Return to search

Avaliação da resistencia, a remoção por tração, de parafusos de titanio inseridos em tibias de coelhos

Orientador : Luis Augusto Passeri / Tese (doutorado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba / Made available in DSpace on 2018-07-28T20:39:42Z (GMT). No. of bitstreams: 1
Garcia_RobsonRodrigues_D.pdf: 2159374 bytes, checksum: af5ca7b5ab774a85bff43685781421c5 (MD5)
Previous issue date: 2001 / Resumo: O propósito deste trabalho foi avaliar a resistência à remoção por tração de parafusos de titânio de 1,5 mm e 2,0 mm de diâmetro, quando inseridos de maneira monocortical e bicortical em fragmentos ósseos, removidos de tíbias de coelhos. Sessenta parafusos de titânio foram testados, divididos em seis grupos. Os grupos M1 e E1 constituíram-se de parafusos de 1,5 mm de diâmetro e inseridos de maneira monocortical. Nos grupos M2 e E2, foram testados parafusos de 2,0 mm
de diâmetro inseridos de maneira monocortical. Já nos grupos M3 e E3, foram testados parafusos de 2,0 mm de diâmetro inseridos de maneira bicortical. Os resultados demonstraram que os parafusos inseridos de maneira monocortical não apresentaram diferenças estatisticamente significantes entre si, apesar do diâmetro e do desenho dos parafusos não serem os mesmos. Os parafusos inseridos de maneira bicortical não diferiram entre si apesar de não terem o mesmo desenho, mas apresentaram resistência à remoção por tração estatisticamente maior que parafusos inseridos de maneira monocortical / Abstract: The aim of this study was to evaluate the pullout strength of 1,5 mm and
2,0 mm diameter titanium screws, when they were inserted in bone fragments of rabbits tibias, monocortically and bicortically. Sixty titanium screws were divided into six groups and tested. The groups M1 and E1 had 1.5 mm screws that were monocortically inserted. In the groups M2 and E2 2.0 mm monocortical screws were tested. And in the groups M3 and E3 2.0 mm bicortical screws were tested. Results demonstrated that the groups of monocortical screws did not showed significant differences between them, although they had different designs and diameters. The groups of bicortical screws did not showed significant differences between them too, although they had different designs, but they showed significantly higher strength than the groups of screws that were monocortically inserted. / Doutorado / Cirurgia / Doutor em Clínica Odontológica

Identiferoai:union.ndltd.org:IBICT/oai:repositorio.unicamp.br:REPOSIP/289682
Date02 October 2001
CreatorsGarcia, Robson Rodrigues
ContributorsUNIVERSIDADE ESTADUAL DE CAMPINAS, Passeri, Luis Augusto, 1957-, Albergaria-Barbosa, José Ricardo de, Moreira, Roger William Fernandes, Freitas, Gersinei Carlos de
Publisher[s.n.], Universidade Estadual de Campinas. Faculdade de Odontologia de Piracicaba, Programa de Pós-Graduação em Clínica Odontológica
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/doctoralThesis
Format82p. : il., application/pdf
Sourcereponame:Repositório Institucional da Unicamp, instname:Universidade Estadual de Campinas, instacron:UNICAMP
Rightsinfo:eu-repo/semantics/openAccess