State University of New York at Buffalo
Low History VMAT Monte Carlo Dose Matrix Denoising with Digital Smoothing Filters
Abstract
dc:description.abstractPurpose: To develop a fast and accurate in-house secondary MU check method based on the combination of Monte Carlo (MC) simulation and digital smoothing filters for independent verification of VMAT treatment planning system dose calculations. Methods: For a VMAT patient treatment plan created for a TrueBeam linac, DICOM information was exported locally from Eclipse after anonymization. The input files for Monte Carlo simulation were then generated with the user code and uploaded to the EGSnrc directory on the CCR. A total of 16 VMAT plans (4 brains, 4 prostates, 4 lungs, 4 H&N) were simulated with 15 different histories respectively, from 10^6 to 10^10, of which 10^10 is used as the benchmark in the follow-up. All the plans were parallelized to run on the computer cluster. The resulting 3D dose matrix were then convolved in MATLAB with 4 different digital smoothing filters (2 Gaussian filters and 2 Savitzky–Golay filters) except for the 10^10 simulation. These filtered and unfiltered results are then used to calculate the absolute discrepancy from the standard result at different threshold. The dose distribution and dose volume histogram comparisons were also done with the open-source software CERR. At last, the overall calculation time was obtained by summing up all the steps. Results: The overall calculation time of 10^6 histories can be compressed to less than 10 min, in some cases even up to 5 min, including all the processing and calculation steps. For 5*10^6 histories, the time can be compressed to within 15 min. At 10^6 history, the absolute average discrepancy between filtered (G2, S-G 1, S-G 2) and the standard results were ~4% for brain cases, ~4.5% for prostate cases, ~3% for lung case, and ~3% for head and neck cases at 90% threshold. Among them, the result obtained by filter G-1 is the worst, generally 1% to 3% higher than the above results, depending on the simulation site. For two S-G filters, the filtered dose distributions with 10^6 histories are comparable to the standard results for brain cases, while for the other 3 sites 5*10^6 histories are required to be comparable to the standard. Gaussian filters perform poorly in dose distribution comparison, especially G-1 filter whose image was still noisy. The dose volume histogram of low history simulation with filters applied also gets closer to the standard compared to without filtering. Conclusion: Combining Monte Carlo simulation and digital smoothing filter is a clinically feasible method for the secondary monitor unit check of the VMAT plans due to its high accuracy (<5%) and low calculation times (<15 min). As computer clusters or cloud computing become more common, this approach will be competitive in most clinical settings.
Degree
thesis:*- Grantor dc:publisher
- State University of New York at Buffalo
- Year dc:date.issued
- 2023
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Guan, Xian; 0009-0009-8751-5181
- Contributors dc:contributor
-
- Nazareth, Daryl
- Radiology
Subjects
dc:subject × 1Rights
dc:rights- Statement dc:rights
-
- Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/10477/89128