{"id":{"repo_id":"whiterose","oai_identifier":"oai:etheses.whiterose.ac.uk:19913"},"canonical_url":"https://search.ndltd.org/etd/whiterose/oai:etheses.whiterose.ac.uk:19913","repository":{"repo_id":"whiterose","name":"White Rose University Consortium","base_url":"https://etheses.whiterose.ac.uk/cgi/oai2"},"display":{"title":"Modelling of Dynamic Edge Loading in Total Hip Replacements with ceramic on polyethylene Bearings","abstract":"The performance of total hip replacement (THR) devices can be affected by various factors such as quality of the tissues surrounding the joint, mismatch of the component centres or the cup positioning during hip replacement surgery. Experimental studies have shown that these factors can cause the separation of the two components during the walking cycle (dynamic separation) and the contact of the femoral head with the rim of the acetabular liner (edge loading), which can lead to increased wear and shortened implant lifespan. There is a need for flexible pre-clinical testing tools which allow THR devices to be assessed under these adverse conditions. In this work, a novel dynamic finite element model was developed that is able to generate dynamic separation as it occurs during the gait cycle. In addition, the ability to interrogate contact mechanics and material strain under separation conditions provides a unique means of assessing the severity of edge loading. This study demonstrates these model capabilities for a range of simulated surgical translational mismatch values, cup inclination angles and swing phase loads for ceramic-on-polyethylene implants. The computational model was developed to replicate one station of the Leeds II hip simulator that mimic in vitro adverse conditions. Firstly, a computational sensitivity model was developed under standard conditions for a stable computational contact. The mechanism of separation was also added. The finite element model was able to predict medial-lateral separation as it occurred dynamically in the gait cycle, including cases where the femoral head was in contact with the rim of the cup. The increase in medial-lateral separation with increased translational mismatch, cup inclination angle and decreased swing phase load were in broad agreement with existing experimental data. The factors that increased the separation level, also increased the permeant deformation on the cup. However, steep cup inclination angle resulted in a higher number of conditions with permanent deformation than the standard cup inclination angle. Moreover, despite the low axial load during swing phase, under some separation conditions, reduced contact area created stress value higher than those at the peak axial load. The developed computational tool can be used to understand the effect of various factors on the separation and contact mechanics simultaneously. As separation is a multi-factorial phenomenon, this model can assist to focus on the selected factors that affect the separation experimentally. Moreover, the effect of components specifications such as materials, geometry, and the cup thickness can be investigated with this model.","abstract_html":"The performance of total hip replacement (THR) devices can be affected by various factors such as quality of the tissues surrounding the joint, mismatch of the component centres or the cup positioning during hip replacement surgery. Experimental studies have shown that these factors can cause the separation of the two components during the walking cycle (dynamic separation) and the contact of the femoral head with the rim of the acetabular liner (edge loading), which can lead to increased wear and shortened implant lifespan. There is a need for flexible pre-clinical testing tools which allow THR devices to be assessed under these adverse conditions. In this work, a novel dynamic finite element model was developed that is able to generate dynamic separation as it occurs during the gait cycle. In addition, the ability to interrogate contact mechanics and material strain under separation conditions provides a unique means of assessing the severity of edge loading. This study demonstrates these model capabilities for a range of simulated surgical translational mismatch values, cup inclination angles and swing phase loads for ceramic-on-polyethylene implants. The computational model was developed to replicate one station of the Leeds II hip simulator that mimic in vitro adverse conditions. Firstly, a computational sensitivity model was developed under standard conditions for a stable computational contact. The mechanism of separation was also added. The finite element model was able to predict medial-lateral separation as it occurred dynamically in the gait cycle, including cases where the femoral head was in contact with the rim of the cup. The increase in medial-lateral separation with increased translational mismatch, cup inclination angle and decreased swing phase load were in broad agreement with existing experimental data. The factors that increased the separation level, also increased the permeant deformation on the cup. However, steep cup inclination angle resulted in a higher number of conditions with permanent deformation than the standard cup inclination angle. Moreover, despite the low axial load during swing phase, under some separation conditions, reduced contact area created stress value higher than those at the peak axial load. The developed computational tool can be used to understand the effect of various factors on the separation and contact mechanics simultaneously. As separation is a multi-factorial phenomenon, this model can assist to focus on the selected factors that affect the separation experimentally. Moreover, the effect of components specifications such as materials, geometry, and the cup thickness can be investigated with this model.","abstract_has_math":false,"creators":["Jahani, Faezeh"],"institution":"University of Leeds","degree_name":"Ph.D","degree_level":"doctoral","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Jones, Alison","Barton , David","Fisher, John","Wilcox, Ruth"],"committee_chairs":[],"committee_members":[],"year":2017,"date_issued":"2017-09","date_published":"2017-09","updated_at":"2026-11-18T22:25:53Z","subjects":[],"languages":["eng"],"rights":[],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier","label":"Identifier","values":["uk.bl.ethos.739781"],"render_values":[{"text":"uk.bl.ethos.739781","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":null,"outbound_label":null,"outbound_source":null,"record_url":null,"record_label":null,"record_kind":null,"record_source":null,"download_url":"https://etheses.whiterose.ac.uk/id/eprint/19913/7/Jahani_F_Mechanical_PhD_2017.pdf","download_label":"Download PDF","download_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://etheses.whiterose.ac.uk/cgi/oai2?verb=GetRecord&metadataPrefix=uketd_dc&identifier=oai%3Aetheses.whiterose.ac.uk%3A19913","prefix":"uketd_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Jones, Alison","Barton , David","Fisher, John","Wilcox, Ruth"]},{"key":"dc:contributor.sponsor","label":"Sponsor","values":["EPSRC","DePuy Synthes"]},{"key":"dc:creator","label":"Author","values":["Jahani, Faezeh"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2017-09"]},{"key":"dc:date.issued","label":"Date","values":["2017-09"]},{"key":"dc:publisher.commercial","label":"Dc Publisher Commercial","values":["University of Leeds"]},{"key":"dc:publisher.department","label":"Dc Publisher Department","values":["Institute of Medical and Biological Engineering (iMBE)(Leeds)"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University of Leeds"]},{"key":"dc:relation.isreferencedby","label":"Dc Relation Isreferencedby","values":["https://etheses.whiterose.ac.uk/id/eprint/19913/"]},{"key":"dc:relation.references","label":"Dc Relation References","values":["Abaqus 6.12 Documentation (2012). Dassault Systèmes. Available at : http://abaqus.software.polimi.it/v6.12/index.html Abaqus 6.14 Documentation (2014). Dassault Systèmes. Available at : http://abaqus.software.polimi.it/v6.14/index.html Affatato S., Leardini W., Zavalloni M. (2006) Hip Joint Simulators: State of the Art. In: Benazzo F., Falez F., Dietrich M. (eds) Bioceramics and Alternative Bearings in Joint Arthroplasty. Ceramics in Orthopaedics. Washington D.C.: Steinkopff. Affatato, S., Spinelli, M., Zavalloni, M., Mazzega-Fabbro, C. and Viceconti, M. (2008) ‘Tribology and total hip joint replacement: Current concepts in mechanical simulation’, Medical Engineering and Physics, 30(10), pp. 1305–1317. doi: 10.1016/j.medengphy.2008.07.006. Affatato, S., Torrecillas, R., Taddei, P., Rocchi, M., Fagnano, C., Ciapetti, G. and Toni, A. (2006) ‘Advanced nanocomposite materials for orthopaedic applications. I. A long-termin vitro wear study of zirconia-toughened alumina’, Journal of Biomedical Materials Research Part B: Applied Biomaterials, 78B(1), pp. 76–82. doi: 10.1002/jbm.b.30462. Al-Hajjar, M., Fisher, J., Tipper, J. L., Williams, S. and Jennings, L. M. (2013) ‘Wear of 36-mm BIOLOX (R) delta ceramic-on-ceramic bearing in total hip replacements under edge loading conditions’, Proceedings Of the Institution Of Mechanical Engineers Part H-Journal Of Engineering In Medicine, 227(H5), pp. 535–542. doi: Doi 10.1177/0954411912474613. Al-Hajjar, M., Leslie, I. J., Tipper, J., Williams, S., Fisher, J. and Jennings, L. M. (2010) 230 ‘Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX delta ceramic-on-ceramic total hip replacement’, Journal of Biomedical Materials Research - Part B Applied Biomaterials, 95 B(2), pp. 263–268. doi: 10.1002/jbm.b.31708. Ali, M., Al-Hajjar, M., Jennings, L. M. and Fisher, J. (2016) ‘Variations in Component Positioning Leading to Dynamic Separation and Edge Loading Influences the Deformation and Wear of Metal-on-Polyethylene Hip Replacement Bearings’, in ORS. Available at: http://www.imbe.leeds.ac.uk/news/documents/ORS2017 Web/Poster_1032_MA_ORS2017.pdf. Bachus, K. N., DeMarco, A. L., Judd, K. T., Horwitz, D. S. and Brodke, D. S. (2006) ‘Measuring contact area, force, and pressure for bioengineering applications: using Fuji Film and TekScan systems.’, Medical Engineering & Physics, 28(5), pp. 483–8. doi: 10.1016/j.medengphy.2005.07.022. Baker R (2013) Measuring walking: a handbook of clinical gait analysis. London: Mac Keith Press. Baldwin, M. A., Clary, C., Maletsky, L. P. and Rullkoetter, P. J. (2009) ‘Verification of predicted specimen-specific natural and implanted patellofemoral kinematics during simulated deep knee bend’, Journal of Biomechanics. Elsevier, 42(14), pp. 2341–2348. doi: 10.1016/j.jbiomech.2009.06.028. Barbour, P. S. M., Stone, M. H. and Fisher, J. (1999) ‘A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 213(6), pp. 455–467. doi: 10.1243/0954411991535068. 231 Bargar, W. L., Bauer, A. and Börner, M. (1998) ‘Primary and revision total hip replacement using the Robodoc system.’, Clinical Orthopaedics and Related Research, (354), pp. 82–91. Bartel, D. L., Burstein, a H., Toda, M. D. and Edwards, D. L. (1985) ‘The effect of conformity and plastic thickness on contact stresses in metal-backed plastic implants.’, Journal of Biomechanical Engineering, 107(3), pp. 193–9. doi: 10.1115/1.3138543. Bergmann, G. et al. (2001) ‘Hip forces and gait patterns from rountine activities’, Journal of Biomechanics, 34, pp. 859–871. doi: http://dx.doi.org/10.1016/S0021-9290(01)00040-9. Bergström, J. S., Rimnac, C. M. and Kurtz, S. M. (2005) ‘Molecular chain stretch is a multiaxial failure criterion for conventional and highly crosslinked UHMWPE’, Journal of Orthopaedic Research, 23(2), pp. 367–375. doi: 10.1016/j.orthres.2004.08.014. Besong, a a, Lee, R., Farrar, R. and Jin, Z. M. (2001) ‘Contact mechanics of a novel metal-on-metal total hip replacement.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine, 215(6), pp. 543–548. doi: 10.1243/0954411011536145. Blumenfeld, T. J., Glaser, D. A., Bargar, W. L., Langston, G. D., Mahfouz, M. R. and Komistek, R. D. (2011) ‘In Vivo Assessment of Total Hip Femoral Head Separation from the Acetabular Cup During 4 Common Daily Activities’, Orthopedics, 34(6). doi: 10.3928/01477447-20110427-06. Bono, J. V, Sanford, L. and Toussaint, J. T. (1994) ‘Severe polyethylene wear in total hip arthroplasty’, The Journal of Arthroplasty, 9, p. 119. 232 Boutin, P. (1972) ‘Total arthroplasty of the hip by sintered aluminum prosthesis. Experimental study and 1st clinical applications.’, Revue de Chirurgie Orthopédique et Traumatologique, 58, pp. 229–246. Bracco, P., Bellare, A., Bistolfi, A., Affatato, S. (2017) ' Ultra-High Molecular Weight Polyethylene: Influence of the Chemical, Physical and Mechanical Properties on the Wear Behavior. A Review', Materials, 10(7), 791. Bragdon, C. R., Jasty, M., Muratoglu, O. K., O’Connor, D. O. and Harris, W. H. (2003) ‘Third-body wear of highly cross-linked polyethylene in a hip simulator.’, The Journal of Arthroplasty, 18(5), pp. 553–61. Brockett, C., Williams, S., Jin, Z., Isaac, G. and Fisher, J. (2006) ‘Friction of Total Hip Replacements With Different Bearings and Loading Conditions’, Journal of Biomedical Materials Research Part B-Applied Biomaterials, (81B), pp. 508–515. doi: 10.1002/jbmb. Brown, S. S. and Clarke, I. C. (2006) ‘A Review of Lubrication Conditions for Wear Simulation in Artificial Hip Replacements’, Tribology Transactions, 49(1), pp. 72–78. doi: 10.1080/05698190500519223. Buechel, F. F. and Pappas, M. J. (2012) Principles of Human Joint Replacement; Design and clinical application. Berlin: Springer. Carter, D. R., Vasu, R. and Harris, W. H. (1982) ‘Stress distributions in the acetabular region-II. Effects of cement thickness and metal backing of the total hip acetabular component’, Journal of Biomechanics, 15(3), pp. 165–170. doi: 10.1016/0021-9290(82)90248-2. Coombs, R., Gristina, A. and Hungerfird, D. (1990) Joint replacement: state of the art. London: Butler & Tanner Ltd. 233 Dandy, D. J. and Edwards, D. J. (2009) Essential orthopaedics and trauma. Edinburgh: Churchill Livingstone. doi: 10.1016/B978-0-443-06718-1.00005-2. Dattani, R. (2007) ‘Femoral osteolysis following total hip replacement.’, Postgraduate medical journal, 83(979), pp. 312–6. doi: 10.1136/pgmj.2006.053215. Delbert, P. (1919) ‘Resultat eloigne d’un visage pour fracture transcervicale du femur.’, Bulletin Societe Chirurgie, 45, p. 305. Dennis, D. A., Komistek, R. D., Northcut, E. J., Ochoa, J. A. and Ritchie, A. (2001) ‘“In vivo” determination of hip joint separation and the forces generated due to impact loading conditions’, Journal of Biomechanics, 34(5), pp. 623–629. doi: 10.1016/S0021-9290(00)00239-6. Dorr, L. D. and Wan, Z. (1998) ‘Causes of and treatment protocol for instability of total hip replacement.’, Clinical orthopaedics and related research, (355), pp. 144–51. Available at: http://www.ncbi.nlm.nih.gov/pubmed/9917599. Dowson, D. (2001) ‘New joints for the Millennium: Wear control in total replacement hip joints’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(4), pp. 335–358. doi: 10.1243/0954411011535939. Dowson, D., McNie, C. M. and Goldsmith, a a J. (2000) ‘Direct experimental evidence of lubrication in a metal-on-metal total hip replacement tested in a joint simulator’, Journal of Mechanical Engineering Science , 214, pp. 75–86. doi: 10.1243/0954406001522822. Dowson, D. and Wright, V. (1973) ‘Bio-tribology. Proceedings of the Conference on The Rheology of Lubrication’, The Institution of Mechanical Engineers and the British Society 234 of Rheology, pp. 81–85. El-Deen, M., Garcia-Finana, M. and Jin, Z. M. (2006) ‘Effect of ultra-high molecular weight polyethylene thickness on contact mechanics in total knee replacement.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine, 220(7), pp. 733–742. Estupinan, J. A., Bartel, D. L. and Wright, T. M. (1998) ‘Residual stresses in ultra-high molecular weight polyethylene loaded cyclically by a rigid moving indenter in nonconforming geometries’, Journal of Orthopaedic Research, 16(1), pp. 80–88. doi: 10.1002/jor.1100160114. Fisher, J. (2011) ‘Bioengineering reasons for the failure of metal-on-metal hip prostheses’, Journal of Bone and Joint Surgery [Br], 9393(8), pp. 1001–4. doi: 10.1302/0301-620X.93B8. Fisher, J., Al Hajjar, M., Williams, S., Tipper, J., Ingham, E. and Jennings, L. (2012) ‘(v) Simulation and measurement of wear in metal-on-metal bearings in vitro- understanding the reasons for increased wear’, Orthopaedics and Trauma. Elsevier Ltd, 26(4), pp. 253–258. doi: 10.1016/j.mporth.2012.05.005. Fisher, J., Jin, Z., Tipper, J., Stone, M. and Ingham, E. (2006) ‘Tribology of alternative bearings.’, Clinical orthopaedics and related research, 453(453), pp. 25–34. doi: 10.1097/01.blo.0000238871.07604.49. Galanis, N. I. and Manolakos, D. E. (2011) ‘Design of a hip joint simulator according to the ISO 14242’, Proceedings of the World Congress on Engineering, III, pp. 2–7. Gao, Y., Jin, Z., Wang, L. and Wang, M. (2015) ‘Finite element analysis of sliding distance 235 and contact mechanics of hip implant under dynamic walking conditions.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 229(6), pp. 469–474. doi: 10.1177/0954411915585380. García-Rey, E. and García-Cimbrelo, E. (2010) ‘Polyethylene in total hip arthroplasty: half a century in the limelight’, Journal of Orthopaedics and Traumatology, 11(2), pp. 67–72. doi: 10.1007/s10195-010-0091-1. Germain, M. A., Hatton, A., Williams, S., Matthews, J. B., Stone, M. H., Fisher, J. and Ingham, E. (2003) ‘Comparison of the cytotoxicity of clinically relevant cobalt-chromium and alumina ceramic wear particles in vitro’, Biomaterials, 24(3), pp. 469–479. doi: 10.1016/S0142-9612(02)00360-5. Gluck, T. (1890) ‘Autoplastik-Transplantation: Implantation von Fremdkörpern.’, Berl Klin Wschr 19, 27, pp. 421–427. Godest, A. C., Beaugonin, M., Haug, E., Taylor, M. and Gregson, P. J. (2002) ‘Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis’, Journal of Biomechanics, 35(2), pp. 267–275. doi: 10.1016/S0021-9290(01)00179-8. Granchi, D. et al. (2003) ‘Immunological changes in patients with primary osteoarthritis of the hip after total joint replacement.’, Journal of Bone and Joint Surgery - Series B, 85(5), pp. 758–764. doi: 10.1302/0301-620X.85B5.13729. Hall, R. M., Bankes, M. J. K. and Blunn, G. (2001) ‘Biotribology for joint replacement’, Current Orthopaedics, 15(4), pp. 281–290. doi: 10.1054/cuor.2001.0197. Halley, D., Glassman, A. and Crowninshield, R. D. (2004) ‘Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head. A case report.’, The 236 Journal of Bone and Joint Surgery. American volume, 86–A(4), pp. 827–30. Available at: http://www.ncbi.nlm.nih.gov/pubmed/15069152. Halloran, J. P., Petrella, A. J. and Rullkoetter, P. J. (2005) ‘Explicit finite element modeling of total knee replacement mechanics’, Journal of Biomechanics, 38(2), pp. 323–331. doi: 10.1016/j.jbiomech.2004.02.046. Hanna, S. A., Somerville, L., McCalden, R. W., Naudie, D. D. and MacDonald, S. J. (2016) ‘Highly cross-linked polyethylene decreases the rate of revision of total hip arthroplasty compared with conventional polyethylene at 13 years follow-up’, The Bone & Joint Journal, 98–B(1), pp. 28–32. doi: 10.1302/0301-620X.98B1.36527. Hashimoto, N., Ando, M., Yayama, T., Uchida, K., Kobayashi, S., Negoro, K. and Baba, H. (2005) ‘Dynamic analysis of the resultant force acting on the hip joint during level walking’, Artificial Organs, 29(5), pp. 387–392. doi: 10.1111/j.1525-1594.2005.29065.x. Hatton, a, Nevelos, J. E., Matthews, J. B., Fisher, J. and Ingham, E. (2003) ‘Effects of clinically relevant alumina ceramic wear particles on TNF-alpha production by human peripheral blood mononuclear phagocytes.’, Biomaterials, 24(7), pp. 1193–204. doi: 10.1016/S0142-9612(02)00510-0. Hertzberg, R. W. and Manson, J. A. (1980) Fatigue of engineering plastics. New York: Academice press. Hua, X., Li, J., Wang, L., Jin, Z., Wilcox, R. and Fisher, J. (2014) ‘Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions’, Journal of Biomechanics. Elsevier, 47(13), pp. 3303–3309. doi: 10.1016/j.jbiomech.2014.08.015. 237 Hua, X., Wroblewski, B. M., Jin, Z. and Wang, L. (2012) ‘The effect of cup inclination and wear on the contact mechanics and cement fixation for ultra high molecular weight polyethylene total hip replacements’, Medical Engineering and Physics. Institute of Physics and Engineering in Medicine, 34(3), pp. 318–325. doi: 10.1016/j.medengphy.2011.07.026. Ingham, E. and Fisher, J. (2000) ‘Biological reactions to wear debris in total joint replacement’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 214(1), pp. 21–37. doi: 10.1243/0954411001535219. Ingram, J. H., Stone, M., Fisher, J. and Ingham, E. (2004) ‘The influence of molecular weight, crosslinking and counterface roughness on TNF-alpha production by macrophages in response to ultra high molecular weight polyethylene particles’, Biomaterials, 25(17), pp. 3511–3522. doi: 10.1016/j.biomaterials.2003.10.054. Iosa, M., Fusco, A., Marchetti, F., Morone, G., Caltagirone, C., Paolucci, S. and Peppe, A. (2013) ‘The golden ratio of gait harmony: Repetitive proportions of repetitive gait phases’, BioMed Research International, 2013. doi: 10.1155/2013/918642. Isaac, G. H., Brockett, C., Breckon, A., van der Jagt, D., Williams, S., Hardaker, C., Fisher, J., Schepers, A. and Surgeon, O. (2009) ‘ceramic-on-metal bearings in total hip replacement whole blood metal ion levels and analysis of retrieved components’, The Journal of Bone and Joint Surgery [Br], 91, pp. 1134–41. doi: 10.1302/0301-620X.91B9. Jacobs, J., Shanbhag, A., TT, G., Black, J. and JO, G. (1994) ‘Wear debris in total joint replacements’, The Journal of the American Academy of Orthopaedic Surgeons, 2(2), pp. 212–220. Jagatia, M. and Jin, Z. M. (2001) ‘Elastohydrodynamic lubrication analysis of metal-on- 238 metal hip prostheses under steady state entraining motion’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(6), pp. 531–541. doi: 10.1243/0954411011536136. Jalali-Vahid, D., Jagatia, M., Jin, Z. M. and Dowson, D. (2001) ‘Prediction of lubricating \"lm thickness in UHMWPE hip joint replacements’, Journal of Biomechanics, 34, pp. 261–266. doi: 10.1016/S0021-9290(00)00181-0. Jassim, S. S., Vanhegan, I. S. and Haddad, F. S. (2012) ‘The epidemiology of total hip arthroplasty in england and wales’, Seminars in Arthroplasty. Elsevier, 23(4), pp. 197–199. doi: 10.1053/j.sart.2012.12.001. Jennings, L. M., Al-Hajjar, M., Brockett, C. L., Williams, S., Tipper, J. L., Ingham, E. and Fisher, J. (2012) ‘(iv) Enhancing the safety and reliability of joint replacement implants’, Orthopaedics and Trauma. Elsevier Ltd, 26(4), pp. 246–252. doi: 10.1016/j.mporth.2012.05.006. Jin, Z. M., Dowson, D. and Fisher, J. (1994) ‘A parametric analysis of the contact stress in ultra-high molecular weight polyethylene acetabular cups.’, Medical engineering & physics, 16(5), pp. 398–405. doi: http://dx.doi.org/10.1016/1350-4533(90)90006-T. Jin, Z. M., Fisher, J. and Ingham, E. (2006) ‘Biotribology: MaterialDesign, Lubrication,and Wear in ArtificialHip Joints’, in Handbook of lubrication and tribology. CRC Press. Jin, Z. M., Heng, S. M., Ng, H. W. and Auger, D. D. (1999) ‘An axisymmetric contact model of ultra high molecular weight polyethylene cups against metallic femoral heads for artificial hip joint replacements’, Proceedings of the Institution of Mechanical Engineers Part H-Journal of Engineering in Medicine, 213(H4), pp. 317–327. doi: 10.1243/0954411991535158. 239 Jin, Z. M., Stone, M., Ingham, E. and Fisher, J. (2006) ‘Biotribology’, Current Orthopaedics, 20, pp. 32–40. Johnson, K. l. (1985) Contact mechanics. Cambridge University Press. Johnston, R. C. and Smidt, G. L. (1969) ‘Measurement of hip-joint motion during walking’, Journal of Bone and Joint Surgery, 6(6), pp. 1083–1094. Kaddick, C. and Wimmer, M. A. (2001) ‘Hip simulator wear testing according to the newly introduced standard ISO 14242’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(5), pp. 429–442. doi: 10.1016/j.jbiomech.2004.04.003. Kayabasi, O. and Ekici, B. (2007) ‘The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method’, Materials & Design, 28(8), pp. 2269–2277. doi: 10.1016/j.matdes.2006.08.012. Kennon Robert E. (2008) Hip and knee surgery: a patient’s guide to hip replacement, hip resurfacing, Knee replacement, & Knee Arthroscopy. Middlebury, CT : Orthopaedic Surgery, P.C. Knahr, K. (2011) ‘Tribology in Total Hip Arthroplasty’, in Tribology in Total Hip Arthroplasty. Heidelberg: Springer. doi: 10.1007/978-3-642-19429-0. Komistek, R. D., Dennis, D. A., Ochoa, J. A., Haas, B. D. and Hammill, C. (2002) ‘In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty’, The Journal of bone and joint surgery. American Volume, 84–A(10), pp. 1836–1841. doi: 327. Komistek, R. D., Stiehl, J. B., Dennis, D. a, Paxson, R. D. and Soutas-Little, R. W. (1998) 240 ‘Mathematical model of the lower extremity joint reaction forces using Kane’s method of dynamics.’, Journal of Biomechanics, 31(2), pp. 185–9. doi: 10.1016/S0021-9290(97)00128-0. Korhonen, R. K., Koistinen, A., Konttinen, Y. T., Santavirta, S. S. and Lappalainen, R. (2005) ‘The effect of geometry and abduction angle on the stresses in cemented UHMWPE acetabular cups – finite element simulations and experimental tests’, BioMedical Engineering OnLine, 4(1 %@ 1475-925X), p. 32. doi: 10.1186/1475-925X-4-32 %U http://dx.doi.org/10.1186/1475-925X-4-32. Kurtz, S. M. (2004) The UHMWPE Handbook, The UHMWPE Handbook. Academic Press. doi: 10.1016/B978-012429851-4/50003-6. Kurtz, S. M., Edidin, A. A. and Bartel, D. L. (1997) ‘Thickness on the Contact Stresses in Metal-Backed’, Bone, 30(6). Kurtz, S. M., Ochoa, J. A., White, C. V., Srivastav, S. and Cournoyer, J. (1998) ‘Backside nonconformity and locking restraints affect liner/shell load transfer mechanisms and relative motion in modular acetabular components for total hip replacement’, Journal of Biomechanics, 31(5), pp. 431–437. doi: 10.1016/S0021-9290(98)00030-X. Kurtz, S. M. and Patel, J. D. (2016) UHMWPE Biomaterials Handbook. Amsterdam, The Netherlands: Elsevier. Labek, G., Janda, W., Agreiter, M., Schuh, R. and Böhler, N. (2011) ‘Organisation, data evaluation, interpretation and effect of arthroplasty register data on the outcome in terms of revision rate in total hip arthroplasty’, International Orthopaedics, 35(2), pp. 157–163. doi: 10.1007/s00264-010-1131-4. 241 Leng, J., Al-Hajjar, M., Wilcox, R., Jones, A., Barton, D. and Fisher, J. (2017) ‘Dynamic virtual simulation of the occurrence and severity of edge loading in hip replacements associated with variation in the rotational and translational surgical position’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 231(4), pp. 299–306. doi: 10.1177/0954411917693261. Leslie, I., Williams, S., Brown, C., Isaac, G., Jin, Z., Ingham, E. and Fisher, J. (2008) ‘Effect of bearing size on the long-term wear, wear debris, and ion levels of large diameter metal-on-metal hip replacements - An in vitro study’, Journal of Biomedical Materials Research - Part B Applied Biomaterials, 87(1), pp. 163–172. doi: 10.1002/jbm.b.31087. Lewinnek, B. Y. G. E., From, J. J., Tarr, M. and Compere, I. C. L. (1978) ‘Dislocations after Total Arthroplasties’, Journal of Bone and Joint Surgery, 60(2), pp. 217–220. Liu, F., Leslie, I., Williams, S., Fisher, J. and Jin, Z. (2008) ‘Development of computational wear simulation of metal-on-metal hip resurfacing replacements’, Journal of Biomechanics, 41(3), pp. 686–694. doi: 10.1016/j.jbiomech.2007.09.020. Liu, F., Udofia, I. J., Jin, Z. M., Hirt, F., Rieker, C., Roberts, P. and Grigoris, P. (2005) ‘Comparison of Contact Mechanics between a Total Hip Replacement and a Hip Resurfacing with a Metal-On-Metal Articulation’, Proceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science, 219(7), pp. 727–732. doi: 10.1243/095440605X31490. Liu, F., Williams, S., Jin, Z. and Fisher, J. (2013) ‘Effect of head contact on the rim of the cup on the offset loading and torque in hip joint replacement’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 227(11), pp. 1147–1154. doi: 10.1177/0954411913496016. 242 Liu, H., Ellison, P. J., Xu, H. and Jin, Z. (2010) ‘Coupling of dynamics and contact mechanics of artificial hip joints in a pendulum model’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 224(8), pp. 989–1003. doi: 10.1243/09544119JEIM687. Lombardi., A. V, Mallory, T. H., Dennis, D. A., Komistek, R. D., Fada, R. A. and Northcut, E. J. (2000) ‘An in vivo determination of total hip arthroplasty pistoning during activity’, The Journal of Arthroplasty, 15(6), pp. 702–709. doi: 10.1054/arth.2000.6637. Mak, M., Jin, Z., Fisher, J. and Stewart, T. D. (2011) ‘Influence of Acetabular Cup Rim Design on the Contact Stress During Edge Loading in Ceramic-on-Ceramic Hip Prostheses’, Journal of Arthroplasty. Elsevier Inc., 26(1), pp. 131–136. doi: 10.1016/j.arth.2009.10.019. Mak, M. M. and Jin, Z. M. (2002) ‘Analysis of contact mechanics in ceramic-on-ceramic hip joint replacements.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 216(4), pp. 231–6. doi: 10.1243/095441102321032193. Manaka, M., Clarke, I. C., Yamamoto, K., Shishido, T., Gustafson, A. and Imakiire, A. (2004) ‘Stripe Wear Rates in Alumina THR—Comparison of Microseparation Simulator Study with Retrieved Implants’, Journal of Biomedical Materials Research, 69, pp. 149–157. doi: 10.1002/jbm.b.20033. Mattei, L., Di Puccio, F., Piccigallo, B. and Ciulli, E. (2011) ‘Lubrication and wear modelling of artificial hip joints: A review’, Tribology International. Elsevier, 44(5), pp. 532–549. doi: 10.1016/j.triboint.2010.06.010. Maxian, T. A., Brown, T. D., Pedersen, D. R. and Callaghan, J. J. (1996) ‘A sliding-distance- 243 coupled finite element formulation for polyethylene wear in total hip arthroplasty.’, Journal of Biomechanics, 29(5), pp. 687–92. Available at: http://www.ncbi.nlm.nih.gov/pubmed/8707799. Mccollum, D. and Gray, W. (1990) ‘Dislocation After Total Hip Arthroplasty Causes and Prevention.’, Clinical Orthopaedics and Related Research, 261, pp. 159–172. McKee, G. K. and Watson-Farrar, J. (1966) ‘Replacement of arthritic hips by the McKee-Farrar prosthesis.’, The Journal of Bone and Joint Surgery. British volume, 48(2), pp. 245–259. McKellop, H., Shen, F., Lu, B., Campbell, P. and Salovey, R. (1999) ‘Development of an Extremely Wear-Resistant Ultra High Molecular Weight Polyethylene for Total Hip Replacements’, Joiirrial of Orthupaedic Research, 17, pp. 157–167. Meng, Q., Liu, F., Fisher, J. and Jin, Z. (2013) ‘Contact mechanics and lubrication analyses of ceramic-on-metal total hip replacements’, Tribology International. Elsevier, 63, pp. 51–60. doi: 10.1016/j.triboint.2012.02.012. Miki, H., Sugano, N., Yonenobu, K., Tsuda, K., Hattori, M. and Suzuki, N. (2013) ‘Detecting cause of dislocation after total hip arthroplasty by patient-specific four-dimensional motion analysis’, Clinical Biomechanics. Elsevier Ltd, 28(2), pp. 182–186. doi: 10.1016/j.clinbiomech.2012.11.009. National Joint Registry (2016). Available at: http://www.njrcentre.org.uk/njrcentre/Portals/0/Documents/England/Reports/13th Annual Report/07950 NJR Annual Report 2016 ONLINE REPORT.pdf. Nevelos, J. E., Ingham, E., Doyle, C., Nevelos, A. B. and Fisher, J. (2001) ‘Wear of HIPed 244 and non-HIPed alumina-alumina hip joints under standard and severe simulator testing conditions’, Biomaterials, 22(16), pp. 2191–2197. doi: 10.1016/S0142-9612(00)00361-6. Nevelos, J., Ingham, E., Doyle, C., Streicher, R., Nevelos, A., Walter, W. and Fisher, J. (2000) ‘Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns’, The Journal of Arthroplasty, 15(6), pp. 7–795. doi: 10.1054/arth.2000.8100. O’Dwyer Lancaster-Jones, O., Williams, S., Jennings, L. M., Thompson, J., Isaac, G. H., Fisher, J. and Al-Hajjar, M. (2017) ‘An in vitro simulation model to assess the severity of edge loading and wear, due to variations in component positioning in hip joint replacements’, Journal of Biomedical Materials Research Part B: Applied Biomaterials. doi: 10.1002/jbm.b.33991. Oral, E., Christensen, S. D., Malhi, A. S., Wannomae, K. K. and Muratoglu, O. K. (2006) ‘Wear Resistance and Mechanical Properties of Highly Cross-linked, Ultrahigh–Molecular Weight Polyethylene Doped With Vitamin E’, The Journal of Arthroplasty, 21(4), pp. 580–591. doi: 10.1016/j.arth.2005.07.009. Pare, P., Medley, J. B., Chan, F. W. and Young, S. K. (2003) ‘On the role of the lambda parameter in simulator wear of metal-on-metal hip implants’, Tribology Series, 41(Mc), pp. 281–290. Parratte, S., Pagnano, M. W., Coleman-Wood, K., Kaufman, K. R. and Berry, D. J. (2009) ‘The 2008 frank stinchfield award: Variation in postoperative pelvic tilt may confound the accuracy of hip navigation systems’, Clinical Orthopaedics and Related Research, 467(1), pp. 43–49. doi: 10.1007/s11999-008-0521-z. 245 Partridge, S. (2016) Wear and rim damage of UHMWPE acetabular cups in total hip replacement. University of Leeds. Parvizi, J., Kim, K.-I., Goldberg, G., Mallo, G. and Hozack, W. J. (2006) ‘Recurrent instability after total hip arthroplasty: beware of subtle component malpositioning’, Clinical Orthopaedics and Related Research®, 447(447), pp. 60–65. doi: 10.1097/01.blo.0000218749.37860.7c. Paul, J. P. (1976) ‘Force actions transmitted by joints in the human body.’, Proceedings of the Royal Society of London. Series B, 192(1107), pp. 163–172. doi: 10.1098/rspb.1976.0004. Peterson, M. (1980) Clasification of wear process. Edited by W. W. Peterson MB. New York: ASME. Plank, G., Estok, D., Muratoglu, O., O’Connor, D., Burroughs, B. and Harris, W. (2005) ‘Contact Stress Assessment of Conventional and Highly Crosslinked Ultra High Molecular Weight Polyethylene Acetabular Liners with Finite Element Analysis and Pressure Sensitive Film’, Journal of biomedical materials research. Part B, Applied biomaterials, 80B, pp. 1–10. doi: 10.1002/jbmb. Reeves, E. a, Barton, D. C., FitzPatrick, D. P. and Fisher, J. (1998) ‘A two-dimensional model of cyclic strain accumulation in ultra-high molecular weight polyethylene knee replacements.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 212(June 1997), pp. 189–198. doi: 10.1243/0954411981533971. Reinders, J., Sonntag, R., Heisel, C., Reiner, T., Vot, L. and Kretzer, J. P. (2013) ‘Wear Performance of Ceramic-On-Metal Hip Bearings’, PLoS ONE, 8(8), pp. 1–7. doi: 246 10.1371/journal.pone.0073252. Revell, P. A. (2008) Joint replacement technology. Cambridge, England: Woodhead Pub. Ries, M. D. and Link, T. M. (2012) ‘Monitoring and Risk of Progression of Osteolysis After Total Hip Arthroplasty’, The Journal of Bone & Joint Surgery, 94–A, pp. 2097–105. Saikko, V. (2005) ‘A 12-station anatomic hip joint simulator.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 219(6), pp. 437–48. doi: 10.1243/095441105X34419. Sariali, E., Stewart, T., Jin, Z. and Fisher, J. (2012) ‘Effect of cup abduction angle and head lateral microseparation on contact stresses in ceramic-on-ceramic total hip arthroplasty’, Journal of Biomechanics. Elsevier, 45(2), pp. 390–393. doi: 10.1016/j.jbiomech.2011.10.033. Schmalzried, T. P. (2009) ‘The Importance of Proper Acetabular Component Positioning and the Challenges to Achieving It’, Operative Techniques in Orthopaedics. Elsevier Inc., 19(3), pp. 132–136. doi: 10.1053/j.oto.2009.09.008. Scholes, S. C. and Unsworth, A. (2006) ‘The Effects of Proteins on the Friction and Lubrication of Artificial Joints’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 220(6), pp. 687–693. doi: 10.1243/09544119JEIM21. Smith, S. L. and Unsworth, A. (2001) ‘A five-station hip joint simulator’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(1), pp. 61–64. doi: 10.1243/0954411011533535. Stewart, T. D., Tipper, J. L., Insley, G., Streicher, R. M., Ingham, E. and Fisher, J. (2003) 247 ‘Long-term wear of ceramic matrix composite materials for hip prostheses under severe swing phase microseparation’, Journal Of Biomedical Materials Research Part B-Applied Biomaterials, 66B(2), pp. 567–573. doi: Doi 10.1002/Jbm.B.10035. Tank, P. W. and Gest, T. R. (2008) Lippincott Williams and Wilkins Atlas of Anatomy. Lippincott Williams & Wilkins,US. Taylor, S. J. G., Perry, J. S., Meswania, J. M., Donaldson, N., Walker, P. S. and Cannon, S. R. (1997) ‘Telemetry of forces from proximal femoral replacements and relevance to fixation’, Journal of Biomechanics, 30(3), pp. 225–234. doi: 10.1016/S0021-9290(96)00141-8. Tortora, G. and Derrickson, B. (2009) Principles of human anatomy. Hoboken, NJ : John Wiley & Sons. Walter, W. L., Insley, G. M., Walter, W. K. and Tuke, M. A. (2004) ‘Edge loading in third generation alumina ceramic-on-ceramic bearings: Stripe wear’, Journal of Arthroplasty, 19(4), pp. 402–413. doi: 10.1016/j.arth.2003.09.018. Wang, A., Polineni, V. K., Stark, C. and Dumbleton, J. H. (1998) ‘Effect of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene acetabular cups.’, The Journal of Arthroplasty, 13(6), pp. 615–20. Available at: http://www.ncbi.nlm.nih.gov/pubmed/9741435. Willert, H.-G. (2005) ‘Metal-on-Metal Bearings and Hypersensitivity in Patients with Artificial Hip Joints A Clinical and Histomorphological Study’, The Journal of Bone and Joint Surgery (American), 87(1), p. 28. doi: 10.2106/JBJS.A.02039pp. Williams, S., Al-Hajjar, M., Isaac, G. H. and Fisher, J. (2013) ‘Comparison of ceramic-on- 248 metal and metal-on-metal hip prostheses under adverse conditions’, Journal of Biomedical Materials Research - Part B Applied Biomaterials, 101 B(5), pp. 770–775. doi: 10.1002/jbm.b.32880. Williams, S., Butterfield, M., Stewart, T., Ingham, E., Stone, M. and Fisher, J. (2003) ‘Wear and deformation of ceramic-on-polyethylene total hip replacements with joint laxity and swing phase microseparation.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine, 217(2), pp. 147–153. doi: 10.1243/09544110360579367. Williams, S., Leslie, I., Isaac, G., Jin, Z., Ingham, E. and Fisher, J. (2008) ‘Tribology and Wear of Metal-on-Metal Hip Prostheses: Influence of Cup Angle and Head Position’, The Journal of Bone and Joint Surgery-American Volume, 90(Suppl 3), pp. 111–117. doi: 10.2106/JBJS.H.00485. Williams, S., Schepers, A., Isaac, G., Hardaker, C., Ingham, E., van der Jagt, D., Breckon, A. and Fisher, J. (2007) ‘The 2007 Otto Aufranc Award. Ceramic-on-metal hip arthroplasties: a comparative in vitro and in vivo study.’, Clinical Orthopaedics and Related Research, 465(465), pp. 23–32. doi: 10.1097/BLO.0b013e31814da946. Wright, T. and Goodman, S. (1995) ‘Implant wear: the future of total joint replacement. Rosemont’, in USA: AAOS."]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["doctoral"]},{"key":"dc:type.qualificationname","label":"Dc Type Qualificationname","values":["Ph.D"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.dcterms_iso639_2","label":"Language (ISO 639-2)","values":["eng"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["uk.bl.ethos.739781"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://etheses.whiterose.ac.uk/id/eprint/19913/7/Jahani_F_Mechanical_PhD_2017.pdf"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["The performance of total hip replacement (THR) devices can be affected by various factors such as quality of the tissues surrounding the joint, mismatch of the component centres or the cup positioning during hip replacement surgery. Experimental studies have shown that these factors can cause the separation of the two components during the walking cycle (dynamic separation) and the contact of the femoral head with the rim of the acetabular liner (edge loading), which can lead to increased wear and shortened implant lifespan. There is a need for flexible pre-clinical testing tools which allow THR devices to be assessed under these adverse conditions. In this work, a novel dynamic finite element model was developed that is able to generate dynamic separation as it occurs during the gait cycle. In addition, the ability to interrogate contact mechanics and material strain under separation conditions provides a unique means of assessing the severity of edge loading. This study demonstrates these model capabilities for a range of simulated surgical translational mismatch values, cup inclination angles and swing phase loads for ceramic-on-polyethylene implants. The computational model was developed to replicate one station of the Leeds II hip simulator that mimic in vitro adverse conditions. Firstly, a computational sensitivity model was developed under standard conditions for a stable computational contact. The mechanism of separation was also added. The finite element model was able to predict medial-lateral separation as it occurred dynamically in the gait cycle, including cases where the femoral head was in contact with the rim of the cup. The increase in medial-lateral separation with increased translational mismatch, cup inclination angle and decreased swing phase load were in broad agreement with existing experimental data. The factors that increased the separation level, also increased the permeant deformation on the cup. However, steep cup inclination angle resulted in a higher number of conditions with permanent deformation than the standard cup inclination angle. Moreover, despite the low axial load during swing phase, under some separation conditions, reduced contact area created stress value higher than those at the peak axial load. The developed computational tool can be used to understand the effect of various factors on the separation and contact mechanics simultaneously. As separation is a multi-factorial phenomenon, this model can assist to focus on the selected factors that affect the separation experimentally. Moreover, the effect of components specifications such as materials, geometry, and the cup thickness can be investigated with this model."]},{"key":"dc:format","label":"Dc Format","values":["text"]},{"key":"dcterms:accessrights","label":"Dcterms Accessrights","values":["public"]},{"key":"dc:title","label":"Title","values":["Modelling of Dynamic Edge Loading in Total Hip Replacements with ceramic on polyethylene Bearings"]}]}],"canonical_facts":{"dc:contributor.advisor":["Jones, Alison","Barton , David","Fisher, John","Wilcox, Ruth"],"dc:contributor.sponsor":["EPSRC","DePuy Synthes"],"dc:creator":["Jahani, Faezeh"],"dc:date":["2017-09"],"dc:date.issued":["2017-09"],"dc:description.abstract":["The performance of total hip replacement (THR) devices can be affected by various factors such as quality of the tissues surrounding the joint, mismatch of the component centres or the cup positioning during hip replacement surgery. Experimental studies have shown that these factors can cause the separation of the two components during the walking cycle (dynamic separation) and the contact of the femoral head with the rim of the acetabular liner (edge loading), which can lead to increased wear and shortened implant lifespan. There is a need for flexible pre-clinical testing tools which allow THR devices to be assessed under these adverse conditions. In this work, a novel dynamic finite element model was developed that is able to generate dynamic separation as it occurs during the gait cycle. In addition, the ability to interrogate contact mechanics and material strain under separation conditions provides a unique means of assessing the severity of edge loading. This study demonstrates these model capabilities for a range of simulated surgical translational mismatch values, cup inclination angles and swing phase loads for ceramic-on-polyethylene implants. The computational model was developed to replicate one station of the Leeds II hip simulator that mimic in vitro adverse conditions. Firstly, a computational sensitivity model was developed under standard conditions for a stable computational contact. The mechanism of separation was also added. The finite element model was able to predict medial-lateral separation as it occurred dynamically in the gait cycle, including cases where the femoral head was in contact with the rim of the cup. The increase in medial-lateral separation with increased translational mismatch, cup inclination angle and decreased swing phase load were in broad agreement with existing experimental data. The factors that increased the separation level, also increased the permeant deformation on the cup. However, steep cup inclination angle resulted in a higher number of conditions with permanent deformation than the standard cup inclination angle. Moreover, despite the low axial load during swing phase, under some separation conditions, reduced contact area created stress value higher than those at the peak axial load. The developed computational tool can be used to understand the effect of various factors on the separation and contact mechanics simultaneously. As separation is a multi-factorial phenomenon, this model can assist to focus on the selected factors that affect the separation experimentally. Moreover, the effect of components specifications such as materials, geometry, and the cup thickness can be investigated with this model."],"dc:format":["text"],"dc:identifier":["uk.bl.ethos.739781"],"dc:identifier.uri":["https://etheses.whiterose.ac.uk/id/eprint/19913/7/Jahani_F_Mechanical_PhD_2017.pdf"],"dc:language.dcterms_iso639_2":["eng"],"dc:publisher.commercial":["University of Leeds"],"dc:publisher.department":["Institute of Medical and Biological Engineering (iMBE)(Leeds)"],"dc:publisher.institution":["University of Leeds"],"dc:relation.isreferencedby":["https://etheses.whiterose.ac.uk/id/eprint/19913/"],"dc:relation.references":["Abaqus 6.12 Documentation (2012). Dassault Systèmes. Available at : http://abaqus.software.polimi.it/v6.12/index.html Abaqus 6.14 Documentation (2014). Dassault Systèmes. Available at : http://abaqus.software.polimi.it/v6.14/index.html Affatato S., Leardini W., Zavalloni M. (2006) Hip Joint Simulators: State of the Art. In: Benazzo F., Falez F., Dietrich M. (eds) Bioceramics and Alternative Bearings in Joint Arthroplasty. Ceramics in Orthopaedics. Washington D.C.: Steinkopff. Affatato, S., Spinelli, M., Zavalloni, M., Mazzega-Fabbro, C. and Viceconti, M. (2008) ‘Tribology and total hip joint replacement: Current concepts in mechanical simulation’, Medical Engineering and Physics, 30(10), pp. 1305–1317. doi: 10.1016/j.medengphy.2008.07.006. Affatato, S., Torrecillas, R., Taddei, P., Rocchi, M., Fagnano, C., Ciapetti, G. and Toni, A. (2006) ‘Advanced nanocomposite materials for orthopaedic applications. I. A long-termin vitro wear study of zirconia-toughened alumina’, Journal of Biomedical Materials Research Part B: Applied Biomaterials, 78B(1), pp. 76–82. doi: 10.1002/jbm.b.30462. Al-Hajjar, M., Fisher, J., Tipper, J. L., Williams, S. and Jennings, L. M. (2013) ‘Wear of 36-mm BIOLOX (R) delta ceramic-on-ceramic bearing in total hip replacements under edge loading conditions’, Proceedings Of the Institution Of Mechanical Engineers Part H-Journal Of Engineering In Medicine, 227(H5), pp. 535–542. doi: Doi 10.1177/0954411912474613. Al-Hajjar, M., Leslie, I. J., Tipper, J., Williams, S., Fisher, J. and Jennings, L. M. (2010) 230 ‘Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX delta ceramic-on-ceramic total hip replacement’, Journal of Biomedical Materials Research - Part B Applied Biomaterials, 95 B(2), pp. 263–268. doi: 10.1002/jbm.b.31708. Ali, M., Al-Hajjar, M., Jennings, L. M. and Fisher, J. (2016) ‘Variations in Component Positioning Leading to Dynamic Separation and Edge Loading Influences the Deformation and Wear of Metal-on-Polyethylene Hip Replacement Bearings’, in ORS. Available at: http://www.imbe.leeds.ac.uk/news/documents/ORS2017 Web/Poster_1032_MA_ORS2017.pdf. Bachus, K. N., DeMarco, A. L., Judd, K. T., Horwitz, D. S. and Brodke, D. S. (2006) ‘Measuring contact area, force, and pressure for bioengineering applications: using Fuji Film and TekScan systems.’, Medical Engineering & Physics, 28(5), pp. 483–8. doi: 10.1016/j.medengphy.2005.07.022. Baker R (2013) Measuring walking: a handbook of clinical gait analysis. London: Mac Keith Press. Baldwin, M. A., Clary, C., Maletsky, L. P. and Rullkoetter, P. J. (2009) ‘Verification of predicted specimen-specific natural and implanted patellofemoral kinematics during simulated deep knee bend’, Journal of Biomechanics. Elsevier, 42(14), pp. 2341–2348. doi: 10.1016/j.jbiomech.2009.06.028. Barbour, P. S. M., Stone, M. H. and Fisher, J. (1999) ‘A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 213(6), pp. 455–467. doi: 10.1243/0954411991535068. 231 Bargar, W. L., Bauer, A. and Börner, M. (1998) ‘Primary and revision total hip replacement using the Robodoc system.’, Clinical Orthopaedics and Related Research, (354), pp. 82–91. Bartel, D. L., Burstein, a H., Toda, M. D. and Edwards, D. L. (1985) ‘The effect of conformity and plastic thickness on contact stresses in metal-backed plastic implants.’, Journal of Biomechanical Engineering, 107(3), pp. 193–9. doi: 10.1115/1.3138543. Bergmann, G. et al. (2001) ‘Hip forces and gait patterns from rountine activities’, Journal of Biomechanics, 34, pp. 859–871. doi: http://dx.doi.org/10.1016/S0021-9290(01)00040-9. Bergström, J. S., Rimnac, C. M. and Kurtz, S. M. (2005) ‘Molecular chain stretch is a multiaxial failure criterion for conventional and highly crosslinked UHMWPE’, Journal of Orthopaedic Research, 23(2), pp. 367–375. doi: 10.1016/j.orthres.2004.08.014. Besong, a a, Lee, R., Farrar, R. and Jin, Z. M. (2001) ‘Contact mechanics of a novel metal-on-metal total hip replacement.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine, 215(6), pp. 543–548. doi: 10.1243/0954411011536145. Blumenfeld, T. J., Glaser, D. A., Bargar, W. L., Langston, G. D., Mahfouz, M. R. and Komistek, R. D. (2011) ‘In Vivo Assessment of Total Hip Femoral Head Separation from the Acetabular Cup During 4 Common Daily Activities’, Orthopedics, 34(6). doi: 10.3928/01477447-20110427-06. Bono, J. V, Sanford, L. and Toussaint, J. T. (1994) ‘Severe polyethylene wear in total hip arthroplasty’, The Journal of Arthroplasty, 9, p. 119. 232 Boutin, P. (1972) ‘Total arthroplasty of the hip by sintered aluminum prosthesis. Experimental study and 1st clinical applications.’, Revue de Chirurgie Orthopédique et Traumatologique, 58, pp. 229–246. Bracco, P., Bellare, A., Bistolfi, A., Affatato, S. (2017) ' Ultra-High Molecular Weight Polyethylene: Influence of the Chemical, Physical and Mechanical Properties on the Wear Behavior. A Review', Materials, 10(7), 791. Bragdon, C. R., Jasty, M., Muratoglu, O. K., O’Connor, D. O. and Harris, W. H. (2003) ‘Third-body wear of highly cross-linked polyethylene in a hip simulator.’, The Journal of Arthroplasty, 18(5), pp. 553–61. Brockett, C., Williams, S., Jin, Z., Isaac, G. and Fisher, J. (2006) ‘Friction of Total Hip Replacements With Different Bearings and Loading Conditions’, Journal of Biomedical Materials Research Part B-Applied Biomaterials, (81B), pp. 508–515. doi: 10.1002/jbmb. Brown, S. S. and Clarke, I. C. (2006) ‘A Review of Lubrication Conditions for Wear Simulation in Artificial Hip Replacements’, Tribology Transactions, 49(1), pp. 72–78. doi: 10.1080/05698190500519223. Buechel, F. F. and Pappas, M. J. (2012) Principles of Human Joint Replacement; Design and clinical application. Berlin: Springer. Carter, D. R., Vasu, R. and Harris, W. H. (1982) ‘Stress distributions in the acetabular region-II. Effects of cement thickness and metal backing of the total hip acetabular component’, Journal of Biomechanics, 15(3), pp. 165–170. doi: 10.1016/0021-9290(82)90248-2. Coombs, R., Gristina, A. and Hungerfird, D. (1990) Joint replacement: state of the art. London: Butler & Tanner Ltd. 233 Dandy, D. J. and Edwards, D. J. (2009) Essential orthopaedics and trauma. Edinburgh: Churchill Livingstone. doi: 10.1016/B978-0-443-06718-1.00005-2. Dattani, R. (2007) ‘Femoral osteolysis following total hip replacement.’, Postgraduate medical journal, 83(979), pp. 312–6. doi: 10.1136/pgmj.2006.053215. Delbert, P. (1919) ‘Resultat eloigne d’un visage pour fracture transcervicale du femur.’, Bulletin Societe Chirurgie, 45, p. 305. Dennis, D. A., Komistek, R. D., Northcut, E. J., Ochoa, J. A. and Ritchie, A. (2001) ‘“In vivo” determination of hip joint separation and the forces generated due to impact loading conditions’, Journal of Biomechanics, 34(5), pp. 623–629. doi: 10.1016/S0021-9290(00)00239-6. Dorr, L. D. and Wan, Z. (1998) ‘Causes of and treatment protocol for instability of total hip replacement.’, Clinical orthopaedics and related research, (355), pp. 144–51. Available at: http://www.ncbi.nlm.nih.gov/pubmed/9917599. Dowson, D. (2001) ‘New joints for the Millennium: Wear control in total replacement hip joints’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(4), pp. 335–358. doi: 10.1243/0954411011535939. Dowson, D., McNie, C. M. and Goldsmith, a a J. (2000) ‘Direct experimental evidence of lubrication in a metal-on-metal total hip replacement tested in a joint simulator’, Journal of Mechanical Engineering Science , 214, pp. 75–86. doi: 10.1243/0954406001522822. Dowson, D. and Wright, V. (1973) ‘Bio-tribology. Proceedings of the Conference on The Rheology of Lubrication’, The Institution of Mechanical Engineers and the British Society 234 of Rheology, pp. 81–85. El-Deen, M., Garcia-Finana, M. and Jin, Z. M. (2006) ‘Effect of ultra-high molecular weight polyethylene thickness on contact mechanics in total knee replacement.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine, 220(7), pp. 733–742. Estupinan, J. A., Bartel, D. L. and Wright, T. M. (1998) ‘Residual stresses in ultra-high molecular weight polyethylene loaded cyclically by a rigid moving indenter in nonconforming geometries’, Journal of Orthopaedic Research, 16(1), pp. 80–88. doi: 10.1002/jor.1100160114. Fisher, J. (2011) ‘Bioengineering reasons for the failure of metal-on-metal hip prostheses’, Journal of Bone and Joint Surgery [Br], 9393(8), pp. 1001–4. doi: 10.1302/0301-620X.93B8. Fisher, J., Al Hajjar, M., Williams, S., Tipper, J., Ingham, E. and Jennings, L. (2012) ‘(v) Simulation and measurement of wear in metal-on-metal bearings in vitro- understanding the reasons for increased wear’, Orthopaedics and Trauma. Elsevier Ltd, 26(4), pp. 253–258. doi: 10.1016/j.mporth.2012.05.005. Fisher, J., Jin, Z., Tipper, J., Stone, M. and Ingham, E. (2006) ‘Tribology of alternative bearings.’, Clinical orthopaedics and related research, 453(453), pp. 25–34. doi: 10.1097/01.blo.0000238871.07604.49. Galanis, N. I. and Manolakos, D. E. (2011) ‘Design of a hip joint simulator according to the ISO 14242’, Proceedings of the World Congress on Engineering, III, pp. 2–7. Gao, Y., Jin, Z., Wang, L. and Wang, M. (2015) ‘Finite element analysis of sliding distance 235 and contact mechanics of hip implant under dynamic walking conditions.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 229(6), pp. 469–474. doi: 10.1177/0954411915585380. García-Rey, E. and García-Cimbrelo, E. (2010) ‘Polyethylene in total hip arthroplasty: half a century in the limelight’, Journal of Orthopaedics and Traumatology, 11(2), pp. 67–72. doi: 10.1007/s10195-010-0091-1. Germain, M. A., Hatton, A., Williams, S., Matthews, J. B., Stone, M. H., Fisher, J. and Ingham, E. (2003) ‘Comparison of the cytotoxicity of clinically relevant cobalt-chromium and alumina ceramic wear particles in vitro’, Biomaterials, 24(3), pp. 469–479. doi: 10.1016/S0142-9612(02)00360-5. Gluck, T. (1890) ‘Autoplastik-Transplantation: Implantation von Fremdkörpern.’, Berl Klin Wschr 19, 27, pp. 421–427. Godest, A. C., Beaugonin, M., Haug, E., Taylor, M. and Gregson, P. J. (2002) ‘Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis’, Journal of Biomechanics, 35(2), pp. 267–275. doi: 10.1016/S0021-9290(01)00179-8. Granchi, D. et al. (2003) ‘Immunological changes in patients with primary osteoarthritis of the hip after total joint replacement.’, Journal of Bone and Joint Surgery - Series B, 85(5), pp. 758–764. doi: 10.1302/0301-620X.85B5.13729. Hall, R. M., Bankes, M. J. K. and Blunn, G. (2001) ‘Biotribology for joint replacement’, Current Orthopaedics, 15(4), pp. 281–290. doi: 10.1054/cuor.2001.0197. Halley, D., Glassman, A. and Crowninshield, R. D. (2004) ‘Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head. A case report.’, The 236 Journal of Bone and Joint Surgery. American volume, 86–A(4), pp. 827–30. Available at: http://www.ncbi.nlm.nih.gov/pubmed/15069152. Halloran, J. P., Petrella, A. J. and Rullkoetter, P. J. (2005) ‘Explicit finite element modeling of total knee replacement mechanics’, Journal of Biomechanics, 38(2), pp. 323–331. doi: 10.1016/j.jbiomech.2004.02.046. Hanna, S. A., Somerville, L., McCalden, R. W., Naudie, D. D. and MacDonald, S. J. (2016) ‘Highly cross-linked polyethylene decreases the rate of revision of total hip arthroplasty compared with conventional polyethylene at 13 years follow-up’, The Bone & Joint Journal, 98–B(1), pp. 28–32. doi: 10.1302/0301-620X.98B1.36527. Hashimoto, N., Ando, M., Yayama, T., Uchida, K., Kobayashi, S., Negoro, K. and Baba, H. (2005) ‘Dynamic analysis of the resultant force acting on the hip joint during level walking’, Artificial Organs, 29(5), pp. 387–392. doi: 10.1111/j.1525-1594.2005.29065.x. Hatton, a, Nevelos, J. E., Matthews, J. B., Fisher, J. and Ingham, E. (2003) ‘Effects of clinically relevant alumina ceramic wear particles on TNF-alpha production by human peripheral blood mononuclear phagocytes.’, Biomaterials, 24(7), pp. 1193–204. doi: 10.1016/S0142-9612(02)00510-0. Hertzberg, R. W. and Manson, J. A. (1980) Fatigue of engineering plastics. New York: Academice press. Hua, X., Li, J., Wang, L., Jin, Z., Wilcox, R. and Fisher, J. (2014) ‘Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions’, Journal of Biomechanics. Elsevier, 47(13), pp. 3303–3309. doi: 10.1016/j.jbiomech.2014.08.015. 237 Hua, X., Wroblewski, B. M., Jin, Z. and Wang, L. (2012) ‘The effect of cup inclination and wear on the contact mechanics and cement fixation for ultra high molecular weight polyethylene total hip replacements’, Medical Engineering and Physics. Institute of Physics and Engineering in Medicine, 34(3), pp. 318–325. doi: 10.1016/j.medengphy.2011.07.026. Ingham, E. and Fisher, J. (2000) ‘Biological reactions to wear debris in total joint replacement’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 214(1), pp. 21–37. doi: 10.1243/0954411001535219. Ingram, J. H., Stone, M., Fisher, J. and Ingham, E. (2004) ‘The influence of molecular weight, crosslinking and counterface roughness on TNF-alpha production by macrophages in response to ultra high molecular weight polyethylene particles’, Biomaterials, 25(17), pp. 3511–3522. doi: 10.1016/j.biomaterials.2003.10.054. Iosa, M., Fusco, A., Marchetti, F., Morone, G., Caltagirone, C., Paolucci, S. and Peppe, A. (2013) ‘The golden ratio of gait harmony: Repetitive proportions of repetitive gait phases’, BioMed Research International, 2013. doi: 10.1155/2013/918642. Isaac, G. H., Brockett, C., Breckon, A., van der Jagt, D., Williams, S., Hardaker, C., Fisher, J., Schepers, A. and Surgeon, O. (2009) ‘ceramic-on-metal bearings in total hip replacement whole blood metal ion levels and analysis of retrieved components’, The Journal of Bone and Joint Surgery [Br], 91, pp. 1134–41. doi: 10.1302/0301-620X.91B9. Jacobs, J., Shanbhag, A., TT, G., Black, J. and JO, G. (1994) ‘Wear debris in total joint replacements’, The Journal of the American Academy of Orthopaedic Surgeons, 2(2), pp. 212–220. Jagatia, M. and Jin, Z. M. (2001) ‘Elastohydrodynamic lubrication analysis of metal-on- 238 metal hip prostheses under steady state entraining motion’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(6), pp. 531–541. doi: 10.1243/0954411011536136. Jalali-Vahid, D., Jagatia, M., Jin, Z. M. and Dowson, D. (2001) ‘Prediction of lubricating \"lm thickness in UHMWPE hip joint replacements’, Journal of Biomechanics, 34, pp. 261–266. doi: 10.1016/S0021-9290(00)00181-0. Jassim, S. S., Vanhegan, I. S. and Haddad, F. S. (2012) ‘The epidemiology of total hip arthroplasty in england and wales’, Seminars in Arthroplasty. Elsevier, 23(4), pp. 197–199. doi: 10.1053/j.sart.2012.12.001. Jennings, L. M., Al-Hajjar, M., Brockett, C. L., Williams, S., Tipper, J. L., Ingham, E. and Fisher, J. (2012) ‘(iv) Enhancing the safety and reliability of joint replacement implants’, Orthopaedics and Trauma. Elsevier Ltd, 26(4), pp. 246–252. doi: 10.1016/j.mporth.2012.05.006. Jin, Z. M., Dowson, D. and Fisher, J. (1994) ‘A parametric analysis of the contact stress in ultra-high molecular weight polyethylene acetabular cups.’, Medical engineering & physics, 16(5), pp. 398–405. doi: http://dx.doi.org/10.1016/1350-4533(90)90006-T. Jin, Z. M., Fisher, J. and Ingham, E. (2006) ‘Biotribology: MaterialDesign, Lubrication,and Wear in ArtificialHip Joints’, in Handbook of lubrication and tribology. CRC Press. Jin, Z. M., Heng, S. M., Ng, H. W. and Auger, D. D. (1999) ‘An axisymmetric contact model of ultra high molecular weight polyethylene cups against metallic femoral heads for artificial hip joint replacements’, Proceedings of the Institution of Mechanical Engineers Part H-Journal of Engineering in Medicine, 213(H4), pp. 317–327. doi: 10.1243/0954411991535158. 239 Jin, Z. M., Stone, M., Ingham, E. and Fisher, J. (2006) ‘Biotribology’, Current Orthopaedics, 20, pp. 32–40. Johnson, K. l. (1985) Contact mechanics. Cambridge University Press. Johnston, R. C. and Smidt, G. L. (1969) ‘Measurement of hip-joint motion during walking’, Journal of Bone and Joint Surgery, 6(6), pp. 1083–1094. Kaddick, C. and Wimmer, M. A. (2001) ‘Hip simulator wear testing according to the newly introduced standard ISO 14242’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(5), pp. 429–442. doi: 10.1016/j.jbiomech.2004.04.003. Kayabasi, O. and Ekici, B. (2007) ‘The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method’, Materials & Design, 28(8), pp. 2269–2277. doi: 10.1016/j.matdes.2006.08.012. Kennon Robert E. (2008) Hip and knee surgery: a patient’s guide to hip replacement, hip resurfacing, Knee replacement, & Knee Arthroscopy. Middlebury, CT : Orthopaedic Surgery, P.C. Knahr, K. (2011) ‘Tribology in Total Hip Arthroplasty’, in Tribology in Total Hip Arthroplasty. Heidelberg: Springer. doi: 10.1007/978-3-642-19429-0. Komistek, R. D., Dennis, D. A., Ochoa, J. A., Haas, B. D. and Hammill, C. (2002) ‘In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty’, The Journal of bone and joint surgery. American Volume, 84–A(10), pp. 1836–1841. doi: 327. Komistek, R. D., Stiehl, J. B., Dennis, D. a, Paxson, R. D. and Soutas-Little, R. W. (1998) 240 ‘Mathematical model of the lower extremity joint reaction forces using Kane’s method of dynamics.’, Journal of Biomechanics, 31(2), pp. 185–9. doi: 10.1016/S0021-9290(97)00128-0. Korhonen, R. K., Koistinen, A., Konttinen, Y. T., Santavirta, S. S. and Lappalainen, R. (2005) ‘The effect of geometry and abduction angle on the stresses in cemented UHMWPE acetabular cups – finite element simulations and experimental tests’, BioMedical Engineering OnLine, 4(1 %@ 1475-925X), p. 32. doi: 10.1186/1475-925X-4-32 %U http://dx.doi.org/10.1186/1475-925X-4-32. Kurtz, S. M. (2004) The UHMWPE Handbook, The UHMWPE Handbook. Academic Press. doi: 10.1016/B978-012429851-4/50003-6. Kurtz, S. M., Edidin, A. A. and Bartel, D. L. (1997) ‘Thickness on the Contact Stresses in Metal-Backed’, Bone, 30(6). Kurtz, S. M., Ochoa, J. A., White, C. V., Srivastav, S. and Cournoyer, J. (1998) ‘Backside nonconformity and locking restraints affect liner/shell load transfer mechanisms and relative motion in modular acetabular components for total hip replacement’, Journal of Biomechanics, 31(5), pp. 431–437. doi: 10.1016/S0021-9290(98)00030-X. Kurtz, S. M. and Patel, J. D. (2016) UHMWPE Biomaterials Handbook. Amsterdam, The Netherlands: Elsevier. Labek, G., Janda, W., Agreiter, M., Schuh, R. and Böhler, N. (2011) ‘Organisation, data evaluation, interpretation and effect of arthroplasty register data on the outcome in terms of revision rate in total hip arthroplasty’, International Orthopaedics, 35(2), pp. 157–163. doi: 10.1007/s00264-010-1131-4. 241 Leng, J., Al-Hajjar, M., Wilcox, R., Jones, A., Barton, D. and Fisher, J. (2017) ‘Dynamic virtual simulation of the occurrence and severity of edge loading in hip replacements associated with variation in the rotational and translational surgical position’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 231(4), pp. 299–306. doi: 10.1177/0954411917693261. Leslie, I., Williams, S., Brown, C., Isaac, G., Jin, Z., Ingham, E. and Fisher, J. (2008) ‘Effect of bearing size on the long-term wear, wear debris, and ion levels of large diameter metal-on-metal hip replacements - An in vitro study’, Journal of Biomedical Materials Research - Part B Applied Biomaterials, 87(1), pp. 163–172. doi: 10.1002/jbm.b.31087. Lewinnek, B. Y. G. E., From, J. J., Tarr, M. and Compere, I. C. L. (1978) ‘Dislocations after Total Arthroplasties’, Journal of Bone and Joint Surgery, 60(2), pp. 217–220. Liu, F., Leslie, I., Williams, S., Fisher, J. and Jin, Z. (2008) ‘Development of computational wear simulation of metal-on-metal hip resurfacing replacements’, Journal of Biomechanics, 41(3), pp. 686–694. doi: 10.1016/j.jbiomech.2007.09.020. Liu, F., Udofia, I. J., Jin, Z. M., Hirt, F., Rieker, C., Roberts, P. and Grigoris, P. (2005) ‘Comparison of Contact Mechanics between a Total Hip Replacement and a Hip Resurfacing with a Metal-On-Metal Articulation’, Proceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science, 219(7), pp. 727–732. doi: 10.1243/095440605X31490. Liu, F., Williams, S., Jin, Z. and Fisher, J. (2013) ‘Effect of head contact on the rim of the cup on the offset loading and torque in hip joint replacement’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 227(11), pp. 1147–1154. doi: 10.1177/0954411913496016. 242 Liu, H., Ellison, P. J., Xu, H. and Jin, Z. (2010) ‘Coupling of dynamics and contact mechanics of artificial hip joints in a pendulum model’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 224(8), pp. 989–1003. doi: 10.1243/09544119JEIM687. Lombardi., A. V, Mallory, T. H., Dennis, D. A., Komistek, R. D., Fada, R. A. and Northcut, E. J. (2000) ‘An in vivo determination of total hip arthroplasty pistoning during activity’, The Journal of Arthroplasty, 15(6), pp. 702–709. doi: 10.1054/arth.2000.6637. Mak, M., Jin, Z., Fisher, J. and Stewart, T. D. (2011) ‘Influence of Acetabular Cup Rim Design on the Contact Stress During Edge Loading in Ceramic-on-Ceramic Hip Prostheses’, Journal of Arthroplasty. Elsevier Inc., 26(1), pp. 131–136. doi: 10.1016/j.arth.2009.10.019. Mak, M. M. and Jin, Z. M. (2002) ‘Analysis of contact mechanics in ceramic-on-ceramic hip joint replacements.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 216(4), pp. 231–6. doi: 10.1243/095441102321032193. Manaka, M., Clarke, I. C., Yamamoto, K., Shishido, T., Gustafson, A. and Imakiire, A. (2004) ‘Stripe Wear Rates in Alumina THR—Comparison of Microseparation Simulator Study with Retrieved Implants’, Journal of Biomedical Materials Research, 69, pp. 149–157. doi: 10.1002/jbm.b.20033. Mattei, L., Di Puccio, F., Piccigallo, B. and Ciulli, E. (2011) ‘Lubrication and wear modelling of artificial hip joints: A review’, Tribology International. Elsevier, 44(5), pp. 532–549. doi: 10.1016/j.triboint.2010.06.010. Maxian, T. A., Brown, T. D., Pedersen, D. R. and Callaghan, J. J. (1996) ‘A sliding-distance- 243 coupled finite element formulation for polyethylene wear in total hip arthroplasty.’, Journal of Biomechanics, 29(5), pp. 687–92. Available at: http://www.ncbi.nlm.nih.gov/pubmed/8707799. Mccollum, D. and Gray, W. (1990) ‘Dislocation After Total Hip Arthroplasty Causes and Prevention.’, Clinical Orthopaedics and Related Research, 261, pp. 159–172. McKee, G. K. and Watson-Farrar, J. (1966) ‘Replacement of arthritic hips by the McKee-Farrar prosthesis.’, The Journal of Bone and Joint Surgery. British volume, 48(2), pp. 245–259. McKellop, H., Shen, F., Lu, B., Campbell, P. and Salovey, R. (1999) ‘Development of an Extremely Wear-Resistant Ultra High Molecular Weight Polyethylene for Total Hip Replacements’, Joiirrial of Orthupaedic Research, 17, pp. 157–167. Meng, Q., Liu, F., Fisher, J. and Jin, Z. (2013) ‘Contact mechanics and lubrication analyses of ceramic-on-metal total hip replacements’, Tribology International. Elsevier, 63, pp. 51–60. doi: 10.1016/j.triboint.2012.02.012. Miki, H., Sugano, N., Yonenobu, K., Tsuda, K., Hattori, M. and Suzuki, N. (2013) ‘Detecting cause of dislocation after total hip arthroplasty by patient-specific four-dimensional motion analysis’, Clinical Biomechanics. Elsevier Ltd, 28(2), pp. 182–186. doi: 10.1016/j.clinbiomech.2012.11.009. National Joint Registry (2016). Available at: http://www.njrcentre.org.uk/njrcentre/Portals/0/Documents/England/Reports/13th Annual Report/07950 NJR Annual Report 2016 ONLINE REPORT.pdf. Nevelos, J. E., Ingham, E., Doyle, C., Nevelos, A. B. and Fisher, J. (2001) ‘Wear of HIPed 244 and non-HIPed alumina-alumina hip joints under standard and severe simulator testing conditions’, Biomaterials, 22(16), pp. 2191–2197. doi: 10.1016/S0142-9612(00)00361-6. Nevelos, J., Ingham, E., Doyle, C., Streicher, R., Nevelos, A., Walter, W. and Fisher, J. (2000) ‘Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns’, The Journal of Arthroplasty, 15(6), pp. 7–795. doi: 10.1054/arth.2000.8100. O’Dwyer Lancaster-Jones, O., Williams, S., Jennings, L. M., Thompson, J., Isaac, G. H., Fisher, J. and Al-Hajjar, M. (2017) ‘An in vitro simulation model to assess the severity of edge loading and wear, due to variations in component positioning in hip joint replacements’, Journal of Biomedical Materials Research Part B: Applied Biomaterials. doi: 10.1002/jbm.b.33991. Oral, E., Christensen, S. D., Malhi, A. S., Wannomae, K. K. and Muratoglu, O. K. (2006) ‘Wear Resistance and Mechanical Properties of Highly Cross-linked, Ultrahigh–Molecular Weight Polyethylene Doped With Vitamin E’, The Journal of Arthroplasty, 21(4), pp. 580–591. doi: 10.1016/j.arth.2005.07.009. Pare, P., Medley, J. B., Chan, F. W. and Young, S. K. (2003) ‘On the role of the lambda parameter in simulator wear of metal-on-metal hip implants’, Tribology Series, 41(Mc), pp. 281–290. Parratte, S., Pagnano, M. W., Coleman-Wood, K., Kaufman, K. R. and Berry, D. J. (2009) ‘The 2008 frank stinchfield award: Variation in postoperative pelvic tilt may confound the accuracy of hip navigation systems’, Clinical Orthopaedics and Related Research, 467(1), pp. 43–49. doi: 10.1007/s11999-008-0521-z. 245 Partridge, S. (2016) Wear and rim damage of UHMWPE acetabular cups in total hip replacement. University of Leeds. Parvizi, J., Kim, K.-I., Goldberg, G., Mallo, G. and Hozack, W. J. (2006) ‘Recurrent instability after total hip arthroplasty: beware of subtle component malpositioning’, Clinical Orthopaedics and Related Research®, 447(447), pp. 60–65. doi: 10.1097/01.blo.0000218749.37860.7c. Paul, J. P. (1976) ‘Force actions transmitted by joints in the human body.’, Proceedings of the Royal Society of London. Series B, 192(1107), pp. 163–172. doi: 10.1098/rspb.1976.0004. Peterson, M. (1980) Clasification of wear process. Edited by W. W. Peterson MB. New York: ASME. Plank, G., Estok, D., Muratoglu, O., O’Connor, D., Burroughs, B. and Harris, W. (2005) ‘Contact Stress Assessment of Conventional and Highly Crosslinked Ultra High Molecular Weight Polyethylene Acetabular Liners with Finite Element Analysis and Pressure Sensitive Film’, Journal of biomedical materials research. Part B, Applied biomaterials, 80B, pp. 1–10. doi: 10.1002/jbmb. Reeves, E. a, Barton, D. C., FitzPatrick, D. P. and Fisher, J. (1998) ‘A two-dimensional model of cyclic strain accumulation in ultra-high molecular weight polyethylene knee replacements.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 212(June 1997), pp. 189–198. doi: 10.1243/0954411981533971. Reinders, J., Sonntag, R., Heisel, C., Reiner, T., Vot, L. and Kretzer, J. P. (2013) ‘Wear Performance of Ceramic-On-Metal Hip Bearings’, PLoS ONE, 8(8), pp. 1–7. doi: 246 10.1371/journal.pone.0073252. Revell, P. A. (2008) Joint replacement technology. Cambridge, England: Woodhead Pub. Ries, M. D. and Link, T. M. (2012) ‘Monitoring and Risk of Progression of Osteolysis After Total Hip Arthroplasty’, The Journal of Bone & Joint Surgery, 94–A, pp. 2097–105. Saikko, V. (2005) ‘A 12-station anatomic hip joint simulator.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 219(6), pp. 437–48. doi: 10.1243/095441105X34419. Sariali, E., Stewart, T., Jin, Z. and Fisher, J. (2012) ‘Effect of cup abduction angle and head lateral microseparation on contact stresses in ceramic-on-ceramic total hip arthroplasty’, Journal of Biomechanics. Elsevier, 45(2), pp. 390–393. doi: 10.1016/j.jbiomech.2011.10.033. Schmalzried, T. P. (2009) ‘The Importance of Proper Acetabular Component Positioning and the Challenges to Achieving It’, Operative Techniques in Orthopaedics. Elsevier Inc., 19(3), pp. 132–136. doi: 10.1053/j.oto.2009.09.008. Scholes, S. C. and Unsworth, A. (2006) ‘The Effects of Proteins on the Friction and Lubrication of Artificial Joints’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 220(6), pp. 687–693. doi: 10.1243/09544119JEIM21. Smith, S. L. and Unsworth, A. (2001) ‘A five-station hip joint simulator’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 215(1), pp. 61–64. doi: 10.1243/0954411011533535. Stewart, T. D., Tipper, J. L., Insley, G., Streicher, R. M., Ingham, E. and Fisher, J. (2003) 247 ‘Long-term wear of ceramic matrix composite materials for hip prostheses under severe swing phase microseparation’, Journal Of Biomedical Materials Research Part B-Applied Biomaterials, 66B(2), pp. 567–573. doi: Doi 10.1002/Jbm.B.10035. Tank, P. W. and Gest, T. R. (2008) Lippincott Williams and Wilkins Atlas of Anatomy. Lippincott Williams & Wilkins,US. Taylor, S. J. G., Perry, J. S., Meswania, J. M., Donaldson, N., Walker, P. S. and Cannon, S. R. (1997) ‘Telemetry of forces from proximal femoral replacements and relevance to fixation’, Journal of Biomechanics, 30(3), pp. 225–234. doi: 10.1016/S0021-9290(96)00141-8. Tortora, G. and Derrickson, B. (2009) Principles of human anatomy. Hoboken, NJ : John Wiley & Sons. Walter, W. L., Insley, G. M., Walter, W. K. and Tuke, M. A. (2004) ‘Edge loading in third generation alumina ceramic-on-ceramic bearings: Stripe wear’, Journal of Arthroplasty, 19(4), pp. 402–413. doi: 10.1016/j.arth.2003.09.018. Wang, A., Polineni, V. K., Stark, C. and Dumbleton, J. H. (1998) ‘Effect of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene acetabular cups.’, The Journal of Arthroplasty, 13(6), pp. 615–20. Available at: http://www.ncbi.nlm.nih.gov/pubmed/9741435. Willert, H.-G. (2005) ‘Metal-on-Metal Bearings and Hypersensitivity in Patients with Artificial Hip Joints A Clinical and Histomorphological Study’, The Journal of Bone and Joint Surgery (American), 87(1), p. 28. doi: 10.2106/JBJS.A.02039pp. Williams, S., Al-Hajjar, M., Isaac, G. H. and Fisher, J. (2013) ‘Comparison of ceramic-on- 248 metal and metal-on-metal hip prostheses under adverse conditions’, Journal of Biomedical Materials Research - Part B Applied Biomaterials, 101 B(5), pp. 770–775. doi: 10.1002/jbm.b.32880. Williams, S., Butterfield, M., Stewart, T., Ingham, E., Stone, M. and Fisher, J. (2003) ‘Wear and deformation of ceramic-on-polyethylene total hip replacements with joint laxity and swing phase microseparation.’, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine, 217(2), pp. 147–153. doi: 10.1243/09544110360579367. Williams, S., Leslie, I., Isaac, G., Jin, Z., Ingham, E. and Fisher, J. (2008) ‘Tribology and Wear of Metal-on-Metal Hip Prostheses: Influence of Cup Angle and Head Position’, The Journal of Bone and Joint Surgery-American Volume, 90(Suppl 3), pp. 111–117. doi: 10.2106/JBJS.H.00485. Williams, S., Schepers, A., Isaac, G., Hardaker, C., Ingham, E., van der Jagt, D., Breckon, A. and Fisher, J. (2007) ‘The 2007 Otto Aufranc Award. Ceramic-on-metal hip arthroplasties: a comparative in vitro and in vivo study.’, Clinical Orthopaedics and Related Research, 465(465), pp. 23–32. doi: 10.1097/BLO.0b013e31814da946. Wright, T. and Goodman, S. (1995) ‘Implant wear: the future of total joint replacement. Rosemont’, in USA: AAOS."],"dcterms:accessrights":["public"],"dc:title":["Modelling of Dynamic Edge Loading in Total Hip Replacements with ceramic on polyethylene Bearings"],"dc:type":["Thesis"],"dc:type.qualificationlevel":["doctoral"],"dc:type.qualificationname":["Ph.D"]},"updated_at":"2026-11-18T22:25:53Z"}