Back to search

Ajou University

Correction of target-controlled infusion following the wrong selection of emulsion concentration of propofol

Abstract

dc:description.abstract

Background: We investigated the correction methods following wrong-settings of emulsion concentrations of propofol as a countermeasure against erroneous target-controlled infusions (TCI). Methods: TCIs were started with targeting 4.0 μg/ml of effect-site concentration (Ceff) of propofol, and the emulsion concentrations were selected for 2.0% instead of 1.0% (FALSE1- 2, n = 24), or 1.0% instead of 2.0% (FALSE2-1, n = 24). These wrong TCIs were corrected at 3 min after infusion start. During FALSE1-2, the deficit was filled up while injecting after equilibrium (n = 12), or while overriding (n = 12). During FALSE2-1, the overdose was evacuated while targeting Ceff (n = 12) or targeting plasma concentration (Cp) (n = 12). The gravimetrical measurements of TCI reproduced the Cp and Ceff using simulations. The reproduced Ceff at 3 min (Ceff-3min) and the time to be normalized within ± 5% of target Ceff (T±5%), were compared between the correction methods. Results: During the wrong TCI, Ceff-3min was 1.98 ± 0.01 μg/ml in FALSE1-2, and 7.99 ± 0.05 μg/ml in FALSE2-1. In FALSE1-2, T±5% was significantly shorter when corrected while overriding (3.9 ± 0.25 min), than corrected after equilibrium (6.9 ± 0.05 min) (P < 0.001). In FALSE2-1, T±5% was significantly shorter during targeting Cp (3.6 ± 0.04 min) than targeting Ceff (6.7 ± 0.15 min) (P < 0.001). Conclusions: The correction methods, based on the pharmacokinetic and pharmacodynamic characteristics, could effectively and rapidly normalize the wrong TCI following erroneously selections of the emulsion concentration of propofol.

Degree

thesis:*
Department dc:contributor.department
대학원 의학과
Year dc:date.issued
2015

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • 이, 원일
Advisor dc:contributor.advisor
  • 채, 윤정
Person Identifier Other person:identifier.other
  • 201224381

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Dc Identifier Other
000000019017

Chain of custody

source
Harvested from
Ajou University
Base URL
repository.ajou.ac.kr/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

이, 원일. Correction of target-controlled infusion following the wrong selection of emulsion concentration of propofol. 2015. http://repository.ajou.ac.kr/handle/201003/11841