Return to search

A Novel Power Management Technique Applied in Non- Contact Vital Sign Detection System

This paper presents a novel power management analysis method to reduce the power consumption for the non-contact vital sign sensor. The sensor consisting of the class-E power amplifier (PA), low noise amplifier (LNA), single pole double through (SPDT) switch, and circularly polarized antenna (CPA) is integrated on the Flame Retardant Class 4 (FR-4) epoxy-glass laminate substrate. The appropriate pulse width and pulse period are determined to decrease the power consumption and accurately detect the human physiological signals (respiration and heartbeat). A simple direct down-conversion architecture with a tunable phase shifter is utilized to eliminate the null detection point and the direct current (DC) offset. The overall power consumption of the proposed sensor with the novel power management technique is only 40 % of the conventional system with the DC bias, which can be utilized for the green energy application.

Identiferoai:union.ndltd.org:NSYSU/oai:NSYSU:etd-0131112-184231
Date31 January 2012
CreatorsChen, Jhih-jie
ContributorsChie-In Lee, Ken-Huang Lin, Chih-Wen Kuo, Tzyy-Sheng Horng
PublisherNSYSU
Source SetsNSYSU Electronic Thesis and Dissertation Archive
LanguageCholon
Detected LanguageEnglish
Typetext
Formatapplication/pdf
Sourcehttp://etd.lib.nsysu.edu.tw/ETD-db/ETD-search/view_etd?URN=etd-0131112-184231
Rightsuser_define, Copyright information available at source archive

Page generated in 0.0018 seconds