Accuracy of High-Frequency Ultrasound in the Detection of White Spot Lesions

Abstract

dc:description.abstract

Objectives: The aim of this study was 1) to determine the accuracy of ultrasound (US) by comparing repeated depth measurements of white spot lesions (WSL) with those of a gold standard of micro-CT (μ-CT) and 2) to assess the bias of the US measurements. Materials and Methods: This in-vitro study included 120 bovine maxillary incisors divided into two groups (SWSL= Shallow WSL and DWSL= Deep WSL). Artificial WSL were created on a 4x4 mm area of the facial surface of each incisor. The SWSL group and the DWSL underwent a 2-day and a 4-day pH cycling protocol, respectively. All samples were imaged with both µ-CT and US using second harmonic imaging at 12/24 MHz. The depth of the WSL was measured three times at the same location on each image, and the average was calculated. Enamel thickness was assessed on the µ-CT images using the same methodology. Results: The average depths for the SWSL were 138.33 µm (±17.84) and 169.22 µm (±37.33) using µ-CT and US, respectively. Average depths in the DWSL were 299.93 µm (±47.67) and 309.28 µm (±75.66), respectively. The absolute mean differences between US and µ-CT were significantly different from zero, suggesting US is not accurate compared to the gold standard (p<0.001). US measures deeper on average than μ-CT in SWSL and DWSL, with a significant difference in SWSL (p< 0.001), reflecting bias. Conclusions: US demonstrates lower accuracy compared to μ-CT in measuring the depth of WSL across varying lesion severities. Furthermore, US overestimates the depths of SWSL and DWSL, with a significant amount in the SWSL group. The clinical implication of this study’s findings is that, currently, US should be used with caution in measuring the depth of WSL. However, unlike µ-CT, US is clinically feasible and there is potential for improved outcomes, warranting further investigation.

Degree

thesis:*
Grantor dc:publisher
State University of New York at Buffalo
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Toothman Sulkowski, Taylor; 0000-0001-5854-3428
Contributors dc:contributor
  • Al-Jewair, Thikriat
  • Orthodontics

Subjects

dc:subject × 6

Rights

dc:rights
Statement dc:rights
  • Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/10477/90280

Chain of custody

source
Harvested from
Buffalo
Base URL
ubir.buffalo.edu/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Toothman Sulkowski, Taylor; 0000-0001-5854-3428. Accuracy of High-Frequency Ultrasound in the Detection of White Spot Lesions. State University of New York at Buffalo, 2025. http://hdl.handle.net/10477/90280