Return to search

Estudo prospectivo comparativo dos ciclos de ultra-som 14% e 67% do WhiteStar na cirurgia de catarata por facoemulsificação com a técnica Nuclear Preslice

Tese (doutorado)—Universidade de Brasília, Faculdade de Medicina, 2007. / Submitted by wesley oliveira leite (leite.wesley@yahoo.com.br) on 2009-10-02T14:12:54Z
No. of bitstreams: 1
tese-doutorado_orig.pdf: 2981836 bytes, checksum: 00ebfbc9e84526f58d03d3387b3509b7 (MD5) / Approved for entry into archive by Marília Freitas(marilia@bce.unb.br) on 2010-02-09T00:06:53Z (GMT) No. of bitstreams: 1
tese-doutorado_orig.pdf: 2981836 bytes, checksum: 00ebfbc9e84526f58d03d3387b3509b7 (MD5) / Made available in DSpace on 2010-02-09T00:06:53Z (GMT). No. of bitstreams: 1
tese-doutorado_orig.pdf: 2981836 bytes, checksum: 00ebfbc9e84526f58d03d3387b3509b7 (MD5)
Previous issue date: 2007 / Objetivos: Comparar os ciclos de 14% e 67% de ultra-som ligado, módulo contínuo,
com controle manual do pulso, no aparelho de facoemulsificação Sovereign com
sistema WhiteStar, para realização de cirurgia de catarata, utilizando-se a técnica
de facoemulsificação Nuclear Preslice, nos quesitos: tempo efetivo de ultra-som,
percentual de ultra-som, volume de infusão e perda de células endoteliais.
Métodos: Estudo prospectivo, comparativo, incluindo 32 pacientes (38 olhos), com
catarata nuclear senil NO3 NC3 e NO4 NC4 (LOCS III), que foram aleatoriamente
divididos, por meio de sorteio, em dois grupos: A e B, sendo realizada a cirurgia de
catarata por facoemulsificação com a técnica Nuclear Preslice, com controle de
pulso manual, no grupo A (n= 21) com ciclo 14%; e no grupo B (n= 17), com ciclo
67%. Os grupos A e B foram subdivididos, segundo o grau de opacidade do
cristalino em dois subgrupos. No pré-operatório foi realizado exame oftalmológico
completo, biometria, paquimetria e microscopia especular, no pós-operatório, além
do exame ocular, com avaliação da acuidade visual e paquimetria nas 24 horas após
a cirurgia, foi feita a microscopia especular com 3 meses.
Resultados: As cirurgias pelo ciclo de ultra-som 67% apresentaram tempo efetivo
de ultra-som significantemente maior do que as cirurgias pelo ciclo de ultra-som
14%, em cataratas mais densas. O percentual de ultra-som no ciclo de ultra-som
67%, em diferentes densidades de cataratas, foi significantemente maior do que no
ciclo de ultra-som 14%. Não houve diferenças quanto ao volume de infusão utilizado
nas cirurgias. Não se observou diferença estatisticamente significante na perda de
células endoteliais nos diferentes ciclos de ultra-som. Houve correlação significante
entre a perda de células endoteliais e as variáveis: tempo efetivo de ultra-som (EPT)
e volume de infusão utilizado na cirurgia, em ambos os ciclos.
Conclusões: O tempo efetivo de ultra-som e o percentual de ultra-som foram
menores nas cirurgias com ciclo de ultra-som 14%, em cataratas mais densas. O
volume de infusão utilizado nas cirurgias e a perda de células endoteliais foram
semelhantes em ambos os ciclos. O tempo efetivo de ultra-som e o volume de
infusão utilizado na cirurgia foram correlacionados à perda de células endoteliais. ____________________________________________________________________________________ ABSTRACT / Purpose: To compare the 14% and 67% duty cycles of ultrasound power, continuous
module, pulse manual control in the Sovereign phacoemulsification system with
WhiteStar power modulation for cataract surgery. The phacoemulsification Nuclear
Preslice technique was used to evaluate ultrasound effective time, ultrasound
percentage, infusion volume amount and endothelial cell loss.
Methods: This was a prospective and comparative clinical trial in 32 patients (38 eyes)
with senile nuclear cataract NO3 and NO4 NC4 (LOCS III) divided in two groups: A and
B. To perform the cataract surgery the phacoemulsification with Nuclear Preslice
technique was used with manual pulse control, for group A with 14% duty cycle (n=21)
and group B with 67% duty cycle (n=17). Both groups were subdivided according to
cristaline opaque level in two sub-groups. Preoperative measures included complete
ophthalmologic exam, biometry, pachymetry and specular microscopy, postoperative 24
hours pachymetry and visual acuity, specular microscopy was mensured at 3 months
follow up.
Results: The surgeries that was used the 67% duty cycle had significantly more
ultrasound effective time than the 14% group in more dense cataracts. The ultrasound
percentage in the 67% duty cycle, in different cataract densities, was significantly more
than in the 14% duty cycle. There was no difference in the infusion volume amount used
in the surgeries. There was no difference statistically significant in endothelial cell loss in
both ultrasound cycles. There was statistically significant correlation between the
endothelial cell loss and the variables: ultrasound effective time (EPT) and infusion
volume amount used in surgeries in both cycles.
Conclusions: This clinical trial showed less effective time of ultrasound and ultrasound
percentage in the 14% duty cycle for more dense cataracts. The infusion volume
amount used and the endothelial cell loss were the same in both cycles. The ultrasound
effective time and the infusion volume amount used in the surgery were correlated to
endothelial cell loss.

Identiferoai:union.ndltd.org:IBICT/oai:repositorio.unb.br:10482/3577
Date January 2007
CreatorsPereira, Antonio Eduardo
ContributorsÁvila, Marcos Pereira de
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/doctoralThesis
Sourcereponame:Repositório Institucional da UnB, instname:Universidade de Brasília, instacron:UNB
Rightsinfo:eu-repo/semantics/openAccess

Page generated in 0.0018 seconds