Two batches of CdSe QDs with different sizes were synthesized for immobilizing in polyisoprene (PI), polymethylmethacrylate (PMMA), and low-density polyethylene (LDPE). The combinations of QDs and polymer substrates were evaluated for their analytical fit-for-use in applicable immunoassays. Hydrogen peroxide standards were injected into the flow injection analyzer (FIA) constructed to simulate enzyme-generated hydrogen peroxide reacting with bis-(2,4,6-trichlorophenyl) oxalate.
Linear correlations between hydrogen peroxide and chemilumenscent intensities yielded regression values greater than 0.9750 for hydrogen peroxide concentrations between 1.0 x 10-4 M and 1.0 x 10-1 M. The developed technique’s LOD was approximately 10 ppm. Variability of the prepared QD-polymer products was as low as 3.2% throughout all preparations.Stability of the preparations was tested during a 30-day period that displayed up to a four-fold increase in the first 10 days. The preparations were decently robust to the FIA system demonstrating up to a 15.20% intensity loss after twenty repetitive injections.
Identifer | oai:union.ndltd.org:ETSU/oai:dc.etsu.edu:etd-2338 |
Date | 01 May 2013 |
Creators | Moore, Christopher S |
Publisher | Digital Commons @ East Tennessee State University |
Source Sets | East Tennessee State University |
Detected Language | English |
Type | text |
Format | application/pdf |
Source | Electronic Theses and Dissertations |
Rights | Copyright by the authors. |
Page generated in 0.001 seconds