77(9), p.1195. Fattahi, P. et al., 2014. A review of organic and inorganic biomaterials for neural interfaces. Advanced Materials, 26(12), pp.1846–1885. Fidler, P.S. et al., 1999. Comparing astrocytic cell lines that are inhibitory or permissive for axon growth: the major axon-inhibitory proteoglycan is NG2. The Journal of neuroscience : the official journal of the Society for Neuroscience, 19(20), pp.8778– Mungo Elms 200688934 82 8788. Frackowiak, E. & Beguin, F., 2001. Carbon materials for the electrochemical storage of energy in capacitors. PERGAMON Carbon, 39, pp.937–950. Frantz, C., Stewart, K.M. & Weaver, V.M., 2010. The extracellular matrix at a glance. Journal of cell science, 123, pp.4195–4200. Frumkin, A.N., 1960. The Double Layer in Electrochemistry. J. Electrochem. Soc, 107(5), pp.461–472. García, A.J., 2005. Get a grip: Integrins in cell-biomaterial interactions. Biomaterials, 26(36), pp.7525–7529. García-Gabaldón, M. et al., 2011. Electrochemical study of a simulated spent pickling solution. International Journal of Electrochemical Science, 6(2), pp.506–519. Ghazni, N.F., Cahill, C.M. & Stroman, P.W., 2010. Tactile sensory and pain networks in the human spinal cord and brain stem mapped by means of functional MR imaging. American Journal of Neuroradiology, 31(4), pp.661–667. Hai, A. et al., 2009. Spine-shaped gold protrusions improve the adherence and electrical coupling of neurons with the surface of micro-electronic devices. Journal of the Royal Society, Interface / the Royal Society, 6(41), pp.1153–65. Hamelin, A. & Martins, A.M., 1996. Cyclic voltammetry at gold single-crystal surfaces. Part 2. Behaviour of high-index faces. Joumal of Electroanalytical Chemistry, 407, pp.13–21. Heim, M. et al., 2012. Combined macro-/mesoporous microelectrode arrays for low-noise extracellular recording of neural networks. Journal of Neurophysiology, 108(6), pp.1793–1803. Hu, Z. et al., 2006. Nanopowder molding method for creating implantable high-aspect-ratio electrodes on thin flexible substrates. Biomaterials, 27(9), pp.2009–2017. Johnson, M.D., Otto, K.J. & Kipke, D.R., 2005. Repeated voltage biasing improves unit recordings by reducing resistive tissue impedances. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 13(2), pp.160–165. Kerner, Z., 1998. Impedance of rough capacitive electrodes : the role of surface disorder. , Mungo Elms 200688934 83 448, pp.139–142. Kerner, Z. & Pajkossy, T., 2000. On the origin of capacitance dispersion of rough electrodes. Electrochimica Acta, 46(2-3), pp.207–211. Kim, S.A., Kim, E.T. & Kim, S.J., 2009. Highly Efficient Cold Sputtered Iridium Oxide Films for Polyimide based Neural Stimulation Electrodes. Journal of Biomedical Engineering Research, 30(3), pp.199–204. Kocherbitov, V., 2015. Surface and Physical Chemistry. , p.Debye Screening Length. Available at: http://www.surfchem.info/calculate/Debye/. Koklu, A., Sabuncu, A.C. & Beskok, A., 2016. Rough Gold Electrodes for Decreasing Impedance at the Electrolyte/Electrode Interface. Electrochimica Acta. Kolb, D.M. & Schneider, J., 1986. Surface reconstruction in electrochemistry: Au(100-(5 × 20), Au(111)-(1 × 23) and Au(110)-(1 × 2). Electrochimica Acta, 31(8), pp.929–936. Kozai, T.D. et al., 2010. Reduction of neurovascular damage resulting from microelectrode insertion into the cerebral cortex using in vivo two-photon mapping. J Neural Eng, 7(4), p.46011. Kozai, T.D.Y. et al., 2016. Chronic In Vivo Evaluation of PEDOT / CNT for Stable Neural Recordings. IEEE Trans Biomed Eng, 63(1), pp.111–119. Kozai, T.D.Y. et al., 2014. Chronic tissue response to carboxymethyl cellulose based dissolvable insertion needle for ultra-small neural probes. Biomaterials, 35(34), pp.9255–9268. Kozai, T.D.Y., Vazquez, A.L., et al., 2012. In vivo two-photon microscopy reveals immediate microglial reaction to implantation of microelectrode through extension of processes. Journal of Neural Engineering, 9(6), p.066001. Kozai, T.D.Y., Langhals, N.B., et al., 2012. Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. Nature materials, 11(12), pp.1065–73. Kurtulus, O. & Seker, E., 2012. Nanotopography Effects on Astrocyte Attachment to Nanoporous Gold Surfaces. In IEEE EMBS. pp. 6568–6571. Mungo Elms 200688934 84 Lim, J.Y. et al., 2005. Human foetal osteoblastic cell response to polymer-demixed nanotopographic interfaces. Journal of the Royal Society, Interface / the Royal Society, 2(2), pp.97–108. Malaga, K.A. et al., 2016. Data-driven model comparing the effects of glial scarring and tip metallization loss on chronic neural recordings. Neural Engineering, 13(1), p.16010. Martic, S., Rains, M.K. & Kraatz, H.B., 2013. Probing copper/tau protein interactions electrochemically. Analytical Biochemistry, 442(2), pp.130–137. McConnell, G.C. et al., 2009. Implanted neural electrodes cause chronic, local inflammation that is correlated with local neurodegeneration. Journal of neural engineering, 6(5), p.056003. Meijs, S. et al., 2016. Influence of fibrous encapsulation on electro-chemical properties of TiN electrodes. Medical Engineering and Physics, 38(5), pp.468–476. Michel, E., Hernandez, D. & Lee, S.Y., 2016. Electrical conductivity and permittivity maps of brain tissues derived from water content based on T 1 -weighted acquisition. Magnetic Resonance in Medicine. Michel, R. et al., 2008. The Influence of PEG Architecture on Protein Adsorption and Conformation. Langmuir, 21(26), pp.12327–12332. Moulton, S.E. et al., 2003. Investigation of protein adsorption and electrochemical behavior at a gold electrode. Journal of Colloid and Interface Science, 261(2), pp.312–319. Moulton, S.E. et al., 2004. Studies of double layer capacitance and electron transfer at a gold electrode exposed to protein solutions. Electrochimica Acta, 49(24), pp.4223–4230. Negi, S. et al., 2010. In vitro comparison of sputtered iridium oxide and platinum-coated neural implantable microelectrode arrays. Biomed. Mater, 5, pp.15007–9. Newbold, C. et al., 2010. Changes in biphasic electrode impedance with protein adsorption and cell growth. Journal of neural engineering, 7(5), p.056011. Ngoka, L.C., 2008. Sample prep for proteomics of breast cancer: proteomics and gene ontology reveal dramatic differences in protein solubilization preferences of Mungo Elms 200688934 85 radioimmunoprecipitation assay and urea lysis buffers. Proteome science, 6, p.30. Oehrlein, G.S. & Kurogi, Y., 1998. Sidewall surface chemistry in directional etching processes. Materials Science and Engineering: R: Reports, 24(4), pp.153–183. Oehrlein, G.S. & Lee, Y.H., 1987. Reactive ion etching related Si surface residues and subsurface damage: Their relationship to fundamental etching mechanisms. Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films, 5(4), p.1585. Ohno, Y. et al., 2009. Electrolyte-Gated Graphene Field-Effect Transistors for Detecting pH and Protein Adsorption. American Chemical Society, 9(9), pp.2–6. Oldham, K.B., 1981. Edge effects in semiinfinite diffusion. Journal of Electroanalytical Chemistry, 122(C), pp.1–17. Oldham, K.B. & Zoski, C.G., 1988. Comparison of voltammetric steady states at hemispherical and disc microelectrodes. Journal of Electroanalytical Chemistry, 256(1), pp.11–19. Otto, K.J., Johnson, M.D. & Kipke, D.R., 2006. Voltage pulses change neural interface properties and improve unit recordings with chronically implanted microelectrodes. IEEE Transactions on Biomedical Engineering, 53(2), pp.333–340. Pajkossy, T., 1997. Capacitance dispersion on solid electrodes: anion adsorption studies on gold single crystal electrodes. Solid State Ionics, 94(1-4), pp.123–129. Pajkossy, T., 1991. Electrochemistry at fractal surfaces. Journal of Electroanalytical Chemistry, 300(1-2), pp.1–11. Pajkossy, T., 2005. Impedance spectroscopy at interfaces of metals and aqueous solutions - Surface roughness, CPE and related issues. Solid State Ionics, 176(25-28), pp.1997–2003. Park, J.H. et al., 2003. The correlation of the complex dielectric constant and blood glucose at low frequency. Biosensors and Bioelectronics, 19(4), pp.321–324. Park, J.M. et al., 1992. Effects of protein charge heterogeneity in protein-polyelectrolyte complexation. Macromolecules, 25(1), pp.290–295. Patel, J. et al., 2013. Electrochemical properties of nanostructured porous gold electrodes in Mungo Elms 200688934 86 biofouling solutions. Analytical Chemistry, 85(23), pp.11610–11618. Patel, P.R. et al., 2015. Insertion of linear 8.4 μ m diameter 16 channel carbon fiber electrode arrays for single unit recordings. Journal of Neural Engineering, 12(4), p.046009. a Powell, E.M. et al., 1997. Mechanisms of astrocyte-directed neurite guidance. Cell and Tissue Research, 290(2), pp.385–393. b Powell, E.M., Fawcett, J.W. & Geller, H.M., 1997. Proteoglycans provide neurite guidance at an astrocyte boundary. Molecular and cellular neurosciences, 10(1-2), pp.27–42. Prasad, A. et al., 2012. Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants. J. Neural Eng. J. Neural Eng, 9(9). Roitbak, T. & Syková, E., 1999. Diffusion barriers evoked in the rat cortex by reactive astrogliosis. GLIA, 28(1), pp.40–48. Rushworth, J. V. et al., 2014. A label-free electrical impedimetric biosensor for the specific detection of Alzheimer’s amyloid-beta oligomers. Biosensors and Bioelectronics, 56, pp.83–90. Sakurai-Yamashita, Y. et al., 2006. Expression of MCP-1 in the hippocampus of SHRSP with ischemia-related delayed neuronal death. Cellular and Molecular Neurobiology, 26(4-6), pp.823–831. Salakhutdinov, I. et al., 2008. Fibronectin adsorption to nanopatterned silicon surfaces. Journal of Nanomaterials, 2008(1). Schaefer, L. & Schaefer, R.M., 2010. Proteoglycans: From structural compounds to signaling molecules. Cell and Tissue Research, 339(1), pp.237–246. Seker, E. et al., 2012. Microfabrication-compatible nanoporous gold foams as biomaterials for drug delivery. Advanced Healthcare Materials, 1(2), pp.172–176. Seker, E. et al., 2010. The fabrication of low-impedance nanoporous gold multiple-electrode arrays for neural electrophysiology studies. Nanotechnology, 21(12), p.125504. Selhuber-Unkel, C. et al., 2008. Cooperativity in Adhesion Cluster Formation during Initial Cell Adhesion. Biophysical Journal, 95(11), pp.5424–5431. Mungo Elms 200688934 87 Smith-Thomas, L.C. et al., 1995. Increased axon regeneration in astrocytes grown in the presence of proteoglycan synthesis inhibitors. Journal of cell science, 108 ( Pt 3, pp.1307–1315. Sommakia, S. et al., 2014. Resistive and reactive changes to the impedance of intracortical microelectrodes can be mitigated with polyethylene glycol under acute in vitro and in vivo settings. Frontiers in neuroengineering, 7(August), p.33. Sommakia, S., Rickus, J.L. & Otto, K.J., 2009. Effects of adsorbed proteins, an antifouling agent and long-duration DC voltage pulses on the impedance of silicon-based neural microelectrodes. In Proceedings of the 31st Annual International Conference of the IEEE Engineering in Medicine and Biology Society: Engineering the Future of Biomedicine, EMBC 2009. pp. 7139–7142. Szarowski, D.H. et al., 2003. Brain responses to micro-machined silicon devices. Brain Research, 983(1-2), pp.23–35. Thomas, G.M., 2012. Does Controlled Roughening of a Surface Increase Its Capacitance ? Tian, Z.Q., Ren, B. & Wu, D.Y., 2002. Surface-enhanced Raman scattering: From noble to transition metals and from rough surfaces to ordered nanostructures. Journal of Physical Chemistry B, 106(37), pp.9463–9483. Triplett, J.W. & Pavalko, F.M., 2006. Disruption of alpha-actinin-integrin interactions at focal adhesions renders osteoblasts susceptible to apoptosis. American journal of physiology. Cell physiology, 291(5), pp.C909–21. Tura, A. et al., 2007. Impedance spectroscopy of solutions at physiological glucose concentrations. Biophysical Chemistry, 129(2-3), pp.235–241. Vahidpour, F. et al., 2016. All-diamond functional surface micro-electrode arrays for brain-slice neural analysis. Physica Status Solidi (a), 8, pp.1–8. Venkatraman, S. et al., 2009. PEDOT coated microelectrode arrays for chronic neural recording and stimulation. 2009 4th International IEEE/EMBS Conference on Neural Engineering, NER ’09, 4, pp.383–386. Volterra, A. & Meldolesi, J., 2009. Astroctes, from brain glue to communication elements: Mungo Elms 200688934 88 The revelution continues. Dialogues in Clinical Neuroscience, 11(3), pp.281–296. Walsh, E. et al., 1992. Neuronal survival and neurite extension supported by astrocytes co-cultured in transwells. Neuroscience Letters, 138(1), pp.103–106. Ward, M.D. & Hammer, D.A., 1993. A theoretical analysis for the effect of focal contact formation on cell-substrate attachment strength. Biophysical Journal, 64(3), pp.936–959. Westergren-Thorsson, G. et al., 1998. Differential expressions of mRNA for proteoglycans, collagens and transforming growth factor-?? in the human cervix during pregnancy and involution. Biochimica et Biophysica Acta - Molecular Basis of Disease, 1406(2), pp.203–213. Wight, T.N., 2002. Versican: A versatile extracellular matrix proteoglycan in cell biology. Current Opinion in Cell Biology, 14(5), pp.617–623. Williams, D.F., Askill, I.N. & Smith, R., 1985. Protein absorption and desorption phenomena on clean metal surfaces. Journal of biomedical materials research, 19, pp.313–320. Xue, P. et al., 2014. Protein covalently conjugated SU-8 surface for the enhancement of mesenchymal stem cell adhesion and proliferation. Langmuir, 30(11), pp.3110–3117. Ying, P. et al., 2004. Adsorption of human serum albumin onto gold: A combined electrochemical and ellipsometric study. Journal of Colloid and Interface Science, 279(1), pp.95–99. Zaidi, K.F. et al., 2010. In vitro Models for Measuring Charge Storage Capacity. IFMBE Proceedings, 32, pp.97–100. Zhang, H. et al., 2012. Layered nanocomposites from gold nanoparticles for neural prosthetic devices. Nano Letters, 12(7), pp.3391–3398. Zhang, J. et al., 2012. Label-Free Electrochemical Detection of Tetracycline by an Aptamer Nano-Biosensor. Analytical Letters, 45(9), pp.986–992. Zhao, H. et al., 1999. Surface-roughness effect on capacitance and leakage current of an insulating film. Zheng, Y. et al., 2013. Electrical measurement of red blood cell deformability on a Mungo Elms 200688934 89 microfluidic device. Lab on a chip, 13(16), pp.3275–83."]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["masters"]},{"key":"dc:type.qualificationname","label":"Dc Type Qualificationname","values":["M.Sc by research"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.dcterms_iso639_2","label":"Language (ISO 639-2)","values":["eng"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://etheses.whiterose.ac.uk/id/eprint/17952/1/Mungo%20Elms%20Masters%20By%20Research%20Thesis%20FINAL.pdf"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Devices called neural electrodes are generally used to record and / or stimulate neural activity whilst extracellularly interfacing with neurons. The ultimate goal of the field is to develop neural electrodes that can continuously record and / or stimulate neural activity for long periods of time (10+ years). Electrodes require low electrode impedance to enable good signal to noise ratio and a high capacity for chemically stable injection of charge (CSC) for stimulating neural activity. Electrode function typically deteriorates after a few months in-vivo due two factors: biofouling and local cellular response (glial scarring). Biofouling is theorised to be an accumulation of multiple proteins at the electrode surface. However this has only been backed up experimentally by single protein models in literature. Electrodes need to be able to combat the effects of biofouling and glial scarring. This study uses nanometre scale roughened gold (Au) electrodes. Electrode roughening has previously been shown to lower impedance, increase CSC and reduce glial response and the effects of biofouling. We compared multiple degrees of roughness with the aim of finding the optimal degree for improving impedance, CSC, biofouling and cellular response. We found that surface roughening increased impedance and only increased CSC for two only degrees of roughness. To find the optimal degree of roughness across conditions, we suspect a larger range of roughness may be necessary to lower impedance with our fabrication technique. This study is the first to use a multiprotein biofouling model. Contrary to literature, we found that incubation with protein decreased impedance, likely due to protein-protein interactions not accounted for in single protein models. Biocompatibility was improved for two degrees of Au roughness. Roughening of SU8, a polymer used to surround the electrodes, decreased biocompatibility. We also used artificial cerebral spinal fluid (aCSF) as an electrolyte, which is more chemically similar to in-vivo than commonly used phosphate buffered saline solution (PBS). The use aCSF as a medium was significant as the measures in aCSF were different from that in PBS."]},{"key":"dc:format","label":"Dc Format","values":["text"]},{"key":"dcterms:accessrights","label":"Dcterms Accessrights","values":["public"]},{"key":"dc:title","label":"Title","values":["Comparison of Multiple Degrees of Electrode Surface Roughness: Impedance, Charge Storage Capacity, Biofouling and Biocompatibility"]}]}],"canonical_facts":{"dc:contributor.advisor":["Chakrabarty, Samit","Steenson, Paul"],"dc:creator":["Elms, Mungo Arthur"],"dc:date":["2017-02"],"dc:date.issued":["2017-02"],"dc:description.abstract":["Devices called neural electrodes are generally used to record and / or stimulate neural activity whilst extracellularly interfacing with neurons. The ultimate goal of the field is to develop neural electrodes that can continuously record and / or stimulate neural activity for long periods of time (10+ years). Electrodes require low electrode impedance to enable good signal to noise ratio and a high capacity for chemically stable injection of charge (CSC) for stimulating neural activity. Electrode function typically deteriorates after a few months in-vivo due two factors: biofouling and local cellular response (glial scarring). Biofouling is theorised to be an accumulation of multiple proteins at the electrode surface. However this has only been backed up experimentally by single protein models in literature. Electrodes need to be able to combat the effects of biofouling and glial scarring. This study uses nanometre scale roughened gold (Au) electrodes. Electrode roughening has previously been shown to lower impedance, increase CSC and reduce glial response and the effects of biofouling. We compared multiple degrees of roughness with the aim of finding the optimal degree for improving impedance, CSC, biofouling and cellular response. We found that surface roughening increased impedance and only increased CSC for two only degrees of roughness. To find the optimal degree of roughness across conditions, we suspect a larger range of roughness may be necessary to lower impedance with our fabrication technique. This study is the first to use a multiprotein biofouling model. Contrary to literature, we found that incubation with protein decreased impedance, likely due to protein-protein interactions not accounted for in single protein models. Biocompatibility was improved for two degrees of Au roughness. Roughening of SU8, a polymer used to surround the electrodes, decreased biocompatibility. We also used artificial cerebral spinal fluid (aCSF) as an electrolyte, which is more chemically similar to in-vivo than commonly used phosphate buffered saline solution (PBS). The use aCSF as a medium was significant as the measures in aCSF were different from that in PBS."],"dc:format":["text"],"dc:identifier.uri":["https://etheses.whiterose.ac.uk/id/eprint/17952/1/Mungo%20Elms%20Masters%20By%20Research%20Thesis%20FINAL.pdf"],"dc:language.dcterms_iso639_2":["eng"],"dc:publisher.commercial":["University of Leeds"],"dc:publisher.department":["Institute of Membrane and Systems Biology (Leeds)"],"dc:publisher.institution":["University of Leeds"],"dc:relation.isreferencedby":["https://etheses.whiterose.ac.uk/id/eprint/17952/"],"dc:relation.references":["Alexander, C.K. & Sadiku, M.N.O., 2001. Fundamentals of Electric Circuits 3rd ed., New York: McGraw-Hill Higher Education. Altuna, A. et al., 2013. SU-8 based microprobes for simultaneous neural depth recording and drug delivery in the brain. Lab on a chip, 13(7), pp.1422–30. Altuna, A. et al., 2012. SU-8 based microprobes with integrated planar electrodes for enhanced neural depth recording. Biosensors and Bioelectronics. Altuna, A., Berganzo, J. & Fernández, L.J., 2015. Polymer SU-8-Based Microprobes for Neural Recording and Drug Delivery. Frontiers in Materials, 2(June), pp.1–5. Aurian-Blajeni, B., Kimball, A.G. & Robblee, L.S., 1987. Correlation Between Charge Storage Capacity and Morphology. Journal of The Electrochemical Society, pp.2637–2638. Babcock, A. a. et al., 2003. Chemokine Expression by Glial Cells Directs Leukocytes to Sites of Axonal Injury in the CNS. J. Neurosci., 23(21), pp.7922–7930. Balaban, N.Q. et al., 2001. Force and focal adhesion assembly: a close relationship studied using elastic micropatterned substrates. Nature cell biology, 3(5), pp.466–472. Barrese, J.C. et al., 2013. Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates. Journal of Neural Engineering, 10(6), p.66014. Bershadsky, A.D. et al., 2006. Assembly and mechanosensory function of focal adhesions: Experiments and models. European Journal of Cell Biology, 85(3-4), pp.165–173. Bhandari, R., Negi, S. & Solzbacher, F., 2011. A novel surface modification method to achieve low impedance neural microelectrode arrays. 2011 16th International Solid-State Sensors, Actuators and Microsystems Conference, TRANSDUCERS’11, pp.2196–2199. Bidola, E.D., Bulhoes, L.O.S. & Rocha-Filho, R.C., 1994. Pt / HCl04 Interface CPE: Influence of Surface Roughness and Electrolyte Concentration. Electrochimica Acta, 39(5), pp.763–769. Mungo Elms 200688934 79 Biggs, M.J.P., Richards, R.G. & Dalby, M.J., 2010. Nanotopographical modification: a regulator of cellular function through focal adhesions. Nanomedicine: Nanotechnology, Biology, and Medicine, 6(5), pp.619–633. Biran, R., Martin, D.C. & Tresco, P.A., 2005. Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays. Experimental Neurology, 195(1), pp.115–126. Bjornsson, C.S. et al., 2006. Effects of insertion conditions on tissue strain and vascular damage during neuroprosthetic device insertion. Journal of neural engineering, 3(3), pp.196–207. Boehler, C. et al., 2015. Iridium Oxide ( IrOx ) serves as adhesion promoter for conducting polymers on neural microelectrodes. IEEE, 49, pp.22–24. Bosman, F.T. & Stamenkovic, I., 2003. Functional structure and composition of the extracellular matrix. Journal of Pathology, 200(4), pp.423–428. Brownson, D.A.C. & Banks, C.E., 2014. The Handbook of Graphene Electrochemistry 1st ed., London: Springer. Buffo, A., Rolando, C. & Ceruti, S., 2010. Astrocytes in the damaged brain: Molecular and cellular insights into their reactive response and healing potential. Biochemical Pharmacology, 79(2), pp.77–89. Burridge, K., 1988. Focal Adhesions: Transmembrane Junctions Between The Extracellular Matrix And The Cytoskeleton. Annual Review of Cell and Developmental Biology, 4(1), pp.487–525. Castagnola, E. et al., 2015. Pedot-CNT-coated low-impedance, ultra-flexible, and brain-conformable micro-ECoG arrays. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 23(3), pp.342–350. Cellot, G. et al., 2016. PEDOT:PSS interfaces support the development of neuronal synaptic networks with reduced neuroglia response in vitro. Frontiers in neuroscience, 9(521). Chapman, C.A.R. et al., 2016. Mechanisms of Reduced Astrocyte Surface Coverage in Cortical Neuron-Glia Co-cultures on Nanoporous Gold Surfaces. Cellular and Molecular Mungo Elms 200688934 80 Bioengineering, 9(3), pp.433–442. Chapman, C.A.R. et al., 2015. Nanoporous gold as a neural interface coating: Effects of topography, surface chemistry, and feature size. ACS Applied Materials and Interfaces, 7(13), pp.7093–7100. Chen, H. et al., 2011. Multilayered polypyrrole-coated carbon nanotubes to improve functional stability and electrical properties of neural electrodes. Journal of Physical Chemistry C, 115(13), pp.5492–5499. Chen, Z. et al., 2005. Impedance immunosensor based on receptor protein adsorbed directly on porous gold film. Analytica Chimica Acta, 553(1-2), pp.190–195. Chesler, M., 2008. Regulation and Modulation of pH in the Brain. Brain, pp.1183–1221. Cho, S.H. et al., 2008. Biocompatible SU-8-based microprobes for recording neural spike signals from regenerated peripheral nerve fibers. IEEE Sensors Journal, 8(11), pp.1830–1836. Chung, T., Wang, J.Q., Wang, J., et al., 2015. Electrode modifications to lower electrode impedance and improve neural signal recording sensitivity. Journal of Neural Engineering, 12(5), p.056018. Chung, T., Wang, J.Q., Cao, B., et al., 2015. Electrode modifications to lower electrode impedance and improve neural signal recording sensitivity. Journal of Neural Engineering, 12(5), p.056018. Cogan, S.F., 2008. Neural stimulation and recording electrodes. Annual review of biomedical engineering, 10, pp.275–309. Cohen, M. et al., 2004. Spatial and temporal sequence of events in cell adhesion: From molecular recognition to focal adhesion assembly. ChemBioChem, 5(10), pp.1393–1399. Compton, R.G. & Banks, C.E., 2011. Understanding Voltammetry 1st ed., Singapore: World Scientific. Cragg, P., Patterson, L. & Purves, M., 1977. The pH of brain extracellular fluid in the cat. The Journal of Physiology, 272(1), p.137. Mungo Elms 200688934 81 Cui, X. et al., 2003. In vivo studies of polypyrrole/peptide coated neural probes. Biomaterials, 24(5), pp.777–787. Curtis, A.S.G. et al., 2001. Substratum nanotopography and the adhesion of biological cells. Are symmetry or regularity of nanotopography important? Biophysical Chemistry, 94(3), pp.275–283. Daikhin, L., Kornyshev, a. & Urbakh, M., 1996. Double-layer capacitance on a rough metal surface. Physical Review E, 53(6), pp.6192–6199. Daikhin, L.I., Kornyshev, a. a. & Urbakh, M., 1997. Double layer capacitance on a rough metal surface: Surface roughness measured by “Debye ruler.” Electrochimica Acta, 42(19), pp.2853–2860. Debye, P., 1936. Dielectric properties of pure liquids. Chemical Reviews, 19, pp.171–182. Desai, S.A. et al., 2010. Improving impedance of implantable microwire multi-electrode arrays by ultrasonic electroplating of durable platinum black. Frontiers in neuroengineering, 3, p.5. Devore, J.L., Farnum, N.R. & Doi, J.A., 1999. Applied Statistics for Engineers and Scientists 3rd ed. M. Taylor, L. Wheel, & A. Coppola, eds., Stamford: Cengage Learning. Di, L. et al., 2011. Protein adsorption and peroxidation of rat retinas under stimulation of a neural probe coated with polyaniline. Acta Biomaterialia, 7(10), pp.3738–3745. Diener, A. et al., 2005. Control of focal adhesion dynamics by material surface characteristics. Biomaterials, 26(4), pp.383–392. Doña Rodríguez, J.M., Herrera Melián, J. a. & Pérez Peña, J., 2000. Determination of the Real Surface Area of Pt Electrodes by Hydrogen Adsorption Using Cyclic Voltammetry. Journal of Chemical Education, 77(9), p.1195. Fattahi, P. et al., 2014. A review of organic and inorganic biomaterials for neural interfaces. Advanced Materials, 26(12), pp.1846–1885. Fidler, P.S. et al., 1999. Comparing astrocytic cell lines that are inhibitory or permissive for axon growth: the major axon-inhibitory proteoglycan is NG2. The Journal of neuroscience : the official journal of the Society for Neuroscience, 19(20), pp.8778– Mungo Elms 200688934 82 8788. Frackowiak, E. & Beguin, F., 2001. Carbon materials for the electrochemical storage of energy in capacitors. PERGAMON Carbon, 39, pp.937–950. Frantz, C., Stewart, K.M. & Weaver, V.M., 2010. The extracellular matrix at a glance. Journal of cell science, 123, pp.4195–4200. Frumkin, A.N., 1960. The Double Layer in Electrochemistry. J. Electrochem. Soc, 107(5), pp.461–472. García, A.J., 2005. Get a grip: Integrins in cell-biomaterial interactions. Biomaterials, 26(36), pp.7525–7529. García-Gabaldón, M. et al., 2011. Electrochemical study of a simulated spent pickling solution. International Journal of Electrochemical Science, 6(2), pp.506–519. Ghazni, N.F., Cahill, C.M. & Stroman, P.W., 2010. Tactile sensory and pain networks in the human spinal cord and brain stem mapped by means of functional MR imaging. American Journal of Neuroradiology, 31(4), pp.661–667. Hai, A. et al., 2009. Spine-shaped gold protrusions improve the adherence and electrical coupling of neurons with the surface of micro-electronic devices. Journal of the Royal Society, Interface / the Royal Society, 6(41), pp.1153–65. Hamelin, A. & Martins, A.M., 1996. Cyclic voltammetry at gold single-crystal surfaces. Part 2. Behaviour of high-index faces. Joumal of Electroanalytical Chemistry, 407, pp.13–21. Heim, M. et al., 2012. Combined macro-/mesoporous microelectrode arrays for low-noise extracellular recording of neural networks. Journal of Neurophysiology, 108(6), pp.1793–1803. Hu, Z. et al., 2006. Nanopowder molding method for creating implantable high-aspect-ratio electrodes on thin flexible substrates. Biomaterials, 27(9), pp.2009–2017. Johnson, M.D., Otto, K.J. & Kipke, D.R., 2005. Repeated voltage biasing improves unit recordings by reducing resistive tissue impedances. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 13(2), pp.160–165. Kerner, Z., 1998. Impedance of rough capacitive electrodes : the role of surface disorder. , Mungo Elms 200688934 83 448, pp.139–142. Kerner, Z. & Pajkossy, T., 2000. On the origin of capacitance dispersion of rough electrodes. Electrochimica Acta, 46(2-3), pp.207–211. Kim, S.A., Kim, E.T. & Kim, S.J., 2009. Highly Efficient Cold Sputtered Iridium Oxide Films for Polyimide based Neural Stimulation Electrodes. Journal of Biomedical Engineering Research, 30(3), pp.199–204. Kocherbitov, V., 2015. Surface and Physical Chemistry. , p.Debye Screening Length. Available at: http://www.surfchem.info/calculate/Debye/. Koklu, A., Sabuncu, A.C. & Beskok, A., 2016. Rough Gold Electrodes for Decreasing Impedance at the Electrolyte/Electrode Interface. Electrochimica Acta. Kolb, D.M. & Schneider, J., 1986. Surface reconstruction in electrochemistry: Au(100-(5 × 20), Au(111)-(1 × 23) and Au(110)-(1 × 2). Electrochimica Acta, 31(8), pp.929–936. Kozai, T.D. et al., 2010. Reduction of neurovascular damage resulting from microelectrode insertion into the cerebral cortex using in vivo two-photon mapping. J Neural Eng, 7(4), p.46011. Kozai, T.D.Y. et al., 2016. Chronic In Vivo Evaluation of PEDOT / CNT for Stable Neural Recordings. IEEE Trans Biomed Eng, 63(1), pp.111–119. Kozai, T.D.Y. et al., 2014. Chronic tissue response to carboxymethyl cellulose based dissolvable insertion needle for ultra-small neural probes. Biomaterials, 35(34), pp.9255–9268. Kozai, T.D.Y., Vazquez, A.L., et al., 2012. In vivo two-photon microscopy reveals immediate microglial reaction to implantation of microelectrode through extension of processes. Journal of Neural Engineering, 9(6), p.066001. Kozai, T.D.Y., Langhals, N.B., et al., 2012. Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. Nature materials, 11(12), pp.1065–73. Kurtulus, O. & Seker, E., 2012. Nanotopography Effects on Astrocyte Attachment to Nanoporous Gold Surfaces. In IEEE EMBS. pp. 6568–6571. Mungo Elms 200688934 84 Lim, J.Y. et al., 2005. Human foetal osteoblastic cell response to polymer-demixed nanotopographic interfaces. Journal of the Royal Society, Interface / the Royal Society, 2(2), pp.97–108. Malaga, K.A. et al., 2016. Data-driven model comparing the effects of glial scarring and tip metallization loss on chronic neural recordings. Neural Engineering, 13(1), p.16010. Martic, S., Rains, M.K. & Kraatz, H.B., 2013. Probing copper/tau protein interactions electrochemically. Analytical Biochemistry, 442(2), pp.130–137. McConnell, G.C. et al., 2009. Implanted neural electrodes cause chronic, local inflammation that is correlated with local neurodegeneration. Journal of neural engineering, 6(5), p.056003. Meijs, S. et al., 2016. Influence of fibrous encapsulation on electro-chemical properties of TiN electrodes. Medical Engineering and Physics, 38(5), pp.468–476. Michel, E., Hernandez, D. & Lee, S.Y., 2016. Electrical conductivity and permittivity maps of brain tissues derived from water content based on T 1 -weighted acquisition. Magnetic Resonance in Medicine. Michel, R. et al., 2008. The Influence of PEG Architecture on Protein Adsorption and Conformation. Langmuir, 21(26), pp.12327–12332. Moulton, S.E. et al., 2003. Investigation of protein adsorption and electrochemical behavior at a gold electrode. Journal of Colloid and Interface Science, 261(2), pp.312–319. Moulton, S.E. et al., 2004. Studies of double layer capacitance and electron transfer at a gold electrode exposed to protein solutions. Electrochimica Acta, 49(24), pp.4223–4230. Negi, S. et al., 2010. In vitro comparison of sputtered iridium oxide and platinum-coated neural implantable microelectrode arrays. Biomed. Mater, 5, pp.15007–9. Newbold, C. et al., 2010. Changes in biphasic electrode impedance with protein adsorption and cell growth. Journal of neural engineering, 7(5), p.056011. Ngoka, L.C., 2008. Sample prep for proteomics of breast cancer: proteomics and gene ontology reveal dramatic differences in protein solubilization preferences of Mungo Elms 200688934 85 radioimmunoprecipitation assay and urea lysis buffers. Proteome science, 6, p.30. Oehrlein, G.S. & Kurogi, Y., 1998. Sidewall surface chemistry in directional etching processes. Materials Science and Engineering: R: Reports, 24(4), pp.153–183. Oehrlein, G.S. & Lee, Y.H., 1987. Reactive ion etching related Si surface residues and subsurface damage: Their relationship to fundamental etching mechanisms. Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films, 5(4), p.1585. Ohno, Y. et al., 2009. Electrolyte-Gated Graphene Field-Effect Transistors for Detecting pH and Protein Adsorption. American Chemical Society, 9(9), pp.2–6. Oldham, K.B., 1981. Edge effects in semiinfinite diffusion. Journal of Electroanalytical Chemistry, 122(C), pp.1–17. Oldham, K.B. & Zoski, C.G., 1988. Comparison of voltammetric steady states at hemispherical and disc microelectrodes. Journal of Electroanalytical Chemistry, 256(1), pp.11–19. Otto, K.J., Johnson, M.D. & Kipke, D.R., 2006. Voltage pulses change neural interface properties and improve unit recordings with chronically implanted microelectrodes. IEEE Transactions on Biomedical Engineering, 53(2), pp.333–340. Pajkossy, T., 1997. Capacitance dispersion on solid electrodes: anion adsorption studies on gold single crystal electrodes. Solid State Ionics, 94(1-4), pp.123–129. Pajkossy, T., 1991. Electrochemistry at fractal surfaces. Journal of Electroanalytical Chemistry, 300(1-2), pp.1–11. Pajkossy, T., 2005. Impedance spectroscopy at interfaces of metals and aqueous solutions - Surface roughness, CPE and related issues. Solid State Ionics, 176(25-28), pp.1997–2003. Park, J.H. et al., 2003. The correlation of the complex dielectric constant and blood glucose at low frequency. Biosensors and Bioelectronics, 19(4), pp.321–324. Park, J.M. et al., 1992. Effects of protein charge heterogeneity in protein-polyelectrolyte complexation. Macromolecules, 25(1), pp.290–295. Patel, J. et al., 2013. Electrochemical properties of nanostructured porous gold electrodes in Mungo Elms 200688934 86 biofouling solutions. Analytical Chemistry, 85(23), pp.11610–11618. Patel, P.R. et al., 2015. Insertion of linear 8.4 μ m diameter 16 channel carbon fiber electrode arrays for single unit recordings. Journal of Neural Engineering, 12(4), p.046009. a Powell, E.M. et al., 1997. Mechanisms of astrocyte-directed neurite guidance. Cell and Tissue Research, 290(2), pp.385–393. b Powell, E.M., Fawcett, J.W. & Geller, H.M., 1997. Proteoglycans provide neurite guidance at an astrocyte boundary. Molecular and cellular neurosciences, 10(1-2), pp.27–42. Prasad, A. et al., 2012. Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants. J. Neural Eng. J. Neural Eng, 9(9). Roitbak, T. & Syková, E., 1999. Diffusion barriers evoked in the rat cortex by reactive astrogliosis. GLIA, 28(1), pp.40–48. Rushworth, J. V. et al., 2014. A label-free electrical impedimetric biosensor for the specific detection of Alzheimer’s amyloid-beta oligomers. Biosensors and Bioelectronics, 56, pp.83–90. Sakurai-Yamashita, Y. et al., 2006. Expression of MCP-1 in the hippocampus of SHRSP with ischemia-related delayed neuronal death. Cellular and Molecular Neurobiology, 26(4-6), pp.823–831. Salakhutdinov, I. et al., 2008. Fibronectin adsorption to nanopatterned silicon surfaces. Journal of Nanomaterials, 2008(1). Schaefer, L. & Schaefer, R.M., 2010. Proteoglycans: From structural compounds to signaling molecules. Cell and Tissue Research, 339(1), pp.237–246. Seker, E. et al., 2012. Microfabrication-compatible nanoporous gold foams as biomaterials for drug delivery. Advanced Healthcare Materials, 1(2), pp.172–176. Seker, E. et al., 2010. The fabrication of low-impedance nanoporous gold multiple-electrode arrays for neural electrophysiology studies. Nanotechnology, 21(12), p.125504. Selhuber-Unkel, C. et al., 2008. Cooperativity in Adhesion Cluster Formation during Initial Cell Adhesion. Biophysical Journal, 95(11), pp.5424–5431. Mungo Elms 200688934 87 Smith-Thomas, L.C. et al., 1995. Increased axon regeneration in astrocytes grown in the presence of proteoglycan synthesis inhibitors. Journal of cell science, 108 ( Pt 3, pp.1307–1315. Sommakia, S. et al., 2014. Resistive and reactive changes to the impedance of intracortical microelectrodes can be mitigated with polyethylene glycol under acute in vitro and in vivo settings. Frontiers in neuroengineering, 7(August), p.33. Sommakia, S., Rickus, J.L. & Otto, K.J., 2009. Effects of adsorbed proteins, an antifouling agent and long-duration DC voltage pulses on the impedance of silicon-based neural microelectrodes. In Proceedings of the 31st Annual International Conference of the IEEE Engineering in Medicine and Biology Society: Engineering the Future of Biomedicine, EMBC 2009. pp. 7139–7142. Szarowski, D.H. et al., 2003. Brain responses to micro-machined silicon devices. Brain Research, 983(1-2), pp.23–35. Thomas, G.M., 2012. Does Controlled Roughening of a Surface Increase Its Capacitance ? Tian, Z.Q., Ren, B. & Wu, D.Y., 2002. Surface-enhanced Raman scattering: From noble to transition metals and from rough surfaces to ordered nanostructures. Journal of Physical Chemistry B, 106(37), pp.9463–9483. Triplett, J.W. & Pavalko, F.M., 2006. Disruption of alpha-actinin-integrin interactions at focal adhesions renders osteoblasts susceptible to apoptosis. American journal of physiology. Cell physiology, 291(5), pp.C909–21. Tura, A. et al., 2007. Impedance spectroscopy of solutions at physiological glucose concentrations. Biophysical Chemistry, 129(2-3), pp.235–241. Vahidpour, F. et al., 2016. All-diamond functional surface micro-electrode arrays for brain-slice neural analysis. Physica Status Solidi (a), 8, pp.1–8. Venkatraman, S. et al., 2009. PEDOT coated microelectrode arrays for chronic neural recording and stimulation. 2009 4th International IEEE/EMBS Conference on Neural Engineering, NER ’09, 4, pp.383–386. Volterra, A. & Meldolesi, J., 2009. Astroctes, from brain glue to communication elements: Mungo Elms 200688934 88 The revelution continues. Dialogues in Clinical Neuroscience, 11(3), pp.281–296. Walsh, E. et al., 1992. Neuronal survival and neurite extension supported by astrocytes co-cultured in transwells. Neuroscience Letters, 138(1), pp.103–106. Ward, M.D. & Hammer, D.A., 1993. A theoretical analysis for the effect of focal contact formation on cell-substrate attachment strength. Biophysical Journal, 64(3), pp.936–959. Westergren-Thorsson, G. et al., 1998. Differential expressions of mRNA for proteoglycans, collagens and transforming growth factor-?? in the human cervix during pregnancy and involution. Biochimica et Biophysica Acta - Molecular Basis of Disease, 1406(2), pp.203–213. Wight, T.N., 2002. Versican: A versatile extracellular matrix proteoglycan in cell biology. Current Opinion in Cell Biology, 14(5), pp.617–623. Williams, D.F., Askill, I.N. & Smith, R., 1985. Protein absorption and desorption phenomena on clean metal surfaces. Journal of biomedical materials research, 19, pp.313–320. Xue, P. et al., 2014. Protein covalently conjugated SU-8 surface for the enhancement of mesenchymal stem cell adhesion and proliferation. Langmuir, 30(11), pp.3110–3117. Ying, P. et al., 2004. Adsorption of human serum albumin onto gold: A combined electrochemical and ellipsometric study. Journal of Colloid and Interface Science, 279(1), pp.95–99. Zaidi, K.F. et al., 2010. In vitro Models for Measuring Charge Storage Capacity. IFMBE Proceedings, 32, pp.97–100. Zhang, H. et al., 2012. Layered nanocomposites from gold nanoparticles for neural prosthetic devices. Nano Letters, 12(7), pp.3391–3398. Zhang, J. et al., 2012. Label-Free Electrochemical Detection of Tetracycline by an Aptamer Nano-Biosensor. Analytical Letters, 45(9), pp.986–992. Zhao, H. et al., 1999. Surface-roughness effect on capacitance and leakage current of an insulating film. Zheng, Y. et al., 2013. Electrical measurement of red blood cell deformability on a Mungo Elms 200688934 89 microfluidic device. Lab on a chip, 13(16), pp.3275–83."],"dcterms:accessrights":["public"],"dc:title":["Comparison of Multiple Degrees of Electrode Surface Roughness: Impedance, Charge Storage Capacity, Biofouling and Biocompatibility"],"dc:type":["Thesis"],"dc:type.qualificationlevel":["masters"],"dc:type.qualificationname":["M.Sc by research"]},"updated_at":"2026-11-18T22:22:33Z"}