{"id":{"repo_id":"whiterose","oai_identifier":"oai:etheses.whiterose.ac.uk:30729"},"canonical_url":"https://search.ndltd.org/etd/whiterose/oai:etheses.whiterose.ac.uk:30729","repository":{"repo_id":"whiterose","name":"White Rose University Consortium","base_url":"https://etheses.whiterose.ac.uk/cgi/oai2"},"display":{"title":"Torques in total hip replacements","abstract":"Elevated frictional torques at the articulating interface of total hip replacements (THRs), if transferred, have the potential to affect component fixation and/or induce micro motion linked to fretting and corrosion at modular junctions. Pre-clinical testing of friction behaviour in THRs has predominantly been assessed under simplified test conditions using uniaxial simulators. These methodologies and simulators are not sufficient if assessment of friction under more clinically relevant conditions, such as multi-axis motion and consideration of surgical implant positioning, are to be investigated. To identify a suitable measuring system and develop methodologies for assessing the frictional torques in THRs, studies were conducted whereby comparisons between a newly developed electromechanical multi-axis single station hip simulator (SSHS) and an existing uniaxial hip simulator as well as a custom-built friction measuring subsystem were made. Additionally, analyses of free-body diagrams in static equilibria and functional verification testing of the SSHS was completed. The assessments conducted showed that the SSHS could detect THR frictional torques without the use of the custom-built friction subsystem; and that the sub system did not provide the additional information expected. Methodologies for assessing THR frictional torques were developed on the SSHS and used to assess the effects of translational positioning (i.e. testing when there was offset between the head and cup centres of rotation) on THR frictional torques. Increasing the applied medial-lateral (ML) translation, as might occur in a patient when there is a lateral misalignment between the centres of the femoral head and acetabular cup, under 1kN constant load and biaxial motion increased the measured frictional torques at the bearing interface of 36mm metal-on-polymer bearings, to over 15Nm. Elevated frictional torques, if transferred to the fixation interface of the acetabular cup, may induce shear stresses that could result in cup displacement. To determine whether the measured elevated frictional torques observed during simulator testing could cause potential displacement in an implanted press-fit acetabular cup, a load-to-failure test that applied a simultaneous 300N or 3kN femoral head load was developed. Results indicated that the elevated frictional torques measured under ML translation on the SSHS were not likely to cause cup displacement in 10 and 20pcf Sawbone blocks when either axial load was applied. This thesis reports on developed new methodologies for assessing THR frictional torques under variations in surgical positioning and for assessing torques required to cause cup loosening under axial loading in load-to-failure tests. It was demonstrated for the first time using a multi-axis hip simulator that surgical implant malpositioning can result in elevated frictional torques at the bearing interface of THRs. However, the impact of these elevated frictional torques on cup fixation requires further work to examine surgical implant malpositioning in THRs under more representative loading and motion conditions.","abstract_html":"Elevated frictional torques at the articulating interface of total hip replacements (THRs), if transferred, have the potential to affect component fixation and/or induce micro motion linked to fretting and corrosion at modular junctions. Pre-clinical testing of friction behaviour in THRs has predominantly been assessed under simplified test conditions using uniaxial simulators. These methodologies and simulators are not sufficient if assessment of friction under more clinically relevant conditions, such as multi-axis motion and consideration of surgical implant positioning, are to be investigated. To identify a suitable measuring system and develop methodologies for assessing the frictional torques in THRs, studies were conducted whereby comparisons between a newly developed electromechanical multi-axis single station hip simulator (SSHS) and an existing uniaxial hip simulator as well as a custom-built friction measuring subsystem were made. Additionally, analyses of free-body diagrams in static equilibria and functional verification testing of the SSHS was completed. The assessments conducted showed that the SSHS could detect THR frictional torques without the use of the custom-built friction subsystem; and that the sub system did not provide the additional information expected. Methodologies for assessing THR frictional torques were developed on the SSHS and used to assess the effects of translational positioning (i.e. testing when there was offset between the head and cup centres of rotation) on THR frictional torques. Increasing the applied medial-lateral (ML) translation, as might occur in a patient when there is a lateral misalignment between the centres of the femoral head and acetabular cup, under 1kN constant load and biaxial motion increased the measured frictional torques at the bearing interface of 36mm metal-on-polymer bearings, to over 15Nm. Elevated frictional torques, if transferred to the fixation interface of the acetabular cup, may induce shear stresses that could result in cup displacement. To determine whether the measured elevated frictional torques observed during simulator testing could cause potential displacement in an implanted press-fit acetabular cup, a load-to-failure test that applied a simultaneous 300N or 3kN femoral head load was developed. Results indicated that the elevated frictional torques measured under ML translation on the SSHS were not likely to cause cup displacement in 10 and 20pcf Sawbone blocks when either axial load was applied. This thesis reports on developed new methodologies for assessing THR frictional torques under variations in surgical positioning and for assessing torques required to cause cup loosening under axial loading in load-to-failure tests. It was demonstrated for the first time using a multi-axis hip simulator that surgical implant malpositioning can result in elevated frictional torques at the bearing interface of THRs. However, the impact of these elevated frictional torques on cup fixation requires further work to examine surgical implant malpositioning in THRs under more representative loading and motion conditions.","abstract_has_math":false,"creators":["Frimpong, Ama Amponsah"],"institution":"University of Leeds","degree_name":"Integrated Ph.D and Master","degree_level":"doctoral","degree_discipline":null,"degree_department":null,"school":null,"contributors":[],"advisors":["Williams, Sophie","Wilcox, Ruth"],"committee_chairs":[],"committee_members":[],"year":2021,"date_issued":"2021-11","date_published":"2021-11","updated_at":"2026-11-18T23:18:02Z","subjects":[],"languages":["eng"],"rights":[],"rights_urls":[],"identifier_entries":[{"key":"dc:identifier","label":"Identifier","values":["uk.bl.ethos.858631"],"render_values":[{"text":"uk.bl.ethos.858631","href":null,"code":true}]}],"person_identifier_entries":[]},"links":{"outbound_url":null,"outbound_label":null,"outbound_source":null,"record_url":null,"record_label":null,"record_kind":null,"record_source":null,"download_url":"https://etheses.whiterose.ac.uk/id/eprint/30729/1/AAFrimpong_Thesis.pdf","download_label":"Download PDF","download_source":"dc:identifier.uri","access_status":null,"embargo_until":null,"access_note":null},"source_record":{"url":"https://etheses.whiterose.ac.uk/cgi/oai2?verb=GetRecord&metadataPrefix=uketd_dc&identifier=oai%3Aetheses.whiterose.ac.uk%3A30729","prefix":"uketd_dc"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Williams, Sophie","Wilcox, Ruth"]},{"key":"dc:contributor.sponsor","label":"Sponsor","values":["EPSRC","University of Leeds","DePuy Synthes"]},{"key":"dc:creator","label":"Author","values":["Frimpong, Ama Amponsah"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2021-11"]},{"key":"dc:date.issued","label":"Date","values":["2021-11"]},{"key":"dc:publisher.department","label":"Dc Publisher Department","values":["School of Mechanical Engineering (Leeds)"]},{"key":"dc:publisher.institution","label":"Dc Publisher Institution","values":["University of Leeds"]},{"key":"dc:relation.isreferencedby","label":"Dc Relation Isreferencedby","values":["https://etheses.whiterose.ac.uk/id/eprint/30729/"]},{"key":"dc:relation.references","label":"Dc Relation References","values":["Adler, E., Stuchin, S.A. and Kummer, F.J. 1992. Stability of press-fit acetabular cups. J Arthroplasty. 7(3), pp.295–301. Affatato, S. 2012. Wear of orthopaedic implants and artificial joints. Elsevier. Al‐Hajjar, M. et al. (2010) ‘Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic‐on‐ceramic total hip replacement’, Journal of Biomedical Materials Research Part B: Applied Biomaterials, 95(2), pp. 263–268. Al-Hajjar, M. et al. (2013) ‘Edge Loading in Total Hip Replacements Due to Translational Surgical Mal-Positioning: In Vitro Techniques’, Bone & Joint Journal Orthopaedic Proceedings Supplement, 95-B(SUPP 34), p. 2. Available at: http://www.bjjprocs.boneandjoint.org.uk/content/95-B/SUPP_34/2.abstract. Al-Hajjar, M. et al. (2014) ‘Wear of Large Diameter Ceramic-on-Ceramic Bearings in Total Hip Replacements under Edge Loading Conditions’, in ORS Annual Meeting 2014. New Orleans. Al-Hajjar, M. et al. (2015) ‘Increased Torque in Hip Replacement Bearings due to Translational Surgical Mal-Positioning; ORS Poster No. 1759’, in. Al-Hajjar, M., Clarkson, P., Williams, S., Jennings, L., Thompson, J. and Fisher, J. 2013. Edge Loading in Total Hip Replacements Due to Translational Surgical Mal-Positioning: In Vitro Techniques. Bone & Joint Journal Orthopaedic Proceedings Supplement. 95-B(SUPP 34), p.2. Al-Hajjar, M., Fisher, J., Hardaker, C., Kuring, G., Thompson, J. and Wiliams, S. 2014. Wear of Large Diameter Ceramic-on-Ceramic Bearings in Total Hip Replacements under Edge Loading Conditions In: ORS Annual Meeting 2014. New Orleans. Al‐Hajjar, M., Leslie, I.J., Tipper, J., Williams, S., Fisher, J. and Jennings, L.M. 2010. Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic‐on‐ceramic total hip replacement. Journal of Biomedical Materials Research Part B: Applied Biomaterials. 95(2), pp.263–268. Ali, M. et al. (2016) ‘Wear and Deformation of Metal-on-Polyethylene Hip Replacements under Variations in Surgical Positioning’, in. Ali, M. et al. (2017) ‘Variations in Component Positioning Leading to Dynamic Separation and Edge Loading Influences the Deformation and Wear of Metal-on-Polyethylene Hip Replacement Bearings’, ORS 2017 Annual Meeting, 4(5). Available at: http://www.ors.org/Transactions/63/1032.pdf (Accessed: 14 September 2017). Antoniades, G., Smith, E.J., Deakin, A.H., Wearing, S.C. and Sarungi, M. 2013. Primary stability of two uncementedacetabular components of different geometry: hemispherical or peripherallyenhanced? Bone & joint research. 2(12), pp.264–9. Apostu, D., Lucaciu, O., Berce, C., Lucaciu, D. and Cosma, D. 2018. Current methods of preventing aseptic loosening and improving osseointegration of titanium implants in cementless total hip arthroplasty: a review. Journal of International Medical Research. 46(6), pp.2104–2119. Apostu, D., Lucaciu, O., Lucaciu, G.D.O., Crisan, B., Crisan, L., Baciut, M., Onisor, F., Baciut, G., Câmpian, R.S. and Bran, S. 2017. Systemic drugs that influence titanium implant osseointegration. Drug Metabolism Reviews. 49(1), pp.92–104. Auger, D. D. et al. (1993) ‘Friction and lubrication in cushion form bearings for artificial hip joints’, ARCHIVE: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine 1989-1996 (vols 203-210), 207(18), pp. 25–33. doi: 10.1243/PIME_PROC_1993_207_265_02. Barrack, R. L. et al. (2001) ‘Virtual reality computer animation of the effect of component position and design on stability after total hip arthroplasty’, Orthopedic Clinics of North America, 32(4), pp. 569–577. doi: https://doi.org/10.1016/S0030-5898(05)70227-3. Bergmann, G. et al. (2012) ‘High-Tech Hip Implant for Wireless Temperature Measurements In Vivo’, PLoS ONE. Edited by H. A. Awad, 7(8), p. e43489. doi: 10.1371/journal.pone.0043489. Berry, D.J. and Lieberman, J. 2012. Surgery of the Hip. Elsevier Health Sciences. Bishop, N. E., Hothan, A. and Morlock, M. M. (2013) ‘High friction moments in large hard-on-hard hip replacement bearings in conditions of poor lubrication’, J Orthop Res. 2012/12/15, 31(5), pp. 807–813. doi: 10.1002/jor.22255. Bishop, N. E., Waldow, F. and Morlock, M. M. (2008) ‘Friction moments of large metal-on-metal hip joint bearings and other modern designs’, Med Eng Phys, 30(8), pp. 1057–1064. doi: DOI 10.1016/j.medengphy.2008.01.001. Bishop, N.E., Hothan, A. and Morlock, M.M. 2013. High friction moments in large hard-on-hard hip replacement bearings in conditions of poor lubrication. J Orthop Res. 31(5), pp.807–813. Bowsher, J. G. and Shelton, J. C. (2001) ‘A hip simulator study of the influence of patient activity level on the wear of crosslinked polyethylene under smooth and roughened femoral conditions’, Wear, 250(1–12), pp. 167–179. doi: http://dx.doi.org/10.1016/S0043-1648(01)00619-6. Bowsher, J.G., Hussain, A., Williams, P.A. and Shelton, J.C. 2005. Large head diameters have the potential to reduce ion release in metal-on metal hip wear simulations. ORS Washington DC., p.1626. Brockett, C. L. et al. (2008) ‘The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements’, Journal of Materials Science: Materials in Medicine, 19(4), pp. 1575–1579. Brockett, C. L. et al. (2013) ‘Squeaking hip arthroplasties: a tribological phenomenon’, J Arthroplasty. 2012/04/07, 28(1), pp. 90–97. doi: 10.1016/j.arth.2012.01.023. Brockett, C., Williams, S., Jin, Z., Isaac, G. and Fisher, J. 2007. Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res B Appl Biomater. 81(2), pp.508–515. Callaghan, J.J., Rosenberg, A.G. and Rubash, H.E. 2007. The Adult Hip. Lippincott Williams & Wilkins. Cameron, H.U., Pilliar, R.M. and MacNab, I. 1973. The effect of movement on the bonding of porous metal to bone. J Biomed Mater Res. 7(4), pp.301–311. CeramTec (2012) BIOLOX®delta. Available at: https://www.ceramtec.com/ceramic-materials/biolox/delta/ (Accessed: 4 July 2017). Charnley, J. (1961) ‘Arthroplasty of Hip - a New Operation’, Lancet, 1(718), pp. 1129-. Charnley, J., Mckee, G.K., Coltart, W.D. and Scales, J.T. 1961. Arthroplasty of the Hip by the Low-Friction Technique. Journal of Bone and Joint Surgery-British Volume. 43(3), p.601. Cherian, J.J., Jauregui, J.J., Banerjee, S., Pierce, T. and Mont, M.A. 2015. What Host Factors Affect Aseptic Loosening After THA and TKA? Clinical Orthopaedics and Related Research. 473(8), pp.2700–2709. Clarke, H.J., Jinnah, R.H., Warden, K.E., Cox, Q.G. and Curtis, M.J. 1991. Evaluation of acetabular stability in uncemented prostheses. J Arthroplasty. 6(4), pp.335–340. Cowin, S.C. (Stephen C. 2001. Bone mechanics handbook. CRC Press. Crosnier, E.A., Keogh, P.S. and Miles, A.W. 2014. A novel method to assess primary stability of press-fit acetabular cups. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 228(11), pp.1126–1134. Crowe, J. F., Sculco, T. P. and Kahn, B. (2003) ‘Revision Total Hip Arthroplasty’, Clinical Orthopaedics and Related Research, 413(413), pp. 175–182. doi: 10.1097/01.blo.0000072469.32680.b6. Curtis, M.J., Jinnah, R.H., Wilson, V.D. and Hungerford, D.S. 1992. The initial stability of uncemented acetabular components. J Bone Joint Surg Br. 74(3), pp.372–376. Damm, P. et al. (2013) ‘Friction in total hip joint prosthesis measured in vivo during walking’, PLoS One, 8(11), p. e78373. doi: 10.1371/journal.pone.0078373. Davim, J.P. 2013. Biotribology. Wiley. Dennis, D. A. et al. (2001) ‘\"In vivo\" determination of hip joint separation and the forces generated due to impact loading conditions.’, Journal of biomechanics, 34(5), pp. 623–9. Available at: http://www.ncbi.nlm.nih.gov/pubmed/11311703 (Accessed: 13 September 2017). Dowson, D. et al. (1985) ‘Influence of Counterface Topography on the Wear of Ultra High Molecular Weight Polyethylene Under Wet or Dry Conditions’, in, pp. 171–187. doi: 10.1021/bk-1985-0287.ch012. Dowson, D., Dalmaz, G. and Lubrecht, A. A. (2003) Tribological Research and Design for Engineering Systems: Proceedings of the 29th Leeds-Lyon Symposium. Elsevier Science. Dowson, D., Hardaker, C., Flett, M. and Isaac, G.H. 2004. A hip joint simulator study of the performance of metal-on-metal joints: Part II: design. J Arthroplasty. 19(8 Suppl 3), pp.124–130. Fisher, J. 2011. Bioengineering reasons for the failure of metal-on-metal hip prostheses: an engineer’s perspective. J Bone Joint Surg Br. 93(8), pp.1001–1004. Fisher, J. and Dowson, D. 1991. Tribology of total artificial joints. Proc Inst Mech Eng H. 205(2), p.73. Gore, T. A., Higginson, G. R. and Kornberg, R. E. (1981) ‘Some evidence of squeeze-film lubrication in hip prostheses’. Available at: http://journals.sagepub.com/doi/pdf/10.1243/EMED_JOUR_1981_010_026_02 (Accessed: 5 July 2017). Haider, H., Weisenburger, J. N. and Garvin, K. L. (2016) ‘Simultaneous measurement of friction and wear in hip simulators’, Proc Inst Mech Eng H, 230(5), pp. 373–388. doi: 10.1177/0954411916644476. Hall, R.M. and Unsworth, A. 1997. Friction in hip prostheses. Biomaterials. 18(15), pp.1017–1026. Harris, W.H. 2012. Edge Loading Has a Paradoxical Effect on Wear in Metal-on-Polyethylene Total Hip Arthroplasties. Clinical Orthopaedics and Related Research. 470(11), pp.3077–3082. Helgason, B., Perilli, E., Schileo, E., Taddei, F., Brynjó Lfsson, S. and Viceconti, M. 2008. Mathematical relationships between bone density and mechanical properties: A literature review. Hong, D., Li, C. and Jeong, J. (2012) ‘A Crosstalk Compensation of a Multi-axis ForceTorque Sensor Based on the Least Squares Method Using LabVIEW’, in 2012 Fourth International Conference on Computational and Information Sciences, pp. 1127–1130. doi: 10.1109/ICCIS.2012.12. Hua, X. et al. (2014) ‘Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions’, Journal of biomechanics, 47(13), pp. 3303–3309. Hua, X., Li, J., Wang, L., Jin, Z., Wilcox, R. and Fisher, J. 2014. Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions. Journal of biomechanics. 47(13), pp.3303–3309. Huber, W.O. and Noble, P.C. 2014. Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins. International Orthopaedics. 38(4), pp.725–731. Jin, Z.M., Dowson, D. and Fisher, J. 1994. A parametric analysis of the contact stress in ultra-high molecular weight polyethylene acetabular cups. Medical Engineering & Physics. 16(5), pp.398–405. Jin, Z.M., Meakins, S., Morlock, M.M., Parsons, P., Hardaker, C., Flett, M. and Isaac, G. 2006. Deformation of press-fitted metallic resurfacing cups. Part 1: Experimental simulation. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 220(2), pp.299–309. Jin, Z.M., Stone, M., Ingham, E. and Fisher, J. 2006. (v) Biotribology. Current Orthopaedics. 20(1), pp.32–40. Johnson, A. T. (Arthur T. (2011) Biology for engineers. CRC Press. Kody, M.H., Kabo, J.M., Markolf, K.L., Dorey, F.J. and Amstutz, H.C. 1990. Strength of initial mechanical fixation of screw ring acetabular components. Clin Orthop Relat Res. (257), pp.146–153. Lachiewicz, P.F., Suh, P.B. and Gilbert, J.A. 1989. In vitro initial fixation of porous-coated acetabular total hip components. J Arthroplasty. 4(3), pp.201–205. Leng, J. et al. (2017) ‘Dynamic virtual simulation of the occurrence and severity of edge loading in hip replacements associated with variation in the rotational and translational surgical position’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 231(4), pp. 299–306. doi: 10.1177/0954411917693261. Lewinnek, G. E. et al. (1978) ‘Dislocations after total hip-replacement arthroplasties’, J Bone Joint Surg Am, 60(2), pp. 217–220. Liao, Y.-S. et al. (2003) ‘The effect of frictional heating and forced cooling on the serum lubricant and wear of UHMW polyethylene cups against cobalt–chromium and zirconia balls’, Biomaterials, 24(18), pp. 3047–3059. Liu, F. et al. (2013) ‘Effect of head contact on the rim of the cup on the offset loading and torque in hip joint replacement’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, p. 0954411913496016. Livermore, J., Ilstrup, D. and Morrey, B. 1990. Effect of femoral head size on wear of the polyethylene acetabular component. The Journal of Bone & Joint Surgery. 72(4), pp.518–528. Lombardi, A. V. et al. (2000) ‘An in vivo determination of total hip arthroplasty pistoning during activity’, The Journal of Arthroplasty, 15(6), pp. 702–709. doi: 10.1054/arth.2000.6637. Longaray, J. et al. (2013) ‘How Do Bearing Size and Bearing Material Affect Frictional Torque? a Hip Simulator Evaluation’, Bone & Joint Journal Orthopaedic Proceedings Supplement, 95(SUPP 34), p. 121. Macdonald, W., Carlsson, L. V, Charnley, G.J. and Jacobsson, C.M. 1999. Press-fit acetabular cup fixation: principles and testing. Proc Inst Mech Eng H. 213(1), pp.33–39. Malik, S. S. (2015) Orthopaedic biomechanics made easy. McCollum, D. E. and Gray, W. J. (1990) ‘Dislocation after total hip arthroplasty. Causes and prevention.’, Clinical orthopaedics and related research, (261), pp. 159–170. McKellop, H. et al. (1981) ‘Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening device’, Journal of Biomedical Materials Research, 15(5), pp. 619–653. doi: 10.1002/jbm.820150503. McMinn, D. J. W. (2009) Modern Hip Resurfacing. Springer. Meneghini, R.M., Meyer, C., Buckley, C.A., Hanssen, A.D. and Lewallen, D.G. 2010. Mechanical Stability of Novel Highly Porous Metal Acetabular Components in Revision Total Hip Arthroplasty. J Arthroplasty. 25(3), pp.337–341. Miller, M.D., Thompson, S.R. and Hart, J. 2012. Review of Orthopaedics. Elsevier Health Sciences. Miyakawa, S., Kawamura, H., Mishima, H. and Yasumoto, J. 2004. Grit-blasted and hydroxyapatite-coated total hip arthroplasty: An 11- to 14-year follow-up study. Journal of Orthopaedic Science. 9(5), pp.462–467. Morlock, M., Bishop, N. and Huber, G. (2011) ‘Biomechanics of Hip Arthroplasty’, in Knahr, K. (ed.) Tribology in Total Hip Arthroplasty. Springer Berlin Heidelberg, pp. 11–24. doi: 10.1007/978-3-642-19429-0_2. Nevelos, J. E. et al. (1999) ‘Analysis of retrieved alumina ceramic components from Mittelmeier total hip prostheses’, Biomaterials, 20(19), pp. 1833–1840. doi: Doi 10.1016/S0142-9612(99)00081-2. Nevelos, J. E. et al. (2001) ‘The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator’, Journal of Materials Science: Materials in Medicine, 12(2), pp. 141–144. Nevelos, J., Ingham, E., Doyle, C., Streicher, R., Nevelos, A., Walter, W. and Fisher, J. 2000. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 15(6), pp.793–795. NJR (2021) ‘National Joint Registry for England, Wales and Northern Ireland’. O’Dwyer Lancaster-Jones, O. et al. (2017) ‘An in vitro simulation model to assess the severity of edge loading and wear, due to variations in component positioning in hip joint replacements’, Journal of Biomedical Materials Research Part B: Applied Biomaterials. doi: 10.1002/jbm.b.33991. Paul, J. (1966) ‘Forces transmitted by joints in the human body’, Proceedings of the Institution of Mechanical. Available at: http://journals.sagepub.com/doi/abs/10.1243/PIME_CONF_1966_181_201_02 (Accessed: 30 June 2017). Rajesh, M. 2012. Mastering Orthopedic Techniques Total Hip Arthroplasty. Jaypee Brothers Pvt. Ltd. Revell, P.A. 2008. Joint replacement technology Preface. Joint Replacement Technology., Rieker, C.B., Schön, R., Konrad, R., Liebentritt, G., Gnepf, P., Shen, M., Roberts, P. and Grigoris, P. 2005. Influence of the clearance on in-vitro tribology of large diameter metal-on-metal articulations pertaining to resurfacing hip implants. Orthopedic Clinics of North America. 36(2), pp.135–142. Saikko, V. (1992) ‘A simulator study of friction in total replacement hip joints’, Proc Inst Mech Eng H. 1992/01/01, 206(4), pp. 201–211. Available at: http://www.ncbi.nlm.nih.gov/pubmed/1300112. Saikko, V. (1993) ‘Wear and friction properties of prosthetic joint materials evaluated on a reciprocating pin-on-flat apparatus’, Wear, 166(2), pp. 169–178. doi: 10.1016/0043-1648(93)90259-O. Saikko, V. 2009. Friction measurement in the biaxial rocking motion hip joint simulator. Journal of tribology. 131(1), p.11201. Saikko, V.O. 1996. A three-axis hip joint simulator for wear and friction studies on total hip prostheses. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 210(3), pp.175–185. Saleh, K. J. et al. (2003) ‘Functional Outcome After Revision Hip Arthroplasty’, Clinical Orthopaedics and Related Research, 416(416), pp. 254–264. doi: 10.1097/01.blo.0000093006.90435.43. Saleh, K.J., Bear, B., Bostrom, M., Wright, T. and Sculco, T.P. 2008. Initial stability of press-fit acetabular components: an in vitro biomechanical study. Am J Orthop (Belle Mead NJ). 37(10), pp.519–522. Sariali, E. et al. (2010) ‘In vitro investigation of friction under edge‐loading conditions for ceramic‐on‐ceramic total hip prosthesis’, Journal of Orthopaedic Research, 28(8), pp. 979–985. Schleichert, J. and Fröhlich, T. (2015) ‘A1.4 - Dynamic Calibration of a Multi-Component Force-/Torque Transducer’, in, pp. 46–49. doi: 10.5162/sensor2015/A1.4. Scholes, S. C. and Unsworth, A. (2000) ‘Comparison of friction and lubrication of different hip prostheses’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 214(1), pp. 49–57. Scholes, S. C. and Unsworth, A. (2006) ‘The effects of proteins on the friction and lubrication of artificial joints’, Proceedings of the Institution of Mechanical Engineers Part H-Journal of Engineering in Medicine, 220(H6), pp. 687–693. doi: Doi 10.1243/09544119jeim21. Scholes, S. C. et al. (2000) ‘The effects of material combination and lubricant on the friction of total hip prostheses’, Wear, 241(2), pp. 209–213. doi: Doi 10.1016/S0043-1648(00)00377-X. Scholes, S. C., Unsworth, A. and Goldsmith, A. A. (2000) ‘A frictional study of total hip joint replacements’, Phys Med Biol, 45(12), pp. 3721–3735. Available at: http://www.ncbi.nlm.nih.gov/pubmed/11131195. Scholl, L. et al. (2016) ‘Friction in modern total hip arthroplasty bearings: Effect of material, design, and test methodology’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 230(1), pp. 50–57. doi: 10.1177/0954411915619452. Small, Scott R, Berend Md, M.E., Howard Bs, L.A., Rogge Phd, R.D., Buckley Phd, C.A. and Ritter, M.A. 2013. High Initial Stability in Porous Titanium Acetabular Cups: A Biomechanical Study. Journal of Arthroplasty. 28, pp.510–516. Smith, A.J., Dieppe, P., Vernon, K., Porter, M., Blom, A.W., National Joint Registry of, E. and Wales 2012. Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet. 379(9822), pp.1199–1204. Smith, S.L., Dowson, D. and Goldsmith, A.A.J. 2001. The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 215(2), pp.161–170. Soballe, K., Hansen, E.S., H, B.R., Jorgensen, P.H. and Bunger, C. 1992. Tissue ingrowth into titanium and hydroxyapatite-coated implants during stable and unstable mechanical conditions. J Orthop Res. 10(2), pp.285–299. Sonntag, R., Braun, S., Al-Salehi, L., Reinders, J., Mueller, U. and Kretzer, J.P. 2017. Three-dimensional friction measurement during hip simulation J. L. Williams, ed. PLOS ONE. 12(9), p.e0184043. Sonntag, R., Reinders, J., Rieger, J.S., Heitzmann, D.W.W. and Kretzer, J.P. 2013. Hard-on-Hard Lubrication in the Artificial Hip under Dynamic Loading Conditions. PLoS One. 8(8), p.e71622. Stiehl, J.B., MacMillan, E. and Skrade, D.A. 1991. Mechanical stability of porous-coated acetabular components in total hip arthroplasty. J Arthroplasty. 6(4), pp.295–300. Taniguchi, N., Fujibayashi, S., Takemoto, M., Sasaki, K., Otsuki, B., Nakamura, T., Matsushita, T., Kokubo, T. and Matsuda, S. 2016. Effect of pore size on bone ingrowth into porous titanium implants fabricated by additive manufacturing: An in vivo experiment. Materials Science and Engineering C. 59, pp.690–701. Totten, G.E. 2006. Handbook of Lubrication and Tribology: Volume I Application and Maintenance, Second Edition. Taylor & Francis. Unsworth, A. (1978) ‘The effects of lubrication in hip prostheses’, Physics in Medicine and Biology, 23(2), p. 004. doi: 10.1088/0031-9155/23/2/004. Unsworth, A. and Pearcy, M. J. (1988) ‘Frictional properties of artificial hip joints’. Available at: http://journals.sagepub.com/doi/pdf/10.1243/EMED_JOUR_1988_017_029_02 (Accessed: 5 July 2017). Wang, A., Essner, A. and Klein, R. 2001. Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 215(2), pp.133–139. Weisenburger, J., Garvin, K. and Haider, H. (2013) ‘Friction Factors of Various Metal on Plastic Hip Replacement Designs With Scratched Femoral Heads Captured During Testing on a Multi-Station Hip Simulator’, Bone & Joint Journal Orthopaedic Proceedings Supplement, 95(SUPP 34), p. 584. Weisenburger, J.N., Naylor, M.G., Schroeder, D.W., White, B.F., Unsworth, A., Garvin, K.L. and Haider, H. 2011. A884. Correlation of friction measurements with wear characteristics during multi-station hip simulator wear tests. Journal of Bone & Joint Surgery, British Volume. 93(SUPP IV), p.448. Williams, S. et al. (2006) ‘Effect of swing phase load on metal-on-metal hip lubrication, friction and wear’, Journal of biomechanics, 39(12), pp. 2274–2281. Williams, S. et al. (2008) ‘Friction of ceramic-on-ceramic and metal-on-metal total hip replacements tested under different lubrication regimes’, Journal of Bone & Joint Surgery, British Volume, 90(SUPP I), p. 189. Williams, S., Butterfield, M., Stewart, T., Ingham, E., Stone, M. and Fisher, J. 2003. Wear and deformation of ceramic-on-polyethylene total hip replacements with joint laxity and swing phase microseparation. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 217(2), pp.147–153. Wimmer, M. A. et al. (2001) ‘Friction and wear properties of metal/metal hip joints - Application of a novel testing and analysis method’, Materialwissenschaft Und Werkstofftechnik, 32(12), pp. 891–896. doi: Doi 10.1002/1521-4052(200112)32:12<891::Aid-Mawe891>3.0.Co;2-Y. Xue, T., Attarilar, S., Liu, S., Liu, J., Song, X., Li, L., Zhao, B. and Tang, Y. 2020. Surface modification techniques of titanium and its alloys to functionally optimize their biomedical properties: Thematic review. Frontiers in Bioengineering and Biotechnology. 8, p.1261. Yan, Y. et al. (2009) ‘The influence of swing phase load on the electrochemical response, friction, and ion release of metal-on-metal hip prostheses in a friction simulator’. doi: 10.1243/13506501JET568. Zivkovic, I., Gonzalez, M. and Amirouche, F. 2010. The Effect of Under-Reaming on the Cup/Bone Interface of a Press Fit Hip Replacement. Journal of Biomechanical Engineering. 132(4), p.041008."]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"dc:type.qualificationlevel","label":"Dc Type Qualificationlevel","values":["doctoral"]},{"key":"dc:type.qualificationname","label":"Dc Type Qualificationname","values":["Integrated Ph.D and Master"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.dcterms_iso639_2","label":"Language (ISO 639-2)","values":["eng"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["uk.bl.ethos.858631"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["https://etheses.whiterose.ac.uk/id/eprint/30729/1/AAFrimpong_Thesis.pdf"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["Elevated frictional torques at the articulating interface of total hip replacements (THRs), if transferred, have the potential to affect component fixation and/or induce micro motion linked to fretting and corrosion at modular junctions. Pre-clinical testing of friction behaviour in THRs has predominantly been assessed under simplified test conditions using uniaxial simulators. These methodologies and simulators are not sufficient if assessment of friction under more clinically relevant conditions, such as multi-axis motion and consideration of surgical implant positioning, are to be investigated. To identify a suitable measuring system and develop methodologies for assessing the frictional torques in THRs, studies were conducted whereby comparisons between a newly developed electromechanical multi-axis single station hip simulator (SSHS) and an existing uniaxial hip simulator as well as a custom-built friction measuring subsystem were made. Additionally, analyses of free-body diagrams in static equilibria and functional verification testing of the SSHS was completed. The assessments conducted showed that the SSHS could detect THR frictional torques without the use of the custom-built friction subsystem; and that the sub system did not provide the additional information expected. Methodologies for assessing THR frictional torques were developed on the SSHS and used to assess the effects of translational positioning (i.e. testing when there was offset between the head and cup centres of rotation) on THR frictional torques. Increasing the applied medial-lateral (ML) translation, as might occur in a patient when there is a lateral misalignment between the centres of the femoral head and acetabular cup, under 1kN constant load and biaxial motion increased the measured frictional torques at the bearing interface of 36mm metal-on-polymer bearings, to over 15Nm. Elevated frictional torques, if transferred to the fixation interface of the acetabular cup, may induce shear stresses that could result in cup displacement. To determine whether the measured elevated frictional torques observed during simulator testing could cause potential displacement in an implanted press-fit acetabular cup, a load-to-failure test that applied a simultaneous 300N or 3kN femoral head load was developed. Results indicated that the elevated frictional torques measured under ML translation on the SSHS were not likely to cause cup displacement in 10 and 20pcf Sawbone blocks when either axial load was applied. This thesis reports on developed new methodologies for assessing THR frictional torques under variations in surgical positioning and for assessing torques required to cause cup loosening under axial loading in load-to-failure tests. It was demonstrated for the first time using a multi-axis hip simulator that surgical implant malpositioning can result in elevated frictional torques at the bearing interface of THRs. However, the impact of these elevated frictional torques on cup fixation requires further work to examine surgical implant malpositioning in THRs under more representative loading and motion conditions."]},{"key":"dc:format","label":"Dc Format","values":["text"]},{"key":"dcterms:accessrights","label":"Dcterms Accessrights","values":["public"]},{"key":"dc:title","label":"Title","values":["Torques in total hip replacements"]}]}],"canonical_facts":{"dc:contributor.advisor":["Williams, Sophie","Wilcox, Ruth"],"dc:contributor.sponsor":["EPSRC","University of Leeds","DePuy Synthes"],"dc:creator":["Frimpong, Ama Amponsah"],"dc:date":["2021-11"],"dc:date.issued":["2021-11"],"dc:description.abstract":["Elevated frictional torques at the articulating interface of total hip replacements (THRs), if transferred, have the potential to affect component fixation and/or induce micro motion linked to fretting and corrosion at modular junctions. Pre-clinical testing of friction behaviour in THRs has predominantly been assessed under simplified test conditions using uniaxial simulators. These methodologies and simulators are not sufficient if assessment of friction under more clinically relevant conditions, such as multi-axis motion and consideration of surgical implant positioning, are to be investigated. To identify a suitable measuring system and develop methodologies for assessing the frictional torques in THRs, studies were conducted whereby comparisons between a newly developed electromechanical multi-axis single station hip simulator (SSHS) and an existing uniaxial hip simulator as well as a custom-built friction measuring subsystem were made. Additionally, analyses of free-body diagrams in static equilibria and functional verification testing of the SSHS was completed. The assessments conducted showed that the SSHS could detect THR frictional torques without the use of the custom-built friction subsystem; and that the sub system did not provide the additional information expected. Methodologies for assessing THR frictional torques were developed on the SSHS and used to assess the effects of translational positioning (i.e. testing when there was offset between the head and cup centres of rotation) on THR frictional torques. Increasing the applied medial-lateral (ML) translation, as might occur in a patient when there is a lateral misalignment between the centres of the femoral head and acetabular cup, under 1kN constant load and biaxial motion increased the measured frictional torques at the bearing interface of 36mm metal-on-polymer bearings, to over 15Nm. Elevated frictional torques, if transferred to the fixation interface of the acetabular cup, may induce shear stresses that could result in cup displacement. To determine whether the measured elevated frictional torques observed during simulator testing could cause potential displacement in an implanted press-fit acetabular cup, a load-to-failure test that applied a simultaneous 300N or 3kN femoral head load was developed. Results indicated that the elevated frictional torques measured under ML translation on the SSHS were not likely to cause cup displacement in 10 and 20pcf Sawbone blocks when either axial load was applied. This thesis reports on developed new methodologies for assessing THR frictional torques under variations in surgical positioning and for assessing torques required to cause cup loosening under axial loading in load-to-failure tests. It was demonstrated for the first time using a multi-axis hip simulator that surgical implant malpositioning can result in elevated frictional torques at the bearing interface of THRs. However, the impact of these elevated frictional torques on cup fixation requires further work to examine surgical implant malpositioning in THRs under more representative loading and motion conditions."],"dc:format":["text"],"dc:identifier":["uk.bl.ethos.858631"],"dc:identifier.uri":["https://etheses.whiterose.ac.uk/id/eprint/30729/1/AAFrimpong_Thesis.pdf"],"dc:language.dcterms_iso639_2":["eng"],"dc:publisher.department":["School of Mechanical Engineering (Leeds)"],"dc:publisher.institution":["University of Leeds"],"dc:relation.isreferencedby":["https://etheses.whiterose.ac.uk/id/eprint/30729/"],"dc:relation.references":["Adler, E., Stuchin, S.A. and Kummer, F.J. 1992. Stability of press-fit acetabular cups. J Arthroplasty. 7(3), pp.295–301. Affatato, S. 2012. Wear of orthopaedic implants and artificial joints. Elsevier. Al‐Hajjar, M. et al. (2010) ‘Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic‐on‐ceramic total hip replacement’, Journal of Biomedical Materials Research Part B: Applied Biomaterials, 95(2), pp. 263–268. Al-Hajjar, M. et al. (2013) ‘Edge Loading in Total Hip Replacements Due to Translational Surgical Mal-Positioning: In Vitro Techniques’, Bone & Joint Journal Orthopaedic Proceedings Supplement, 95-B(SUPP 34), p. 2. Available at: http://www.bjjprocs.boneandjoint.org.uk/content/95-B/SUPP_34/2.abstract. Al-Hajjar, M. et al. (2014) ‘Wear of Large Diameter Ceramic-on-Ceramic Bearings in Total Hip Replacements under Edge Loading Conditions’, in ORS Annual Meeting 2014. New Orleans. Al-Hajjar, M. et al. (2015) ‘Increased Torque in Hip Replacement Bearings due to Translational Surgical Mal-Positioning; ORS Poster No. 1759’, in. Al-Hajjar, M., Clarkson, P., Williams, S., Jennings, L., Thompson, J. and Fisher, J. 2013. Edge Loading in Total Hip Replacements Due to Translational Surgical Mal-Positioning: In Vitro Techniques. Bone & Joint Journal Orthopaedic Proceedings Supplement. 95-B(SUPP 34), p.2. Al-Hajjar, M., Fisher, J., Hardaker, C., Kuring, G., Thompson, J. and Wiliams, S. 2014. Wear of Large Diameter Ceramic-on-Ceramic Bearings in Total Hip Replacements under Edge Loading Conditions In: ORS Annual Meeting 2014. New Orleans. Al‐Hajjar, M., Leslie, I.J., Tipper, J., Williams, S., Fisher, J. and Jennings, L.M. 2010. Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic‐on‐ceramic total hip replacement. Journal of Biomedical Materials Research Part B: Applied Biomaterials. 95(2), pp.263–268. Ali, M. et al. (2016) ‘Wear and Deformation of Metal-on-Polyethylene Hip Replacements under Variations in Surgical Positioning’, in. Ali, M. et al. (2017) ‘Variations in Component Positioning Leading to Dynamic Separation and Edge Loading Influences the Deformation and Wear of Metal-on-Polyethylene Hip Replacement Bearings’, ORS 2017 Annual Meeting, 4(5). Available at: http://www.ors.org/Transactions/63/1032.pdf (Accessed: 14 September 2017). Antoniades, G., Smith, E.J., Deakin, A.H., Wearing, S.C. and Sarungi, M. 2013. Primary stability of two uncementedacetabular components of different geometry: hemispherical or peripherallyenhanced? Bone & joint research. 2(12), pp.264–9. Apostu, D., Lucaciu, O., Berce, C., Lucaciu, D. and Cosma, D. 2018. Current methods of preventing aseptic loosening and improving osseointegration of titanium implants in cementless total hip arthroplasty: a review. Journal of International Medical Research. 46(6), pp.2104–2119. Apostu, D., Lucaciu, O., Lucaciu, G.D.O., Crisan, B., Crisan, L., Baciut, M., Onisor, F., Baciut, G., Câmpian, R.S. and Bran, S. 2017. Systemic drugs that influence titanium implant osseointegration. Drug Metabolism Reviews. 49(1), pp.92–104. Auger, D. D. et al. (1993) ‘Friction and lubrication in cushion form bearings for artificial hip joints’, ARCHIVE: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine 1989-1996 (vols 203-210), 207(18), pp. 25–33. doi: 10.1243/PIME_PROC_1993_207_265_02. Barrack, R. L. et al. (2001) ‘Virtual reality computer animation of the effect of component position and design on stability after total hip arthroplasty’, Orthopedic Clinics of North America, 32(4), pp. 569–577. doi: https://doi.org/10.1016/S0030-5898(05)70227-3. Bergmann, G. et al. (2012) ‘High-Tech Hip Implant for Wireless Temperature Measurements In Vivo’, PLoS ONE. Edited by H. A. Awad, 7(8), p. e43489. doi: 10.1371/journal.pone.0043489. Berry, D.J. and Lieberman, J. 2012. Surgery of the Hip. Elsevier Health Sciences. Bishop, N. E., Hothan, A. and Morlock, M. M. (2013) ‘High friction moments in large hard-on-hard hip replacement bearings in conditions of poor lubrication’, J Orthop Res. 2012/12/15, 31(5), pp. 807–813. doi: 10.1002/jor.22255. Bishop, N. E., Waldow, F. and Morlock, M. M. (2008) ‘Friction moments of large metal-on-metal hip joint bearings and other modern designs’, Med Eng Phys, 30(8), pp. 1057–1064. doi: DOI 10.1016/j.medengphy.2008.01.001. Bishop, N.E., Hothan, A. and Morlock, M.M. 2013. High friction moments in large hard-on-hard hip replacement bearings in conditions of poor lubrication. J Orthop Res. 31(5), pp.807–813. Bowsher, J. G. and Shelton, J. C. (2001) ‘A hip simulator study of the influence of patient activity level on the wear of crosslinked polyethylene under smooth and roughened femoral conditions’, Wear, 250(1–12), pp. 167–179. doi: http://dx.doi.org/10.1016/S0043-1648(01)00619-6. Bowsher, J.G., Hussain, A., Williams, P.A. and Shelton, J.C. 2005. Large head diameters have the potential to reduce ion release in metal-on metal hip wear simulations. ORS Washington DC., p.1626. Brockett, C. L. et al. (2008) ‘The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements’, Journal of Materials Science: Materials in Medicine, 19(4), pp. 1575–1579. Brockett, C. L. et al. (2013) ‘Squeaking hip arthroplasties: a tribological phenomenon’, J Arthroplasty. 2012/04/07, 28(1), pp. 90–97. doi: 10.1016/j.arth.2012.01.023. Brockett, C., Williams, S., Jin, Z., Isaac, G. and Fisher, J. 2007. Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res B Appl Biomater. 81(2), pp.508–515. Callaghan, J.J., Rosenberg, A.G. and Rubash, H.E. 2007. The Adult Hip. Lippincott Williams & Wilkins. Cameron, H.U., Pilliar, R.M. and MacNab, I. 1973. The effect of movement on the bonding of porous metal to bone. J Biomed Mater Res. 7(4), pp.301–311. CeramTec (2012) BIOLOX®delta. Available at: https://www.ceramtec.com/ceramic-materials/biolox/delta/ (Accessed: 4 July 2017). Charnley, J. (1961) ‘Arthroplasty of Hip - a New Operation’, Lancet, 1(718), pp. 1129-. Charnley, J., Mckee, G.K., Coltart, W.D. and Scales, J.T. 1961. Arthroplasty of the Hip by the Low-Friction Technique. Journal of Bone and Joint Surgery-British Volume. 43(3), p.601. Cherian, J.J., Jauregui, J.J., Banerjee, S., Pierce, T. and Mont, M.A. 2015. What Host Factors Affect Aseptic Loosening After THA and TKA? Clinical Orthopaedics and Related Research. 473(8), pp.2700–2709. Clarke, H.J., Jinnah, R.H., Warden, K.E., Cox, Q.G. and Curtis, M.J. 1991. Evaluation of acetabular stability in uncemented prostheses. J Arthroplasty. 6(4), pp.335–340. Cowin, S.C. (Stephen C. 2001. Bone mechanics handbook. CRC Press. Crosnier, E.A., Keogh, P.S. and Miles, A.W. 2014. A novel method to assess primary stability of press-fit acetabular cups. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 228(11), pp.1126–1134. Crowe, J. F., Sculco, T. P. and Kahn, B. (2003) ‘Revision Total Hip Arthroplasty’, Clinical Orthopaedics and Related Research, 413(413), pp. 175–182. doi: 10.1097/01.blo.0000072469.32680.b6. Curtis, M.J., Jinnah, R.H., Wilson, V.D. and Hungerford, D.S. 1992. The initial stability of uncemented acetabular components. J Bone Joint Surg Br. 74(3), pp.372–376. Damm, P. et al. (2013) ‘Friction in total hip joint prosthesis measured in vivo during walking’, PLoS One, 8(11), p. e78373. doi: 10.1371/journal.pone.0078373. Davim, J.P. 2013. Biotribology. Wiley. Dennis, D. A. et al. (2001) ‘\"In vivo\" determination of hip joint separation and the forces generated due to impact loading conditions.’, Journal of biomechanics, 34(5), pp. 623–9. Available at: http://www.ncbi.nlm.nih.gov/pubmed/11311703 (Accessed: 13 September 2017). Dowson, D. et al. (1985) ‘Influence of Counterface Topography on the Wear of Ultra High Molecular Weight Polyethylene Under Wet or Dry Conditions’, in, pp. 171–187. doi: 10.1021/bk-1985-0287.ch012. Dowson, D., Dalmaz, G. and Lubrecht, A. A. (2003) Tribological Research and Design for Engineering Systems: Proceedings of the 29th Leeds-Lyon Symposium. Elsevier Science. Dowson, D., Hardaker, C., Flett, M. and Isaac, G.H. 2004. A hip joint simulator study of the performance of metal-on-metal joints: Part II: design. J Arthroplasty. 19(8 Suppl 3), pp.124–130. Fisher, J. 2011. Bioengineering reasons for the failure of metal-on-metal hip prostheses: an engineer’s perspective. J Bone Joint Surg Br. 93(8), pp.1001–1004. Fisher, J. and Dowson, D. 1991. Tribology of total artificial joints. Proc Inst Mech Eng H. 205(2), p.73. Gore, T. A., Higginson, G. R. and Kornberg, R. E. (1981) ‘Some evidence of squeeze-film lubrication in hip prostheses’. Available at: http://journals.sagepub.com/doi/pdf/10.1243/EMED_JOUR_1981_010_026_02 (Accessed: 5 July 2017). Haider, H., Weisenburger, J. N. and Garvin, K. L. (2016) ‘Simultaneous measurement of friction and wear in hip simulators’, Proc Inst Mech Eng H, 230(5), pp. 373–388. doi: 10.1177/0954411916644476. Hall, R.M. and Unsworth, A. 1997. Friction in hip prostheses. Biomaterials. 18(15), pp.1017–1026. Harris, W.H. 2012. Edge Loading Has a Paradoxical Effect on Wear in Metal-on-Polyethylene Total Hip Arthroplasties. Clinical Orthopaedics and Related Research. 470(11), pp.3077–3082. Helgason, B., Perilli, E., Schileo, E., Taddei, F., Brynjó Lfsson, S. and Viceconti, M. 2008. Mathematical relationships between bone density and mechanical properties: A literature review. Hong, D., Li, C. and Jeong, J. (2012) ‘A Crosstalk Compensation of a Multi-axis ForceTorque Sensor Based on the Least Squares Method Using LabVIEW’, in 2012 Fourth International Conference on Computational and Information Sciences, pp. 1127–1130. doi: 10.1109/ICCIS.2012.12. Hua, X. et al. (2014) ‘Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions’, Journal of biomechanics, 47(13), pp. 3303–3309. Hua, X., Li, J., Wang, L., Jin, Z., Wilcox, R. and Fisher, J. 2014. Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions. Journal of biomechanics. 47(13), pp.3303–3309. Huber, W.O. and Noble, P.C. 2014. Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins. International Orthopaedics. 38(4), pp.725–731. Jin, Z.M., Dowson, D. and Fisher, J. 1994. A parametric analysis of the contact stress in ultra-high molecular weight polyethylene acetabular cups. Medical Engineering & Physics. 16(5), pp.398–405. Jin, Z.M., Meakins, S., Morlock, M.M., Parsons, P., Hardaker, C., Flett, M. and Isaac, G. 2006. Deformation of press-fitted metallic resurfacing cups. Part 1: Experimental simulation. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 220(2), pp.299–309. Jin, Z.M., Stone, M., Ingham, E. and Fisher, J. 2006. (v) Biotribology. Current Orthopaedics. 20(1), pp.32–40. Johnson, A. T. (Arthur T. (2011) Biology for engineers. CRC Press. Kody, M.H., Kabo, J.M., Markolf, K.L., Dorey, F.J. and Amstutz, H.C. 1990. Strength of initial mechanical fixation of screw ring acetabular components. Clin Orthop Relat Res. (257), pp.146–153. Lachiewicz, P.F., Suh, P.B. and Gilbert, J.A. 1989. In vitro initial fixation of porous-coated acetabular total hip components. J Arthroplasty. 4(3), pp.201–205. Leng, J. et al. (2017) ‘Dynamic virtual simulation of the occurrence and severity of edge loading in hip replacements associated with variation in the rotational and translational surgical position’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 231(4), pp. 299–306. doi: 10.1177/0954411917693261. Lewinnek, G. E. et al. (1978) ‘Dislocations after total hip-replacement arthroplasties’, J Bone Joint Surg Am, 60(2), pp. 217–220. Liao, Y.-S. et al. (2003) ‘The effect of frictional heating and forced cooling on the serum lubricant and wear of UHMW polyethylene cups against cobalt–chromium and zirconia balls’, Biomaterials, 24(18), pp. 3047–3059. Liu, F. et al. (2013) ‘Effect of head contact on the rim of the cup on the offset loading and torque in hip joint replacement’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, p. 0954411913496016. Livermore, J., Ilstrup, D. and Morrey, B. 1990. Effect of femoral head size on wear of the polyethylene acetabular component. The Journal of Bone & Joint Surgery. 72(4), pp.518–528. Lombardi, A. V. et al. (2000) ‘An in vivo determination of total hip arthroplasty pistoning during activity’, The Journal of Arthroplasty, 15(6), pp. 702–709. doi: 10.1054/arth.2000.6637. Longaray, J. et al. (2013) ‘How Do Bearing Size and Bearing Material Affect Frictional Torque? a Hip Simulator Evaluation’, Bone & Joint Journal Orthopaedic Proceedings Supplement, 95(SUPP 34), p. 121. Macdonald, W., Carlsson, L. V, Charnley, G.J. and Jacobsson, C.M. 1999. Press-fit acetabular cup fixation: principles and testing. Proc Inst Mech Eng H. 213(1), pp.33–39. Malik, S. S. (2015) Orthopaedic biomechanics made easy. McCollum, D. E. and Gray, W. J. (1990) ‘Dislocation after total hip arthroplasty. Causes and prevention.’, Clinical orthopaedics and related research, (261), pp. 159–170. McKellop, H. et al. (1981) ‘Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening device’, Journal of Biomedical Materials Research, 15(5), pp. 619–653. doi: 10.1002/jbm.820150503. McMinn, D. J. W. (2009) Modern Hip Resurfacing. Springer. Meneghini, R.M., Meyer, C., Buckley, C.A., Hanssen, A.D. and Lewallen, D.G. 2010. Mechanical Stability of Novel Highly Porous Metal Acetabular Components in Revision Total Hip Arthroplasty. J Arthroplasty. 25(3), pp.337–341. Miller, M.D., Thompson, S.R. and Hart, J. 2012. Review of Orthopaedics. Elsevier Health Sciences. Miyakawa, S., Kawamura, H., Mishima, H. and Yasumoto, J. 2004. Grit-blasted and hydroxyapatite-coated total hip arthroplasty: An 11- to 14-year follow-up study. Journal of Orthopaedic Science. 9(5), pp.462–467. Morlock, M., Bishop, N. and Huber, G. (2011) ‘Biomechanics of Hip Arthroplasty’, in Knahr, K. (ed.) Tribology in Total Hip Arthroplasty. Springer Berlin Heidelberg, pp. 11–24. doi: 10.1007/978-3-642-19429-0_2. Nevelos, J. E. et al. (1999) ‘Analysis of retrieved alumina ceramic components from Mittelmeier total hip prostheses’, Biomaterials, 20(19), pp. 1833–1840. doi: Doi 10.1016/S0142-9612(99)00081-2. Nevelos, J. E. et al. (2001) ‘The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator’, Journal of Materials Science: Materials in Medicine, 12(2), pp. 141–144. Nevelos, J., Ingham, E., Doyle, C., Streicher, R., Nevelos, A., Walter, W. and Fisher, J. 2000. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty. 15(6), pp.793–795. NJR (2021) ‘National Joint Registry for England, Wales and Northern Ireland’. O’Dwyer Lancaster-Jones, O. et al. (2017) ‘An in vitro simulation model to assess the severity of edge loading and wear, due to variations in component positioning in hip joint replacements’, Journal of Biomedical Materials Research Part B: Applied Biomaterials. doi: 10.1002/jbm.b.33991. Paul, J. (1966) ‘Forces transmitted by joints in the human body’, Proceedings of the Institution of Mechanical. Available at: http://journals.sagepub.com/doi/abs/10.1243/PIME_CONF_1966_181_201_02 (Accessed: 30 June 2017). Rajesh, M. 2012. Mastering Orthopedic Techniques Total Hip Arthroplasty. Jaypee Brothers Pvt. Ltd. Revell, P.A. 2008. Joint replacement technology Preface. Joint Replacement Technology., Rieker, C.B., Schön, R., Konrad, R., Liebentritt, G., Gnepf, P., Shen, M., Roberts, P. and Grigoris, P. 2005. Influence of the clearance on in-vitro tribology of large diameter metal-on-metal articulations pertaining to resurfacing hip implants. Orthopedic Clinics of North America. 36(2), pp.135–142. Saikko, V. (1992) ‘A simulator study of friction in total replacement hip joints’, Proc Inst Mech Eng H. 1992/01/01, 206(4), pp. 201–211. Available at: http://www.ncbi.nlm.nih.gov/pubmed/1300112. Saikko, V. (1993) ‘Wear and friction properties of prosthetic joint materials evaluated on a reciprocating pin-on-flat apparatus’, Wear, 166(2), pp. 169–178. doi: 10.1016/0043-1648(93)90259-O. Saikko, V. 2009. Friction measurement in the biaxial rocking motion hip joint simulator. Journal of tribology. 131(1), p.11201. Saikko, V.O. 1996. A three-axis hip joint simulator for wear and friction studies on total hip prostheses. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 210(3), pp.175–185. Saleh, K. J. et al. (2003) ‘Functional Outcome After Revision Hip Arthroplasty’, Clinical Orthopaedics and Related Research, 416(416), pp. 254–264. doi: 10.1097/01.blo.0000093006.90435.43. Saleh, K.J., Bear, B., Bostrom, M., Wright, T. and Sculco, T.P. 2008. Initial stability of press-fit acetabular components: an in vitro biomechanical study. Am J Orthop (Belle Mead NJ). 37(10), pp.519–522. Sariali, E. et al. (2010) ‘In vitro investigation of friction under edge‐loading conditions for ceramic‐on‐ceramic total hip prosthesis’, Journal of Orthopaedic Research, 28(8), pp. 979–985. Schleichert, J. and Fröhlich, T. (2015) ‘A1.4 - Dynamic Calibration of a Multi-Component Force-/Torque Transducer’, in, pp. 46–49. doi: 10.5162/sensor2015/A1.4. Scholes, S. C. and Unsworth, A. (2000) ‘Comparison of friction and lubrication of different hip prostheses’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 214(1), pp. 49–57. Scholes, S. C. and Unsworth, A. (2006) ‘The effects of proteins on the friction and lubrication of artificial joints’, Proceedings of the Institution of Mechanical Engineers Part H-Journal of Engineering in Medicine, 220(H6), pp. 687–693. doi: Doi 10.1243/09544119jeim21. Scholes, S. C. et al. (2000) ‘The effects of material combination and lubricant on the friction of total hip prostheses’, Wear, 241(2), pp. 209–213. doi: Doi 10.1016/S0043-1648(00)00377-X. Scholes, S. C., Unsworth, A. and Goldsmith, A. A. (2000) ‘A frictional study of total hip joint replacements’, Phys Med Biol, 45(12), pp. 3721–3735. Available at: http://www.ncbi.nlm.nih.gov/pubmed/11131195. Scholl, L. et al. (2016) ‘Friction in modern total hip arthroplasty bearings: Effect of material, design, and test methodology’, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 230(1), pp. 50–57. doi: 10.1177/0954411915619452. Small, Scott R, Berend Md, M.E., Howard Bs, L.A., Rogge Phd, R.D., Buckley Phd, C.A. and Ritter, M.A. 2013. High Initial Stability in Porous Titanium Acetabular Cups: A Biomechanical Study. Journal of Arthroplasty. 28, pp.510–516. Smith, A.J., Dieppe, P., Vernon, K., Porter, M., Blom, A.W., National Joint Registry of, E. and Wales 2012. Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet. 379(9822), pp.1199–1204. Smith, S.L., Dowson, D. and Goldsmith, A.A.J. 2001. The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 215(2), pp.161–170. Soballe, K., Hansen, E.S., H, B.R., Jorgensen, P.H. and Bunger, C. 1992. Tissue ingrowth into titanium and hydroxyapatite-coated implants during stable and unstable mechanical conditions. J Orthop Res. 10(2), pp.285–299. Sonntag, R., Braun, S., Al-Salehi, L., Reinders, J., Mueller, U. and Kretzer, J.P. 2017. Three-dimensional friction measurement during hip simulation J. L. Williams, ed. PLOS ONE. 12(9), p.e0184043. Sonntag, R., Reinders, J., Rieger, J.S., Heitzmann, D.W.W. and Kretzer, J.P. 2013. Hard-on-Hard Lubrication in the Artificial Hip under Dynamic Loading Conditions. PLoS One. 8(8), p.e71622. Stiehl, J.B., MacMillan, E. and Skrade, D.A. 1991. Mechanical stability of porous-coated acetabular components in total hip arthroplasty. J Arthroplasty. 6(4), pp.295–300. Taniguchi, N., Fujibayashi, S., Takemoto, M., Sasaki, K., Otsuki, B., Nakamura, T., Matsushita, T., Kokubo, T. and Matsuda, S. 2016. Effect of pore size on bone ingrowth into porous titanium implants fabricated by additive manufacturing: An in vivo experiment. Materials Science and Engineering C. 59, pp.690–701. Totten, G.E. 2006. Handbook of Lubrication and Tribology: Volume I Application and Maintenance, Second Edition. Taylor & Francis. Unsworth, A. (1978) ‘The effects of lubrication in hip prostheses’, Physics in Medicine and Biology, 23(2), p. 004. doi: 10.1088/0031-9155/23/2/004. Unsworth, A. and Pearcy, M. J. (1988) ‘Frictional properties of artificial hip joints’. Available at: http://journals.sagepub.com/doi/pdf/10.1243/EMED_JOUR_1988_017_029_02 (Accessed: 5 July 2017). Wang, A., Essner, A. and Klein, R. 2001. Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 215(2), pp.133–139. Weisenburger, J., Garvin, K. and Haider, H. (2013) ‘Friction Factors of Various Metal on Plastic Hip Replacement Designs With Scratched Femoral Heads Captured During Testing on a Multi-Station Hip Simulator’, Bone & Joint Journal Orthopaedic Proceedings Supplement, 95(SUPP 34), p. 584. Weisenburger, J.N., Naylor, M.G., Schroeder, D.W., White, B.F., Unsworth, A., Garvin, K.L. and Haider, H. 2011. A884. Correlation of friction measurements with wear characteristics during multi-station hip simulator wear tests. Journal of Bone & Joint Surgery, British Volume. 93(SUPP IV), p.448. Williams, S. et al. (2006) ‘Effect of swing phase load on metal-on-metal hip lubrication, friction and wear’, Journal of biomechanics, 39(12), pp. 2274–2281. Williams, S. et al. (2008) ‘Friction of ceramic-on-ceramic and metal-on-metal total hip replacements tested under different lubrication regimes’, Journal of Bone & Joint Surgery, British Volume, 90(SUPP I), p. 189. Williams, S., Butterfield, M., Stewart, T., Ingham, E., Stone, M. and Fisher, J. 2003. Wear and deformation of ceramic-on-polyethylene total hip replacements with joint laxity and swing phase microseparation. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 217(2), pp.147–153. Wimmer, M. A. et al. (2001) ‘Friction and wear properties of metal/metal hip joints - Application of a novel testing and analysis method’, Materialwissenschaft Und Werkstofftechnik, 32(12), pp. 891–896. doi: Doi 10.1002/1521-4052(200112)32:12<891::Aid-Mawe891>3.0.Co;2-Y. Xue, T., Attarilar, S., Liu, S., Liu, J., Song, X., Li, L., Zhao, B. and Tang, Y. 2020. Surface modification techniques of titanium and its alloys to functionally optimize their biomedical properties: Thematic review. Frontiers in Bioengineering and Biotechnology. 8, p.1261. Yan, Y. et al. (2009) ‘The influence of swing phase load on the electrochemical response, friction, and ion release of metal-on-metal hip prostheses in a friction simulator’. doi: 10.1243/13506501JET568. Zivkovic, I., Gonzalez, M. and Amirouche, F. 2010. The Effect of Under-Reaming on the Cup/Bone Interface of a Press Fit Hip Replacement. Journal of Biomechanical Engineering. 132(4), p.041008."],"dcterms:accessrights":["public"],"dc:title":["Torques in total hip replacements"],"dc:type":["Thesis"],"dc:type.qualificationlevel":["doctoral"],"dc:type.qualificationname":["Integrated Ph.D and Master"]},"updated_at":"2026-11-18T23:18:02Z"}