Return to search

Avaliação de dose administrada no tratamento de neoplasia ginecológica por ressonância de spin eletrônico com L-alanina / Dose evaluation administered in the treatment of gynecological câncer by eléctron spin resonance with L-alanine

Neoplasias ginecológicas são os tumores de maior incidência no público feminino. Como alternativa para o tratamento destes tumores existe a radioterapia, em que a sua eficácia depende prioritariamente da correta administração da dose prescrita ao tumor, com acurácia de ±5%, conforme estabelecido pela American Association of Physicists in Medicine (AAPM). Para verificar a administração de doses existem diversas técnicas dosimétricas, entre elas a ressonância de spin eletrônico (RSE) com L-alanina, material tecido-equivalente, que relaciona dose com a quantidade de radicais livres formados pela radiação. Portanto, o objetivo deste estudo é a verificação da dose administrada na região tumoral de pacientes diagnosticadas com câncer ginecológico submetidas à teleterapia, utilizando L-alanina e a RSE. Para comparação entre as modalidades de radioterapia, foi realizada braquiterapia ginecológica com objeto simulador, verificando a dose no reto durante o tratamento, que deve ser inferior a 65% da dose prescrita ao tumor, conforme a International Commission on Radiation Units (ICRU). Para a análise de teleterapia foi construído um aparato, para a inserção vaginal do material radio-sensível, que foi utilizado durante 5 das 25 sessões de tratamento. As doses foram determinadas com o sistema de planejamento do tratamento mediante as imagens da tomografia computadorizada, em conjunto com softwares de planejamento, e a existência de marcadores radiopacos no aparato permitiu a confirmação das doses, conforme referência óssea através de imagens de portal feitas no momento anterior ao tratamento. Após a irradiação foi feita a leitura das amostras com o espectrômetro de RSE, em que a amplitude pico a pico da linha principal do sinal, normalizado pela massa, determinou a dose entregue nas regiões de interesse. Previamente ao estudo in vivo foi efetuada a teleterapia com objeto simulador, nas mesmas condições de tratamentos in vivo, apresentando desvio médio em relação ao planejamento de (0,2 ± 3,5)%, viabilizando o estudo com pacientes. Para a primeira paciente o desvio médio obtido foi de (0,5 ± 3,7)% e para a segunda de (-0,7 ± 2,8)%. Na simulação da braquiterapia o desvio médio observado foi de (-1,3 ± 9,2)%. Por causa do pequeno número amostral de pacientes, esta pesquisa é classificada como um estudo de viabilidade, em que a utilização de L-alanina com a RSE se mostrou eficiente para o objetivo proposto, fazendo com que os resultados da teleterapia estivessem de acordo com o estipulado pela AAPM e na braquiterapia a dose administrada no reto obedecesse ao relatório 38 da ICRU. / Gynecological malignancies are the most incident female tumors. As an alternative treatment for these tumors exist the radiotherapy, which their effectiveness mainly depends on the right administration of the prescribed dose to the tumor, with an accuracy of ±5%, as established by the American Association of Physicists in Medicine (AAPM). There are a lot of dosimetric techniques for dose verification, including electron spin resonance (ESR) with L-alanine, a tissue-equivalent material, which relates dose with the amount of free radicals formed by radiation. Therefore, the aim of this study is to verify the dose administered in the tumor of patients diagnosed with gynecological cancer subjected to external beam radiation therapy using L-alanine and ESR. For comparison between the radiotherapy modalities was performed gynecological brachytherapy in phantom, verifying the dose to the rectum during treatment, which must be less than 65% of the prescribed dose to the tumor, according to the International Commission on Radiation Units (ICRU). For teletherapy analysis was constructed an apparatus for vaginal insertion of the radio-sensitive material, which was used in 5 of the 25 treatment fractions. The doses were determined with the treatment planning system with the computed tomography images, together with the planning software, and the existence of radiopaque markers in the apparatus allowed the dose confirmation, according to the bone reference through portal images made before treatment. After the irradiation the samples was read with the ESR spectrometer, wherein the peak to peak amplitude of the signal main line, normalized by weight, determined the dose delivered into the regions of interest. Prior to the in vivo study was performed the teletherapy with phantom, under the same conditions of in vivo treatments, presenting mean deviation to the planning dose of (0.2 ± 3.5)%, enabling the study with patients. For the first patient, the mean deviation obtained was of (0.5 ± 3.7)% and for the second one of (-0.7 ± 2.8)%. In the brachytherapy simulation the mean deviation observed was of (-1.3 ± 9.2)%. Due to the small sample of patients, this research is classified as a feasibility study, which the use of L-alanine with the ESR was efficient for the objective proposed, making the teletherapy results consistent with the stipulated by the AAPM, and in the brachytherapy the dose to the rectum obey the ICRU report 38.

Identiferoai:union.ndltd.org:IBICT/oai:teses.usp.br:tde-01082012-145158
Date29 June 2012
CreatorsAmanda Burg Rech
ContributorsOswaldo Baffa Filho, Patrícia Nicolucci
PublisherUniversidade de São Paulo, Física Aplicada à Medicina e Biologia, USP, BR
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/masterThesis
Sourcereponame:Biblioteca Digital de Teses e Dissertações da USP, instname:Universidade de São Paulo, instacron:USP
Rightsinfo:eu-repo/semantics/openAccess

Page generated in 0.0016 seconds