Cal Poly
Comparison of Scanning Electron Microscopy and Confocal Laser Microscopy for Tissue Surface Roughness Characterization
Abstract
dc:description.abstractIt was found that the measurements captured by confocal microscopy and scanning electron microscopy had a statistically significant difference for bovine tissue. There was not a statistically significant for porcine and poultry tissue. The intent of the study is to perform a comparative study to examine efficacy of two distinct technologies for a singular purpose: tissue surface roughness characterization. The two technologies compared are a confocal reflectance microscope and a scanning electron microscope. The comparison was made by comparing two surface roughness parameters [Ra and Rq] within ImageJ. The study examined three different animal species [porcine, bovine, and poultry] to highlight if different tissues presented alternative conclusions for the efficiency of either technology. Additional analysis was produced comparing two cutting methods [Kleen Kut versus conventional], as well as six different poultry processing technique combinations.
Degree
thesis:*- Name thesis:degree_name
- MS in Biomedical Engineering
- Discipline thesis:degree_discipline
- Biomedical Engineering
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Dhaliwal, Tarnvir
- Contributors dc:contributor
-
- Lily Laiho
- Biomedical Engineering
- College of Engineering
Subjects
dc:subject × 7Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.calpoly.edu/theses/2764