Abstract

dc:description.abstract

Purpose: To model the Varian TrueBeam linear accelerator (linac) in Monte Carlo dose calculation code, EGSnrc, using the phase space files provided by Varian. To broaden the applicability of Enhanced Optimization (EO) onto TrueBeam treatment plans (delivered with 6X/6X FFF beams, and HD MLCs), and by using focused EO, an objective(s) modification method. Methods: Varian’s TrueBeam linac component information is proprietary, so a complete BEAMnrc accelerator modeling is not possible; instead, Varian provides phase space files. Previously developed Trilogy-based EGSnrc processing tools were modified for use with the True-Beam phase space files, which required modifications to the DOSXYZnrc and BEAMnrc modules. After processing the simulation data for plans and beamlets, EO was applied to plans to produce new plan parameters (i.e., MLC leaf position), which were imported back into Eclipse, to generate a new treatment plan for comparison with original configuration in AAA and AXB calculated distributions. Four Eclipse VMAT TrueBeam HD MLC treatment plans were evaluated (from MC simulation to EO application): two with 6X and two with 6X FFF energies. Results: For homogenous cases, PTV DVH discrepancies between Eclipse and MC distributions are minor, with AXB-MC providing better agreement than AAA-MC. In cases with heterogeneities (e.g., lung), both comparisons showed significant discrepancies, although AXB-MC was still in better agreement. The majority of reference point dose calculations showed less than 2% discrepancy for both AXB-MC and AAA-MC comparisons, with the lung case exceeding 3%. OAR DVHs showed far greater differences than PTV DVHs for both comparisons (e.g., Brain 2 MC-AAA had DVH discrepancy value of 0.580% for PTV and 1.308% for OAR) and depended on adjacent tissue density. With the use of focused EO, qualitative improvement to OAR(s) doses were observed within MC distribution. Quantitively, the objective score (calculated with base objectives set) was found to decrease in all cases except for one brain case. Plans with significant heterogeneities had less drastic score reduction (~7%) in contrast to homogeneous cases (e.g., brains, ~36%). Also, only flattened beam profile cases were EO responsive, with FFF plans sometimes failing to replicate the improvements to Eclipse calculated EO plan. Conclusion: EO can be now applied to TrueBeam-delivered treatment plans, and novel delivery techniques, such as HD MLCs, FFF beam profile, and SBRT plans were explored. The OAR objectives modification was expanded from Mathews et al. to the concept of Focused EO. The results may be of interest to physicists and physicians with regards to dose calculation accuracy and critical structure sparing, and follow up work is planned to explore such topics in greater depth. The work undertaken here will form the foundation for carrying those works out in future.

Degree

thesis:*
Grantor dc:publisher
State University of New York at Buffalo
Year dc:date.issued
2021

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Pant, Ankit
Contributors dc:contributor
  • Nazareth, Daryl
  • Radiology

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions.
  • Copyright retained by author.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/10477/87570

Chain of custody

source
Harvested from
Buffalo
Base URL
ubir.buffalo.edu/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Pant, Ankit. Enhanced Optimization with VMAT Treatment Plan Delivered Using Varian TrueBeam System. State University of New York at Buffalo, 2021. http://hdl.handle.net/10477/87570