Back to search

State University of New York at Buffalo

Physics and Geometry Input-Based Neural Network Dose Engine with Control Point Input Generator

Abstract

dc:description.abstract

Purpose: This dissertation aims for three purposes: apply Monte Carlo (MC) methods for monitor unit verification (MUV) workflow for Varian TrueBeam VMAT plans; create an open source control point (2-degree sub-arcs) geometry based network input generator framework; and develop a transformer-based deep learning network framework for predicting VMAT dose distributions using this control point-based framework. Methods: For the MC-MUV process, EGSnrc was used to calculate 13 TrueBeam VMAT plans and then the MC and commercial software, MUCheck’s calculations were compared to the treatment planning system calculations using reference point dose; for select plans/arcs, application of a Gaussian filter post-processing to MC doses (2x107 histories only) was also explored. For each control point (CP), the control point aperture dose (CARD) calculation was performed by averaging the ray-tracing of two adjacent aperture shapes to the patient volume, along with first-principles dose calculations. The patient plan CT was cropped/mapped to the dose matrix and HU values also converted to relative electron density and physical density. MC calculations with CP geometry were used as ground truth. Finally, a transformer network architecture was trained on the mapping, y=f(x,r), where x represents the physical density, r represents the CARD and y is the corresponding output dose distribution (the model’s prediction corresponding to ground truth), using a modified version of the Improved Dose Transformer Algorithm (iDoTA). Training was performed on control point datasets obtained from 6 MV clinical brain and lung cases. Evaluations were performed using new brain and lung plans control point datasets. Additional evaluation was performed by comparing the predicted dose with Eclipse Acuros XB calculations. Results: MC-MUV simulation with 5 × 107 histories per arc offered good agreement with Eclipse and a reasonable average calculation time of 9–18 min per full plan. The average absolute dose difference was 3.0%; in contrast to MUCheck’s, 8.4%. Filter application allowed for comparable point dose evaluation to 5x107 (increase of 2-3x histories from non-filtered). The deep learning framework evaluation showed that the input pairs could correlate to output dose with inference times of 15-45ms per control point. The average gamma pass rate (2%, 2 mm, 10% global threshold) for arc dose predictions were above 95%; for (3%, 3 mm, 10%), the GPR was above 90% for control point dose predictions for both brain and lung cases. Comparison with Eclipse Acuros XB showed that most of the arcs predictions had 90% gamma passing rate (for 5%, 0mm criteria), thereby having potential for an MC-MUV framework alternative…

Degree

thesis:*
Grantor dc:publisher
State University of New York at Buffalo
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Pant, Ankit; 0009-0009-9581-8721
Contributors dc:contributor
  • Nazareth, Daryl
  • Radiology

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/10477/90366

Chain of custody

source
Harvested from
Buffalo
Base URL
ubir.buffalo.edu/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Pant, Ankit; 0009-0009-9581-8721. Physics and Geometry Input-Based Neural Network Dose Engine with Control Point Input Generator. State University of New York at Buffalo, 2025. http://hdl.handle.net/10477/90366