Boston University
Optimizing solvent evaporation parameters with the N-EVAP nitrogen evaporator: assessing the impact of heated sample preparation on drug analysis
Abstract
dc:description.abstractSample preparation is a critical step in many analytical workflows, as the quality of subsequent instrumental analysis depends heavily on its effectiveness. Nitrogen-assisted solvent evaporation is a common laboratory technique; however, limited data exists regarding the effects of experimental parameters on evaporation performance. This study evaluated the evaporation rates of various solvents using an N-EVAP(TM) nitrogen evaporator (Organomation(TM), Berlin, Massachusetts, United States) under different conditions and initiated an investigation into how heated evaporation may influence the quantitative analysis of drugs. The first phase of the study included evaporation trials that were conducted at ambient temperature and at elevated temperatures utilizing a water bath to assess the effect of heating on evaporation rates. The elevated-temperature trials were performed at temperatures slightly below each solvent's boiling point. Additional experiments examinedsolvent mixtures commonly used in laboratory settings, the influence of tube type and size (test tubes and microcentrifuge tubes), and the performance of different needle configurations (standard, bent, and pseudo-spiral) across varying flow rates, temperatures, and needle gauges. Results demonstrated that heating substantially increased evaporation rates for all solvents tested, although the magnitude of improvement varied among solvents. Even modest increases in water bath temperature produced notable increases in evaporation rate. The introduction of bent needle designs allowed for higher nitrogen flow rates without causing sample splashing, resulting in slight improvements in evaporation under ambient conditions but minimal additional benefit under heated conditions. Furthermore, larger tube sizes exhibited significantly greater evaporation rates under heated conditions compared to smaller tubes. Overall, these findings indicate that controlled heating and optimized gas delivery can improve evaporation efficiency, reduce sample preparation time, and increase laboratory throughput. The ability to adjust these parameters to meet specific sample preparation requirements makes this approach applicable across a wide range of laboratory settings. The second part of this experiment aimed to evaluate the quantitative effects of heated and unheated nitrogen dry-down evaporation. Unheated trials were to be conducted at room temperature, while heated trials were planned at a series of increasing temperatures. This portion of the study was designed to mimic a solid-phase extraction procedure using a spiked elution solution, followed by sample reconstitution and analysis using a QSight® Liquid Chromatography–Mass Spectrometry (LC-MS) system. However, due to instrument-related issues, the experiment could not be completed beyond the initial trials used to establish LC-MS parameters. As a result, further data collection and quantitative analysis were not possible. This represented a significant limitation of the study, as it prevented evaluation of the potential effects of heated nitrogen evaporation on drug quantitation and instrumental performance.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Biomedical Forensic Sciences
- Grantor
- Boston University
- Year dc:date.issued
- 2026
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Briggs, Emma E.
- Advisor dc:contributor.advisor
-
- Wareing, Celeste
Subjects
dc:subject × 2Rights
- Language dc:language.iso
- en_US
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/2144/53543