Comparison of 4DCT and 4DCBCT in measuring motion trajectory of moving targets in phantom

Abstract

dc:description.abstract

Stereotactic Body Radiation Therapy (SBRT) is one of the treatment regimens for lung cancer in which a high dose of radiation is delivered to the tumor in a few fractions. The involuntary movement of lungs during breathing results in the movement of lung tumor during the treatment. Thus, SBRT necessitates the use of motion resolving imaging methods both for treatment planning and image guidance. Planning 4D Computed Tomography (4DCT) and the treatment localization 4D Cone Beam Computed Tomography (4DCBCT) can resolve tumor motion. However, due to the difference in the implementation of 4DCT on a CT scanner and 4DCBCT on radiation delivery system, it is important to ensure measured motion by two systems are in congruence. Therefore, the main purpose of this thesis is to compare the measured motion trajectory of moving target in phantom between 4DCT and 4DCBCT. We imaged the moving target, undergoing regular sinusoidal and irregular patient-like motions separately, and determined the measured motion using 4DCT and 4DCBCT. We also investigated the target enlargement due to motion in 3DCBCT which does not resolve motion. The differences between the programmed and measured target motion amplitudes were determined. The results indicate that, for regular breathing waveforms, 4DCT has the smallest median difference and inter-quartile range. 4DCBCT advanced and basic methods accurately measure motion but exhibit greater variability in measured motion differences as compared to 4DCT. Superior-inferior motion measured by basic 4DCBCT is closer to that measured by 4DCT, while advanced 4DCBCT reconstruction measures smaller motion than 4DCT by nearly 1 mm on average. The natural averaging happening in the 3DCBCT is consistent with the target motion, however, the extracted motion amplitude has larger variability compared to both 4DCT and 4DCBCT. Target volume measured in the average images of 4DCT and 4DCBCT and in 3DCBCT, are nearly equal. Similarly, the target volume measured in the maximum intensity projections of 4DCT and 4DCBCT are nearly equal. In cases of irregular patient breathing patterns, 4DCT tends to measure motion more accurately compared to the programmed waveform than 4DCBCT, although this is not consistently observed across all cases. 4DCT images faster than the 4DCBCT and does not see the entire irregularity in patient waveforms. While in some cases, 4DCBCT has poor image quality due to smaller number of projections per motion phase which affects the motion measurement accuracy. Both 4DCT and 4DCBCT are capable of accurately measuring tumor motion for regularly breathing targets, with 4DCT showing slightly higher precision. Since the projected blurring due to motion in 3DCBCT has poor confidence in motion measurement, 3DCBCT should be used carefully, especially when the motion is large and expected to vary between planning and treatment sessions. For irregular breathing, performance varies depending on waveform characteristics, though 4DCBCT’s artifacts remain a limitation. These findings support the complementary roles of both modalities in SBRT planning and delivery.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
Master's
Discipline thesis:degree_discipline
Medical Physics
Grantor
University of Alberta
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Handa, Bhumika
Advisor dc:contributor.advisor
  • Rathee, Satyapal (Medical Physics)

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • This thesis is made available by the University of Alberta Library with permission of the copyright owner solely for non-commercial purposes. This thesis, or any portion thereof, may not otherwise be copied or reproduced without the written consent of the copyright owner, except to the extent permitted by Canadian copyright law.
Language dc:language.iso
en

Identifiers

dc:identifier.*
DOI dc:identifier.doi
https://doi.org/10.7939/83260

Chain of custody

source
Harvested from
University of Alberta
Base URL
era.library.ualberta.ca/oai
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Handa, Bhumika. Comparison of 4DCT and 4DCBCT in measuring motion trajectory of moving targets in phantom. Master's thesis, University of Alberta, 2025. https://doi.org/10.7939/83260