This paper presents a method of parameter identification for a finite-element model of the human middle ear. The parameter values are estimated using a characterization of the difference in natural frequencies and mode shapes of the tympanic membrane between the model and the specimens. Experimental results were obtained from temporal bone specimens under sound excitation (300–3,000 Hz). The first 3 modes of the tympanic membrane could be observed with a laser scanning vibrometer and were used to estimate the stiffness parameters for the orthotropic finite-element model of the eardrum. A further point of discussion is the parameter sensitivity and its implication for the identification process. / Dieser Beitrag ist mit Zustimmung des Rechteinhabers aufgrund einer (DFG-geförderten) Allianz- bzw. Nationallizenz frei zugänglich.