Διόρδωση σκέδασης σε τομογράφο εκπομπής ποζιτρονίων / Scatter correction in 3D PET

In 3D positron emission tomography the scatter effect is a significant physical
factor degrading image quality. The advancements in computing that occurred the
last decades al lowed us to simulate the scatter coincidences fast and ef ficiently.
The main concern now is how accurately do we simulate the scatter events. The
scope of this project is the implementation and the evaluation of a scatter
simulation algorithm that would be able to simulate the scat ter ef fect more
precisely than the existing ones.
One way to simulate the scatter distribut ion is with an algorithm, first published by
Watson and Ollinger, that is based on the Klein-Nishina formula. These methods
have been implemented taking into account only the single scatter events (where a
photon scatters once in the at tenuation medium). Multiple scatter is generally taken
into account by some scaling or filtering procedure. Their main advantages are
short computational time and relatively good precision compared to previous more
heuristical methods. Although these single scatter algorithms have been effective
there are cases where their results are not that accurate. For low energy thresholds
and large at tenuation mediums multiple scatter is increased. Given that a
significant percentage of people are over -weight (thus the at tenuation medium has
large volume) we should consider introducing multiple scatter events in our
simulations. Moreover, the distribut ion of all scatter events is broader than the one
of single scatter events therefore even if the single scatter distribution is scaled it
will not match the total scatter one.
In previous work by C. Tsoumpas et al, a new scatter simulation algorithm was
developed that attempts to approximate the total scatter distribution by taking into
account the case where the one annihilated photon is scattered twice and the case
where both annihilated photons are scattered once. These two cases describe the
double scatter events and by introducing them into our scatter simulation algorithm
we aim to obtain a better approximation of the total scatter distribution. In this
thesis we have improved this double scatter simulation algorithm in two important
ways. When both annihilated photons scatter they acquire a favourable polarization
direction with respect to each other and this influences thei r detect ion
probabilities, especially when low energy photons are detected. In the algorithm
that we implemented we considered this effect by using the polarized Klein-Nishina
formula for this case. In addition, we investigated and validated the need to
introduce extra solid angle factors in the implementation.
The whole implementation is based on the STIR library (Software for Tomographic
Image Reconstruction) written in the C++ programming language.
Scatter events can also be simulated by Monte Carlo simulation packages such as
SimSET. SimSET is a public domain package designed to simulate positron
emission tomography (PET) (and Single Photon Emission Tomography) and was
used extensively in this project. Monte Carlo packages because of their ability to
exclude any unknown physical parameter they can simulate physical processes like
the ones that take place in PET very accurately. Thus they were essential for the evaluation of our scatter correction algorithm. The reason why Monte Carlo
packages are not used inclinical practice instead of the model-based methods is
that they demand a large computational time.
Besides Monte Carlo packages we also per formed a series of experimental scans in
order to evaluate our scatter simulation algorithm. The tomograph used for the
experiments was the ECAT 962 used in a 3D mode. / -

Identiferoai:union.ndltd.org:upatras.gr/oai:nemertes:10889/1209
Date23 December 2008
CreatorsΔίκαιος, Νικόλαος
ContributorsΝικήτα, Κωνσταντίνα Σ., Dikaios, Nikolaos, Νικήτα, Κωνσταντίνα Σ., Ουζούνογλου, Νικόλαος, Κουτσούρης, Δημήτριος
Source SetsUniversity of Patras
Languagegr
Detected LanguageEnglish
TypeThesis
Rights0
RelationΗ ΒΥΠ διαθέτει αντίτυπο της διατριβής σε έντυπη μορφή στο βιβλιοστάσιο διδακτορικών διατριβών που βρίσκεται στο ισόγειο του κτιρίου της.

Page generated in 0.0018 seconds