Return to search

Sistema dosimétrico portátil sem fio para uso em radiologia

Submitted by Isaac Francisco de Souza Dias (isaac.souzadias@ufpe.br) on 2016-06-13T19:13:46Z
No. of bitstreams: 2
license_rdf: 1232 bytes, checksum: 66e71c371cc565284e70f40736c94386 (MD5)
Dissertacao_Charles.pdf: 4536947 bytes, checksum: 646198e97695c41d56a17eb67e2ca63b (MD5) / Made available in DSpace on 2016-06-13T19:13:46Z (GMT). No. of bitstreams: 2
license_rdf: 1232 bytes, checksum: 66e71c371cc565284e70f40736c94386 (MD5)
Dissertacao_Charles.pdf: 4536947 bytes, checksum: 646198e97695c41d56a17eb67e2ca63b (MD5)
Previous issue date: 2015-08-21 / Para medir a dose em procedimentos que utilizam raios X foi desenvolvido um sistema dosimétrico.
O equipamento desenvolvido consiste de um dosímetro portátil de leitura direta com
base em detectores semicondutores e um sistema computacional de processamento de dados
instalado em um computador. A comunicação entre o dosímetro portátil e o sistema computacional
pode ser realizada via rede sem fio ou conexão USB. Diversos dosímetros podem
ser conectados a um mesmo sistema computacional. Os dados que trafegam pela rede são
transmitidos em tempo real para todos os usuários através de um protocolo de comunicação
desenvolvido especialmente para este projeto. O sistema tem potencial de aplicação tanto na
dosimetria de paciente como na monitoração ocupacional. O sistema dosimétrico foi testado
nas qualidades de radiodiagnóstico definidas pela IEC 61267 denominadas de RQR3, RQR5,
RQR8, RQRI0. Na monitoração ocupacional foram utilizadas as qualidades de radioproteção
definidas pela ISO 4037 conhecidas como 40, N60, N80, NI00 e N120. Além disso, o dosímetro
foi também avaliado com energias dos raios gama do Cs-137 e Co-60. Com base nos
resultados, pode-se observar que a resposta do dosímetro possui coeficiente de variação menor
1 % e desvio padrão inferior a 10 % nos testes de reprodutibilidade. O sistema apresenta uma
resposta linear quando exposto ao kerma no ar nas qualidades de radiação estudadas. / A dosimetric system has been developed for dose measurement in procedures which makes
use of X-rays. The developed instrument consists of a semiconductor-detector based portable
wireless dosimeter and a computational data processing system installed in a computer. Communication
between the portable dosimeter and the system installed on the computer can be
performed via wireless network or USB. Several portable dosimeters can be connected to one
computational system. The data that travel over the network are transmitted in real time
to all users through a communication protocol specifically designed for this application. The
system has potential application in both for patient dosimetry and occupational monitoring.
The dosimetric system has been tested at the IEC 61267 radiation qualities RQR3, RQR5,
RQR8 and RQR10, for radiology, and, ISO 4037 N40, N60, N80, N100 and N120, for occupational
monitoring. Moreover, the dosimeter has also been evaluated in gamma ray energies of
Cs-137 and Co-60. Based on the results, one can observe that the dosimeter response has a
coefficient of variation below 1 %, and a standard deviation lower than 10 % for reproducibility
tests. The system displays a linear response when exposed to Air Kerma for the studied
radiation qualities.

Identiferoai:union.ndltd.org:IBICT/oai:repositorio.ufpe.br:123456789/17089
Date21 August 2015
CreatorsOLIVEIRA, Charles Nilton do Prado.
ContributorsSANTOS, Edval José Pinheiro
PublisherUniversidade Federal de Pernambuco, Programa de Pos Graduacao em Engenharia Biomedica, UFPE, Brasil
Source SetsIBICT Brazilian ETDs
LanguageBreton
Detected LanguagePortuguese
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/masterThesis
Sourcereponame:Repositório Institucional da UFPE, instname:Universidade Federal de Pernambuco, instacron:UFPE
RightsAttribution-NonCommercial-NoDerivs 3.0 Brazil, http://creativecommons.org/licenses/by-nc-nd/3.0/br/, info:eu-repo/semantics/openAccess

Page generated in 0.0018 seconds