Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study

Abstract

dc:description.abstract

Introduction: The Lubo TM collar is a cervical motion restriction device featuring a unique external jaw-thrust mechanism designed to provide non-invasive airway patency. In addition, tracheal intubation is facilitated by releasing an anterior chin strap; this allows better mouth opening than the previous generation of semi-rigid cervical collars. This study aimed to compare tracheal intubation using the Lubo TM collar combined with manual in-line stabilization (MILS) to intubation with MILS alone. The primary outcome was the time to successful intubation. Secondary outcomes compared intubation success rate, Cormack-Lehane grade, ease of intubation and dental trauma. Methods: A randomized, cross-over, equivalence study was performed. Eighty full-time physician anaesthesia providers were recruited. Participants performed tracheal intubation using direct laryngoscopy on a manikin under two different scenarios: with the Lubo TM collar and MILS applied, and with MILS and no cervical collar. The time to successful intubation was measured and compared using two-one-sided and paired t-tests. Results: Intubation times fell well within the a priori equivalence limits of 10 seconds, with a mean difference (95% CI) of 0.52 seconds (-1.30 to 2.56). There was no significant difference in intubation time with the Lubo TM collar (mean [SD] 19.2 [4.5] seconds) compared to the MILS alone group (19.7 [5.2] seconds). The overall success rate was 98.7% in the Lubo group and 100% in the MILS group. Adequate laryngoscopy views (Cormack-Lehane grades I to IIb) were equivalent between groups (Lubo 92.5% versus MILS alone 93.7%). Conclusion: In this manikin-based study, the time to intubation with the Lubo TM collar and MILS applied was equivalent to time to intubation with MILS alone, with similar intubating conditions. Thus, the Lubo TM collar and MILS may simplify airway management by reducing the number of steps required to perform intubation in patients requiring cervical motion restriction.

Degree

thesis:*
Grantor
Department of Anaesthesia and Perioperative Medicine
Year dc:date.issued
2022

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Behari, Dinell
Advisor dc:contributor.advisor
  • Hofmeyr, Ross

Subjects

dc:subject × 1

Rights

Language dc:language.rfc3066
eng

Identifiers

dc:identifier.*
Dc Identifier Apacitation
Behari, D. (2022). <i>Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study</i>. (). ,Faculty of Health Sciences ,Department of Anaesthesia and Perioperative Medicine. Retrieved from http://hdl.handle.net/11427/37417
Dc Identifier Chicagocitation
Behari, Dinell. <i>"Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study."</i> ., ,Faculty of Health Sciences ,Department of Anaesthesia and Perioperative Medicine, 2022. http://hdl.handle.net/11427/37417
Dc Identifier Citation
Behari, D. 2022. Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study. . ,Faculty of Health Sciences ,Department of Anaesthesia and Perioperative Medicine. http://hdl.handle.net/11427/37417
Dc Identifier Ris
TY - Master Thesis AU - Behari, Dinell AB - Introduction: The Lubo TM collar is a cervical motion restriction device featuring a unique external jaw-thrust mechanism designed to provide non-invasive airway patency. In addition, tracheal intubation is facilitated by releasing an anterior chin strap; this allows better mouth opening than the previous generation of semi-rigid cervical collars. This study aimed to compare tracheal intubation using the Lubo TM collar combined with manual in-line stabilization (MILS) to intubation with MILS alone. The primary outcome was the time to successful intubation. Secondary outcomes compared intubation success rate, Cormack-Lehane grade, ease of intubation and dental trauma. Methods: A randomized, cross-over, equivalence study was performed. Eighty full-time physician anaesthesia providers were recruited. Participants performed tracheal intubation using direct laryngoscopy on a manikin under two different scenarios: with the Lubo TM collar and MILS applied, and with MILS and no cervical collar. The time to successful intubation was measured and compared using two-one-sided and paired t-tests. Results: Intubation times fell well within the a priori equivalence limits of 10 seconds, with a mean difference (95% CI) of 0.52 seconds (-1.30 to 2.56). There was no significant difference in intubation time with the Lubo TM collar (mean [SD] 19.2 [4.5] seconds) compared to the MILS alone group (19.7 [5.2] seconds). The overall success rate was 98.7% in the Lubo group and 100% in the MILS group. Adequate laryngoscopy views (Cormack-Lehane grades I to IIb) were equivalent between groups (Lubo 92.5% versus MILS alone 93.7%). Conclusion: In this manikin-based study, the time to intubation with the Lubo TM collar and MILS applied was equivalent to time to intubation with MILS alone, with similar intubating conditions. Thus, the Lubo TM collar and MILS may simplify airway management by reducing the number of steps required to perform intubation in patients requiring cervical motion restriction. DA - 2022_ DB - OpenUCT DP - University of Cape Town KW - Anaesthesia and Perioperative Medicine LK - https://open.uct.ac.za PY - 2022 T1 - Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study TI - Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study UR - http://hdl.handle.net/11427/37417 ER -
Dc Identifier Vancouvercitation
Behari D. Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study. []. ,Faculty of Health Sciences ,Department of Anaesthesia and Perioperative Medicine, 2022 [cited yyyy month dd]. Available from: http://hdl.handle.net/11427/37417

Chain of custody

source
Harvested from
University of Cape Town
Base URL
open.uct.ac.za/oai/request
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Behari, Dinell. Intubation during spinal motion restriction using the Lubo TM cervical collar a minikin simulation study. Department of Anaesthesia and Perioperative Medicine, 2022. https://hdl.handle.net/11427/37417