Quantitative Skin 23Na-MRI: A Novel Approach for Measuring Accurate Tissue Sodium Concentrations in Healthy Adults
Abstract
dc:description.abstractResearchers have quantified sodium (23Na), the second most abundant element in the body, non-invasively using MRI as a potential biomarker of tissue viability. Recent advances in sodium physiology have identified skin as a major sodium reservoir, where sodium ions bind to negatively charged glycosaminoglycans without associated water retention. Several 23Na-MRI studies have attempted to quantify skin tissue sodium concentration (TSC) in healthy individuals and in various pathologies such as hypertension, cardiovascular diseases, type-2 diabetes, and chronic kidney disease, often reporting elevated sodium levels compared to healthy controls. Most of these studies, however, employed gradient-echo (GRE) sequences with relatively long echo times (TE ~ 2 ms) and 3×3×30 mm3 large voxels using a volume coil at 3 T. These acquisitions typically underestimated skin TSC, yielding ~20 mM in healthy adults – much lower than the expected ~85 mM reported using atomic absorption spectroscopy. These underestimated TSC measurements are biased by the large voxel size partial volume effects, and few studies have used surface coils, yielding smaller voxels to mitigate these effects. Recently, skin TSC was measured using a surface coil that can yield 0.4×4×4 mm3 voxels at 4.7 T, yielding ~35 mM in healthy adults, indicating persistent bias. This underestimation primarily arises from sodium’s rapid biexponential transverse (T2) relaxation in skin, which leads to substantial signal loss in current quantitative sodium imaging methods. Therefore, this thesis aims to measure skin TSC accurately by compensating for the unavoidable signal loss associated with very rapid transverse biexponential relaxation. Sodium is a spin-3/2 nucleus that exhibit quadrupolar interaction with the local electric field gradients resulting in sodium’s rapid biexponential relaxation in skin. Current methods cannot capture rapidly decaying sodium signal even with ultrashort TE as there is a large power spectral density component around 0 Hz resulting in very short values of T2f for sodium that contributes 60% of the total signal. To compensate this signal loss, accurate estimation of T2 relaxation parameters is required while accounting for relaxation variations with sex, age, ethnicity, and pathology. Here, a novel imaging approach was developed by varying radio frequency (RF) pulse durations at a constant echo time (TE =2 ms) to capture the skin sodium’s rapid biexponential relaxation signal loss dependency on RF pulse length. A total of eleven scans with various contrasts were acquired in 33 minutes: seven scans with varying 90 degrees RF pulse durations 70-900 µs and constant TE = 2 ms, two scans with extended TE = 6 and 10 ms, and two scans with flip angles 60 and 120 degrees. The resulting skin image intensities were fitted to a set of pre-computed relative signal values (a signal dictionary) to attain sodium’s fast component of transverse relaxation – T2f (60%), slow component of transverse relaxation – T2s(40%), relative B1 (rB1), and subsequently compensate for the relaxation-induced signal loss to yield accurate TSC measurements. The methodology was validated by using a gellan gum phantom with known sodium concentration and then applied in vivo by scanning calf skin of 20 healthy volunteers using a custom-built surface coil at 4.7 T. Because skin contains a small fraction of blood with a high sodium concentration of 145 mM, a blood component was incorporated into the signal dictionary to quantify blood sodium fraction within skin. The fitted parameters T2f, T2s, rB1, and blood sodium fraction resulted in compensated TSC value of 89 ± 3 mM – approximately 4x greater than the previously reported values and consistent with those reported by atomic absorption spectroscopy – the gold standard, thereby validating the robustness and accuracy of the proposed methodology. The findings from this work introduce a new quantitative approach for accurately assessing skin tissue sodium concentration (TSC) by compensating for sodium’s rapid biexponential signal loss – a factor that was previously overlooked in 23Na-MRI studies of skin. This methodology establishes a foundation for extending it to other tissues exhibiting similar relaxation characteristics. While the current implementation requires a long scan time of approximately 33 minutes, future studies should aim for faster acquisition while integrating the method with clinically accessible pulse sequences, such as GRE at 3 T field strength. By addressing these challenges, this approach has the potential to emerge as a reliable tool in future for clinical research monitoring disease progression using skin TSC as a biomarker.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- Master's
- Grantor
- University of Alberta
- Year dc:date.issued
- 2026
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Singh, Gurparkash
- Advisor dc:contributor.advisor
-
- Beaulieu, Christian (Biomedical Engineering)
Subjects
dc:subject × 4Rights
dc:rights- Statement dc:rights
-
- This thesis is made available by the University of Alberta Library with permission of the copyright owner solely for non-commercial purposes. This thesis, or any portion thereof, may not otherwise be copied or reproduced without the written consent of the copyright owner, except to the extent permitted by Canadian copyright law.
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- DOI dc:identifier.doi
- https://doi.org/10.7939/84024