Return to search

Integrated electronics design for high-frequency intravascular ultrasound imaging

Close integration of front-end electronics and the transducer array within the catheter is critical for successful implementation of CMUT-based intravascular ultrasound (IVUS) imaging catheters to enable next generation imaging tools. Therefore, this research developed and implemented custom-designed electronic circuits and systems integrated with an IC compatible transducer technology for realization of miniature IVUS imaging catheters operating at 10-50 MHz frequency range.
In one path of this research, an IC is custom designed in a 0.35-um CMOS process to monolithically integrate with a CMUT array (CMUT-on-CMOS) to realize a single-chip, highly-flexible, forward-looking (FL) IVUS imaging system. The amplifiers that are custom-designed achieved transducer thermal-mechanical noise dominated receive performance in a CMUT-on-CMOS implementation. In parallel to the FL-IVUS effort, for realization of a side-looking IVUS catheter based on an annular phased array, a dynamic receive beamformer IC is custom designed also in a 0.35-um CMOS process.
Overall, the circuits and systems developed as part of this dissertation form a critical step in the translation of the research on CMUT-based IVUS catheters into real clinical applications for better management of coronary arterial diseases.

Identiferoai:union.ndltd.org:GATECH/oai:smartech.gatech.edu:1853/45860
Date19 October 2011
CreatorsGurun, Gokce
PublisherGeorgia Institute of Technology
Source SetsGeorgia Tech Electronic Thesis and Dissertation Archive
Detected LanguageEnglish
TypeDissertation

Page generated in 0.0016 seconds