Spelling suggestions: "subject:"intratracheal anesthesia"" "subject:"intratracheale anesthesia""
1 |
Endotracheal suctioning via an adaptor effects on oxygenation and airway pressuresDavid, Deborah Sue. January 1984 (has links)
Thesis (M.S.)--University of Wisconsin--Madison, 1984. / Typescript. eContent provider-neutral record in process. Description based on print version record. Includes bibliographical references (leaves 62-64).
|
2 |
The effects of hyperoxygenation and hyperinflation on arterial blood gases prior to endotracheal suctionConverse, Amy Lees. January 1980 (has links)
Thesis (M.S.)--University of Wisconsin--Madison. / Typescript. eContent provider-neutral record in process. Description based on print version record. Includes bibliographical references (leaves 81-84).
|
3 |
The effectiveness of a double-lumen oxygen-suction catheter during endotracheal suctioning on arterial blood gas parametersWashburn, Susan C. January 1979 (has links)
Thesis (M.S.)--University of Wisconsin--Madison. / Typescript. eContent provider-neutral record in process. Description based on print version record. Includes bibliographical references (leaves 81-84).
|
4 |
A prospective comparative study of continuous and intermittent endotracheal tube cuff pressure measurement in an adult intensive care unitMemela, Mduduzi Emmanuel January 2010 (has links)
Submitted in fulfilment of the Master's Degree in Clinical Technology, Durban University of Technology, 2010. / Introduction: The aim of this study was to establish the most reliable standard
method for monitoring endotracheal tube cuff pressure in an intensive care unit.
Methodology: The study was conducted at King Edward VIII Hospital ICU on adult
patients undergoing prolonged intubation of more than 24 hours. Consent was
obtained from the patient’s next of kin. The patient’s Pcuff for this study was
recorded in two ways simultaneously for a period of 12 hours during the day. The
principal investigator recorded the Pcuff thrice during the study period using the
Posey cufflator®. Continuous recording was done using a pressure transducer
connected to the Nihon Kohden BSM®. Factors causing changes in Pcuff were also
documented.
Results: Thirty-five critically ill adult patients were enrolled into the study. Nineteen
(54.3%) of the subjects were male. Seventeen out of 35 subjects were studied for
the entire 720 minute period. The mean time of study of the group was 667 minutes
with the lowest period being 135 minutes for one patient. The group mean ±
Standard deviation (SD) was 26.6 8.7 with a 95% confidence index of 9.2 – 44.0
and the median value was 25 for continuous readings. For the entire group, 13% of
the time was spent in the low pressure range (< 20 cmH2O), while 23% was spent in
the high pressure (> 30 cmH2O). A mean of 64% of the time was spent in the normal
pressure range. Overall, the most frequently encountered events that caused
pressure changes were body movement, coughing, head movement and suctioning
accounting for 26.2%, 20.1%, 19.2% and 9.4% respectively. For intermittent
readings, the mean ± SD of all patients for T0 was 25.3 ± 6.9; for T6 25.9 ± 8.7 and
for T12 24.8 ± 3.8. The overall mean ± SD for all readings was 25.6 ± 7.1. For the
entire group, 12% of the time was spent in the low pressure range (< 20 cmH2O),
while 5% was spent in the high pressure (> 30 cmH2O). A mean of 83% of the time
was spent in the normal pressure range. The correlation between intermittent
pressure and the continuous reading at the same time was r = 0.87.
iii
Discussion: Continuous monitoring of Pcuff indicated that the endotracheal cuff
pressure varies extensively during mechanical ventilation in critically ill patients, such
variation being noted both between patients and within an individual patient. In an
attempt to compare intermittent and continuous monitoring of endotracheal cuff
pressures, a good correlation between the two measurements was demonstrated.
However, the variations in pressures noted for an individual patient would not have
been detected if endotracheal cuff pressures were monitored intermittently. Hence,
with continuous monitoring the pressure changes may be detected early.
Conclusion: Continuous monitoring of cuff pressure during mechanical ventilation in
intensive care units is thus recommended for all patients. If intermittent monitoring is
performed, it should be more frequently than eight-hourly. It is recommended that a
pressure range of 20-30 cmH2O still be used as the normal range. The role of self
adjusting pressure devices, although needing further exploration, holds much
promise. / DUT Postgraduate Development Services.
|
Page generated in 0.0781 seconds