Yes / The impact of spectral filtration in digital (‘cine’) acquisition was investigated using a flat panel cardiac interventional X-ray imaging system. A 0.1-mm copper (Cu) and 1.0-mm aluminium (Al) filter added to the standard acquisition mode created the filtered mode for comparison. Image sequences of 35 patients were acquired, a double-blind subjective image quality assessment was completed and dose–area product (DAP) rates were calculated. Entrance surface dose (ESD) and effective dose (E) rates were determined for 20- and 30-cm phantoms. Phantom ESD fell by 28 and 41 % and E by 1 and 0.7 %, for the 20- and 30-cm phantoms, respectively, when using the filtration. Patient DAP rates fell by 43 % with no statistically significant difference in clinical image quality. Adding 0.1-mm Cu and 1.0-mm Al filtration in acquisition substantially reduces patient ESD and DAP, with no significant change in E or clinical image quality. / Supported in part by a research grant from Philips Healthcare, The Netherlands.
Identifer | oai:union.ndltd.org:BRADFORD/oai:bradscholars.brad.ac.uk:10454/16963 |
Date | 20 February 2014 |
Creators | Davies, A.G., Gislason-Lee, Amber J., Cowen, A.R., Kengyelics, S.M., Lupton, M., Moore, J., Gaffney, C., Sparrow, T., Irving, H., Ali, S., Middleton, R., Campbell, S., Ackroyd,,, Sivananthan, M. |
Source Sets | Bradford Scholars |
Language | English |
Detected Language | English |
Type | Article, Accepted manuscript |
Rights | © The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com. This is a pre-copyedited, author-produced version of an article accepted for publication in Radiation Protection Dosimetry following peer review. The version of record [Davies AG, Gislason-Lee AJ, Cowen AR et al (2014) Does the use of additional x-ray beam filtration during cine acquisition reduce clinical image quality and effective dose in cardiac interventional imaging? Radiation Protection Dosimetry. 162(4): 597-604.] is available online at: https://doi.org/10.1093/rpd/ncu020, Unspecified |
Page generated in 0.0021 seconds