Return to search

Avaliação biomecanica in vitro de tecnicas de fixação rigida metalica para osteotomia sagital do ramo mandibular em movimentos de avanço e recuo / An in vitro biomechanical evaluation of fixation techniques for sagittal split ramus osteotomy in mandibular advancement and setback

Orientador: Luis Augusto Passeri / Tese doutorado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba / Made available in DSpace on 2018-08-09T20:14:50Z (GMT). No. of bitstreams: 1
Brasileiro_BernardoFerreira_D.pdf: 9990302 bytes, checksum: ddce5ea3896243c2b52242b595e7946d (MD5)
Previous issue date: 2007 / Resumo: Um melhor entendimento biomecânico da fixação interna rígida (FIR) na osteotomia sagital dos ramos mandibulares (OSRM) pode ser baseado em pesquisas com o intuito de investigar sua função in vitro antes da aplicação clínica. Assim, este trabalho teve como objetivo avaliar comparativamente a resistência de três diferentes técnicas de FIR em réplicas de hemimandíbulas humanas de poliuretano simulando a OSRM para movimentos de 5 mm em avanço e recuo. As técnicas de FIR incluíram uma miniplaca ponte com 4 furos e 4 parafusos monocorticais (grupo miniplaca), uma miniplaca com 4 furos e 4 parafusos monocorticais mais um parafuso bicortical posicional (grupo híbrido) e 3 parafusos bicortical posicionais em configuração em ¿L¿ invertido (grupo ¿L¿ invertido). Os parafusos e as miniplacas utilizados eram de titânio e pertencentes ao sistema de 2,0 mm. Estes foram avaliados quanto à padronização dimensional por meio do coeficiente de variação, que evidenciou alta semelhança macroscópica. As hemimandíbulas foram submetidas a testes de carregamento linear vertical e lateral por uma unidade de testes mecânicos Instron 4411 para registro da carga de pico aos deslocamentos de 1 mm, 3 mm, 5 mm e 10 mm. Médias e desvio padrão foram avaliados aplicando-se a Análise de Variância e o teste de Tukey em nível de significância de 5%. O grupo miniplaca apresentou menores valores de carga de pico ao deslocamento (p<0,01) quando comparada com as outras técnicas de fixação considerando-se igual tipo de movimento e direção de carga. O grupo ¿L¿ invertido foi ainda superior em resistência (p<0,01) ao grupo híbrido em movimento de avanço e com aplicação de carga verticalmente. Concluiu-se que a utilização da técnica de FIR para a OSRM baseada em 3 parafusos bicorticais em ¿L¿ invertido foi a mais estável, e que a colocação de um parafuso bicortical pode otimizar significativamente a resistência da fixação com miniplacas e parafusos monocorticais / Abstract: A better understanding of the biomechanics of sagittal split ramus osteotomy (SSRO) rigid internal fixation (RIF) can be based on research aiming to evaluate its function in vitro before clinical application. Thus, the goal of this investigation was to evaluate comparatively the resistance of three different RIF techniques applied to polyurethane synthetic hemi-mandible replicas simulating the SSRO for 5 mm advancement and setback. RIF techniques included a 4-hole miniplate and 4 monocortical screws (miniplate group), a 4-hole plate and 4 monocortical screws with one additional bicortical positional screw (hybrid group), and 3 bicortical positional screws in a traditional inverted-L pattern (inverted-L group). Screws and miniplates were made of titanium and from 2.0 mm system. These were analyzed according to dimensional standardization by a coefficient of
variation, which revealed high macroscopic similarity. The hemi-mandibles were submitted lateral and vertical loading tests in an Instron 4411 mechanical testing unit for recording of peak loading at 1 mm, 3 mm, 5 mm and 10 mm of displacement. Means and standard deviation were analyzed using Analysis of Variance and Tukey test with a 5% level of significance. Miniplate group showed lower load peak scores (p<.01) when compared to the others fixation techniques regarding equal type of movement and force direction. Inverted-L group demonstrated higher resistance (p<.01) than hybrid group during advancement and with application of vertical load. It was concluded the RIF technique for SSRO based on 3 bicortical screws in the inverted-L pattern was the most stable, and the installation of a bicortical screw may significantly optimize the resistance of the miniplate and monocortical screws fixation / Doutorado / Cirurgia e Traumatologia Buco-Maxilo-Faciais / Doutor em Clínica Odontológica

Identiferoai:union.ndltd.org:IBICT/oai:repositorio.unicamp.br:REPOSIP/289676
Date26 October 2007
CreatorsBrasileiro, Bernardo Ferreira
ContributorsUNIVERSIDADE ESTADUAL DE CAMPINAS, Passeri, Luis Augusto, 1957-, Gabrielli, Marisa Aparecida Cabrini, Sverzut, Cassio Evard, Filho, Valfrido Antonio Pereira, Trivellato, Alexandre Elias
Publisher[s.n.], Universidade Estadual de Campinas. Faculdade de Odontologia de Piracicaba, Programa de Pós-Graduação em Clínica Odontológica
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/doctoralThesis
Format144p. : il., application/pdf
Sourcereponame:Repositório Institucional da Unicamp, instname:Universidade Estadual de Campinas, instacron:UNICAMP
Rightsinfo:eu-repo/semantics/openAccess

Page generated in 0.0017 seconds