• Refine Query
  • Source
  • Publication year
  • to
  • Language
  • 1
  • Tagged with
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • About
  • The Global ETD Search service is a free service for researchers to find electronic theses and dissertations. This service is provided by the Networked Digital Library of Theses and Dissertations.
    Our metadata is collected from universities around the world. If you manage a university/consortium/country archive and want to be added, details can be found on the NDLTD website.
1

En bildkvalitésutvärdering av två datortomografer i syfte att rättfärdiga ett inköp av en ny datortomograf : En fantomstudie / An Image Quality Analysis of Two CT Scanners for The Purpose of Justifying a Purchase of a New CT Scanner : A Phantom Study

Burke, Molly, Gustafsson, Linnéa January 2022 (has links)
Antal datortomografiundersökningar har ökat under flera år i Sverige tack vare tekniska utvecklingar och ökad tillgänglighet på sjukvård. Södertälje sjukhus röntgenavdelningen är i behov av att byta ut en utdaterad datortomograf (eng: Computed tomography, CT) och avdelningen för medicinsk teknik har föreslagit ett inköp av en CT med fotonräknande-detektor. Bilddata framställdes genom en fantomstudie för att påvisa förhållandet mellanstråldosparametern CTDIvol och kontrast-brus-förhållandet (CNR) hos CT-systemen: SOMATOM Drive och NAEOTOM Alpha. Den genererade datan påvisade att det finns en väsentlig skillnad i CNR-CTDIvol-förhållandet mellan SOMATOM Drive och NAEOTOM Alpha. Resultaten tydliggör att NAEOTOM Alpha kan producera bilder med betydligt mindre brus vid lägre stråldoser. Ett inköp av en fotonräknande detektor CT skulle kunna rättfärdigas utifrån bildkvalitéförbättringen som systemet kan erbjuda. / The number of computed tomography (CT) scans has increased during the past years in Sweden due to technical advancements and increased availability of healthcare. The x-ray department at Södertälje hospital is in need of replacing an outdated computed tomography and the departmentof clinical engineering has proposed a purchase of a photon-counting detector CT. Image data was produced through a phantom study to demonstrate the relationship between the parameter CTDIvol radiation dose and the contrast-to-noise ratio (CNR) of the CT systems: SOMATOM Drive and NAEOTOM Alpha. The generated data demonstrated that there is a substantial difference in the CNR-CTDIvol relationship between SOMATOM Drive and NAEOTOM Alpha. The results entail that NAEOTOM Alpha can produce images with considerably less noise at lower radiation doses. The purchase of a photon-counting CT could be justified by the improved image quality it can offer.

Page generated in 0.1927 seconds