A study of the inter-test reliability of manual and automatic threshold measures obtained from a selected work force.
Abstract
dc:description.abstractPurpose of Study: The purpose of this study was to determine the inter-test reliability between the manual and automated audiometric procedures among a typical industrial work force. Procedure: Each of the 55 subjects used for the study were administered both the automated and manual audiological tests at frequencies of 500, 1000, 2000, 3000, 4000, and 6000 Hz. All tests were administered and scored by the same tester in a soundproof audiological booth. The order of the manual-automated test presentation was counterbalanced to reduce the influence of learning effects. Manual thresholds were obtained with a Model 1701 Grason-Stadler Audiometer using the ascending method. Automatic thresholds were determined with a Tracor Rudmose Model ARJ-4 Audiometer with an attenuation rate of 4.34 dB per second. Each automated tracing was measured in millimeters and the midpoint markings were averaged to give the threshold for each frequency. A Randomized Block Factorial Analysis of Variance was employed to analyze the data. The factors considered were subjects, test frequencies, and the two methods of obtaining thresholds. Findings of the Study: The difference between the automated and manual testing methods was found to be significant beyond the .05 level of significance. The frequency and test method-frequency interaction was found to be significant beyond the .001 level of significance. The manual testing method yielded an over-all threshold difference of 5.34 dB lower than the automated testing method. In reference to frequencies tested, the manual method consistently produced lower threshold averages with a range of 1.08 dB at 6000 Hz to 8.90 dB at 500 Hz. Use of automated testing did not appear to be a practical threshold measure for all subjects because 45% of the subjects did not produce valid automated audiograms. The results of the study showed little learning took place between ears using the automated method. However, with the use of a pre-test period, a small percentage of subjects would have possibly produced valid automated audiograms. Significant threshold differences were found at 500, 1000, and 2000 Hz between the two testing methods. At this time research is not adequate to predict the differences between thresholds and will depend on further research. Therefore, at this time, the two methods can not reliably be interchanged.
Degree
thesis:*- Name thesis:degree_name
- M.A.
- Level thesis:degree_level
- Masters
- Department thesis:degree_department
- Dept. of Oral Communication.
- Grantor
- Baylor University.
- Year dc:date.issued
- 1976
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- May, Berna LaRue, 1953-
- Advisor dc:contributor.advisor
-
- Barnett, Madalene R.
Subjects
dc:subject × 1Rights
dc:rights.accessrights- Statement dc:rights.accessrights
-
- Worldwide access
- Baylor University works are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. Contact libraryquestions@baylor.edu for inquiries about permission.
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/2104/14267