Design and Development of a Dynamic Knee Injury SimulatorCassidy, Karla January 2009 (has links)
The knee is one of the most complex joints in the body, relying entirely on
ligaments and muscles for stabilization. With the rise in people participating in
sports, including a significant increase in female athletes, the prevalence of anterior
cruciate ligament (ACL) injuries is very evident. With recent research showing
that ACL injuries lead to osteoarthritis 10-20 years after the injury, determining
the cause of these injuries to be able to prevent them is crucial.
To date, both in-vivo and in-vitro techniques have been used to analyze the
influences of the ACL injury including neuromuscular, anatomical, and kinematic.
In-vivo techniques used to investigate knee kinematics is limited by the inability
to take real ACL strain measurements while in-vitro techniques used to investigate
anatomical considerations is limited by the inability to apply true muscular and
kinematic forces.
The purpose of the present thesis is to show the design and validation of a dynamic knee injury simulator. The simulator puts a cadaver knee, original ligaments
and patellar tendon still attached, through motions which put the ACL at a high-risk of injury with realistic influence of muscles. The muscular influences are
applied with actuators pulling the same force pro les as natural muscles. To get
realistic muscle pro files, Anybody Software is used. Anybody Software is a modeling
software which puts a skeletal system through prescribed motions and using an
optimization algorithm calculates the muscle force pro file. The motion of the knee
in the sagittal plane is simulated with actuators.
The simulator consists of four actuators which are used in force control mode to
add the muscle influence to the knee. Another two belt actuators are used for the
joint motions, one each for the hip and ankle. The hip will move along the resultant
Z motion and the ankle will move along the resultant Y motion. Simple gait is
used for initial validation, the actuators chosen have speed and force capabilities
for high-risk motions.
The gait was successfully simulated and muscle force versus time profi le tracked
the input well. The regression coeffcient study shows very good comparison. The
hamstring muscle group is the only one which does not show very good comparison
however this is only due to the jumpy nature of the hamstring profi le. The ACL
strain fell within a similar range to published gait ACL strain data. The validation was successful, and with greater available force and speeds in the actuators, showing the use of this simulator during high-risk motions is possible.
|
2 |
Design and Development of a Dynamic Knee Injury SimulatorCassidy, Karla January 2009 (has links)
The knee is one of the most complex joints in the body, relying entirely on
ligaments and muscles for stabilization. With the rise in people participating in
sports, including a significant increase in female athletes, the prevalence of anterior
cruciate ligament (ACL) injuries is very evident. With recent research showing
that ACL injuries lead to osteoarthritis 10-20 years after the injury, determining
the cause of these injuries to be able to prevent them is crucial.
To date, both in-vivo and in-vitro techniques have been used to analyze the
influences of the ACL injury including neuromuscular, anatomical, and kinematic.
In-vivo techniques used to investigate knee kinematics is limited by the inability
to take real ACL strain measurements while in-vitro techniques used to investigate
anatomical considerations is limited by the inability to apply true muscular and
kinematic forces.
The purpose of the present thesis is to show the design and validation of a dynamic knee injury simulator. The simulator puts a cadaver knee, original ligaments
and patellar tendon still attached, through motions which put the ACL at a high-risk of injury with realistic influence of muscles. The muscular influences are
applied with actuators pulling the same force pro les as natural muscles. To get
realistic muscle pro files, Anybody Software is used. Anybody Software is a modeling
software which puts a skeletal system through prescribed motions and using an
optimization algorithm calculates the muscle force pro file. The motion of the knee
in the sagittal plane is simulated with actuators.
The simulator consists of four actuators which are used in force control mode to
add the muscle influence to the knee. Another two belt actuators are used for the
joint motions, one each for the hip and ankle. The hip will move along the resultant
Z motion and the ankle will move along the resultant Y motion. Simple gait is
used for initial validation, the actuators chosen have speed and force capabilities
for high-risk motions.
The gait was successfully simulated and muscle force versus time profi le tracked
the input well. The regression coeffcient study shows very good comparison. The
hamstring muscle group is the only one which does not show very good comparison
however this is only due to the jumpy nature of the hamstring profi le. The ACL
strain fell within a similar range to published gait ACL strain data. The validation was successful, and with greater available force and speeds in the actuators, showing the use of this simulator during high-risk motions is possible.
|
3 |
Design of a Cyclic Sliding, Dynamically Loaded Wear Testing Device for the Evaluation of Total Knee Replacement MaterialsThompson, Matthew Thomas 08 August 2001 (has links)
During normal walking, the relative motion of the human knee involves flexion/extension, anterior/posterior sliding, and medial/lateral rotation. As well, the knee experiences a complex, dynamic loading curve with a peak of up to seven times body weight. However, most wear testing machines that have been used to evaluate total knee replacement materials are unidirectional and/or apply only static force. This thesis presents an alternate wear testing device capable of simulating the most prevalent motions of the knee, and applying physiologically-correct loading to the material interface. By incorporating a CoCr disc, an UHMWPE block, stepping motors, pneumatic components, computer control, and linear tables in an x-y configuration, the device is capable of quickly screening new and alternative materials to UHMWPE before evaluating them on a much more expensive knee simulator. In addition, flexibility of the device allows programming of many different motion and loading configurations permitting materials testing under only certain circumstances, or evaluating the effects on wear of specific motions. Design rationale, development, validation, and future recommendations are presented. / Master of Science
|