Ventilation imaging using 4D-CT is a convenient and cost effective functional imaging methodology which might be of value in radiotherapy treatment planning to spare functional lung volumes. To calculate ventilation imaging from 4D-CT we must use deformable image registration (DIR). This study validates the DIR methods and investigates the dependence of calculated ventilation on DIR methods and ventilation algorithms.
The first hypothesis is if ventilation algorithms are robust then they will be insensitive to the precise DIR used provided the DIR is accurate. The second hypothesis is that the change in Houndsfield Unit (HU) method is less dependent on the DIR used and depends more on the CT image quality due to the inherent noise of HUs in normal CT imaging.
DIR of the normal end expiration and inspiration phases of the 4D-CT images was used to correlate the voxels between the two respiratory phases. All DIR algorithms were validated using a 4D pixel-based and point-validated breathing thorax model, consisting of a 4D-CT image data set along with associated landmarks. Three different DIR algorithms, Optical Flow (OF), Diffeomorphic Demons (DD) and Diffeomorphic Morphons (DM), were retrospectively applied to the same group of 10 esophagus and 10 lung cancer cases all of which had associated 4D-CT image sets that encompassed the entire lung volume. Three different ventilation calculation algorithms were compared (Jacobian, ΔV, and HU) using the Dice similarity coefficient comparison.
In the validation of the DIR algorithms, the average target registration errors with one standard deviation for the DIR algorithms were 1.6 ± 0.7 mm, maximum 3.1 mm for OF, 1.3 ± 0.6 mm, maximum 3.3 mm for DM, 1.3 ± 0.6 mm, maximum 2.8 mm for DD, indicating registration errors were within 2 voxels.
Dependence of ventilation images on the DIR was greater for the ΔV and the Jacobian methods than for the HU method. The Dice similarity coefficient for 20% of low ventilation volume for ΔV was 0.33 ± 0.03 between OF and DM, 0.44 ± 0.05 between OF and DD and 0.51 ± 0.04 between DM and DD. The similarity comparisons for Jacobian was 0.32 ± 0.03, 0.44 ± 0.05 and 0.51 ± 0.04 respectively, and for HU 0.53 ± 0.03, 0.56 ± 0.03 and 0.76 ± 0.04 respectively.
Dependence of ventilation images on the ventilation method used showed good agreement between the ΔV and Jacobian methods but differences between these two and the HU method were significantly greater. Dice similarity coefficient for using OF as DIR was 0.86 ± 0.01 between ΔV and Jacobian, 0.28 ± 0.04 between ΔV and HU and 0.28 ± 0.04 between Jacobian and HU respectively. When using DM or DD as DIR, similar values were obtained when comparing the different ventilation calculation methods. The similarity values for 20% of the high ventilation volume were close to those found for the 20% low ventilation volume.
Mean target registration error for all three DIR methods was within one voxel suggesting that the registration done by either of the methods is quite accurate. Ventilation calculation from 4D-CT demonstrates some degree of dependency on the DIR algorithm employed. Similarities between ΔV and Jacobian are higher than between ΔV and HU and Jacobian and HU. This shows that ΔV and Jacobian are very similar, but HU is a very different ventilation calculation method.
Identifer | oai:union.ndltd.org:USF/oai:scholarcommons.usf.edu:etd-4392 |
Date | 01 January 2011 |
Creators | Latifi, Kujtim |
Publisher | Scholar Commons |
Source Sets | University of South Flordia |
Detected Language | English |
Type | text |
Format | application/pdf |
Source | Graduate Theses and Dissertations |
Rights | default |
Page generated in 0.002 seconds