1 |
A Noninvasive Sizing Method to Choose Fitted Amplatzer Septal Occluder by Transthoracic Echocardiography in Patients with Secundum Atrial Septal DefectsChien, Kuang-Jen 15 June 2006 (has links)
Abstract:
Background: At present, device closure of interatrial communication has become a well established technique in order to adequately treat severe left-to-right shunt associated with ASDs. During the traditional procedure, fluoroscopy with the waist of a compliant balloon is used to determine the appropriate size of the closure device and defect sizing. Choice of adequate closure device using transthoracic echocardiography (TTE) has been hitherto unreported.
Methods & Materials: Between December 2002 and October 2004, 40 patients (15 males, 25 females, mean age; 11.7 ¡Ó 7.8 years ) with secundum ASDs underwent transcatheter closure at our institution. In group 1, 30 patients had the procedure by balloon sizing and TTE sizing. In 10 patients (group 2), TTE sizing was used as the sole too l for selecting device size and the device size was chosen to be based on the Amplatzer septal occluder ( ASO ) size and TTE size ratio in group 1. The procedure was performed under continuous transoesophageal echocardiographic monitor with general anesthesia.
Results: The correlation was found between TTE and stretched balloon sizing diameter SBD ( y= 1.2645x-1.4465; R²=0.9861 ), and between TTE size and ASO size ( y = 1.3412x-1.2864; R²=0.9929 ) in group 1. In group 2, statistical correlation between TTE and ASO ( y=1.3419x-0.1172; R²=0.9934 ) was also found. Good linear regression between TTE size and ASO chosen size was noted in group 1 and group 2 (R²=0.99).In group 2, successful device implantation was accomplished in all patients whose device size was chosen to be based on the ASO and TTE ratio in group 1.
Conclusions: TTE sizing is a safe and ideal method to measure interatrial defect and choose the occluding device respectively. With our experience, the sizing based on the TTE is generally easier than measurement from the balloon sizing.
|