Objective: The purpose of this study is to determine the relationship between cone beam CT (CBCT) voxel size and tooth root fracture detection. Materials and Methods: Vertical and horizontal root fractures were induced in a total of 30 teeth, and 15 teeth were left intact. Teeth were imaged with projection digital radiography and the Kodak 9000 3D CBCT system with a native voxel size of 76 μm. The CBCT voxels were then downsampled to 100 μm, 200 μm and 300 μm. Five blinded observers evaluated both sets of images with a 1 week washout interval between each set of observations. Results: CBCT outperformed the projection images for fracture detection for all voxel sizes except 300 μm (p<0.05). No significant differences were found between the different voxel sizes (p>0.05). Conclusion: Although voxel size does not impact the interpretation of root fractures, in vitro, CBCT outperformed projection imaging for voxel sizes less than 300 μm.
Identifer | oai:union.ndltd.org:TORONTO/oai:tspace.library.utoronto.ca:1807/42669 |
Date | 20 November 2013 |
Creators | Amintavakoli, Niloufar |
Contributors | Lam, Ernest W. |
Source Sets | University of Toronto |
Language | en_ca |
Detected Language | English |
Type | Thesis |
Page generated in 0.0017 seconds