{"id":{"repo_id":"whiterose","oai_identifier":"oai:etheses.whiterose.ac.uk:24417"},"canonical_url":"https://search.ndltd.org/etd/whiterose/oai:etheses.whiterose.ac.uk:24417","repository":{"repo_id":"whiterose","name":"White Rose University Consortium","base_url":"https://etheses.whiterose.ac.uk/cgi/oai2"},"display":{"title":"Characterisation and evaluation of a novel transmission detector for intra-fraction monitoring of radiotherapy.","abstract":"The goal of this thesis was to characterise a novel transmission detector in the context of signal prediction. This was to eliminate the need to collect a baseline signal for the device before treatment. This not only saves time, but, by independently generating the baseline signal, the process is less prone to missing errors. A simple analytical algorithm was designed and was found to be capable of detecting gross errors, however, it was shown not to be accurate enough to detect MLC position errors that could have a clinical effect on the delivery. MU check software was commissioned, however the fluence distribution it produced lacked the complexity for accurate signal prediction. A Monte Carlo model of a linac was built and validated then used to demonstrate that the detector could be modelled as two slabs of Perspex; the signal being proportional to the dose measured in the air between them. Two Monte Carlo models were then made using different systems, these were both evaluated by comparing predicted signals to measured signals for VMAT plans. Both models performed well and were capable of detecting leaf errors ~1mm; the merits of both are discussed with regard to error detection and ease of use.","abstract_html":"The goal of this thesis was to characterise a novel transmission detector in the context of signal prediction. This was to eliminate the need to collect a baseline signal for the device before treatment. This not only saves time, but, by independently generating the baseline signal, the process is less prone to missing errors. A simple analytical algorithm was designed and was found to be capable of detecting gross errors, however, it was shown not to be accurate enough to detect MLC position errors that could have a clinical effect on the delivery. MU check software was commissioned, however the fluence distribution it produced lacked the complexity for accurate signal prediction. A Monte Carlo model of a linac was built and validated then used to demonstrate that the detector could be modelled as two slabs of Perspex; the signal being proportional to the dose measured in the air between them. Two Monte Carlo models were then made using different systems, these were both evaluated by comparing predicted signals to measured signals for VMAT plans. Both models performed well and were capable of detecting leaf errors ~1mm; the merits of both are discussed with regard to error detection and ease of use.","abstract_has_math":false,"creators":["Johnson, Daniel"],"institution":"University of Leeds","degree_name":"Ph.D","degree_level":"doctoral","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Thwaites, D","Buckley, D","Cosgrove, V","Weston, S"],"committee_chairs":[],"committee_members":[],"year":2018,"date_issued":"2018-11","date_published":"2018-11","updated_at":"2026-11-18T22:33:39Z","subjects":[],"languages":["eng"],"rights":[],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier","label":"Identifier","values":["uk.bl.ethos.797960"],"render_values":[{"text":"uk.bl.ethos.797960","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":null,"outbound_label":null,"outbound_source":null,"record_url":null,"record_label":null,"record_kind":null,"record_source":null,"download_url":"https://etheses.whiterose.ac.uk/id/eprint/24417/1/DJ_PhD_Thesis_Final.pdf","download_label":"Download PDF","download_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://etheses.whiterose.ac.uk/cgi/oai2?verb=GetRecord&metadataPrefix=uketd_dc&identifier=oai%3Aetheses.whiterose.ac.uk%3A24417","prefix":"uketd_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Thwaites, D","Buckley, D","Cosgrove, V","Weston, S"]},{"key":"dc:creator","label":"Author","values":["Johnson, Daniel"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2018-11"]},{"key":"dc:date.issued","label":"Date","values":["2018-11"]},{"key":"dc:publisher.commercial","label":"Dc Publisher Commercial","values":["University of Leeds"]},{"key":"dc:publisher.department","label":"Dc Publisher Department","values":["Academic Unit of Medical Physics (Leeds)"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University of Leeds"]},{"key":"dc:relation.isreferencedby","label":"Dc Relation Isreferencedby","values":["https://etheses.whiterose.ac.uk/id/eprint/24417/"]},{"key":"dc:relation.references","label":"Dc Relation References","values":["AAMP (2008), `Task group 106 report: Accelerator beam data commissioning equipment and procedures', Medical Physics 35(9), 4186-4215. Das, Indra J. and Cheng, Chee-Wai and Watts, Ronald J. and Ahnesjö, Anders and Gibbons, John and Li, X. Allen and Lowenstein, Jessica and Mitra, Raj K. and Simon, William E. and Zhu, Timothy C. AAPM (2001), `Task group 50 report: Basic applications of multileaf collimators'. Boyer, Arthur et al. AAPM (2005), `Task group 87 report: Diode in vivo dosimetry for patients receiving external beam radiation therapy'. Yorke, Ellen and Alecu, R and Ding, L and Fontenla, D and Kalend, A and Kaurin, D and Masterson-McGary, ME and Marinello, G and Matzen, T and Saini, A and others. AAPM (2007), `Task group 105 report: Issues associated with clinical implementation of monte carlo-based photon and electron external beam treatment planning', Medical Physics 34(12). Chetty, Indrin J. and Curran, Bruce and Cygler, Joanna E. and DeMarco, John J. and Ezzell, Gary and Faddegon, Bruce A. and Kawrakow, Iwan and Keall, Paul J. and Liu, Helen and Ma, C.-M. Charlie and Rogers, D. W. O. and Seuntjens, Jan and Sheikh-Bagheri, Daryoush and Siebers, Jeffrey V. AAPM (2009), `Task group 142 report: Quality assurance of medical accelerators a', Medical physics 36, 4197-4212. Klein, Eric E and Hanley, Joseph and Bayouth, John and Yin, Fang-Fang and Simon, William and Dresser, Sean and Serago, Christopher and Aguirre, Francisco and Ma, Lijun and Arjomandy, Bijan. AAPM (2011), `Task group 114 report: Verification of monitor unit calculations for non-imrt clinical radiotherapy', Medical Physics 38(1), 504-530. Stern, Robin L. and Heaton, Robert and Fraser, Martin W. and Murty Goddu, S. and Kirby, Thomas H. and Lam, Kwok Leung and Molineu, Andrea and Zhu, Timothy C. AAPM (2014), `Task group 71 report: Monitor unit calculations for external photon and electron beams', Medical physics 41(3). Gibbons, John P and Antolak, John A and Followill, David S and Huq, M Saiful and Klein, Eric E and Lam, Kwok L and Palta, Jatinder R and Roback, Donald M and Reid, Mark and Khan, Faiz M. AAPM (2018), `Task group 218 report: Tolerance limits and methodologies for imrt measurement-based verification qa', Medical physics . Miften, Moyed and Olch, Arthur and Mihailidis, Dimitris and Moran, Jean and Pawlicki, Todd and Molineu, Andrea and Li, Harold and Wijesooriya, Krishni and Shi, Jie and Xia, Ping and others. Adamus-Górka, M., Mavroidis, P., Lind, B. K. & Brahme, A. (2011), `Comparison of dose response models for predicting normal tissue complications from cancer radiotherapy: application in rat spinal cord', Cancers 3(2), 2421-2443. Agnew, A., Agnew, C., Grattan, M., Hounsell, A. & McGarry, C. (2014), `Monitoring daily mlc positional errors using trajectory log files and epid measurements for imrt and vmat deliveries', Physics in Medicine & Biology 59(9), N49. Agostinelli, S., Allison, J., Amako, K. a., Apostolakis, J., Araujo, H., Arce, P., Asai, M., Axen, D., Banerjee, S., Barrand, G. . et al. (2003), `Geant4|a simulation toolkit', Nuclear instruments and methods in physics research section A: Accelerators, Spectrometers, Detectors and Associated Equipment 506, 250-303. Ahnesjö, A. (1989), `Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media', Medical physics 16(4), 577-592. Alber, M., Broggi, S., De Wagter, C., Eichwurzel, I., Engström, P., Fiorino, C., Georg, D., Hartmann, G., Knöös, T., Leal, A. et al. (2008), `Guidelines for the verification of imrt', ESTRO booklet 7. Alber, M. & Nüsslin, F. (1999), `An objective function for radiation treatment optimization based on local biological measures', Physics in Medicine & Biology 44(2), 479. Ash, D. (2007), `Lessons from epinal'. Ash, D. & Bates, T. (1994), `Report on the clinical effects of inadvertent radiation underdosage in 1045 patients', Clinical Oncology 6(4), 214-226. Asnaashari, K., Chow, J. C. & Heydarian, M. (2013), `Dosimetric comparison between two mlc systems commonly used for stereotactic radiosurgery and radiotherapy: a monte carlo and experimental study', Physica medica 29(4), 350-356. Asuni, G., Jensen, J. M. & McCurdy, B. M. C. (2011), `A monte carlo investigation of contaminant electrons due to a novel in vivo transmission detector', Physics in Medicine and Biology 56(4), 1207. Azorín, Nieto, J. (2004), Thermoluminescence dosimetry (tld) and its application in medical physics, in `AIP Conference Proceedings', Vol. 724, AIP, pp. 20-27. Bai, S., Li, G., Wang, M., Jiang, Q., Zhang, Y. & Wei, Y. (2013), `Effect of mlc leaf position, collimator rotation angle, and gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma', Medical Dosimetry 38(2), 143-147. Baker, S., Dahele, M., Lagerwaard, F. J. & Senan, S. (2016), `A critical review of recent developments in radiotherapy for non-small cell lung cancer', Radiation oncology 11(1), 115. Bielajew, A. F., Mohan, R. & Chui, C.-S. (1989), `Improved bremsstrahlung photon angular sampling in the egs4 code system', National Research Council of Canada Report PIRS-0203 . Bielajew, A. F. & Rogers, D. (1986), `Presta: the parameter reduced electron-step transport algorithm for electron monte carlo transport', Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 18(1-6), 165-181. BIPM (2006), `The international system of units (si)', Bureau International des Poids et Mesures, . BIR (2007), `Towards safer radiotherapy'. British Institute of Radiology, Institute of Physics and Engineering in Medicine, National Patient Safety Agency, Society and College of Radiographers, The Royal College of Radiologists. Boesecke, R., Doll, J., Bauer, B., Schlegel, W., Pastyr, O. & Lorenz, W. (1988), `Treatment planning for conformation therapy using a multi-leaf collimator.', Strahlentherapie und Onkologie: Organ der Deutschen Rontgengesellschaft...[et al] 164(3), 151-154. Bogdanich, W. (2010), `Radiation offers new cures, and ways to do harm', The New York Times 23. Bohsung, J., Strahlentherapie, K. f., Berlin, C. & Mitte, S. (2004), `The 2d-array seven29 a new way of dosimetric verification of imrt beams', Application note . Bortfeld, T. (2006), `Imrt: a review and preview', Physics in Medicine & Biology 51(13), R363. Brahme, A. (1984), `Dosimetric precision requirements in radiation therapy', Acta Radiologica: Oncology 23(5), 379-391. Brahme, A. (1987), `Design principles and clinical possibilities with a new generation of radiation therapy equipment: A review', Acta Oncologica 26(6), 403-412. Brahme, A. (1988a), `Optimal setting of multileaf collimators in stationary beam radiation therapy.', Strahlentherapie und Onkologie 164(6), 343-350. Brahme, A. (1988b), `Optimization of stationary and moving beam radiation therapy techniques', Radiotherapy and Oncology 12(2), 129-140. Brahme, A., Roos, J.-E. & Lax, I. (1982), `Solution of an integral equation encountered in rotation therapy', Physics in Medicine & Biology 27(10), 1221. Briesmeister, J. F. et al. (2000), `Mcnptm-a general monte carlo n-particle transport code', Version 4C, LA-13709-M, Los Alamos National Laboratory . British Standards Institution (2000), Transition to ISO 9001: 2000: Guidance on ISO 9001: 2000, Quality Management Systems, Requirements, British Standards Institution. MacNee, Colin and Pyle, Jim and Sutherland, Sandy. Buckley, L. A. & Rogers, D. (2006), `Wall correction factors, pwall, for thimble ionization chambers', Medical physics 33(2), 455-464. Budgell, G., Brown, K., Cashmore, J., Duane, S., Frame, J., Hardy, M., Paynter, D. & Thomas, R. (2016), `IPEM topical report 1: guidance on implementing flattening filter free (FFF) radiotherapy', Physics in Medicine & Biology 61(23), 8360. Budgell, G. J., Mott, J., Williams, P. C. & Brown, K. (2000), `Requirements for leaf position accuracy for dynamic multileaf collimation', Physics in Medicine & Biology 45(5), 1211. Caivano, R., Califano, G., Fiorentino, A., Cozzolino, M., Oliviero, C., Pedicini, P., Clemente, S., Chiumento, C. & Fusco, V. (2014), `Clinically relevant quality assurance for intensity modulated radiotherapy plans: gamma maps and dvh-based evaluation', Cancer investigation 32(3), 85-91. Cancer Research (2019), `Fund gives around 5,800 more cancer patients access to advanced radiotherapy treatment every year | cancer research uk', https://www.cancerresearchuk.org/about-us/cancer-news/press-release/2013-07-11-fund-gives-around-5800-more-cancer-patients-access/-to-advanced-radiotherapy-treatment-every-year#2. (Accessed on 02/08/2019). Celi, S., Costa, E., Wessels, C., Mazal, A., Fourquet, A. & Francois, P. (2016), `Epid based in vivo dosimetry system: clinical experience and results', Journal of applied clinical medical physics 17(3), 262-276. Chalmers, A. J. (2016), `Science in focus: combining radiotherapy with inhibitors of the dna damage response', Clinical Oncology 28(5), 279-282. Chandraraj, V., Stathakis, S., Manickam, R., Esquivel, C., Supe, S. S. & Papanikolaou, N. (2010), `Consistency and reproducibility of the vmat plan delivery using three independent validation methods', Journal of Applied Clinical Medical Physics 12(1), 129-140. Chang, J. Y. (2015), Intensity-modulated radiotherapy, not 3 dimensional conformal, is the preferred technique for treating locally advanced lung cancer, in `Seminars in radiation oncology', Vol. 25, Elsevier, pp. 110-116. Cheng, J. Y., Ning, H., Arora, B. C., Zhuge, Y. & Miller, R. W. (2016), `Output factor comparison of monte carlo and measurement for varian truebeam 6 mv and 10 mv flattening filter-free stereotactic radiosurgery system', Journal of applied clinical medical physics 17(3), 100-110. Chibani, O., Moftah, B. & Ma, C.-M. C. (2011), `On monte carlo modeling of megavoltage photon beams: A revisited study on the sensitivity of beam parameters', Medical physics 38(1), 188-201. Childress, N., Chen, Q. & Rong, Y. (2015), `Parallel/opposed: Imrt qa using treatment log files is superior to conventional measurement-based method', Journal of applied clinical medical physics 16(1), 4-7. Childress, N., Stephens, E., Eklund, D. & Zhang, M. (2012), `Mobius3d white paper: dose calculation algorithm', Houston, TX: Mobius Medical Systems . Cho, B. (2018), `Intensity-modulated radiation therapy: a review with a physics perspective', Radiation oncology journal 36(1), 1. Convery, D. & Rosenbloom, M. (1992), `The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimation', Physics in medicine & biology 37(6), 1359. CRUK (2015), `Cancer statistics for the uk'. (Accessed on 07/03/2018). URL: https://www.cancerresearchuk.org/health-professional/cancer-statisticsfor-the-uk CRUK, N. (2014), `Vision for radiotherapy 2014-24.'. (Accessed on 01/08/2018). URL: https://www.cancerresearchuk.org Cunningham, J., Shrivastava, P. & Wilkinson, J. (1972), `Program irreg| calculation of dose from irregularly shaped radiation beams', Computer Programs in Biomedicine 2(3), 192-199. Dale, R. G., Jones, B. et al. (2007), Radiobiological modelling in radiation oncology, British Institute of Radiology. Das, I. J., Ding, G. X. & Ahnesjö, A. (2008), `Small fields: Nonequilibrium radiation dosimetry', Medical Physics 35(1), 206-215. De Vlamynck, K., Palmans, H., Verhaegen, F., De Wagter, C., De Neve, W. & Thierens, H. (1999), `Dose measurements compared with monte carlo simulations of narrow 6 mv multileaf collimator shaped photon beams', Medical physics 26(9), 1874-1882. Delaby, N., Bouvier, J., Jouyaux, F., Barateau, A. & Lafond, C. (2017), `40. validation of a transit epid device for a clinical use: application to iviewdose (elekta)', Physica Medica: European Journal of Medical Physics 44, 19-20. Department of Health, Donaldson, L. (2006), `On the state of public health', 2006 Annual Report of The Chief Medical Officer. Donaldson, L. (2007), `Reducing harm from radiotherapy', BMJ: British Medical Journal 334(7588), 272. Edwards, C., Hamer, E., Mountford, P. & Moloney, A. (2007), `An update survey of uk in vivo radiotherapy dosimetry practice', The British journal of radiology 80(960), 1011-1014. Edwards, C. & Mountford, P. (2009), `Characteristics of in vivo radiotherapy dosimetry', The British journal of radiology 82(983), 881-883. Ezzell, G. A., Galvin, J. M., Low, D., Palta, J. R., Rosen, I., Sharpe, M. B., Xia, P., Xiao, Y., Xing, L. & Yu, C. X. (2003), `Guidance document on delivery, treatment planning, and clinical implementation of imrt: report of the imrt subcommittee of the aapm radiation therapy committee', Medical physics 30(8), 2089-2115. Farkona, S., Diamandis, E. P. & Blasutig, I. M. (2016), `Cancer immunotherapy: the beginning of the end of cancer?', BMC medicine 14(1), 73. Feldman, A., Edwards, F. M. & Hendee, W. R. (2001), `The routine use of personal patient dosimeters is of little value in detecting therapeutic misadministrations', Medical physics 28(3), 295-297. Feller, W. (2008), An introduction to probability theory and its applications, Vol. 2, John Wiley & Sons. Feng, Z., Wu, H., Zhang, Y., Zhang, Y., Cheng, J. & Su, X. (2015), `Dosimetric comparison between jaw tracking and static jaw techniques in intensity-modulated radiotherapy', Radiation Oncology 10(1), 28. Feygelman, V., Opp, D., Javedan, K., Saini, A. & Zhang, G. (2010), `Evaluation of a 3d diode array dosimeter for helical tomotherapy delivery qa', Medical Dosimetry 35(4), 324-329. Fippel, M. (1999), `Fast monte carlo dose calculation for photon beams based on the vmc electron algorithm', Medical Physics 26(8), 1466-1475. Fippel, M., Haryanto, F., Dohm, O., Nüsslin, F. & Kriesen, S. (2003), `A virtual photon energy fluence model for monte carlo dose calculation', Medical Physics 30(3), 301-311. Fippel, M., Laub, W., Huber, B. & Nüsslin, F. (1999), `Experimental investigation of a fast monte carlo photon beam dose calculation algorithm', Physics in Medicine & Biology 44(12), 3039. Fisher, C. M. & Rabinovitch, R. (2014), `Frontiers in radiotherapy for early-stage invasive breast cancer', Journal of Clinical Oncology 32(26), 2894. Fix, M. K., Keall, P. J., Dawson, K. & Siebers, J. V. (2004), `Monte carlo source model for photon beam radiotherapy: photon source characteristics', Medical physics 31(11), 3106-3121. Fleckenstein, J., Jahnke, L., Lohr, F., Wenz, F. & Hesser, J. (2013), `Development of a geant4 based monte carlo algorithm to evaluate the monaco vmat treatment accuracy', Zeitschrift für Medizinische Physik 23(1), 33-45. Fontenot, J. D. (2014), `Evaluation of a novel secondary check tool for intensitymodulated radiotherapy treatment planning', Journal of applied clinical medical physics 15(5), 207-215. Fraass, B., Doppke, K., Hunt, M., Kutcher, G., Starkschall, G., Stern, R. & Van Dyke, J. (1998), `Task group 53 report: Quality assurance for clinical radiotherapy treatment planning', Medical Physics 25(10), 1773-1829. Fragoso, M., Kawrakow, I., Faddegon, B. A., Solberg, T. D. & Chetty, I. J. (2009), `Fast, accurate photon beam accelerator modeling using beamnrc: A systematic investigation of efficiency enhancing methods and cross-section data', Medical physics 36(12), 5451-5466. Franks, K. & McNair, H. (2012), `Implementation of image-guided radiotherapy', Clinical Oncology 24(9), 625-626. Fredh, A., Scherman, J. B., Fog, L. S. & Munck af Rosenschöld, P. (2013), `Patient qa systems for rotational radiation therapy: a comparative experimental study with intentional errors', Medical physics 40(3), 031716. Galvin, J. M., Smith, A. R. & Lally, B. (1993), `Characterization of a multileaf collimator system', International Journal of Radiation Oncology Biology Physics 25(2), 181-192. Georg, D., Knöös, T. & McClean, B. (2011), `Current status and future perspective of flattening filter free photon beams', Medical physics 38(3), 1280-1293. Gholampourkashi, S., Cygler, J. E., Belec, J., Vujicic, M. & Heath, E. (2019), `Monte carlo and analytic modeling of an elekta infinity linac with agility mlc: Investigating the significance of accurate model parameters for small radiation fields', Journal of Applied Clinical Medical Physics 20(1), 55-67. Godart, J., Korevaar, E., Visser, R., Wauben, D. & van't Veld, A. A. (2011), `Reconstruction of high-resolution 3d dose from matrix measurements: error detection capability of the compass correction kernel method', Physics in Medicine & Biology 56(15), 5029. Greene, D. D. & Williams, P. C. (1997), Linear accelerators for radiation therapy, 2nd ed edn, Bristol : Institute of Physics Pub. Previous ed.: 1985. Greer, P. B., Cadman, P., Lee, C. & Bzdusek, K. (2009), `An energy fluence-convolution model for amorphous silicon epid dose prediction', Medical Physics 36(2), 547-555. Griessbach, I., Lapp, M., Bohsung, J., Gademann, G. & Harder, D. (2005), `Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams', Medical Physics 32(12), 3750-3754. Grofsmid, D., Dirkx, M., Marijnissen, H., Woudstra, E. & Heijmen, B. (2010), `Dosimetric validation of a commercial monte carlo based imrt planning system', Medical physics 37(2), 540-549. Harrison, R. & Morgan, A. (2007), `In vivo dosimetry: hidden dangers?', The British journal of radiology 80(957), 691-692. Heath, E. & Seuntjens, J. (2003), `Development and validation of a beamnrc component module for accurate monte carlo modelling of the varian dynamic millennium multileaf collimator', Physics in Medicine & Biology 48(24), 4045. Heilemann, G., Poppe, B. & Laub, W. (2013), `On the sensitivity of common gamma-index evaluation methods to mlc misalignments in rapidarc quality assurance', Medical physics 40(3). Herwiningsih, S. & Fielding, A. (2016), Focal spot estimation of an elekta dedicated stereotactic linear accelerator monte carlo model, in `Journal of Physics: Conference Series', Vol. 694, IOP Publishing, p. 012013. Higgins, P., Alaei, P., Gerbi, B. & Dusenbery, K. (2003), `In vivo diode dosimetry for routine quality assurance in imrt', Medical physics 30(12), 3118-3123. Hounsell, A. R. (1998), `Monitor chamber backscatter for intensity modulated radiation therapy using multileaf collimators', Physics in Medicine & Biology 43(2), 445. Huq, M. S., Das, I. J., Steinberg, T. & Galvin, J. M. (2002), `A dosimetric comparison of various multileaf collimators', Physics in Medicine & Biology 47(12), N159. Hussein, M., Adams, E. J., Jordan, T. J., Clark, C. H. & Nisbet, A. (2013a), `A critical evaluation of the ptw 2d-array seven29 and octavius ii phantom for imrt and vmat verification', Journal of applied clinical medical physics 14(6), 274-292. Hussein, M., Clark, C. & Nisbet, A. (2017), `Challenges in calculation of the gamma index in radiotherapy - towards good practice', Physica Medica 36, 1-11. Hussein, M., Rowshanfarzad, P., Ebert, M. A., Nisbet, A. & Clark, C. H. (2013b), `A comparison of the gamma index analysis in various commercial imrt/vmat qa systems', Radiotherapy and Oncology 109(3), 370-376. Huyskens, D., Bogaerts, R., Verstraete, J., Lööf, M., Nyström, H., Fiorino, C., Broggi, S., Jornet, N., Ribas, M. & Thwaites, D. (2001), `Practical guidelines for the implementation of in vivo dosimetry with diodes in external radiotherapy with photon beams (entrance dose)'. IAEA (2013), Development of procedures for in vivo dosimetry in radiotherapy., Series IA, International Atomic Energy Agency, Vienna. ICRP (2001), `Prevention of accidents to patients undergoing radiation therapy', International Commission on Radiological Protection 86 30(3). ICRP (2009), `Preventing accidental exposures from new external beam radiation therapy technologies', International Commission on Radiological Protection Publication 112 39(4). P. Ortiz Lopez, J.-M. Cosset, P. Dunscombe, O. Holmberg, J.-C. Rosenwald, L. Pinillos Ashton, J.J. Vilaragut Llanes, S. Vatnitsky. ICRU (1985), `Stopping powers for electrons and positrons', ICRU report 37. Brice, David K. ICRU (1992), `Photon, electron, proton and neutron interaction data for body tissues', ICRU report 46. Bethesda, MD. ICRU (1994), `Prescribing, recording and reporting photon beam therapy', International Commission on Radiation Units and Measurements report 50. Jones, Douglas. ICRU (1999), `Prescribing, recording and reporting photon beam therapy (supplement to icru report 50)', International Commission on Radiation Units and Measurements report 62. T. Landberg J. Chavaudra J. Dobbs J. -P. Gerard G. Hanks J. -C. Horiot K. -A. Johansson T. Möller J. Purdy N. Suntharalingam H. Svensson. IPEM (2006), `Balancing costs and benefits of checking in radiotherapy', Institute of Physics and Engineering in Medicine . McKenzie, Alan. IPEM (2008), `Guidance for the clinical implementation of intensity modulated radiation therapy', Institute of Physics and Engineering in Medicine (IPEM) Report 96. James, H and Beavis, A and Budgell, G and Clark, C and Convery, D and Mott, J and others. IPEM (2018), `Physics aspects of quality control in radiotherapy second edition', Institute of Physics and Engineering in Medicine report 81. Patel, Weston S, Palmer A and others. Islam, M. K., Norrlinger, B. D., Smale, J. R., Heaton, R. K., Galbraith, D., Fan, C. & Jaffray, D. A. (2009), `An integral quality monitoring system for real-time verification of intensity modulated radiation therapy', Medical physics 36(12), 5420-5428. Izewska, J. & Andreo, P. (2000), `The iaea/who tld postal programme for radiotherapy hospitals', Radiotherapy and oncology 54(1), 65-72. Johnson, D., Weston, S. J., Cosgrove, V. P. & Thwaites, D. I. (2014), `A simple model for predicting the signal for a head-mounted transmission chamber system, allowing imrt in-vivo dosimetry without pretreatment linac time', Journal of Applied Clinical Medical Physics 15(4), 270-279. Joiner, M. C. & Van der Kogel, A. (2016), Basic clinical radiobiology, Vol. 2, CRC press. Jones, L., Hoban, P. & Metcalfe, P. (2001), `The use of the linear quadratic model in radiotherapy: a review', Australasian Physics & Engineering Sciences in Medicine 24(3), 132-146. Jordan, T. J. & Williams, P. C. (1994), `The design and performance characteristics of a multileaf collimator', Physics in Medicine & Biology 39(2), 231. Jordan, T. & Williams, P. (1991), Commissioning and use of a multileaf collimator system, in `Proceedings of the 1st biennial ESTRO Meeting on Physics in Clinical Radiotherapy', p. 3. Juste, B., Miro, R., Campayo, J., Diez, S. & Verdu, G. (2008), Photon spectra calculation for an elekta linac beam using experimental scatter measurements and monte carlo techniques., in `Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE', IEEE, pp. 3289-3292. Kairn, T., Crowe, S., Poole, C. & Fielding, A. L. (2009), `Effects of collimator backscatter in an elekta linac by monte carlo simulation', Australasian Physics & Engineering Sciences in Medicine 32(3), 129. Kajaria, A., Sharma, N., Sharma, S., Pradhan, S., Mandal, A. & Aggarwal, L. (2017), `Monte carlo study of a flattening filter-free 6 mv photon beam using the beamnrc code'. Karagoz, G., Zorlu, F., Yeğiner, M., Yıldız, D. & Gürkaynak, M. (2014), `Evaluation of mlc leaf position accuracy for static and dynamic imrt treatments using david in vivo dosimetric system', International Journal of Radiation Oncology* Biology*Physics 90(1). Kawrakow, I., Fippel, M. & Friedrich, K. (1996), `3d electron dose calculation using a voxel based monte carlo algorithm (vmc)', Medical physics 23(4), 445-457. Kawrakow, I., Mainegra-Hing, E., Rogers, D. W. O., Tessier, F. & Walters, B. (2000), The egsnrc code system: Monte carlo simulation of electron and photon transport, Technical report, National Research Council Canada. Kawrakow, I., Rogers, D. & Walters, B. (2004), `Large efficiency improvements in beamnrc using directional bremsstrahlung splitting', Medical physics 31(10), 2883-2898. Khan, F. M. (2010), The physics of radiation therapy, Walters Kluwer, Lippincott Williams & Wilkins. Kim, J.-i., Park, S.-Y., Kim, H. J., Kim, J. H., Ye, S.-J. & Park, J. M. (2014), `The sensitivity of gamma-index method to the positioning errors of high-definition mlc in patient-specific vmat qa for sbrt', Radiation Oncology 9(1), 167. Kinsella, P., Shields, L., McCavana, P., McClean, B. & Langan, B. (2016), `Determination of mlc model parameters for monaco using commercial diode arrays', Journal of applied clinical medical physics 17(4), 37-47. Klein, E. E., Drzymala, R. E., Purdy, J. A. & Michalski, J. (2005), `Errors in radiation oncology: a study in pathways and dosimetric impact', Journal of Applied clinical medical physics 6(3), 81-94. Kung, J., Chen, G. & Kuchnir, F. (2000), `A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assurance', Medical physics 27(10), 2226-2230. Künzler, T., Fotina, I., Stock, M. & Georg, D. (2009), `Experimental verification of a commercial monte carlo-based dose calculation module for high-energy photon beams', Physics in Medicine & Biology 54(24), 7363. Laub, W. U., Kaulich, T. W. & Nüsslin, F. (1999), `A diamond detector in the dosimetry of high-energy electron and photon beams', Physics in Medicine & Biology 44(9), 2183. Lawless, C., Landau, D., Faivre-Finn, C., Boyd, K., Lester, J., Fenwick, J., Maguire, J., McCartney, E., Paul, J., Parsons, E. et al. (2017), `Adscan: a randomised phase ii study of accelerated, dose escalated, sequential chemoradiotherapy in non-small cell lung cancer (nsclc)', Lung Cancer 103, S80-S81. Lea, D. & Catcheside, D. (1942), `The mechanism of the induction by radiation of chromosome aberrations intradescantia', Journal of genetics 44(2-3), 216-245. Leeds University (2018), `Marc1 - advanced research computing', http://arc.leeds.ac.uk/systems/marc1/. (Accessed on 08/14/2018). Leer, J.-W. H., McKenzie, A., Scalliet, P. & Thwaites, D. I. (1998), Practical Guidelines for the Implementation of a Quality System in Radiotherapy: A Project of the ESTRO Quality Assurance Committee Sponsored by\" Europe Against Cancer\", ESTRO. Lillicrap, S. C., Owen, B., Williams, J. R. & Williams, P. C. (1990), `Code of practice for high-energy photon therapy dosimetry based on the npl absorbed dose calibration service', Physics in Medicine and Biology 35(10), 1355-1360. Ling, C. C., Zhang, P., Archambault, Y., Bocanek, J., Tang, G. & LoSasso, T. (2008), `Commissioning and quality assurance of rapidarc radiotherapy delivery system', International Journal of Radiation Oncology* Biology* Physics 72(2), 575-581. Litzenberg, D. W., Moran, J. M. & Fraass, B. A. (2002), `Verification of dynamic and segmental imrt delivery by dynamic log file analysis', Journal of applied clinical medical physics 3(2), 63-72. Liu, H. H., Mackie, T. R. & McCullough, E. C. (2000), `Modeling photon output caused by backscattered radiation into the monitor chamber from collimator jaws using a monte carlo technique', Medical physics 27(4), 737-744. Liu, H. H., Verhaegen, F. & Dong, L. (2001), `A method of simulating dynamic multileaf collimators using monte carlo techniques for intensity-modulated radiation therapy', Physics in Medicine & Biology 46(9), 2283. Lobo, J. & Popescu, I. A. (2010), `Two new dosxyznrc sources for 4d monte carlo simulations of continuously variable beam configurations, with applications to rapidarc, vmat, tomotherapy and cyberknife', Physics in Medicine and Biology 55(16), 4431-4443. Looe, H. K., Harder, D., Rühmann, A., Willborn, K. C. & Poppe, B. (2010), `Enhanced accuracy of the permanent surveillance of imrt deliveries by iterative deconvolution of david chamber signal profiles', Physics in Medicine and Biology 55(14), 3981-3992. LoSasso, T., Chui, C.-S. & Ling, C. C. (1998), `Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy', Medical physics 25(10), 1919-1927. LoSasso, T., Chui, C.-S. & Ling, C. C. (2001), `Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode', Medical physics 28(11), 2209-2219. Low, D. A., Harms, W. B., Mutic, S. & Purdy, J. A. (1998), `A technique for the quantitative evaluation of dose distributions', Medical physics 25(5), 656-661. Lyman, J. T. (1985), `Complication probability as assessed from dose-volume histograms', Radiation Research 104(2s), S13-S19. Ma, C.-m. & Nahum, A. (1993), `Effect of size and composition of the central electrode on the response of cylindrical ionization chambers in high-energy photon and electron beams', Physics in Medicine & Biology 38(2), 267. Ma, C. M. & Rogers, D. W. O. (2018), `Beamdp users manual'. National Research Council of Canada. MacDougall, N. D., Graveling, M., Hansen, V. N., Brownsword, K. & Morgan, A. (2017), `In vivo dosimetry in uk external beam radiotherapy: current and future usage', The British journal of radiology 90(1072), 20160915. Mackay, R. I. & Williams, P. C. (2009), `The cost effectiveness of in vivo dosimetry is not proven', The British journal of radiology 82(976), 265-266. Mani, K. R., Upadhayay, S. & Das, K. M. (2017), `Influence of jaw tracking in intensity-modulated and volumetric-modulated arc radiotherapy for head and neck cancers: a dosimetric study', Radiation oncology journal 35(1), 90. Mans, A., Wendling, M., McDermott, L., Sonke, J.-J., Tielenburg, R., Vijlbrief, R., Mijnheer, B., Van Herk, M. & Stroom, J. (2010), `Catching errors with in vivo epid dosimetry', Medical physics 37, 2638-2644. Mayles, P., Nahum, A. & Rosenwald, J.-C. (2007), Handbook of radiotherapy physics: theory and practice, CRC Press. McDonald, D. G., Jacqmin, D. J., Mart, C. J., Koch, N. C., Peng, J. L., Ashenafi, M. S., Fugal, M. A. & Vanek, K. N. (2017), `Validation of a modern secondcheck dosimetry system using a novel verification phantom', Journal of applied clinical medical physics 18(1), 170-177. McGowan & Faddegon (2013), `Statdose for 3d dose distributions'. National Research Council of Canada. McParland, B. J. (1989), `A derivation of the electron mass scattering power for electron dose calculations', Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 274(3), 592-596. MedCom (2018), `Medcom online: Prosoma core'. (Accessed on 02/19/2018). URL: www.medcom-online.de/clinical-areas-products/teletherapy/ prosomacore/ Meurant, G. (2012), The dosimetry of ionizing radiation, Vol. 3, Elsevier. Mijnheer, B. (2008), `State of the art of in vivo dosimetry', Radiation protection dosimetry 131(1), 117-122. Mijnheer, B., Beddar, S., Izewska, J. & Reft, C. (2013), `In vivo dosimetry in external beam radiotherapy', Medical Physics 40(7). Mishra, B., Mishra, S., Selvam, T. P., Chavan, S. & Pethe, S. (2018), `Comparison of measured and monte carlo calculated dose distributions from indigenously developed 6 mv flattening filter free medical linear accelerator', Journal of medical physics 43(3), 162. Mobius3D (2019), `Mobius medical systems, lp', http://mobiusmed.com/mobius3d/. (Accessed on 01/15/2019). Moeller, R. (1989), New concept for multileaf collimation, in `Proceedings of the Sixth Varian European Users Meeting, San Remo, Italy, 1989', Varian, Palo Alto, CA, pp. 126-131. Mohammed, M., Chakir, E., Boukhal, H., Saeed, M. & El Bardouni, T. (2016), `Evaluation of variance reduction techniques in beamnrc monte carlo simulation to improve the computing efficiency', Journal of Radiation Research and Applied Sciences 9(4), 424-430. Monseux, A., Baltieri, V., Sottiaux, A., Tomsej, M. & Leclercq, C. (2016), `Assessment of portal dosimetry accuracy as a qa tool for vmat clinical treatment plans using dolphin/compass tools', Physica Medica: European Journal of Medical Physics 32, 355-356. Morrison, K., Tree, A., Khoo, V. & Van As, N. J. a. (2018), `The pace trial: International randomised study of laparoscopic prostatectomy vs. stereotactic body radiotherapy (sbrt) and standard radiotherapy vs. sbrt for early stage organ-confined prostate cancer.', Journal of Clinical Oncology 36, TPS153-TPS153. Mu, G., Ludlum, E. & Xia, P. (2007), `Impact of mlc leaf position errors on simple and complex imrt plans for head and neck cancer', Physics in Medicine & Biology 53(1), 77. Munro, A. (2007), `Hidden danger, obvious opportunity: error and risk in the management of cancer', The British journal of radiology 80(960), 955-966. Munro, T. & Gilbert, C. (1961), `The relation between tumour lethal doses and the radiosensitivity of tumour cells', The British journal of radiology 34(400), 246-251. Nakaguchi, Y., Ono, T., Maruyama, M., Shimohigashi, Y. & Kai, Y. (2017), `Validation of a method for in vivo 3d dose reconstruction in sbrt using a new transmission detector', Journal of applied clinical medical physics 18(4), 69-75. NATCANSAT (2019), `National radiotherapy dataset - rtds', http://www.natcansat.nhs.uk/rt/rtds.aspx. (Accessed on 02/08/2019). National Cancer Peer Review Programme (2014), `Radiotherapy service evidence guide', Radiotherapy Service Evidence Guide . Nelson, C., Garcia, M., Calderon, E. & Kirsner, S. (2016), `Su-f-t-306: Validation of mobius 3d and fx for elekta linear accelerators', Medical physics 43(6Part17), 3533-3533. Niemierko, A. (1997), `Reporting and analyzing dose distributions: a concept of equivalent uniform dose', Medical physics 24(1), 103-110. Nithiyanantham, K., Mani, G. K., Subramani, V., Mueller, L., Palaniappan, K. K. & Kataria, T. (2015), `Analysis of direct clinical consequences of mlc positional errors in volumetric-modulated arc therapy using 3d dosimetry system', Journal of applied clinical medical physics 16(5), 296-305. Oborn, B. M., Williams, M., Bailey, M. & Carolan, M. (2014), Imrt treatment monitor unit verification using absolute calibrated beamnrc and geant4 monte carlo simulations, in `Journal of Physics: Conference Series', Vol. 489, IOP Publishing, p. 012020. Oderinde, O. & du Plessis, F. (2016), `O3. accurate monte carlo modelling of an elekta synergy equipped with an agility 160-leaf mlc', Physica Medica 32, 141-142. Ohira, S., Ueda, Y., Isono, M., Masaoka, A., Hashimoto, M., Miyazaki, M., Takashina, M., Koizumi, M. & Teshima, T. (2017), `Can clinically relevant dose errors in patient anatomy be detected by gamma passing rate or modulation complexity score in volumetric-modulated arc therapy for intracranial tumors?', Journal of radiation research 58(5), 685-692. Oliver, M., Gagne, I., Bush, K., Zavgorodni, S., Ansbacher, W. & Beckham, W. (2010), `Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy', Radiotherapy and oncology 97(3), 554-560. Otto, K. (2008), `Volumetric modulated arc therapy: Imrt in a single gantry arc', Medical physics 35(1), 310-317. Padilla-Cabal, F., Pérez-Liva, M., Lara, E., Alfonso, R. & Lopez-Pino, N. (2015), `Monte carlo calculations of an elekta precise sl-25 photon beam model', Journal of Radiotherapy in Practice 14(3), 311-322. Palma, D. A., Verbakel, W. F., Otto, K. & Senan, S. (2010), `New developments in arc radiation therapy: a review', Cancer treatment reviews 36(5), 393-399. Pasler, M., Michel, K., Marrazzo, L., Obenland, M., Pallotta, S., Björnsgard, M. & Lutterbach, J. (2017), `Error detection capability of a novel transmission detector: a validation study for online vmat monitoring', Physics in Medicine & Biology 62(18), 7440. Pena, J., Sánchez-Doblado, F., Capote, R., Terrón, J. & Gómez, F. (2006), `Monte carlo correction factors for a farmer 0.6 cm3 ion chamber dose measurement in the build-up region of the 6 mv clinical beam', Physics in Medicine & Biology 51(6), 1523. Piermattei, A., Fidanzio, A., Azario, L., Grimaldi, L., D'Onofrio, G., Cilla, S., Stimato, G., Gaudino, D., Ramella, S., D'Angelillo, R. et al. (2007), `Application of a practical method for the isocenter point in vivo dosimetry by a transit signal', Physics in Medicine & Biology 52(16), 5101. Pitcher, G. M., Hogstrom, K. R. & Carver, R. L. (2016), `Radiation leakage dose from elekta electron collimation system', Journal of applied clinical medical physics 17(5), 157-176. Pitcher, G. M., Hogstrom, K. R. & Carver, R. L. (2017), `Improved electron collimation system design for elekta linear accelerators', Journal of applied clinical medical physics 18(5), 259-270. Popescu, I. A., Shaw, C., Zavgorodni, S. F. & Beckham, W. A. (2005), `Absolute dose calculations for monte carlo simulations of radiotherapy beams', Physics in Medicine & Biology 50(14), 3375. Poppe, B., Looe, H. K., Chofor, N., Rühmann, A., Harder, D. & Willborn, K. C. (2010), `Clinical performance of a transmission detector array for the permanent supervision of imrt deliveries', Radiotherapy and Oncology 95(2), 158-165. Poppe, B., Thieke, C., Beyer, D., Kollhoff, R., Djouguela, A., Rühmann, A., Willborn, K. C. & Harder, D. (2006), `David|a translucent multi-wire transmission ionization chamber for in-vivo verification of imrt and conformal irradiation techniques', Physics in Medicine and Biology 51(5), 1237-1248. PTW (2010), `A comparison of detectors used in detector arrays'. (Accessed: August 2018. URL: http://www.ptw.de/3095.html Qi, P. & Xia, P. (2013), `Relationship of segment area and monitor unit efficiency in aperture-based imrt optimization', Journal of Applied Clinical Medical Physics 14(3), 232-243. QUANTEC (2010), `Quantitative analyses of normal tissue effects in the clinic (quantec): an introduction to the scientific issues', International Journal of Radiation Oncology Biology Physics 76(3), S3-S9. Bentzen, Søren M and Constine, Louis S and Deasy, Joseph O and Eisbruch, Avi and Jackson, Andrew and Marks, Lawrence B and Ten Haken, Randall K and Yorke, Ellen D. Rangaraj, D., Zhu, M., Yang, D., Palaniswaamy, G., Yaddanapudi, S., Wooten, O. H., Brame, S. & Mutic, S. (2013), `Catching errors with patient-specific pretreatment machine log file analysis', Practical radiation oncology 3(2), 80-90. Rangel, A. & Dunscombe, P. (2009), `Tolerances on mlc leaf position accuracy for imrt delivery with a dynamic mlc', Medical physics 36(7), 3304-3309. RCR (2008a), `Implementing in vivo dosimetry'. Institute of Physics and Engineering in Medicine, National Patient Safety Agency, Society and College of Radiographers, The Royal College of Radiologists. RCR (2008b), `On target: ensuring geometric accuracy in radiotherapy'. Royal College of Radiologists, Institute of Physics and Engineering in Medicine, Society and college of Radiographers, Authors: Barrett, Jane and Hoskin, P and Gaze, M and Greener, T and Kirby, M and McNair, H and Powell, M and Routsis, D and Tait, D. Redpath, A. & Thwaites, D. (1991), `A 3-dimensional scatter correction algorithm for photon beams (dosimetry)', Physics in Medicine & Biology 36(6), 779. Ricketts, K., Navarro, C., Lane, K., Blowfield, C., Cotten, G., Tomala, D., Lord, C., Jones, J. & Adeyemi, A. (2016), `Clinical experience and evaluation of patient treatment verification with a transit dosimeter', International Journal of Radiation Oncology* Biology* Physics 95(5), 1513-1519. Rogers, D., Bielajew, A. et al. (1990), `Monte carlo techniques of electron and photon transport for radiation dosimetry', The dosimetry of ionizing radiation 3, 427-539. Rogers, D. W. O. (2006), `Fifty years of monte carlo simulations for medical physics', Physics in Medicine and Biology 51(13), R287. Rogers, D. W. O., Faddegon, B. A., Ding, G. X., Ma, C.-M., We, J. & Mackie, T. R. (1995), `Beam: A monte carlo code to simulate radiotherapy treatment units', Medical Physics 22(5), 503-524. Rogers, D., Walters, B. & Kawrakow, I. (2001), BEAMnrc users manual, Report, National Research Council Canada. Roussakis, Y. G. (2016), Strategies for adaptive radiotherapy: towards clinically efficient work ows, PhD thesis, University of Birmingham. Salvat, F., Fernández-Varea, J. M. & Sempau, J. (2006), Penelope-2008: A code system for monte carlo simulation of electron and photon transport, in `Workshop Proceedings', Vol. 4. Schreiner, L., Holmes, O. & Salomons, G. (2013), Analysis and evaluation of planned and delivered dose distributions: practical concerns with γ- and χ- evaluations, in `Journal of Physics: Conference Series', Vol. 444, IOP Publishing, p. 012016. SCoR (2015), `Radiotherapy board|intensity modulated radiotherapy (imrt) in the uk: current access and predictions of future access rates.', Society and College of Radiographers, Institute of Physics and Engineering in Medicine, Royal College of Radiologists . Taylor, R and Hassan, S and D'Souza, D. Scottish Executive (2006), Unintended overexposure of patient lisa norris during radiotherapy treatment at the beatson oncology centre, glasgow in january 2006, in `Report of an Investigation by the Inspector Appointed by the Scottish Ministers for the Ionising Radiation (Medical Exposures) Regulations 2000', Scottish Government. Seco, J. & Verhaegen, F. (2016), Monte Carlo Techniques in Radiation Therapy, Imaging in Medical Diagnosis and Therapy, CRC Press. Semenenko, V. A., Reitz, B., Day, E., Qi, X. S., Miften, M. & Li, X. A. (2008), `Evaluation of a commercial biologically based imrt treatment planning system', Medical Physics 35(12), 5851-5860. Seuntjens, J., Kawrakow, I., Borg, J., Hobeila, F. & Rogers, D. (2002), Calculated and measured air-kerma response of ionization chambers in low and medium energy photon beams, in `Recent Developments in Accurate Radiation Dosimetry: Proc. Int. Workshop', pp. 69-84. Sheikh-Bagheri, D. & Rogers, D. (2002a), `Monte carlo calculation of nine megavoltage photon beam spectra using the beam code', Medical physics 29(3), 391-402. Sheikh-Bagheri, D. & Rogers, D. (2002b), `Sensitivity of megavoltage photon beam monte carlo simulations to electron beam and other parameters', Medical physics 29(3), 379-390. Sheikh-Bagheri, D., Rogers, D., Ross, C. K. & Seuntjens, J. P. (2000), `Comparison of measured and monte carlo calculated dose distributions from the nrc linac', Medical physics 27(10), 2256-2266. Siantar, C. H., Walling, R., Daly, T., Faddegon, B., Albright, N., Bergstrom, P., Bielajew, A., Chuang, C., Garrett, D., House, R. et al. (2001), `Description and dosimetric verification of the peregrine monte carlo dose calculation system for photon beams incident on a water phantom', Medical physics 28(7), 1322-1337. Sikora, M. & Alber, M. (2009), `A virtual source model of electron contamination of a therapeutic photon beam', Physics in Medicine & Biology 54(24), 7329. Sikora, M., Dohm, O. & Alber, M. (2007), `A virtual photon source model of an elekta linear accelerator with integrated mini mlc for monte carlo based imrt dose calculation', Physics in Medicine & Biology 52(15), 4449. Smith, K., Balter, P., Duhon, J., White, G. A., Vassy, D. L., Miller, R. A., Serago, C. F. & Fairobent, L. A. (2017), `Aapm medical physics practice guideline 8. a.: linear accelerator performance tests', Journal of applied clinical medical physics 18(4), 23-39. Song, T., Zhou, L. & Jiang, S. (2012), `Su-e-t-496: Monte carlo simulation of a 6mv varian truebeam without flattening filter linac', Medical physics 39(6Part17), 3819-3819. Sontag, M. R. & Cunningham, J. R. (1978), `The equivalent tissue-air ratio method for making absorbed dose calculations in a heterogeneous medium', Radiology 129(3), 787-794. Spezi, E., Angelini, A., Romani, F. & Ferri, A. (2005), `Characterization of a 2d ion chamber array for the verification of radiotherapy treatments', Physics in Medicine & Biology 50(14), 3361. Stasi, M., Attili, A., Brusasco, C., Cirio, R., Denis, J., Donetti, M., Evans, P., Giordanengo, S., Giuliacci, A., Givehchi, N. et al. (2010), `Architecture and development of a detector for on-line verification in imrt.', IBA dosimetry . Stathakis, S., Myers, P., Esquivel, C., Mavroidis, P. & Papanikolaou, N. (2013), `Characterization of a novel 2d array dosimeter for patient-specific quality assurance with volumetric arc therapy', Medical physics 40(7). Stieler, F., Wolff, D., Schmid, H., Welzel, G., Wenz, F. & Lohr, F. (2011), `A comparison of several modulated radiotherapy techniques for head and neck cancer and dosimetric validation of vmat', Radiotherapy and Oncology 101(3), 388-393. Stock, M., Kroupa, B. & Georg, D. (2005), `Interpretation and evaluation of the γ index and the γ index angle for the verification of imrt hybrid plans', Physics in Medicine & Biology 50(3), 399. Storm, L. & Israel, H. I. (1970), `Photon cross sections from 1 kev to 100 mev for elements z= 1 to z= 100', Atomic Data and Nuclear Data Tables 7(6), 565-681. Thoelking, J., Fleckenstein, J., Sekar, Y., Boggula, R., Lohr, F., Wenz, F. & Wertz, H. (2016b), `Patient-specific online dose verification based on transmission detector measurements', Radiotherapy and Oncology 119(2), 351-356. Thoelking, J., Sekar, Y., Fleckenstein, J., Lohr, F., Wenz, F. & Wertz, H. (2016a), `Characterization of a new transmission detector for patient individualized online plan verification and its influence on 6mv x-ray beam characteristics', Zeitschrift für Medizinische Physik 26(3), 200-208. Thompson, C., Weston, S., Cosgrove, V. & Thwaites, D. (2014), `A dosimetric characterization of a novel linear accelerator collimator', Medical physics 41(3). Thwaites, D. (2013), Accuracy required and achievable in radiotherapy dosimetry: have modern technology and techniques changed our views?, in `Journal of Physics: Conference Series', Vol. 444, IOP Publishing, p. 012006. Thwaites, D. I., Mijnheer, B. & Mills, J. A. (2005), `Quality assurance of external beam radiotherapy', Radiation oncology physics: a handbook for teachers and students pp. 470-50. Toft, B. (2004), `Independent review of the circumstances surrounding a serious adverse incident that occurred in the REDACTED'. URL: http://www.who.int/patientsafety/news/RadiotherapyadverseeventToftreport.pdf Tuazon, B., Narayanasamy, G., Papanikolaou, N., Kirby, N., Mavroidis, P. & Stathakis, S. (2018), `Evaluation and comparison of second-check monitor unit calculation software with pinnacle 3 treatment planning system', Physica Medica 45, 186-191. Tyagi, N., Yang, K., Gersten, D. & Yan, D. (2012), `A real time dose monitoring and dose reconstruction tool for patient specific vmat qa and delivery', Medical physics 39(12), 7194-7204. Valdenaire, S., Mailleux, H. & Fau, P. (2016), `Modeling of flattening filter free photon beams with analytical and monte carlo tps', Biomedical Physics & Engineering Express 2(3), 035010. Van de Walle, J., Martens, C., Reynaert, N., Palmans, H., Coghe, M., De Neve, W., De Wagter, C. & Thierens, H. (2003), `Monte carlo model of the elekta sliplus accelerator: validation of a new mlc component module in beam for a 6 mv beam', Physics in Medicine & Biology 48(3), 371. Van Elmpt, W., McDermott, L., Nijsten, S.,Wendling, M., Lambin, P. & Mijnheer, B. (2008), `A literature review of electronic portal imaging for radiotherapy dosimetry', Radiotherapy and Oncology 88(3), 289-309. Van Esch, A., Clermont, C., Devillers, M., Iori, M. & Huyskens, D. P. (2007), `On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2d ion chamber array and the octavius phantom', Medical physics 34(10), 3825-3837. Van Herk, M. (2004), Errors and margins in radiotherapy, in `Seminars in radiation oncology', Vol. 14, Elsevier, pp. 52-64. Vanderstraeten, B., Chin, P. W., Fix, M., Leal, A., Mora, G., Reynaert, N., Seco, J., Soukup, M., Spezi, E., De Neve, W. et al. (2007), `Conversion of ct numbers into tissue parameters for monte carlo dose calculations: a multi-centre study', Physics in Medicine & Biology 52(3), 539. Vatnitsky, S., Lopez, P. O., Izewska, J., Meghzifene, A. & Levin, V. (2001), `The radiation overexposure of radiotherapy patients in panama 15 june 2001', Radiotherapy and Oncology 60(3), 237-238. Vazquez-Quino, L., Huerta-Hernandez, C. & Rangaraj, D. (2017), Clinical experience with mobius fx software for 3d dose verification for prostate vmat plans and comparison with physical measurements, in `Journal of Physics: Conference Series', Vol. 847, IOP Publishing, p. 012060. Venkataraman, S., Malkoske, K. E., Jensen, M., Nakonechny, K. D., Asuni, G. & McCurdy, B. M. (2009), `The influence of a novel transmission detector on 6 mv x-ray beam characteristics', Physics in Medicine & Biology 54(10), 3173. Venselaar, J., Welleweerd, H. & Mijnheer, B. (2001), `Tolerances for the accuracy of photon beam dose calculations of treatment planning systems', Radiotherapy and Oncology 60(2), 191-201. Verhaegen, F. & Seuntjens, J. (2003), `Monte carlo modelling of external radiotherapy photon beams', Physics in medicine & biology 48(21), R107. Verhaegen, F., Symonds-Tayler, R., Liu, H. & Nahum, A. (2000), `Backscatter towards the monitor ion chamber in high-energy photon and electron beams: charge integration versus monte carlo simulation', Physics in Medicine & Biology 45(11), 3159. Visser, R., Wauben, D., De Groot, M., Godart, J., Langendijk, J., van't Veld, A. & Korevaar, E. (2013), `Efficient and reliable 3d dose quality assurance for imrt by combining independent dose calculations with measurements', Medical physics 40(2). Wack, G., Lalande, F. & Seligman, M. (2007), `Summary of asn on the epinal radiotherapy accident', French Nuclear Safety Authority . Waldron, T. (2002), The theory and operation of computer-controlled medical linear accelerator, in `MEDICAL PHYSICS', Vol. 29, AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS STE 1 NO 1, 2 HUNTINGTON QUADRANGLE, MELVILLE, NY 11747-4502 USA, pp. 1277-1277. Walters, B. (2016), `Discussion concerning the observed higher dose in small voxels'. Walters, B., Kawrakow, I. & Rogers, D. (2002), `History by history statistical estimators in the beam code system', Medical Physics 29(12), 2745-2752. Walters, B. & Rogers, D. (2003), DOSXYZnrc users manual, Report, National Research Council Canada. Webb, S. (2003), `The physical basis of imrt and inverse planning', The British journal of radiology 76(910), 678-689. WHO (2008), `Radiotherapy risk profile'. World Health Organization. Williams, M. & McKenzie, A. (2008), `Can we afford not to implement in vivo dosimetry?', The British journal of radiology 81(969), 681-684. Wu, H., Jiang, F., Yue, H., Hu, Q., Zhang, J., Liu, Z., Gong, J., Li, S., Geng, J. & Zhang, Y. (2016), `A comparative study of identical vmat plans with and without jaw tracking technique', Journal of applied clinical medical physics 17(5), 133-141. Wulff, J., Heverhagen, J. & Zink, K. (2008), `Monte-carlo-based perturbation and beam quality correction factors for thimble ionization chambers in high-energy photon beams', Physics in Medicine & Biology 53(11), 2823. Zhen, H., Nelms, B. E. & Tomé, W. A. (2011), `Moving from gamma passing rates to patient dvh-based qa metrics in pretreatment dose qa', Medical physics 38(10), 5477-5489. Zhen, H., Nelms, B. E. & Tomé, W. A. (2013), `On the use of biomathematical models in patient-specific imrt dose qa', Medical physics 40(7)."]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["doctoral"]},{"key":"dc:type.qualificationname","label":"Dc Type Qualificationname","values":["Ph.D"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.dcterms_iso639_2","label":"Language (ISO 639-2)","values":["eng"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["uk.bl.ethos.797960"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://etheses.whiterose.ac.uk/id/eprint/24417/1/DJ_PhD_Thesis_Final.pdf"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["The goal of this thesis was to characterise a novel transmission detector in the context of signal prediction. This was to eliminate the need to collect a baseline signal for the device before treatment. This not only saves time, but, by independently generating the baseline signal, the process is less prone to missing errors. A simple analytical algorithm was designed and was found to be capable of detecting gross errors, however, it was shown not to be accurate enough to detect MLC position errors that could have a clinical effect on the delivery. MU check software was commissioned, however the fluence distribution it produced lacked the complexity for accurate signal prediction. A Monte Carlo model of a linac was built and validated then used to demonstrate that the detector could be modelled as two slabs of Perspex; the signal being proportional to the dose measured in the air between them. Two Monte Carlo models were then made using different systems, these were both evaluated by comparing predicted signals to measured signals for VMAT plans. Both models performed well and were capable of detecting leaf errors ~1mm; the merits of both are discussed with regard to error detection and ease of use."]},{"key":"dc:format","label":"Dc Format","values":["text"]},{"key":"dcterms:accessrights","label":"Dcterms Accessrights","values":["public"]},{"key":"dc:title","label":"Title","values":["Characterisation and evaluation of a novel transmission detector for intra-fraction monitoring of radiotherapy."]}]}],"canonical_facts":{"dc:contributor.advisor":["Thwaites, D","Buckley, D","Cosgrove, V","Weston, S"],"dc:creator":["Johnson, Daniel"],"dc:date":["2018-11"],"dc:date.issued":["2018-11"],"dc:description.abstract":["The goal of this thesis was to characterise a novel transmission detector in the context of signal prediction. This was to eliminate the need to collect a baseline signal for the device before treatment. This not only saves time, but, by independently generating the baseline signal, the process is less prone to missing errors. A simple analytical algorithm was designed and was found to be capable of detecting gross errors, however, it was shown not to be accurate enough to detect MLC position errors that could have a clinical effect on the delivery. MU check software was commissioned, however the fluence distribution it produced lacked the complexity for accurate signal prediction. A Monte Carlo model of a linac was built and validated then used to demonstrate that the detector could be modelled as two slabs of Perspex; the signal being proportional to the dose measured in the air between them. Two Monte Carlo models were then made using different systems, these were both evaluated by comparing predicted signals to measured signals for VMAT plans. Both models performed well and were capable of detecting leaf errors ~1mm; the merits of both are discussed with regard to error detection and ease of use."],"dc:format":["text"],"dc:identifier":["uk.bl.ethos.797960"],"dc:identifier.uri":["https://etheses.whiterose.ac.uk/id/eprint/24417/1/DJ_PhD_Thesis_Final.pdf"],"dc:language.dcterms_iso639_2":["eng"],"dc:publisher.commercial":["University of Leeds"],"dc:publisher.department":["Academic Unit of Medical Physics (Leeds)"],"dc:publisher.institution":["University of Leeds"],"dc:relation.isreferencedby":["https://etheses.whiterose.ac.uk/id/eprint/24417/"],"dc:relation.references":["AAMP (2008), `Task group 106 report: Accelerator beam data commissioning equipment and procedures', Medical Physics 35(9), 4186-4215. Das, Indra J. and Cheng, Chee-Wai and Watts, Ronald J. and Ahnesjö, Anders and Gibbons, John and Li, X. Allen and Lowenstein, Jessica and Mitra, Raj K. and Simon, William E. and Zhu, Timothy C. AAPM (2001), `Task group 50 report: Basic applications of multileaf collimators'. Boyer, Arthur et al. AAPM (2005), `Task group 87 report: Diode in vivo dosimetry for patients receiving external beam radiation therapy'. Yorke, Ellen and Alecu, R and Ding, L and Fontenla, D and Kalend, A and Kaurin, D and Masterson-McGary, ME and Marinello, G and Matzen, T and Saini, A and others. AAPM (2007), `Task group 105 report: Issues associated with clinical implementation of monte carlo-based photon and electron external beam treatment planning', Medical Physics 34(12). Chetty, Indrin J. and Curran, Bruce and Cygler, Joanna E. and DeMarco, John J. and Ezzell, Gary and Faddegon, Bruce A. and Kawrakow, Iwan and Keall, Paul J. and Liu, Helen and Ma, C.-M. Charlie and Rogers, D. W. O. and Seuntjens, Jan and Sheikh-Bagheri, Daryoush and Siebers, Jeffrey V. AAPM (2009), `Task group 142 report: Quality assurance of medical accelerators a', Medical physics 36, 4197-4212. Klein, Eric E and Hanley, Joseph and Bayouth, John and Yin, Fang-Fang and Simon, William and Dresser, Sean and Serago, Christopher and Aguirre, Francisco and Ma, Lijun and Arjomandy, Bijan. AAPM (2011), `Task group 114 report: Verification of monitor unit calculations for non-imrt clinical radiotherapy', Medical Physics 38(1), 504-530. Stern, Robin L. and Heaton, Robert and Fraser, Martin W. and Murty Goddu, S. and Kirby, Thomas H. and Lam, Kwok Leung and Molineu, Andrea and Zhu, Timothy C. AAPM (2014), `Task group 71 report: Monitor unit calculations for external photon and electron beams', Medical physics 41(3). Gibbons, John P and Antolak, John A and Followill, David S and Huq, M Saiful and Klein, Eric E and Lam, Kwok L and Palta, Jatinder R and Roback, Donald M and Reid, Mark and Khan, Faiz M. AAPM (2018), `Task group 218 report: Tolerance limits and methodologies for imrt measurement-based verification qa', Medical physics . Miften, Moyed and Olch, Arthur and Mihailidis, Dimitris and Moran, Jean and Pawlicki, Todd and Molineu, Andrea and Li, Harold and Wijesooriya, Krishni and Shi, Jie and Xia, Ping and others. Adamus-Górka, M., Mavroidis, P., Lind, B. K. & Brahme, A. (2011), `Comparison of dose response models for predicting normal tissue complications from cancer radiotherapy: application in rat spinal cord', Cancers 3(2), 2421-2443. Agnew, A., Agnew, C., Grattan, M., Hounsell, A. & McGarry, C. (2014), `Monitoring daily mlc positional errors using trajectory log files and epid measurements for imrt and vmat deliveries', Physics in Medicine & Biology 59(9), N49. Agostinelli, S., Allison, J., Amako, K. a., Apostolakis, J., Araujo, H., Arce, P., Asai, M., Axen, D., Banerjee, S., Barrand, G. . et al. (2003), `Geant4|a simulation toolkit', Nuclear instruments and methods in physics research section A: Accelerators, Spectrometers, Detectors and Associated Equipment 506, 250-303. Ahnesjö, A. (1989), `Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media', Medical physics 16(4), 577-592. Alber, M., Broggi, S., De Wagter, C., Eichwurzel, I., Engström, P., Fiorino, C., Georg, D., Hartmann, G., Knöös, T., Leal, A. et al. (2008), `Guidelines for the verification of imrt', ESTRO booklet 7. Alber, M. & Nüsslin, F. (1999), `An objective function for radiation treatment optimization based on local biological measures', Physics in Medicine & Biology 44(2), 479. Ash, D. (2007), `Lessons from epinal'. Ash, D. & Bates, T. (1994), `Report on the clinical effects of inadvertent radiation underdosage in 1045 patients', Clinical Oncology 6(4), 214-226. Asnaashari, K., Chow, J. C. & Heydarian, M. (2013), `Dosimetric comparison between two mlc systems commonly used for stereotactic radiosurgery and radiotherapy: a monte carlo and experimental study', Physica medica 29(4), 350-356. Asuni, G., Jensen, J. M. & McCurdy, B. M. C. (2011), `A monte carlo investigation of contaminant electrons due to a novel in vivo transmission detector', Physics in Medicine and Biology 56(4), 1207. Azorín, Nieto, J. (2004), Thermoluminescence dosimetry (tld) and its application in medical physics, in `AIP Conference Proceedings', Vol. 724, AIP, pp. 20-27. Bai, S., Li, G., Wang, M., Jiang, Q., Zhang, Y. & Wei, Y. (2013), `Effect of mlc leaf position, collimator rotation angle, and gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma', Medical Dosimetry 38(2), 143-147. Baker, S., Dahele, M., Lagerwaard, F. J. & Senan, S. (2016), `A critical review of recent developments in radiotherapy for non-small cell lung cancer', Radiation oncology 11(1), 115. Bielajew, A. F., Mohan, R. & Chui, C.-S. (1989), `Improved bremsstrahlung photon angular sampling in the egs4 code system', National Research Council of Canada Report PIRS-0203 . Bielajew, A. F. & Rogers, D. (1986), `Presta: the parameter reduced electron-step transport algorithm for electron monte carlo transport', Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 18(1-6), 165-181. BIPM (2006), `The international system of units (si)', Bureau International des Poids et Mesures, . BIR (2007), `Towards safer radiotherapy'. British Institute of Radiology, Institute of Physics and Engineering in Medicine, National Patient Safety Agency, Society and College of Radiographers, The Royal College of Radiologists. Boesecke, R., Doll, J., Bauer, B., Schlegel, W., Pastyr, O. & Lorenz, W. (1988), `Treatment planning for conformation therapy using a multi-leaf collimator.', Strahlentherapie und Onkologie: Organ der Deutschen Rontgengesellschaft...[et al] 164(3), 151-154. Bogdanich, W. (2010), `Radiation offers new cures, and ways to do harm', The New York Times 23. Bohsung, J., Strahlentherapie, K. f., Berlin, C. & Mitte, S. (2004), `The 2d-array seven29 a new way of dosimetric verification of imrt beams', Application note . Bortfeld, T. (2006), `Imrt: a review and preview', Physics in Medicine & Biology 51(13), R363. Brahme, A. (1984), `Dosimetric precision requirements in radiation therapy', Acta Radiologica: Oncology 23(5), 379-391. Brahme, A. (1987), `Design principles and clinical possibilities with a new generation of radiation therapy equipment: A review', Acta Oncologica 26(6), 403-412. Brahme, A. (1988a), `Optimal setting of multileaf collimators in stationary beam radiation therapy.', Strahlentherapie und Onkologie 164(6), 343-350. Brahme, A. (1988b), `Optimization of stationary and moving beam radiation therapy techniques', Radiotherapy and Oncology 12(2), 129-140. Brahme, A., Roos, J.-E. & Lax, I. (1982), `Solution of an integral equation encountered in rotation therapy', Physics in Medicine & Biology 27(10), 1221. Briesmeister, J. F. et al. (2000), `Mcnptm-a general monte carlo n-particle transport code', Version 4C, LA-13709-M, Los Alamos National Laboratory . British Standards Institution (2000), Transition to ISO 9001: 2000: Guidance on ISO 9001: 2000, Quality Management Systems, Requirements, British Standards Institution. MacNee, Colin and Pyle, Jim and Sutherland, Sandy. Buckley, L. A. & Rogers, D. (2006), `Wall correction factors, pwall, for thimble ionization chambers', Medical physics 33(2), 455-464. Budgell, G., Brown, K., Cashmore, J., Duane, S., Frame, J., Hardy, M., Paynter, D. & Thomas, R. (2016), `IPEM topical report 1: guidance on implementing flattening filter free (FFF) radiotherapy', Physics in Medicine & Biology 61(23), 8360. Budgell, G. J., Mott, J., Williams, P. C. & Brown, K. (2000), `Requirements for leaf position accuracy for dynamic multileaf collimation', Physics in Medicine & Biology 45(5), 1211. Caivano, R., Califano, G., Fiorentino, A., Cozzolino, M., Oliviero, C., Pedicini, P., Clemente, S., Chiumento, C. & Fusco, V. (2014), `Clinically relevant quality assurance for intensity modulated radiotherapy plans: gamma maps and dvh-based evaluation', Cancer investigation 32(3), 85-91. Cancer Research (2019), `Fund gives around 5,800 more cancer patients access to advanced radiotherapy treatment every year | cancer research uk', https://www.cancerresearchuk.org/about-us/cancer-news/press-release/2013-07-11-fund-gives-around-5800-more-cancer-patients-access/-to-advanced-radiotherapy-treatment-every-year#2. (Accessed on 02/08/2019). Celi, S., Costa, E., Wessels, C., Mazal, A., Fourquet, A. & Francois, P. (2016), `Epid based in vivo dosimetry system: clinical experience and results', Journal of applied clinical medical physics 17(3), 262-276. Chalmers, A. J. (2016), `Science in focus: combining radiotherapy with inhibitors of the dna damage response', Clinical Oncology 28(5), 279-282. Chandraraj, V., Stathakis, S., Manickam, R., Esquivel, C., Supe, S. S. & Papanikolaou, N. (2010), `Consistency and reproducibility of the vmat plan delivery using three independent validation methods', Journal of Applied Clinical Medical Physics 12(1), 129-140. Chang, J. Y. (2015), Intensity-modulated radiotherapy, not 3 dimensional conformal, is the preferred technique for treating locally advanced lung cancer, in `Seminars in radiation oncology', Vol. 25, Elsevier, pp. 110-116. Cheng, J. Y., Ning, H., Arora, B. C., Zhuge, Y. & Miller, R. W. (2016), `Output factor comparison of monte carlo and measurement for varian truebeam 6 mv and 10 mv flattening filter-free stereotactic radiosurgery system', Journal of applied clinical medical physics 17(3), 100-110. Chibani, O., Moftah, B. & Ma, C.-M. C. (2011), `On monte carlo modeling of megavoltage photon beams: A revisited study on the sensitivity of beam parameters', Medical physics 38(1), 188-201. Childress, N., Chen, Q. & Rong, Y. (2015), `Parallel/opposed: Imrt qa using treatment log files is superior to conventional measurement-based method', Journal of applied clinical medical physics 16(1), 4-7. Childress, N., Stephens, E., Eklund, D. & Zhang, M. (2012), `Mobius3d white paper: dose calculation algorithm', Houston, TX: Mobius Medical Systems . Cho, B. (2018), `Intensity-modulated radiation therapy: a review with a physics perspective', Radiation oncology journal 36(1), 1. Convery, D. & Rosenbloom, M. (1992), `The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimation', Physics in medicine & biology 37(6), 1359. CRUK (2015), `Cancer statistics for the uk'. (Accessed on 07/03/2018). URL: https://www.cancerresearchuk.org/health-professional/cancer-statisticsfor-the-uk CRUK, N. (2014), `Vision for radiotherapy 2014-24.'. (Accessed on 01/08/2018). URL: https://www.cancerresearchuk.org Cunningham, J., Shrivastava, P. & Wilkinson, J. (1972), `Program irreg| calculation of dose from irregularly shaped radiation beams', Computer Programs in Biomedicine 2(3), 192-199. Dale, R. G., Jones, B. et al. (2007), Radiobiological modelling in radiation oncology, British Institute of Radiology. Das, I. J., Ding, G. X. & Ahnesjö, A. (2008), `Small fields: Nonequilibrium radiation dosimetry', Medical Physics 35(1), 206-215. De Vlamynck, K., Palmans, H., Verhaegen, F., De Wagter, C., De Neve, W. & Thierens, H. (1999), `Dose measurements compared with monte carlo simulations of narrow 6 mv multileaf collimator shaped photon beams', Medical physics 26(9), 1874-1882. Delaby, N., Bouvier, J., Jouyaux, F., Barateau, A. & Lafond, C. (2017), `40. validation of a transit epid device for a clinical use: application to iviewdose (elekta)', Physica Medica: European Journal of Medical Physics 44, 19-20. Department of Health, Donaldson, L. (2006), `On the state of public health', 2006 Annual Report of The Chief Medical Officer. Donaldson, L. (2007), `Reducing harm from radiotherapy', BMJ: British Medical Journal 334(7588), 272. Edwards, C., Hamer, E., Mountford, P. & Moloney, A. (2007), `An update survey of uk in vivo radiotherapy dosimetry practice', The British journal of radiology 80(960), 1011-1014. Edwards, C. & Mountford, P. (2009), `Characteristics of in vivo radiotherapy dosimetry', The British journal of radiology 82(983), 881-883. Ezzell, G. A., Galvin, J. M., Low, D., Palta, J. R., Rosen, I., Sharpe, M. B., Xia, P., Xiao, Y., Xing, L. & Yu, C. X. (2003), `Guidance document on delivery, treatment planning, and clinical implementation of imrt: report of the imrt subcommittee of the aapm radiation therapy committee', Medical physics 30(8), 2089-2115. Farkona, S., Diamandis, E. P. & Blasutig, I. M. (2016), `Cancer immunotherapy: the beginning of the end of cancer?', BMC medicine 14(1), 73. Feldman, A., Edwards, F. M. & Hendee, W. R. (2001), `The routine use of personal patient dosimeters is of little value in detecting therapeutic misadministrations', Medical physics 28(3), 295-297. Feller, W. (2008), An introduction to probability theory and its applications, Vol. 2, John Wiley & Sons. Feng, Z., Wu, H., Zhang, Y., Zhang, Y., Cheng, J. & Su, X. (2015), `Dosimetric comparison between jaw tracking and static jaw techniques in intensity-modulated radiotherapy', Radiation Oncology 10(1), 28. Feygelman, V., Opp, D., Javedan, K., Saini, A. & Zhang, G. (2010), `Evaluation of a 3d diode array dosimeter for helical tomotherapy delivery qa', Medical Dosimetry 35(4), 324-329. Fippel, M. (1999), `Fast monte carlo dose calculation for photon beams based on the vmc electron algorithm', Medical Physics 26(8), 1466-1475. Fippel, M., Haryanto, F., Dohm, O., Nüsslin, F. & Kriesen, S. (2003), `A virtual photon energy fluence model for monte carlo dose calculation', Medical Physics 30(3), 301-311. Fippel, M., Laub, W., Huber, B. & Nüsslin, F. (1999), `Experimental investigation of a fast monte carlo photon beam dose calculation algorithm', Physics in Medicine & Biology 44(12), 3039. Fisher, C. M. & Rabinovitch, R. (2014), `Frontiers in radiotherapy for early-stage invasive breast cancer', Journal of Clinical Oncology 32(26), 2894. Fix, M. K., Keall, P. J., Dawson, K. & Siebers, J. V. (2004), `Monte carlo source model for photon beam radiotherapy: photon source characteristics', Medical physics 31(11), 3106-3121. Fleckenstein, J., Jahnke, L., Lohr, F., Wenz, F. & Hesser, J. (2013), `Development of a geant4 based monte carlo algorithm to evaluate the monaco vmat treatment accuracy', Zeitschrift für Medizinische Physik 23(1), 33-45. Fontenot, J. D. (2014), `Evaluation of a novel secondary check tool for intensitymodulated radiotherapy treatment planning', Journal of applied clinical medical physics 15(5), 207-215. Fraass, B., Doppke, K., Hunt, M., Kutcher, G., Starkschall, G., Stern, R. & Van Dyke, J. (1998), `Task group 53 report: Quality assurance for clinical radiotherapy treatment planning', Medical Physics 25(10), 1773-1829. Fragoso, M., Kawrakow, I., Faddegon, B. A., Solberg, T. D. & Chetty, I. J. (2009), `Fast, accurate photon beam accelerator modeling using beamnrc: A systematic investigation of efficiency enhancing methods and cross-section data', Medical physics 36(12), 5451-5466. Franks, K. & McNair, H. (2012), `Implementation of image-guided radiotherapy', Clinical Oncology 24(9), 625-626. Fredh, A., Scherman, J. B., Fog, L. S. & Munck af Rosenschöld, P. (2013), `Patient qa systems for rotational radiation therapy: a comparative experimental study with intentional errors', Medical physics 40(3), 031716. Galvin, J. M., Smith, A. R. & Lally, B. (1993), `Characterization of a multileaf collimator system', International Journal of Radiation Oncology Biology Physics 25(2), 181-192. Georg, D., Knöös, T. & McClean, B. (2011), `Current status and future perspective of flattening filter free photon beams', Medical physics 38(3), 1280-1293. Gholampourkashi, S., Cygler, J. E., Belec, J., Vujicic, M. & Heath, E. (2019), `Monte carlo and analytic modeling of an elekta infinity linac with agility mlc: Investigating the significance of accurate model parameters for small radiation fields', Journal of Applied Clinical Medical Physics 20(1), 55-67. Godart, J., Korevaar, E., Visser, R., Wauben, D. & van't Veld, A. A. (2011), `Reconstruction of high-resolution 3d dose from matrix measurements: error detection capability of the compass correction kernel method', Physics in Medicine & Biology 56(15), 5029. Greene, D. D. & Williams, P. C. (1997), Linear accelerators for radiation therapy, 2nd ed edn, Bristol : Institute of Physics Pub. Previous ed.: 1985. Greer, P. B., Cadman, P., Lee, C. & Bzdusek, K. (2009), `An energy fluence-convolution model for amorphous silicon epid dose prediction', Medical Physics 36(2), 547-555. Griessbach, I., Lapp, M., Bohsung, J., Gademann, G. & Harder, D. (2005), `Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams', Medical Physics 32(12), 3750-3754. Grofsmid, D., Dirkx, M., Marijnissen, H., Woudstra, E. & Heijmen, B. (2010), `Dosimetric validation of a commercial monte carlo based imrt planning system', Medical physics 37(2), 540-549. Harrison, R. & Morgan, A. (2007), `In vivo dosimetry: hidden dangers?', The British journal of radiology 80(957), 691-692. Heath, E. & Seuntjens, J. (2003), `Development and validation of a beamnrc component module for accurate monte carlo modelling of the varian dynamic millennium multileaf collimator', Physics in Medicine & Biology 48(24), 4045. Heilemann, G., Poppe, B. & Laub, W. (2013), `On the sensitivity of common gamma-index evaluation methods to mlc misalignments in rapidarc quality assurance', Medical physics 40(3). Herwiningsih, S. & Fielding, A. (2016), Focal spot estimation of an elekta dedicated stereotactic linear accelerator monte carlo model, in `Journal of Physics: Conference Series', Vol. 694, IOP Publishing, p. 012013. Higgins, P., Alaei, P., Gerbi, B. & Dusenbery, K. (2003), `In vivo diode dosimetry for routine quality assurance in imrt', Medical physics 30(12), 3118-3123. Hounsell, A. R. (1998), `Monitor chamber backscatter for intensity modulated radiation therapy using multileaf collimators', Physics in Medicine & Biology 43(2), 445. Huq, M. S., Das, I. J., Steinberg, T. & Galvin, J. M. (2002), `A dosimetric comparison of various multileaf collimators', Physics in Medicine & Biology 47(12), N159. Hussein, M., Adams, E. J., Jordan, T. J., Clark, C. H. & Nisbet, A. (2013a), `A critical evaluation of the ptw 2d-array seven29 and octavius ii phantom for imrt and vmat verification', Journal of applied clinical medical physics 14(6), 274-292. Hussein, M., Clark, C. & Nisbet, A. (2017), `Challenges in calculation of the gamma index in radiotherapy - towards good practice', Physica Medica 36, 1-11. Hussein, M., Rowshanfarzad, P., Ebert, M. A., Nisbet, A. & Clark, C. H. (2013b), `A comparison of the gamma index analysis in various commercial imrt/vmat qa systems', Radiotherapy and Oncology 109(3), 370-376. Huyskens, D., Bogaerts, R., Verstraete, J., Lööf, M., Nyström, H., Fiorino, C., Broggi, S., Jornet, N., Ribas, M. & Thwaites, D. (2001), `Practical guidelines for the implementation of in vivo dosimetry with diodes in external radiotherapy with photon beams (entrance dose)'. IAEA (2013), Development of procedures for in vivo dosimetry in radiotherapy., Series IA, International Atomic Energy Agency, Vienna. ICRP (2001), `Prevention of accidents to patients undergoing radiation therapy', International Commission on Radiological Protection 86 30(3). ICRP (2009), `Preventing accidental exposures from new external beam radiation therapy technologies', International Commission on Radiological Protection Publication 112 39(4). P. Ortiz Lopez, J.-M. Cosset, P. Dunscombe, O. Holmberg, J.-C. Rosenwald, L. Pinillos Ashton, J.J. Vilaragut Llanes, S. Vatnitsky. ICRU (1985), `Stopping powers for electrons and positrons', ICRU report 37. Brice, David K. ICRU (1992), `Photon, electron, proton and neutron interaction data for body tissues', ICRU report 46. Bethesda, MD. ICRU (1994), `Prescribing, recording and reporting photon beam therapy', International Commission on Radiation Units and Measurements report 50. Jones, Douglas. ICRU (1999), `Prescribing, recording and reporting photon beam therapy (supplement to icru report 50)', International Commission on Radiation Units and Measurements report 62. T. Landberg J. Chavaudra J. Dobbs J. -P. Gerard G. Hanks J. -C. Horiot K. -A. Johansson T. Möller J. Purdy N. Suntharalingam H. Svensson. IPEM (2006), `Balancing costs and benefits of checking in radiotherapy', Institute of Physics and Engineering in Medicine . McKenzie, Alan. IPEM (2008), `Guidance for the clinical implementation of intensity modulated radiation therapy', Institute of Physics and Engineering in Medicine (IPEM) Report 96. James, H and Beavis, A and Budgell, G and Clark, C and Convery, D and Mott, J and others. IPEM (2018), `Physics aspects of quality control in radiotherapy second edition', Institute of Physics and Engineering in Medicine report 81. Patel, Weston S, Palmer A and others. Islam, M. K., Norrlinger, B. D., Smale, J. R., Heaton, R. K., Galbraith, D., Fan, C. & Jaffray, D. A. (2009), `An integral quality monitoring system for real-time verification of intensity modulated radiation therapy', Medical physics 36(12), 5420-5428. Izewska, J. & Andreo, P. (2000), `The iaea/who tld postal programme for radiotherapy hospitals', Radiotherapy and oncology 54(1), 65-72. Johnson, D., Weston, S. J., Cosgrove, V. P. & Thwaites, D. I. (2014), `A simple model for predicting the signal for a head-mounted transmission chamber system, allowing imrt in-vivo dosimetry without pretreatment linac time', Journal of Applied Clinical Medical Physics 15(4), 270-279. Joiner, M. C. & Van der Kogel, A. (2016), Basic clinical radiobiology, Vol. 2, CRC press. Jones, L., Hoban, P. & Metcalfe, P. (2001), `The use of the linear quadratic model in radiotherapy: a review', Australasian Physics & Engineering Sciences in Medicine 24(3), 132-146. Jordan, T. J. & Williams, P. C. (1994), `The design and performance characteristics of a multileaf collimator', Physics in Medicine & Biology 39(2), 231. Jordan, T. & Williams, P. (1991), Commissioning and use of a multileaf collimator system, in `Proceedings of the 1st biennial ESTRO Meeting on Physics in Clinical Radiotherapy', p. 3. Juste, B., Miro, R., Campayo, J., Diez, S. & Verdu, G. (2008), Photon spectra calculation for an elekta linac beam using experimental scatter measurements and monte carlo techniques., in `Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE', IEEE, pp. 3289-3292. Kairn, T., Crowe, S., Poole, C. & Fielding, A. L. (2009), `Effects of collimator backscatter in an elekta linac by monte carlo simulation', Australasian Physics & Engineering Sciences in Medicine 32(3), 129. Kajaria, A., Sharma, N., Sharma, S., Pradhan, S., Mandal, A. & Aggarwal, L. (2017), `Monte carlo study of a flattening filter-free 6 mv photon beam using the beamnrc code'. Karagoz, G., Zorlu, F., Yeğiner, M., Yıldız, D. & Gürkaynak, M. (2014), `Evaluation of mlc leaf position accuracy for static and dynamic imrt treatments using david in vivo dosimetric system', International Journal of Radiation Oncology* Biology*Physics 90(1). Kawrakow, I., Fippel, M. & Friedrich, K. (1996), `3d electron dose calculation using a voxel based monte carlo algorithm (vmc)', Medical physics 23(4), 445-457. Kawrakow, I., Mainegra-Hing, E., Rogers, D. W. O., Tessier, F. & Walters, B. (2000), The egsnrc code system: Monte carlo simulation of electron and photon transport, Technical report, National Research Council Canada. Kawrakow, I., Rogers, D. & Walters, B. (2004), `Large efficiency improvements in beamnrc using directional bremsstrahlung splitting', Medical physics 31(10), 2883-2898. Khan, F. M. (2010), The physics of radiation therapy, Walters Kluwer, Lippincott Williams & Wilkins. Kim, J.-i., Park, S.-Y., Kim, H. J., Kim, J. H., Ye, S.-J. & Park, J. M. (2014), `The sensitivity of gamma-index method to the positioning errors of high-definition mlc in patient-specific vmat qa for sbrt', Radiation Oncology 9(1), 167. Kinsella, P., Shields, L., McCavana, P., McClean, B. & Langan, B. (2016), `Determination of mlc model parameters for monaco using commercial diode arrays', Journal of applied clinical medical physics 17(4), 37-47. Klein, E. E., Drzymala, R. E., Purdy, J. A. & Michalski, J. (2005), `Errors in radiation oncology: a study in pathways and dosimetric impact', Journal of Applied clinical medical physics 6(3), 81-94. Kung, J., Chen, G. & Kuchnir, F. (2000), `A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assurance', Medical physics 27(10), 2226-2230. Künzler, T., Fotina, I., Stock, M. & Georg, D. (2009), `Experimental verification of a commercial monte carlo-based dose calculation module for high-energy photon beams', Physics in Medicine & Biology 54(24), 7363. Laub, W. U., Kaulich, T. W. & Nüsslin, F. (1999), `A diamond detector in the dosimetry of high-energy electron and photon beams', Physics in Medicine & Biology 44(9), 2183. Lawless, C., Landau, D., Faivre-Finn, C., Boyd, K., Lester, J., Fenwick, J., Maguire, J., McCartney, E., Paul, J., Parsons, E. et al. (2017), `Adscan: a randomised phase ii study of accelerated, dose escalated, sequential chemoradiotherapy in non-small cell lung cancer (nsclc)', Lung Cancer 103, S80-S81. Lea, D. & Catcheside, D. (1942), `The mechanism of the induction by radiation of chromosome aberrations intradescantia', Journal of genetics 44(2-3), 216-245. Leeds University (2018), `Marc1 - advanced research computing', http://arc.leeds.ac.uk/systems/marc1/. (Accessed on 08/14/2018). Leer, J.-W. H., McKenzie, A., Scalliet, P. & Thwaites, D. I. (1998), Practical Guidelines for the Implementation of a Quality System in Radiotherapy: A Project of the ESTRO Quality Assurance Committee Sponsored by\" Europe Against Cancer\", ESTRO. Lillicrap, S. C., Owen, B., Williams, J. R. & Williams, P. C. (1990), `Code of practice for high-energy photon therapy dosimetry based on the npl absorbed dose calibration service', Physics in Medicine and Biology 35(10), 1355-1360. Ling, C. C., Zhang, P., Archambault, Y., Bocanek, J., Tang, G. & LoSasso, T. (2008), `Commissioning and quality assurance of rapidarc radiotherapy delivery system', International Journal of Radiation Oncology* Biology* Physics 72(2), 575-581. Litzenberg, D. W., Moran, J. M. & Fraass, B. A. (2002), `Verification of dynamic and segmental imrt delivery by dynamic log file analysis', Journal of applied clinical medical physics 3(2), 63-72. Liu, H. H., Mackie, T. R. & McCullough, E. C. (2000), `Modeling photon output caused by backscattered radiation into the monitor chamber from collimator jaws using a monte carlo technique', Medical physics 27(4), 737-744. Liu, H. H., Verhaegen, F. & Dong, L. (2001), `A method of simulating dynamic multileaf collimators using monte carlo techniques for intensity-modulated radiation therapy', Physics in Medicine & Biology 46(9), 2283. Lobo, J. & Popescu, I. A. (2010), `Two new dosxyznrc sources for 4d monte carlo simulations of continuously variable beam configurations, with applications to rapidarc, vmat, tomotherapy and cyberknife', Physics in Medicine and Biology 55(16), 4431-4443. Looe, H. K., Harder, D., Rühmann, A., Willborn, K. C. & Poppe, B. (2010), `Enhanced accuracy of the permanent surveillance of imrt deliveries by iterative deconvolution of david chamber signal profiles', Physics in Medicine and Biology 55(14), 3981-3992. LoSasso, T., Chui, C.-S. & Ling, C. C. (1998), `Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy', Medical physics 25(10), 1919-1927. LoSasso, T., Chui, C.-S. & Ling, C. C. (2001), `Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode', Medical physics 28(11), 2209-2219. Low, D. A., Harms, W. B., Mutic, S. & Purdy, J. A. (1998), `A technique for the quantitative evaluation of dose distributions', Medical physics 25(5), 656-661. Lyman, J. T. (1985), `Complication probability as assessed from dose-volume histograms', Radiation Research 104(2s), S13-S19. Ma, C.-m. & Nahum, A. (1993), `Effect of size and composition of the central electrode on the response of cylindrical ionization chambers in high-energy photon and electron beams', Physics in Medicine & Biology 38(2), 267. Ma, C. M. & Rogers, D. W. O. (2018), `Beamdp users manual'. National Research Council of Canada. MacDougall, N. D., Graveling, M., Hansen, V. N., Brownsword, K. & Morgan, A. (2017), `In vivo dosimetry in uk external beam radiotherapy: current and future usage', The British journal of radiology 90(1072), 20160915. Mackay, R. I. & Williams, P. C. (2009), `The cost effectiveness of in vivo dosimetry is not proven', The British journal of radiology 82(976), 265-266. Mani, K. R., Upadhayay, S. & Das, K. M. (2017), `Influence of jaw tracking in intensity-modulated and volumetric-modulated arc radiotherapy for head and neck cancers: a dosimetric study', Radiation oncology journal 35(1), 90. Mans, A., Wendling, M., McDermott, L., Sonke, J.-J., Tielenburg, R., Vijlbrief, R., Mijnheer, B., Van Herk, M. & Stroom, J. (2010), `Catching errors with in vivo epid dosimetry', Medical physics 37, 2638-2644. Mayles, P., Nahum, A. & Rosenwald, J.-C. (2007), Handbook of radiotherapy physics: theory and practice, CRC Press. McDonald, D. G., Jacqmin, D. J., Mart, C. J., Koch, N. C., Peng, J. L., Ashenafi, M. S., Fugal, M. A. & Vanek, K. N. (2017), `Validation of a modern secondcheck dosimetry system using a novel verification phantom', Journal of applied clinical medical physics 18(1), 170-177. McGowan & Faddegon (2013), `Statdose for 3d dose distributions'. National Research Council of Canada. McParland, B. J. (1989), `A derivation of the electron mass scattering power for electron dose calculations', Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 274(3), 592-596. MedCom (2018), `Medcom online: Prosoma core'. (Accessed on 02/19/2018). URL: www.medcom-online.de/clinical-areas-products/teletherapy/ prosomacore/ Meurant, G. (2012), The dosimetry of ionizing radiation, Vol. 3, Elsevier. Mijnheer, B. (2008), `State of the art of in vivo dosimetry', Radiation protection dosimetry 131(1), 117-122. Mijnheer, B., Beddar, S., Izewska, J. & Reft, C. (2013), `In vivo dosimetry in external beam radiotherapy', Medical Physics 40(7). Mishra, B., Mishra, S., Selvam, T. P., Chavan, S. & Pethe, S. (2018), `Comparison of measured and monte carlo calculated dose distributions from indigenously developed 6 mv flattening filter free medical linear accelerator', Journal of medical physics 43(3), 162. Mobius3D (2019), `Mobius medical systems, lp', http://mobiusmed.com/mobius3d/. (Accessed on 01/15/2019). Moeller, R. (1989), New concept for multileaf collimation, in `Proceedings of the Sixth Varian European Users Meeting, San Remo, Italy, 1989', Varian, Palo Alto, CA, pp. 126-131. Mohammed, M., Chakir, E., Boukhal, H., Saeed, M. & El Bardouni, T. (2016), `Evaluation of variance reduction techniques in beamnrc monte carlo simulation to improve the computing efficiency', Journal of Radiation Research and Applied Sciences 9(4), 424-430. Monseux, A., Baltieri, V., Sottiaux, A., Tomsej, M. & Leclercq, C. (2016), `Assessment of portal dosimetry accuracy as a qa tool for vmat clinical treatment plans using dolphin/compass tools', Physica Medica: European Journal of Medical Physics 32, 355-356. Morrison, K., Tree, A., Khoo, V. & Van As, N. J. a. (2018), `The pace trial: International randomised study of laparoscopic prostatectomy vs. stereotactic body radiotherapy (sbrt) and standard radiotherapy vs. sbrt for early stage organ-confined prostate cancer.', Journal of Clinical Oncology 36, TPS153-TPS153. Mu, G., Ludlum, E. & Xia, P. (2007), `Impact of mlc leaf position errors on simple and complex imrt plans for head and neck cancer', Physics in Medicine & Biology 53(1), 77. Munro, A. (2007), `Hidden danger, obvious opportunity: error and risk in the management of cancer', The British journal of radiology 80(960), 955-966. Munro, T. & Gilbert, C. (1961), `The relation between tumour lethal doses and the radiosensitivity of tumour cells', The British journal of radiology 34(400), 246-251. Nakaguchi, Y., Ono, T., Maruyama, M., Shimohigashi, Y. & Kai, Y. (2017), `Validation of a method for in vivo 3d dose reconstruction in sbrt using a new transmission detector', Journal of applied clinical medical physics 18(4), 69-75. NATCANSAT (2019), `National radiotherapy dataset - rtds', http://www.natcansat.nhs.uk/rt/rtds.aspx. (Accessed on 02/08/2019). National Cancer Peer Review Programme (2014), `Radiotherapy service evidence guide', Radiotherapy Service Evidence Guide . Nelson, C., Garcia, M., Calderon, E. & Kirsner, S. (2016), `Su-f-t-306: Validation of mobius 3d and fx for elekta linear accelerators', Medical physics 43(6Part17), 3533-3533. Niemierko, A. (1997), `Reporting and analyzing dose distributions: a concept of equivalent uniform dose', Medical physics 24(1), 103-110. Nithiyanantham, K., Mani, G. K., Subramani, V., Mueller, L., Palaniappan, K. K. & Kataria, T. (2015), `Analysis of direct clinical consequences of mlc positional errors in volumetric-modulated arc therapy using 3d dosimetry system', Journal of applied clinical medical physics 16(5), 296-305. Oborn, B. M., Williams, M., Bailey, M. & Carolan, M. (2014), Imrt treatment monitor unit verification using absolute calibrated beamnrc and geant4 monte carlo simulations, in `Journal of Physics: Conference Series', Vol. 489, IOP Publishing, p. 012020. Oderinde, O. & du Plessis, F. (2016), `O3. accurate monte carlo modelling of an elekta synergy equipped with an agility 160-leaf mlc', Physica Medica 32, 141-142. Ohira, S., Ueda, Y., Isono, M., Masaoka, A., Hashimoto, M., Miyazaki, M., Takashina, M., Koizumi, M. & Teshima, T. (2017), `Can clinically relevant dose errors in patient anatomy be detected by gamma passing rate or modulation complexity score in volumetric-modulated arc therapy for intracranial tumors?', Journal of radiation research 58(5), 685-692. Oliver, M., Gagne, I., Bush, K., Zavgorodni, S., Ansbacher, W. & Beckham, W. (2010), `Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy', Radiotherapy and oncology 97(3), 554-560. Otto, K. (2008), `Volumetric modulated arc therapy: Imrt in a single gantry arc', Medical physics 35(1), 310-317. Padilla-Cabal, F., Pérez-Liva, M., Lara, E., Alfonso, R. & Lopez-Pino, N. (2015), `Monte carlo calculations of an elekta precise sl-25 photon beam model', Journal of Radiotherapy in Practice 14(3), 311-322. Palma, D. A., Verbakel, W. F., Otto, K. & Senan, S. (2010), `New developments in arc radiation therapy: a review', Cancer treatment reviews 36(5), 393-399. Pasler, M., Michel, K., Marrazzo, L., Obenland, M., Pallotta, S., Björnsgard, M. & Lutterbach, J. (2017), `Error detection capability of a novel transmission detector: a validation study for online vmat monitoring', Physics in Medicine & Biology 62(18), 7440. Pena, J., Sánchez-Doblado, F., Capote, R., Terrón, J. & Gómez, F. (2006), `Monte carlo correction factors for a farmer 0.6 cm3 ion chamber dose measurement in the build-up region of the 6 mv clinical beam', Physics in Medicine & Biology 51(6), 1523. Piermattei, A., Fidanzio, A., Azario, L., Grimaldi, L., D'Onofrio, G., Cilla, S., Stimato, G., Gaudino, D., Ramella, S., D'Angelillo, R. et al. (2007), `Application of a practical method for the isocenter point in vivo dosimetry by a transit signal', Physics in Medicine & Biology 52(16), 5101. Pitcher, G. M., Hogstrom, K. R. & Carver, R. L. (2016), `Radiation leakage dose from elekta electron collimation system', Journal of applied clinical medical physics 17(5), 157-176. Pitcher, G. M., Hogstrom, K. R. & Carver, R. L. (2017), `Improved electron collimation system design for elekta linear accelerators', Journal of applied clinical medical physics 18(5), 259-270. Popescu, I. A., Shaw, C., Zavgorodni, S. F. & Beckham, W. A. (2005), `Absolute dose calculations for monte carlo simulations of radiotherapy beams', Physics in Medicine & Biology 50(14), 3375. Poppe, B., Looe, H. K., Chofor, N., Rühmann, A., Harder, D. & Willborn, K. C. (2010), `Clinical performance of a transmission detector array for the permanent supervision of imrt deliveries', Radiotherapy and Oncology 95(2), 158-165. Poppe, B., Thieke, C., Beyer, D., Kollhoff, R., Djouguela, A., Rühmann, A., Willborn, K. C. & Harder, D. (2006), `David|a translucent multi-wire transmission ionization chamber for in-vivo verification of imrt and conformal irradiation techniques', Physics in Medicine and Biology 51(5), 1237-1248. PTW (2010), `A comparison of detectors used in detector arrays'. (Accessed: August 2018. URL: http://www.ptw.de/3095.html Qi, P. & Xia, P. (2013), `Relationship of segment area and monitor unit efficiency in aperture-based imrt optimization', Journal of Applied Clinical Medical Physics 14(3), 232-243. QUANTEC (2010), `Quantitative analyses of normal tissue effects in the clinic (quantec): an introduction to the scientific issues', International Journal of Radiation Oncology Biology Physics 76(3), S3-S9. Bentzen, Søren M and Constine, Louis S and Deasy, Joseph O and Eisbruch, Avi and Jackson, Andrew and Marks, Lawrence B and Ten Haken, Randall K and Yorke, Ellen D. Rangaraj, D., Zhu, M., Yang, D., Palaniswaamy, G., Yaddanapudi, S., Wooten, O. H., Brame, S. & Mutic, S. (2013), `Catching errors with patient-specific pretreatment machine log file analysis', Practical radiation oncology 3(2), 80-90. Rangel, A. & Dunscombe, P. (2009), `Tolerances on mlc leaf position accuracy for imrt delivery with a dynamic mlc', Medical physics 36(7), 3304-3309. RCR (2008a), `Implementing in vivo dosimetry'. Institute of Physics and Engineering in Medicine, National Patient Safety Agency, Society and College of Radiographers, The Royal College of Radiologists. RCR (2008b), `On target: ensuring geometric accuracy in radiotherapy'. Royal College of Radiologists, Institute of Physics and Engineering in Medicine, Society and college of Radiographers, Authors: Barrett, Jane and Hoskin, P and Gaze, M and Greener, T and Kirby, M and McNair, H and Powell, M and Routsis, D and Tait, D. Redpath, A. & Thwaites, D. (1991), `A 3-dimensional scatter correction algorithm for photon beams (dosimetry)', Physics in Medicine & Biology 36(6), 779. Ricketts, K., Navarro, C., Lane, K., Blowfield, C., Cotten, G., Tomala, D., Lord, C., Jones, J. & Adeyemi, A. (2016), `Clinical experience and evaluation of patient treatment verification with a transit dosimeter', International Journal of Radiation Oncology* Biology* Physics 95(5), 1513-1519. Rogers, D., Bielajew, A. et al. (1990), `Monte carlo techniques of electron and photon transport for radiation dosimetry', The dosimetry of ionizing radiation 3, 427-539. Rogers, D. W. O. (2006), `Fifty years of monte carlo simulations for medical physics', Physics in Medicine and Biology 51(13), R287. Rogers, D. W. O., Faddegon, B. A., Ding, G. X., Ma, C.-M., We, J. & Mackie, T. R. (1995), `Beam: A monte carlo code to simulate radiotherapy treatment units', Medical Physics 22(5), 503-524. Rogers, D., Walters, B. & Kawrakow, I. (2001), BEAMnrc users manual, Report, National Research Council Canada. Roussakis, Y. G. (2016), Strategies for adaptive radiotherapy: towards clinically efficient work ows, PhD thesis, University of Birmingham. Salvat, F., Fernández-Varea, J. M. & Sempau, J. (2006), Penelope-2008: A code system for monte carlo simulation of electron and photon transport, in `Workshop Proceedings', Vol. 4. Schreiner, L., Holmes, O. & Salomons, G. (2013), Analysis and evaluation of planned and delivered dose distributions: practical concerns with γ- and χ- evaluations, in `Journal of Physics: Conference Series', Vol. 444, IOP Publishing, p. 012016. SCoR (2015), `Radiotherapy board|intensity modulated radiotherapy (imrt) in the uk: current access and predictions of future access rates.', Society and College of Radiographers, Institute of Physics and Engineering in Medicine, Royal College of Radiologists . Taylor, R and Hassan, S and D'Souza, D. Scottish Executive (2006), Unintended overexposure of patient lisa norris during radiotherapy treatment at the beatson oncology centre, glasgow in january 2006, in `Report of an Investigation by the Inspector Appointed by the Scottish Ministers for the Ionising Radiation (Medical Exposures) Regulations 2000', Scottish Government. Seco, J. & Verhaegen, F. (2016), Monte Carlo Techniques in Radiation Therapy, Imaging in Medical Diagnosis and Therapy, CRC Press. Semenenko, V. A., Reitz, B., Day, E., Qi, X. S., Miften, M. & Li, X. A. (2008), `Evaluation of a commercial biologically based imrt treatment planning system', Medical Physics 35(12), 5851-5860. Seuntjens, J., Kawrakow, I., Borg, J., Hobeila, F. & Rogers, D. (2002), Calculated and measured air-kerma response of ionization chambers in low and medium energy photon beams, in `Recent Developments in Accurate Radiation Dosimetry: Proc. Int. Workshop', pp. 69-84. Sheikh-Bagheri, D. & Rogers, D. (2002a), `Monte carlo calculation of nine megavoltage photon beam spectra using the beam code', Medical physics 29(3), 391-402. Sheikh-Bagheri, D. & Rogers, D. (2002b), `Sensitivity of megavoltage photon beam monte carlo simulations to electron beam and other parameters', Medical physics 29(3), 379-390. Sheikh-Bagheri, D., Rogers, D., Ross, C. K. & Seuntjens, J. P. (2000), `Comparison of measured and monte carlo calculated dose distributions from the nrc linac', Medical physics 27(10), 2256-2266. Siantar, C. H., Walling, R., Daly, T., Faddegon, B., Albright, N., Bergstrom, P., Bielajew, A., Chuang, C., Garrett, D., House, R. et al. (2001), `Description and dosimetric verification of the peregrine monte carlo dose calculation system for photon beams incident on a water phantom', Medical physics 28(7), 1322-1337. Sikora, M. & Alber, M. (2009), `A virtual source model of electron contamination of a therapeutic photon beam', Physics in Medicine & Biology 54(24), 7329. Sikora, M., Dohm, O. & Alber, M. (2007), `A virtual photon source model of an elekta linear accelerator with integrated mini mlc for monte carlo based imrt dose calculation', Physics in Medicine & Biology 52(15), 4449. Smith, K., Balter, P., Duhon, J., White, G. A., Vassy, D. L., Miller, R. A., Serago, C. F. & Fairobent, L. A. (2017), `Aapm medical physics practice guideline 8. a.: linear accelerator performance tests', Journal of applied clinical medical physics 18(4), 23-39. Song, T., Zhou, L. & Jiang, S. (2012), `Su-e-t-496: Monte carlo simulation of a 6mv varian truebeam without flattening filter linac', Medical physics 39(6Part17), 3819-3819. Sontag, M. R. & Cunningham, J. R. (1978), `The equivalent tissue-air ratio method for making absorbed dose calculations in a heterogeneous medium', Radiology 129(3), 787-794. Spezi, E., Angelini, A., Romani, F. & Ferri, A. (2005), `Characterization of a 2d ion chamber array for the verification of radiotherapy treatments', Physics in Medicine & Biology 50(14), 3361. Stasi, M., Attili, A., Brusasco, C., Cirio, R., Denis, J., Donetti, M., Evans, P., Giordanengo, S., Giuliacci, A., Givehchi, N. et al. (2010), `Architecture and development of a detector for on-line verification in imrt.', IBA dosimetry . Stathakis, S., Myers, P., Esquivel, C., Mavroidis, P. & Papanikolaou, N. (2013), `Characterization of a novel 2d array dosimeter for patient-specific quality assurance with volumetric arc therapy', Medical physics 40(7). Stieler, F., Wolff, D., Schmid, H., Welzel, G., Wenz, F. & Lohr, F. (2011), `A comparison of several modulated radiotherapy techniques for head and neck cancer and dosimetric validation of vmat', Radiotherapy and Oncology 101(3), 388-393. Stock, M., Kroupa, B. & Georg, D. (2005), `Interpretation and evaluation of the γ index and the γ index angle for the verification of imrt hybrid plans', Physics in Medicine & Biology 50(3), 399. Storm, L. & Israel, H. I. (1970), `Photon cross sections from 1 kev to 100 mev for elements z= 1 to z= 100', Atomic Data and Nuclear Data Tables 7(6), 565-681. Thoelking, J., Fleckenstein, J., Sekar, Y., Boggula, R., Lohr, F., Wenz, F. & Wertz, H. (2016b), `Patient-specific online dose verification based on transmission detector measurements', Radiotherapy and Oncology 119(2), 351-356. Thoelking, J., Sekar, Y., Fleckenstein, J., Lohr, F., Wenz, F. & Wertz, H. (2016a), `Characterization of a new transmission detector for patient individualized online plan verification and its influence on 6mv x-ray beam characteristics', Zeitschrift für Medizinische Physik 26(3), 200-208. Thompson, C., Weston, S., Cosgrove, V. & Thwaites, D. (2014), `A dosimetric characterization of a novel linear accelerator collimator', Medical physics 41(3). Thwaites, D. (2013), Accuracy required and achievable in radiotherapy dosimetry: have modern technology and techniques changed our views?, in `Journal of Physics: Conference Series', Vol. 444, IOP Publishing, p. 012006. Thwaites, D. I., Mijnheer, B. & Mills, J. A. (2005), `Quality assurance of external beam radiotherapy', Radiation oncology physics: a handbook for teachers and students pp. 470-50. Toft, B. (2004), `Independent review of the circumstances surrounding a serious adverse incident that occurred in the REDACTED'. URL: http://www.who.int/patientsafety/news/RadiotherapyadverseeventToftreport.pdf Tuazon, B., Narayanasamy, G., Papanikolaou, N., Kirby, N., Mavroidis, P. & Stathakis, S. (2018), `Evaluation and comparison of second-check monitor unit calculation software with pinnacle 3 treatment planning system', Physica Medica 45, 186-191. Tyagi, N., Yang, K., Gersten, D. & Yan, D. (2012), `A real time dose monitoring and dose reconstruction tool for patient specific vmat qa and delivery', Medical physics 39(12), 7194-7204. Valdenaire, S., Mailleux, H. & Fau, P. (2016), `Modeling of flattening filter free photon beams with analytical and monte carlo tps', Biomedical Physics & Engineering Express 2(3), 035010. Van de Walle, J., Martens, C., Reynaert, N., Palmans, H., Coghe, M., De Neve, W., De Wagter, C. & Thierens, H. (2003), `Monte carlo model of the elekta sliplus accelerator: validation of a new mlc component module in beam for a 6 mv beam', Physics in Medicine & Biology 48(3), 371. Van Elmpt, W., McDermott, L., Nijsten, S.,Wendling, M., Lambin, P. & Mijnheer, B. (2008), `A literature review of electronic portal imaging for radiotherapy dosimetry', Radiotherapy and Oncology 88(3), 289-309. Van Esch, A., Clermont, C., Devillers, M., Iori, M. & Huyskens, D. P. (2007), `On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2d ion chamber array and the octavius phantom', Medical physics 34(10), 3825-3837. Van Herk, M. (2004), Errors and margins in radiotherapy, in `Seminars in radiation oncology', Vol. 14, Elsevier, pp. 52-64. Vanderstraeten, B., Chin, P. W., Fix, M., Leal, A., Mora, G., Reynaert, N., Seco, J., Soukup, M., Spezi, E., De Neve, W. et al. (2007), `Conversion of ct numbers into tissue parameters for monte carlo dose calculations: a multi-centre study', Physics in Medicine & Biology 52(3), 539. Vatnitsky, S., Lopez, P. O., Izewska, J., Meghzifene, A. & Levin, V. (2001), `The radiation overexposure of radiotherapy patients in panama 15 june 2001', Radiotherapy and Oncology 60(3), 237-238. Vazquez-Quino, L., Huerta-Hernandez, C. & Rangaraj, D. (2017), Clinical experience with mobius fx software for 3d dose verification for prostate vmat plans and comparison with physical measurements, in `Journal of Physics: Conference Series', Vol. 847, IOP Publishing, p. 012060. Venkataraman, S., Malkoske, K. E., Jensen, M., Nakonechny, K. D., Asuni, G. & McCurdy, B. M. (2009), `The influence of a novel transmission detector on 6 mv x-ray beam characteristics', Physics in Medicine & Biology 54(10), 3173. Venselaar, J., Welleweerd, H. & Mijnheer, B. (2001), `Tolerances for the accuracy of photon beam dose calculations of treatment planning systems', Radiotherapy and Oncology 60(2), 191-201. Verhaegen, F. & Seuntjens, J. (2003), `Monte carlo modelling of external radiotherapy photon beams', Physics in medicine & biology 48(21), R107. Verhaegen, F., Symonds-Tayler, R., Liu, H. & Nahum, A. (2000), `Backscatter towards the monitor ion chamber in high-energy photon and electron beams: charge integration versus monte carlo simulation', Physics in Medicine & Biology 45(11), 3159. Visser, R., Wauben, D., De Groot, M., Godart, J., Langendijk, J., van't Veld, A. & Korevaar, E. (2013), `Efficient and reliable 3d dose quality assurance for imrt by combining independent dose calculations with measurements', Medical physics 40(2). Wack, G., Lalande, F. & Seligman, M. (2007), `Summary of asn on the epinal radiotherapy accident', French Nuclear Safety Authority . Waldron, T. (2002), The theory and operation of computer-controlled medical linear accelerator, in `MEDICAL PHYSICS', Vol. 29, AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS STE 1 NO 1, 2 HUNTINGTON QUADRANGLE, MELVILLE, NY 11747-4502 USA, pp. 1277-1277. Walters, B. (2016), `Discussion concerning the observed higher dose in small voxels'. Walters, B., Kawrakow, I. & Rogers, D. (2002), `History by history statistical estimators in the beam code system', Medical Physics 29(12), 2745-2752. Walters, B. & Rogers, D. (2003), DOSXYZnrc users manual, Report, National Research Council Canada. Webb, S. (2003), `The physical basis of imrt and inverse planning', The British journal of radiology 76(910), 678-689. WHO (2008), `Radiotherapy risk profile'. World Health Organization. Williams, M. & McKenzie, A. (2008), `Can we afford not to implement in vivo dosimetry?', The British journal of radiology 81(969), 681-684. Wu, H., Jiang, F., Yue, H., Hu, Q., Zhang, J., Liu, Z., Gong, J., Li, S., Geng, J. & Zhang, Y. (2016), `A comparative study of identical vmat plans with and without jaw tracking technique', Journal of applied clinical medical physics 17(5), 133-141. Wulff, J., Heverhagen, J. & Zink, K. (2008), `Monte-carlo-based perturbation and beam quality correction factors for thimble ionization chambers in high-energy photon beams', Physics in Medicine & Biology 53(11), 2823. Zhen, H., Nelms, B. E. & Tomé, W. A. (2011), `Moving from gamma passing rates to patient dvh-based qa metrics in pretreatment dose qa', Medical physics 38(10), 5477-5489. Zhen, H., Nelms, B. E. & Tomé, W. A. (2013), `On the use of biomathematical models in patient-specific imrt dose qa', Medical physics 40(7)."],"dcterms:accessrights":["public"],"dc:title":["Characterisation and evaluation of a novel transmission detector for intra-fraction monitoring of radiotherapy."],"dc:type":["Thesis"],"dc:type.qualificationlevel":["doctoral"],"dc:type.qualificationname":["Ph.D"]},"updated_at":"2026-11-18T22:33:39Z"}