Back to search

State University of New York at Buffalo

Fracture Load of Cantilever Implant-Supported Zirconia Fixed Partial Dentures: An In Vitro Study

Abstract

dc:description.abstract

Purpose: The purpose of this study is to evaluate the fracture strength of cantilevered implant supported fixed partial dentures (cFPDs) fabricated using two distinct types of zirconia: conventional high-strength monochromatic composition (3Y-TZP - tetragonal zirconia polycrystal) and high-translucency zirconia with polychromic multilayer and uniform compositions (4Y-PSZ and 5Y-PSZ). The study also aims to examine the impact of varying connector dimensions on the mechanical performance of these materials. Materials and methods: Forty cantilevered implant-supported fixed partial dentures (FPDs) were digitally designed and milled from zirconia blanks, then divided into four groups (n=10) based on material type and connector dimensions: Group 1 (4 × 4 mm2, IPS e.max ZirCAD LT), Group 2 (3 × 3 mm2, IPS e.max ZirCAD LT), Group 3 (4 × 4 mm2, IPS e.max ZirCAD Prime Esthetic), and Group 4 (3 × 3 mm2, IPS e.max ZirCAD Prime Esthetic). All specimens were subjected to static loading until catastrophic fracture occurred, indicated by a sudden load drop, to determine maximum load-to-fracture values. Statistical analysis was performed using two-way analysis of variance (ANOVA), followed by the post hoc test (α=.05). Results: Two-way ANOVA demonstrated significant effects for both material type and connector dimension on the load-to-fracture values (P≤.001). Post hoc analysis revealed significant differences among Groups LT 4x4 mm2, LT 3x3 mm2, PE 4x4 mm2, and PE 3x3 mm2(P≤.001), with no significant difference between LT 3x3 mm2 and PE 4x4 mm2 (P= 0.906). Group LT 4x4 mm2 exhibited the highest mean load-to-fracture value (464.3 ± 240.1 N), while Group PE 3x3 mm2 showed the lowest (109.3 ± 33.3 N). The mode of failure for all specimens involved zirconia framework fracture without damage to the abutment screws or implant analogs. Crack propagation predominantly originated at the distal abutment wall, extending toward the cantilever connector region. Conclusion In conclusion, this study highlights the critical role of material selection and connector dimensions in the fracture strength of cantilevered implant-supported fixed partial dentures (cFPDs). Specifically, 3Y-TZP with larger connector dimensions (4×4 mm2) exhibited significantly higher fracture strength compared to other materials and smaller connector sizes. These findings suggest that optimizing material properties and connector dimensions can enhance the durability of cFPDs in clinical applications, potentially improving the long-term success of implant-supported restorations.

Degree

thesis:*
Grantor dc:publisher
State University of New York at Buffalo
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Alsulaiman, Nada
Contributors dc:contributor
  • Li, Rui
  • Oral Biology

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • Copyright retained by author. Users of works found in University at Buffalo Institutional Repository (UBIR) are responsible for identifying and contacting the copyright owner for permission to reuse. University at Buffalo Libraries do not manage rights for copyright-protected works and cannot assist with permissions.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/10477/90535

Chain of custody

source
Harvested from
Buffalo
Base URL
ubir.buffalo.edu/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Alsulaiman, Nada. Fracture Load of Cantilever Implant-Supported Zirconia Fixed Partial Dentures: An In Vitro Study. State University of New York at Buffalo, 2025. http://hdl.handle.net/10477/90535