Abstract

dc:description.abstract

Radiation exposure from interventional fluoroscopy procedures creates concern in medical procedures, especially due to the enormous scatter radiation that affects both patients and healthcare staff resulting in radiation-induced health hazards that include skin burn injury and eye-lens cataract formation as well as increased cancer risk. The occupational dose of the personnel involved in this type of procedure is seen to be the highest among others, making radiation safety measures a critical component in the workflow for fluoroscopically guided medical procedures. To address this, a real-time Scattered Radiation Display System (SDS) was developed by our group. This system displays a color-coded mapping of procedure-room scatter distributions by accessing the digital output updates of technique and geometric parameters from the angiography system through the controller area network (CAN) bus device. The system provides prompt visual feedback to the staff by display of the procedure-room scatter radiation distribution in horizontal planes at selectable heights, providing an awareness of regions in the room with high radiation levels. This dissertation describes enhancements made to the SDS improving and fostering a safer clinical environment during procedures. The system was initially developed for scatter dose distributions in a limited area near the patient table, and with procedures limited to thoracic and cranial regions of the patient. However, this existing setup posed limitations. The small virtual room size did not accurately reflect standard procedure room dimensions and could not provide sufficient feedback of scatter levels for all medical personnel as they moved about in the room. Additionally, the SDS uses a Lookup Table (LUT) to display scatter distribution, which is limited by its high data storage and load time requirements. To address these challenges, the SDS framework was expanded by creating a robust database of scatter dose distributions using Monte Carlo simulations. These simulations incorporated Zubal phantoms for various anatomical regions including the head, neck, chest, abdomen, and legs. To preprocess the phantom data, create scatter detection plane and extract required material and density information, MATLAB tools were used. The extracted materials and associated densities were then converted into a compatible format for input into the DOSXYZnrc GUI for dose mapping simulation. To accurately represent a typical clinical scenario, the simulation volume was increased, with matrix dimensions ranging from 450x450x76 to 650x650x76 and 1250x1250x76 with a resolution of 4 mm per pixel…

Degree

thesis:*
Grantor dc:publisher
State University of New York at Buffalo
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Orji, Martina; 0009-0007-8306-4876
Contributors dc:contributor
  • Bednarek, Daniel
  • Radiology

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/10477/90428

Chain of custody

source
Harvested from
Buffalo
Base URL
ubir.buffalo.edu/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Orji, Martina; 0009-0007-8306-4876. Enhancements to the Scatter Display System (SDS) for Staff Dose Management During Fluoroscopically Guided Interventional Procedures. State University of New York at Buffalo, 2025. http://hdl.handle.net/10477/90428