Brigham Young University
Design and Characterization of a Miniaturized Spectrometer for Wearable Applications
Abstract
dc:description.abstractAs individual health monitors continue to become more widely adopted in helping individuals make informed decisions, new technologies need to be developed to obtain more biometric data. Spectroscopy is a well-known tool to gain biological information. Traditionally spectrometers are large and expensive making personal or wearable health monitors difficult. Here we present the development and characterization of a miniaturized short wavelength infrared spectrometer for wearable applications. We present a carbon nanotube parallel hole collimator can effectively select a narrow set of allowed angles of light to be separated by a linear variable filter and detected at a photodiode array making a spectrometer. We will go over the calibration of the spectrometer showing a resolution of 13 nm at 1300 nm. Improvements on the original collimator data will be discussed, including carbon nanotube growth without infiltration and growth on transparent substrates. We will also show measurements made on human subjects yielding a pulse.
Degree
thesis:*- Name thesis:degree_name
- PhD
- Year dc:date
- 2022
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Westover, Tyler Richard
Subjects
dc:subject × 5Rights
- Licence dc:rights.license
- Language dc:language
- english
Identifiers
dc:identifier.*- Repository record dc:identifier
- https://scholarsarchive.byu.edu/etd/9631