The Application of Beam Angular Correction in the Real-Time Skin Dose Tracking System for Interventional Fluoroscopic Procedures
Abstract
dc:description.abstractFluoroscopically-guided interventional procedures can result in high dose to the skin on the entrance surface of the patient. This high skin dose can cause threshold-dependent cutaneous injury. In some of the procedures, such as coronary intervention and cardiac catheter ablation, the cumulative skin dose is up to several Gy, which exceeds the threshold of deterministic effect on skin and lead to skin injuries, such as erythema, epilation and desquamation. Because of the risk of these effects, accurate determination of skin dose is important to allow dose to be managed with confidence during the procedure. The angle of incidence depends on the gantry angulation and patient contour and this results in the incident beam being non-normal to the skin surface during fluoroscopic procedures. According to the skin dose distribution calculated from the point spread function (SD-PSF) convolution, the skin dose varies with the angle of beam incidence since the path length of the primary photons through the skin layer increases as the angle of incidence theta decreases and since the direction of scatter relative to the skin layer changes. Thus, the corrections for angle of x-ray incidence is needed to increase the accuracy of skin-dose estimation. EGSnrc Monte-Carlo (MC) software was used to calculate the primary plus scatter dose averaged over various depths into the entrance surface of a water phantom at the center of the field as a function of incident beam angle and were normalized to the incident primary dose to obtain the angular correction factors (ACF’s). Validation of the ACF’s was done with Gaf-chromic XR-QA2 film measurement. In order to show the effects of angular correction, the ACF was integrated into skin Dose Tracking System (DTS) for dose comparison. The DTS was updated by applying ACF’s for accurate skin dose calculation. The DTS that we developed provides a real-time, color-coded mapping of the skin dose distribution during fluoroscopically-guide interventional procedures. It indicates the peak skin dose value on the entire surface as well as the peak dose within the current field of view. During the procedure, a CAN bus module transfers physical and exposure parameters for every exposure pulse from the imaging system to the DTS. In addition to allowing dose calculation in real-time, all exposure parameter information is also recorded in a log file that allows post-procedure processing and analysis. Log files from a number of clinical procedures were used to calculate the skin dose distributions with and without angular correction. In addition, a fitting technique was also developed to quickly determine an angular correction factor (ACF) for real-time dose estimation using our DTS during fluoroscopically-guided interventional procedures. The purpose of this work is to show the effect of angular correction on the skin dose calculation in the DTS for clinical fluoroscopically-guided procedures. Accurate estimation of the skin dose is important to evaluate the risk of deterministic cutaneous effects.
Degree
thesis:*- Grantor dc:publisher
- State University of New York at Buffalo
- Year dc:date.issued
- 2021
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Sun, Sheng-Hsuan
- Contributors dc:contributor
-
- Bednarek, Daniel
- Radiology
Subjects
dc:subject × 3Rights
dc:rights- Statement dc:rights
-
- Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions.
- Copyright retained by author.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/10477/87634