Return to search

Comparação dos valores da pressão intraocular obtidos com diferentes tonômetros e suas correlações com dados biométricos oculares no glaucoma congênito / Comparison between intraocular pressure obtained with different tonometers and their correlations with biometric parameters in congenital glaucoma

OBJETIVOS: Comparar os valores da pressão intraocular (PIO), obtidos por intermédio do tonômetro de Perkins (TPK), tonômetro de contorno dinâmico Pascal (TCD) e Tono-Pen (TNP), confrontando-os com o tonômetro de aplanação de Goldmann (TAG), e correlacionar seus valores tonométricos com parâmetros biométricos oculares em pacientes portadores de glaucoma congênito primário. MÉTODOS: Estudo prospectivo, transversal, com inclusão de 46 pacientes (46 olhos) com diagnóstico de glaucoma congênito primário, com idades entre 12 e 40 anos, após obtenção do Termo de Consentimento Livre e Esclarecido. Todos os olhos estudados foram submetidos à tonometria ocular usando os tonômetros de Goldmann, de Perkins, o Tono-Pen e o tonômetro de contorno dinâmico Pascal. A ordem das tonometrias foi aleatória e o tonômetro de Goldmann foi adotado como padrão ouro. Os parâmetros biométricos estudados foram a curvatura (medida pela ceratometria), a espessura central da córnea (paquimetria), o diâmetro corneal (medido por meio de compasso cirúrgico) e o comprimento axial do olho foi obtido pela biometria ultrassônica. As distribuições dos parâmetros biométricos, assim como das tonometrias foram plotadas e analisadas conforme teste de Kolmogorov-Smirnov para aceitação de normalidade. O teste t de Student pareado para amostras independentes foi empregado para comparar as médias tonométricas de cada tonômetro com o padrão ouro. As correlações entre os parâmetros biométricos e as tonometrias foram realizadas através do coeficiente de correlação de Pearson e gráficos de regressão linear. O mesmo procedimento foi feito entre os parâmetros biométricos e as diferenças entre as três distintas tonometrias e a TAG. A concordância entre os tonômetros Perkins, Pascal e Tono-Pen e o de Goldamnn foi realizada pelo teste de concordância de correlação (CCC) e graficamente pelo método de Bland-Altman. O valor de corte de 2 mmHg foi adotado para avaliar a empregabilidade clínica desses tonômetros em pacientes semelhantes aos da amostra. RESULTADOS: O teste de Kolmogorov-Smirnov indicou aceitação da normalidade para todas as distribuições estudadas (TAG, TPK, TCD, TNP, Diâmetros corneal e axial, ceratometria média e paquimetria). Os momentos de correlação de Pearson empregados para realizar estudo das correlações entre os parâmetros biométricos e cada tonometria foram estatisticamente não significativos. As correlações entre os parâmetros biométricos e as diferenças entre tonometrias em relação ao TAG não apresentaram significância em sua maioria, sendo a única exceção a correlação moderada entre a diferença do TAG e o Tono-Pen versus a ceratometria média. No entanto, o coeficiente de determinação evidenciou influência modesta da ceratometria nessas diferenças (r² = 0,16; p = 0,004). O teste t de Student pareado demonstrou diferença significativa entre o TAG e o TP (p < 0,001). A diferença não foi significativa entre o TAG e o Pascal (p = 0.30), ou entre o TAG e o Tono-Pen (p = 0.68). Houve excelente concordância entre o TAG e o TP (CCC = 0,98; intervalo de confiança 95% (IC95%) = 0,97 - 0,99), já entre o TAG e o Pascal (CCC = 0,89; IC95% = 0,82 - 0,94) e entre o TAG e o Tono-Pen (CCC = 0,92; IC95% = 0,87 - 0,95). O gráfico tipo Bland-Altman TAG x Perkins demonstrou diferença média de 0,47 mmHg com intervalo de 95% (I95%) situado entre -0,98 e 1,92 mmHg. A dispersão das diferenças seguiu caráter aleatório. Os outros dois tonômetros também tiveram suas dispersões em caráter aleatório. TAG em relação ao tonômetro de Pascal apresentou diferença média de -0,3 mmHg (I95% = -4,2 a 3,6 mmHg). A diferença média do TAG em relação ao Tono-Pen foi de -0,1 mmHg (I95% = -3,7 a 3,5 mmHg). Em 21% dos pacientes, o tonômetro de Pascal apresentou diferenças maiores de 2 mmHg em relação ao TAG, ao passo que no Tono- Pen essa proporção foi de 17,3%, e o tonômetro de Perkins não apresentou em nenhum dos pacientes diferenças maiores que estes limites. CONCLUSÕES: A ceratometria, paquimetria e os diâmetros axial e corneal não se correlacionaram com a PIO obtida por meio do TAG, tonômetro de Perkins, Pascal ou Tono-Pen. As diferenças tonométricas entre TAG (padrão ouro) e os outros tonômetros também não se correlacionaram com esses parâmetros biométricos, com exceção da ceratometria média, que se correlacionou positiva e moderadamente com a diferença da PIO entre TAG e Tono-Pen. O tonômetro de Perkins apresentou concordância substancial com o TAG, já o Tono-Pen e o TCD apresentaram concordâncias moderadas, sendo a concordância do Tono-Pen maior que a do TCD / OBJECTIVES: To compare the values of intraocular pressure (IOP) obtained by Perkins tonometer (PKT), Pascal dynamic contour tonometer (DCT) and Tono-Pen (TP), comparing then with Goldmann applanation tonometry (GAT), analyzing their correlations with ocular biometric parameters in patients with primary congenital glaucoma. METHODS: Prospective and cross-sectional study, including 46 patients (46 eyes) diagnosed with primary congenital glaucoma, between 12 and 40 years old, after obtaining informed consent. Keratometry was performed, followed by Goldmann applanation tonometry, Perkins tonometry, DCT and TP. The order of tonometries was randomized. The Goldmann tonometer was adopted as the gold standard. Ultrasound pachymetry, ultrasound biometry and corneal diameter measurement with surgical compass were also performed. The distributions of biometric parameters, as well as the tonometries were plotted and analyzed using Kolmogorov-Smirnov for acceptance of normality. Paired Student\'s t test for independent samples was used to compare the means of each tonometry with the gold standard. The correlations between biometric parameters and tonometries were performed by Pearson\'s product moment correlation coefficient and linear regression plots. The same procedure was done between biometric parameters and the differences between the three distinct tonometries and Goldmann tonometry. The agreement between tonometers and the GAT was performed through concordance correlation coefficient (CCC) and graphically by the Bland-Altman method. End point of 2 mmHg was adopted to evaluate clinical employability of these tonometers in patients with similar conditions. RESULTS: The Kolmogorov-Smirnov indicated acceptance of normality for all distributions studied (GAT, PKT, DCT, TP, corneal diameter, axial length, keratometry and pachymetry). All the Pearson´s product moment correlation coefficients between biometric parameters and each tonometry were not significant. The correlations between the biometric parameters and the differences between tonometries compared to the gold standard were not significant in most cases. The only exception was a positive and moderate correlation between the difference of the GAT and Tono-Pen versus the keratometry. The determintation coefficient revealed a considerable, but no large influence of K on the differences between GAT and Tono-Pen (r² = 0.16; p = 0.004). Student\'s paired t test showed a significant difference between GAT and PT (p < 0.001). The difference was not significant between the GAT and Pascal (p = 0.30), or between the GAT and Tono-Pen (p = 0.68). There was excellent agreement between GAT and PT (CCC = 0.98, 95% confidence interval (95% CI) = 0.97 to 0.99), as between GAT and Pascal (CCC = 0.90, 95% CI = 0.82 to 0.94) and between GAT and Tono-Pen (CCC = 0.92, 95% CI = .87 to .95). Bland-Altman plot GAT x Perkins showed a mean difference of 0, 47 mmHg with 95% CI located between -0.98 and 1.92 mmHg. The distribution of the IOP differences was aleatory. The other two differences` distributions also had aleatory characteristics. When comparing GAT with Pascal, the mean difference was -0.3 mmHg (95% CI = -4.2% to 3.6 mmHg). Mean difference between GAT and Tono-Pen was -0.1 mmHg (95% CI = -3,7 to 3.5 mmHg). Pascal tonometer showed a difference greater than 2 mmHg comparing to GAT in 21% of the patients, while with Tono-Pen this ratio was 17.3% and the Perkins tonometer did not present in any patient differences greater than these limits. CONCLUSIONS: The keratometry, pachymetry, corneal diameter and axial length did not correlate with IOP obtained by GAT, Perkins tonometer, Tono-Pen or Pascal. The differences between GAT (gold standard) and the other tonometers also did not correlate with these biometric parameters, with the exception of corneal curvature, which was positive and moderately correlated with the difference in IOP between GAT and Tono-Pen. GAT and Perkins tonometer showed substantial agreement, although Tono-Pen and DCT showed moderate agreement with GAT. The concordance obtained with Tono-Pen was higher than the concordance obtained with DCT

Identiferoai:union.ndltd.org:IBICT/oai:teses.usp.br:tde-16012014-154728
Date01 November 2013
CreatorsMarcio Henrique Mendes
ContributorsAlberto Jorge Betinjane, Norma Allemann, Wilma Lelis Barboza, Ralph Cohen
PublisherUniversidade de São Paulo, Oftalmologia, USP, BR
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/doctoralThesis
Sourcereponame:Biblioteca Digital de Teses e Dissertações da USP, instname:Universidade de São Paulo, instacron:USP
Rightsinfo:eu-repo/semantics/openAccess

Page generated in 0.004 seconds