Back to search

Brigham Young University

Design and Evaluation of a CMOS Contact-Imaging System for Microfluidics

Abstract

dc:description.abstract

A CMOS contact-imaging system for microfluidics is presented. The microsystem integrates a five-layer PDMS microfluidic network and a CMOS image sensor fabricated in a standard 0.18 µ­m technology. The CMOS image sensor consists of two 10×1-pixel array, an amplifier, and a control logic. The imager is able to achieve a low dark signal of 1.67 mV/s, a maximum integration time of 514 s, and a high dynamic range of 75.2 dB at 1 s integration time. The microfluidic device integrates several actuated valve to achieve a fully automated lab-on-a-chip. This work also presents a quantitative comparison of the photomultiplier tube (PMT) and the CMOS contact-imaging system. The CMOS-microfluidic device is validated using an on-chip chemiluminescent analyte.

Degree

thesis:*
Name thesis:degree_name
MS
Year dc:date
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Galan Cherrez, Andres Moroni

Subjects

dc:subject × 3

Rights

Licence dc:rights.license
Language dc:language
english

Identifiers

dc:identifier.*
Repository record dc:identifier
https://scholarsarchive.byu.edu/etd/7219

Chain of custody

source
Harvested from
Brigham Young University
Base URL
scholarsarchive.byu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Galan Cherrez, Andres Moroni. Design and Evaluation of a CMOS Contact-Imaging System for Microfluidics. 2017. https://scholarsarchive.byu.edu/etd/7219