Division of Biomedical Engineering
A qualitative holographic study of hemipelvic and acetabular deformation caused by different hip prostheses
Abstract
dc:description.abstractAseptic loosening of the components is probably the most common long-term complication resulting in failure of Total Hip Arthroplasty. The mechanical behaviour of bone under load is one of the contributory causes of loosening encountered at the prosthesis/cement/bone interface. The present study dealt with a series of invitro experiments conducted on epoxy resin models of human hemi-pelves with different commercially available acetabular components implanted in them. These are used for the construction of simplified models of the artificial hip joint (three-dimensional) and of the prosthesis/cement/bone acetabular interface (two-dimensional). Loading conditions for the models included tensioning of the simulated abductor muscles for the hemi-pelvic and femoral loading for the prosthesis/cement/bone interface study. The experimental method employed was real-time holographic interferometry, a stress analysis technique recently used in the biomechanical field, which permitted whole-field simultaneously inspection of deformation patterns. The holographic interferograms were interpreted in a qualitative rather than a quantitative manner. The models do not exactly represent the in-vivo situation. Since this study identified high stresses both in the hip bone as well as in the interface (prosthesis/bone) it is suggested that these stresses are implicated in the mechanical pathogenesis of loosening. The observed changes in stress levels detected in our models could serve as a guide for future designs of acetabular prostheses as well as guide a in surgical techniques.
Degree
thesis:*- Grantor dc:publisher.institution
- Division of Biomedical Engineering
- Year dc:date.issued
- 1989
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Spirakis, Athanasios Apostolou
- Advisor dc:contributor.advisor
-
- Gryzagoridis, Jasson
Subjects
dc:subject × 7Rights
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Dc Identifier Apacitation
- Spirakis, A. A. (1989). <i>A qualitative holographic study of hemipelvic and acetabular deformation caused by different hip prostheses</i>. (Thesis). University of Cape Town ,Faculty of Health Sciences ,Division of Biomedical Engineering. Retrieved from http://hdl.handle.net/11427/25792
- Dc Identifier Chicagocitation
- Spirakis, Athanasios Apostolou. <i>"A qualitative holographic study of hemipelvic and acetabular deformation caused by different hip prostheses."</i> Thesis., University of Cape Town ,Faculty of Health Sciences ,Division of Biomedical Engineering, 1989. http://hdl.handle.net/11427/25792
- Dc Identifier Citation
- Spirakis, A. 1989. A qualitative holographic study of hemipelvic and acetabular deformation caused by different hip prostheses. University of Cape Town.
- Dc Identifier Ris
- TY - Thesis / Dissertation AU - Spirakis, Athanasios Apostolou AB - Aseptic loosening of the components is probably the most common long-term complication resulting in failure of Total Hip Arthroplasty. The mechanical behaviour of bone under load is one of the contributory causes of loosening encountered at the prosthesis/cement/bone interface. The present study dealt with a series of invitro experiments conducted on epoxy resin models of human hemi-pelves with different commercially available acetabular components implanted in them. These are used for the construction of simplified models of the artificial hip joint (three-dimensional) and of the prosthesis/cement/bone acetabular interface (two-dimensional). Loading conditions for the models included tensioning of the simulated abductor muscles for the hemi-pelvic and femoral loading for the prosthesis/cement/bone interface study. The experimental method employed was real-time holographic interferometry, a stress analysis technique recently used in the biomechanical field, which permitted whole-field simultaneously inspection of deformation patterns. The holographic interferograms were interpreted in a qualitative rather than a quantitative manner. The models do not exactly represent the in-vivo situation. Since this study identified high stresses both in the hip bone as well as in the interface (prosthesis/bone) it is suggested that these stresses are implicated in the mechanical pathogenesis of loosening. The observed changes in stress levels detected in our models could serve as a guide for future designs of acetabular prostheses as well as guide a in surgical techniques. DA - 1989 DB - OpenUCT DP - University of Cape Town LK - https://open.uct.ac.za PB - University of Cape Town PY - 1989 T1 - A qualitative holographic study of hemipelvic and acetabular deformation caused by different hip prostheses TI - A qualitative holographic study of hemipelvic and acetabular deformation caused by different hip prostheses UR - http://hdl.handle.net/11427/25792 ER -
- Dc Identifier Vancouvercitation
- Spirakis AA. A qualitative holographic study of hemipelvic and acetabular deformation caused by different hip prostheses. [Thesis]. University of Cape Town ,Faculty of Health Sciences ,Division of Biomedical Engineering, 1989 [cited yyyy month dd]. Available from: http://hdl.handle.net/11427/25792