University of Adelaide
Development of an external beam porton a 16.5 MeV proton cyclotron for radiobiological experiments
Abstract
dc:description.abstractAn external beam port for radiobiological experiments was successfully developed and characterised on a GE PETtrace cyclotron. The dose-rate was reduced from kGy/s to ~30 Gy/min and the beam diameter was ~20 mm with a uniform flatness of 5% at 30 cm from the beam port. Also, at 30 cm from the beam port, proton energy had reduced from mono energetic 16.5 MeV to a spectrum with the maximum energy peak at ~6.25 MeV. This project also aimed to characterise the behaviour of commercially available optical fibres in proton beams. To do this, silica and PMMA optical fibres were placed in the proton beam and the light output was measured using both a PMT and an optical spectrometer. The light output from the fibres in the proton beam was investigated as a function of dose-rate (beam current) and proton energy. Both fibre types showed a linear response to dose-rate and a proton energy dependence.
Degree
thesis:*- College dc:contributor.school
- University of South Australia. UniSA Stem.
- Year dc:date.issued
- 2020
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Asp, Johan
Subjects
dc:subject × 4Rights
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/11541.2/147374