The design of a reflectance spectrometer and its calibration using SrTiO3 /

Abstract

dc:description.abstract

A new Ultra-High Vacuum (UHV) reflectance spectrometer was successfully designed, making use of a Janis Industries ST-400 sample cryostat, IR Labs bolometer, and Briiker IFS 66 v/S spectrometer. Two of the noteworthy features include an in situ gold evaporator and internal reference path, both of which allow for the experiment to progress with a completely undisturbed sample position. As tested, the system was designed to operate between 4.2 K and 325 K over a frequency range of 60 - 670 cm~^. This frequency range can easily be extended through the addition of appUcable detectors. Tests were performed on SrTiOa, a highly ionic incipient ferroelectric insulator with a well known reflectance. The presence and temperatmre dependence of the lowest frequency "soft" mode were measured, as was the presence of the other two infrared modes. During the structural phase transition from cubic to tetragonal perovskite, the splitting of the second phonon mode was also observed. All of the collected data indicate good agreement with previous measurements, with a minor discrepency between the actual and recorded sample temperatures.

Degree

thesis:*
Name thesis:degree_name
M.Sc. Physics
Level thesis:degree_level
Masters
Discipline thesis:degree_discipline
Faculty of Mathematics and Science
Department dc:contributor.department
Department of Physics
Grantor
Brock University
Year dc:date.issued
2004

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Klassen, J. Robert.

Subjects

dc:subject × 2

Rights

Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/10464/2335

Chain of custody

source
Harvested from
Brock University
Base URL
brocku.scholaris.ca/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Klassen, J. Robert.. The design of a reflectance spectrometer and its calibration using SrTiO3 /. Masters thesis, Brock University, 2004. https://hdl.handle.net/10464/2335