Assessment of and improvements to a stereophotogrammetric patient positioning system for proton therapy
Abstract
dc:description.abstractThis thesis describes the construction and use of the facemask at the National Accelerator Centre (NAC) as used to both immobilise and position patients for precision proton radiotherapy. The precision achieved using the stereophotogrammetric (SPG) positioning system is measured, and the shortcomings and errors in using the facemask by the SPG system are measured and analysed. The implementation of improvements made to the SPG system is reported upon, and alternative means of both supporting the fiducial markers and immobilising the patient are investigated and evaluated. The accuracy of positioning a facemask using the SPG system is 1.4 mm and of positioning a newly designed frame is 1.6 mm. These measurements were made without using a patient. It is estimated that the total uncertainty of positioning a patient's tumour at the isocentre is 1.6 (1SD) mm using the facemask and it is estimated that the precision using the frame will be less than this value. The largest component of this error (1.39 mm) is due to the error in obtaining the CT scanner co-ordinates. These results are comparable to those obtained by other investigators. The movement of patient bony landmarks within the facemask was measured to be 1.0 ± 0.8 mm. Three main recommendations are that the CT scanner co-ordinating procedure be improved, the SPG computer program be rewritten in parts to achieve greater speed and accuracy, and that the new frame be used. The frame is easier to manufacture than the facemask and allows real time monitoring of the position of the patient's head by the SPG system thus allowing faster throughput of patients and better positioning quality control.
Degree
thesis:*- Grantor dc:publisher.institution
- Division of Medical Physics
- Year dc:date.issued
- 1999
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Hough, Jan K
- Advisor dc:contributor.advisor
-
- Vaughan, Christopher Leonard (Kit)
Subjects
dc:subject × 1Rights
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Dc Identifier Apacitation
- Hough, J. K. (1999). <i>Assessment of and improvements to a stereophotogrammetric patient positioning system for proton therapy</i>. (Thesis). University of Cape Town ,Faculty of Health Sciences ,Division of Medical Physics. Retrieved from http://hdl.handle.net/11427/26784
- Dc Identifier Chicagocitation
- Hough, Jan K. <i>"Assessment of and improvements to a stereophotogrammetric patient positioning system for proton therapy."</i> Thesis., University of Cape Town ,Faculty of Health Sciences ,Division of Medical Physics, 1999. http://hdl.handle.net/11427/26784
- Dc Identifier Citation
- Hough, J. 1999. Assessment of and improvements to a stereophotogrammetric patient positioning system for proton therapy. University of Cape Town.
- Dc Identifier Ris
- TY - Thesis / Dissertation AU - Hough, Jan K AB - This thesis describes the construction and use of the facemask at the National Accelerator Centre (NAC) as used to both immobilise and position patients for precision proton radiotherapy. The precision achieved using the stereophotogrammetric (SPG) positioning system is measured, and the shortcomings and errors in using the facemask by the SPG system are measured and analysed. The implementation of improvements made to the SPG system is reported upon, and alternative means of both supporting the fiducial markers and immobilising the patient are investigated and evaluated. The accuracy of positioning a facemask using the SPG system is 1.4 mm and of positioning a newly designed frame is 1.6 mm. These measurements were made without using a patient. It is estimated that the total uncertainty of positioning a patient's tumour at the isocentre is 1.6 (1SD) mm using the facemask and it is estimated that the precision using the frame will be less than this value. The largest component of this error (1.39 mm) is due to the error in obtaining the CT scanner co-ordinates. These results are comparable to those obtained by other investigators. The movement of patient bony landmarks within the facemask was measured to be 1.0 ± 0.8 mm. Three main recommendations are that the CT scanner co-ordinating procedure be improved, the SPG computer program be rewritten in parts to achieve greater speed and accuracy, and that the new frame be used. The frame is easier to manufacture than the facemask and allows real time monitoring of the position of the patient's head by the SPG system thus allowing faster throughput of patients and better positioning quality control. DA - 1999 DB - OpenUCT DP - University of Cape Town LK - https://open.uct.ac.za PB - University of Cape Town PY - 1999 T1 - Assessment of and improvements to a stereophotogrammetric patient positioning system for proton therapy TI - Assessment of and improvements to a stereophotogrammetric patient positioning system for proton therapy UR - http://hdl.handle.net/11427/26784 ER -
- Dc Identifier Vancouvercitation
- Hough JK. Assessment of and improvements to a stereophotogrammetric patient positioning system for proton therapy. [Thesis]. University of Cape Town ,Faculty of Health Sciences ,Division of Medical Physics, 1999 [cited yyyy month dd]. Available from: http://hdl.handle.net/11427/26784