Background Displaced acromial fractures are challenging to treat. Complex bony anatomy, variable fracture morphology and limitations of available implants present challenges in achieving favourable surgical outcomes. We determined to what extent currently available scapular and clavicular plating systems are able to provide adequate fixation options. Methods Patients presenting to an urban trauma centre with acromial fractures sustained from blunt trauma between 2012 and 2016 were identified (n = 15, 14M / 1F). The fracture patterns were categorized according to location (Type I = 13%, Type II = 27%, Type III = 60%). Computed Tomography (CT) scans were reconstructed to produce three-dimensional (3D) printed anatomical models on which a quantitative fit analysis was performed. Measurements were performed twice, by five separate observers, with fit graded as anatomical fit (< 2mm), intermediate fit (> 2mm) or no-fit. Results The anterior clavicle 6 hole plate fitted best in 45.7% of cases. Acromial plates only achieved 27.3%. The acromion short plate together with the lateral clavicle short plates performed the best in Type II fractures. An inter-observer intraclass correlation coefficient (ICC) agreement of 0.974 was obtained. Conclusion The available commercial acromial plating system fails to provide adequate congruency and fit for fixation. Clavicular plates were superior alternative implants. 3D printed anatomical models can be used effectively to assist in templating implants preoperatively.

Degree

thesis:*
Grantor
Department of Surgery
Year dc:date.issued
2019

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Charilaou, Johan
Advisor dc:contributor.advisor
  • Roche, Stephan

Subjects

dc:subject × 1

Rights

Language dc:language.rfc3066
eng

Identifiers

dc:identifier.*
Dc Identifier Apacitation
Charilaou, J. (2019). <i>Quantitative fit analysis of acromion fracture plating systems using three-dimensional anatomical modelling</i>. (). ,Faculty of Health Sciences ,Department of Surgery. Retrieved from http://hdl.handle.net/11427/31486
Dc Identifier Chicagocitation
Charilaou, Johan. <i>"Quantitative fit analysis of acromion fracture plating systems using three-dimensional anatomical modelling."</i> ., ,Faculty of Health Sciences ,Department of Surgery, 2019. http://hdl.handle.net/11427/31486
Dc Identifier Citation
Charilaou, J. 2019. Quantitative fit analysis of acromion fracture plating systems using three-dimensional anatomical modelling. . ,Faculty of Health Sciences ,Department of Surgery. http://hdl.handle.net/11427/31486
Dc Identifier Ris
TY - Thesis / Dissertation AU - Charilaou, Johan AB - Background Displaced acromial fractures are challenging to treat. Complex bony anatomy, variable fracture morphology and limitations of available implants present challenges in achieving favourable surgical outcomes. We determined to what extent currently available scapular and clavicular plating systems are able to provide adequate fixation options. Methods Patients presenting to an urban trauma centre with acromial fractures sustained from blunt trauma between 2012 and 2016 were identified (n = 15, 14M / 1F). The fracture patterns were categorized according to location (Type I = 13%, Type II = 27%, Type III = 60%). Computed Tomography (CT) scans were reconstructed to produce three-dimensional (3D) printed anatomical models on which a quantitative fit analysis was performed. Measurements were performed twice, by five separate observers, with fit graded as anatomical fit (< 2mm), intermediate fit (> 2mm) or no-fit. Results The anterior clavicle 6 hole plate fitted best in 45.7% of cases. Acromial plates only achieved 27.3%. The acromion short plate together with the lateral clavicle short plates performed the best in Type II fractures. An inter-observer intraclass correlation coefficient (ICC) agreement of 0.974 was obtained. Conclusion The available commercial acromial plating system fails to provide adequate congruency and fit for fixation. Clavicular plates were superior alternative implants. 3D printed anatomical models can be used effectively to assist in templating implants preoperatively. DA - 2019 DB - OpenUCT DP - University of Cape Town KW - Orthopaedic Surgery LK - https://open.uct.ac.za PY - 2019 T1 - Quantitative fit analysis of acromion fracture plating systems using three-dimensional anatomical modelling TI - Quantitative fit analysis of acromion fracture plating systems using three-dimensional anatomical modelling UR - http://hdl.handle.net/11427/31486 ER -
Dc Identifier Vancouvercitation
Charilaou J. Quantitative fit analysis of acromion fracture plating systems using three-dimensional anatomical modelling. []. ,Faculty of Health Sciences ,Department of Surgery, 2019 [cited yyyy month dd]. Available from: http://hdl.handle.net/11427/31486

Chain of custody

source
Harvested from
University of Cape Town
Base URL
open.uct.ac.za/oai/request
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Charilaou, Johan. Quantitative fit analysis of acromion fracture plating systems using three-dimensional anatomical modelling. Department of Surgery, 2019. https://hdl.handle.net/11427/31486