Abstract
dc:description.abstractReal-time monitoring of alpha and beta emitting radionuclides in the environment and in mineral-processing has always proved elusive. The short path lengths, low target activities and harsh environments mean a quite specialised sensor is required. Two sensors have been constructed and tested with the sensitivity to measure gross alpha and beta particles below the activity level of 1 Bq/ml using optical fibres made of scintillating polymers. Scintillation allows for direct, real-time, semi-continuous measurements that are not available with current techniques. These polymers are suitable for withstanding the low pH solutions at elevated temperatures in abrasive mineral-processing environments. The new devices were found to be more sensitive by a factor related to the increase in their surface area and the new detection limit is now below 0.1 Bq/ml for the beta sensor and 0.15 for the alpha sensor. The feasibility of a novel sensing technique to specifically detect polonium- 210 (Po-210) is also investigated. The envisioned device would allow in-line sensing of Po-210 specifically by selective uptake onto a coated scintillator.
Degree
thesis:*- College dc:contributor.school
- School of Physical Sciences : Physics
- Year dc:date.issued
- 2020
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Mik, Jarrah Lionel Herbert
- Advisors dc:contributor.advisor
-
- Ottaway, David
- Spooner, Nigel
- Kalnins, Christopher
Rights
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/2440/127453