Return to search

Development of [NH3] Ammonia target for Cyclone-30 at KFSH&RC

Introduction
Nitrogen [13N] NH3 is a liquid radioisotope, produced by medical cyclotrons for nuclear medicine application and widely applied for evaluation of myocardial perfusion in clinical assessments [1,2]. Owing to its short half-life (10 minutes), the unloading procedure of the radio-active solution of [13N]NH3 from the target is crucial in saving the activity produced for patient. Therefore, an efficient technique in un-loading the radioactive solution from the target body was developed using COMSOL Multiphysics. The new design of the target with improved unloading technique resulted in 30% increase of the available 13N activity. In our experiments, 13N was produced by the 16O(p,α)13N reaction. The energy of proton beam was 16.5 MeV.

Material and Methods
A 2D model was developed using COMSOL Multiphysics to simulate the inner geometry of [13N] Ammonia target. In the 2D model, water and aluminum were used as materials for the inner body and outer boundary (walls), respectively. The physical equations used to solve the problem of allocating proper place for the loading/unloading opening is turbulent, k-ε Module being extracted from fluid flow module. FIGURE 1 shows the result of simulating water flow on the target water channels. The entrance of the pushing solution (for unloading) was designed to create a turbulent flow inside the target body and, hence, to collect most of the activity inside the target.
FIGURE 2 shows the setup for 13N production. A peristaltic pump is used to push the solution after irradiation to the hotcell at 6 ml/min flowrate. The distance from the target to the hotcell is approximately 30 meters.
Results and Conclusion
FIGURE 3 presents activity produced in milicurie (mCi) for several patient runs. The activity obtained in some experiments reached up to 330 mCi when we irradiated the target with 25 μA for 15 min. This was satisfactory for delivery to the patient at the nuclear medicine department. Moreover, purity of [13N] purity was above 95 % what meets the standard regulation for administration to a patient.

Identiferoai:union.ndltd.org:DRESDEN/oai:qucosa.de:bsz:d120-qucosa-165921
Date19 May 2015
CreatorsAlrumayan, F., Alghaith, A., Akkam, Q., Marsood, A., AlQhatani, M.
ContributorsKing Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia,, Helmholtz-Zentrum Dresden - Rossendorf,
PublisherHelmholtz-Zentrum Dresden - rossendorf
Source SetsHochschulschriftenserver (HSSS) der SLUB Dresden
LanguageEnglish
Detected LanguageEnglish
Typedoc-type:conferenceObject
Formatapplication/pdf
SourceWTTC15

Page generated in 0.0023 seconds